YSMAN067 Anglais Rév.1 (Hema Star II Service)
YSMAN067 Anglais Rév.1 (Hema Star II Service)
Technical information
YSMAN067
Summary
SUMMARY 2
ELECTRICITY SUPPLY 1. 2. 3. 4. GENERAL REMARKS FIRST POWER SUPPLY SECOND POWER SUPPLY SCREEN
3 3 3 3 3 4 4 4 4 4 4 5 5 6 6 6 7 8 9 10
DESCRIPTION OF ELECTRONIC BOARDS 1. 2. 3. 4. 5. SYSTEM BOARD PREAMPLIFIER BOARD L10-CI-PRA20 PROCESSING BOARD L1O-PR-FIL3O M ECHANICAL / F LUIDIC MANAGEMENT BOARD L10-CI-DIL30 INTERFACE BOARD L1O-CI-INT-AA
SEQUENCING OF MAIN FLUIDIC CYCLES 1. 2. 3. 4. 5. 6. 7. DILUTION SEQUENCING COUNTING CYCLE RINSING AND DRAINAGE AT THE END OF COUNTING CYCLE (DILUTER) RINSING AND DRAINAGE AT THE END OF CYCLE (FLUIDIC ) SEQUENCING FIRST RINSING CYCLE SEQUENCING OF RINSING CYCLE EXCEPT THE FIRST SEQUENCING OF RINSING CYCLES AFTER 5 MINUTES
10
YSMAN067
Electricity supply
1. General remarks The device works on 10w voltage power from two sources in ventilated casing. The main supply element is Iocated at the back of the device preventing any risk of electrocution. Important: These partitioned power supplies do not have any accessible fuses. Contact your technical support. N.B: the presence of the different voltages displayed by leds on the interface board (appendix E3). 2. First power supply PC type power supply giving all boards the following voltages: +12V -12V +5V Power supply linked to the floppy disk driver and the system board, then to all other boards via the expansion slots and cable layers. It works on 50 or 60 Hz and there is a switch between the two EEC jacks for changing to 115 or 230V. 3. Second power supply Works on 30 VI 2 A, feeding the interface board. Accepts from 47 to 440 Hz and 2 voltage ranges, switched over by internal strap or automatic according to the version: - 95 to 132V -187 to 265V 4. Screen Works on fixed voltage 220 V, 50 to 60 Hz. On request, it may be supplied with different power supply voltages.
YSMAN067
YSMAN067
SYRINGE Sample 32 l BLOOD Amount 11 ml SAMPLE EXT. RINSING 3 ml ext. needle OF NEEDLE FIRST DILUTION 2 ml ext. needle 2 ml int. needle 1.5 ml ext. needle 1.5 ml int. needle to init DROP EXTRACTOR SAMPLE IN DILUTION 1
PHASE
NEEDLE Total ascent Descent 40 mm Ascent 15 mm Descent 15 mm Ascent 5 mm Ascent 5 mm Descent 15 mm Ascent 15 mm Descent 23 mm Ascent 29 mm
HOLDER
Init rinsing
Rinsing WBC
Descent 9 mm
6.5 ml ext. needle 1.5 ml int. needle RINSING LIQUID IN CHAMBER WC Amount 7 ml 7 ml ext. needle
Amount 7 ml RINSING 6 ml ext. needle LIQUID IN CHAMBER RC 1 ml int. needle POSITION AWAIT OF COUNTING To switch
YSMAN067
2. Counting cycle
FIRST COUNTING CHART
PHASE
ACTION Pressure lowering with air syringe Injection of a bubble into the capillary Bubble passes in front of the start sensor Bubble passes in front of the end sensor Control of counting time Transfer of results WC, RC, PLT to PC Counting of duration determined by first chart Transfer of results WBC, RBC, PLT to PC Pressure lowering with air syringe Injection of a bubble into the capillary Bubble passes in front of the start sensor Bubble passes in front of the end sensor Control of counting time Transfer of results WC, RC, PLT to PC Counting of duration determined by third chart Hemoglobin measure Transfer of results WBC, RBC, PLT to PC
READY FOR ANOTHER CYCLE 4. Rinsing and drainage at the end of cycle (fluidic) CHAMBER DIL. WBC
Transfer CHAMBER READY Transfer CHAMBER READY Drainage control CHAMBER DIL. RBC CHAMBER WBC Drainage CHAMBER RBC
Drainage
Drainage
YSMAN067
Descent 50 mm
Init rinsing
REQUEST DRAINAGE OF NEEDLE RINSING CHAMBER REQUEST DRAINAGE OF NEEDLE RINSING CHAMBER
Ascent 30 mm Descent 30 mm
Rinsing WBC
Ascent 30 mm Descent 30 mm
WBC RBC
YSMAN067
Init rinsing
2 ml ext. needle 2 ml ext. needle DISTRIBUTION TO CHAMBER WBC Amount 9 ml 6 ml ext. needle 2 ml int. needle DISTRIBUTION TO CHAMBER RBC Amount 9 ml 6 ml ext. needle 2 ml int. needle
Ascent 30 mm
Descent 30 mm
Rinsing WBC
Ascent 30 mm
REQUEST TRANSFER CHAMBER WBC
WBC RBC
Descent 30 mm
YSMAN067
Descent 50 mm
Init WBC
Ascent 30 mm Descent 30 mm WBC RBC
REQUEST TRANSFER CHAMBER RBC REQUEST CAPILLARY RINSING REQUEST CAPILLARY RINSING MECHANICAL INITIALIZATIONS
YSMAN067
Machine block diagram Trimmer location on preamplifier board Trimmer location on interface board Trimmer location on A/D conversion board
10
YSMAN067
Appendix D1
PREAMPLIFIER BOARD L10-CI-PRA20
11
YSMAN067
Appendix D2
A/D CONVERSION BOARD L10-ANA20
12
YSMAN067
Appendix D21
COMPONENT LOCATION ON CONVERSION BOARD
13
YSMAN067
Appendix D3
PROCESSING BOARD L10-CI-FIL30
14
YSMAN067
Appendix D31
COMPONENT LOCATION ON PROCESSING BOARD
15
YSMAN067
Appendix D4
MECHANICAL / FLUIDIC MANAGEMENT BOARD L10-CI-DIL30
16
YSMAN067
Appendix D41
COMPONENT LOCATION ON MECHANICAL / FLUIDIC MANAGEMENT BOARD
17
YSMAN067
Appendix D5
INTERFACE BOARD L10-CI-INT-AA
18
YSMAN067
Appendix D51
COMPONENT LOCATION ON INTERFACE BOARD
19
YSMAN067
Appendix E1
MACHINE BLOCK DIAGRAM
20
YSMAN067
Appendix E22
TRIMMER LOCATION ON PREAMPLIFIER BOARD
21
YSMAN067
Appendix E3
TRIMMER LOCATION ON INTERFACE BOARD
22
YSMAN067
Appendix E44
TRIMMER LOCATION ON A/D CONVERSION BOARD
23
YSMAN067
Appendix F1
FLUIDIC CABLING
24
YSMAN067
Appendix F2
FLUIDIC SKELETON DIAGRAM
25
YSMAN067
Appendix F21
WBC DILUTION TRANSFER
Tubes concerns
EV contrles
26
YSMAN067
Appendix F22
RBC DILUTION TRANSFER
Tubes concerns
EV contrles
27
YSMAN067
Appendix F23
VOLUMETRIC SENSOR RINSING (DETERGENT CIRCUIT)
Tubes concerns
EV contrles
28
YSMAN067
Appendix F24
COUNTING
Tubes concerns
EV contrles
29
YSMAN067
Appendix F25
DRAINAGE OF COUNTING CHAMBERS
Tubes concerns
EV contrles
30
YSMAN067
Appendix F26
DILUTER SEQUENCES
Diluent loading
31
Appendix F3
DISMOUNTING OF RUBY APERTURES
Rubis S -A / rv.2
Appendix F4
AIR AND LIQUID SYRINGE GREASING
33