COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4
THE LOWEST FORCE OR TORQUE POINT USED IN THE LOADING RANGE
PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 200 LBF 0.001% 9/26/13
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1 TRANSDUCER UNCERTAINTY IN % FOR LOWEST FORCE APPLIED K =2.4 TRANSDUCERUNCERTAINTY IN % FOR LOWEST FORCE APPLIED K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
ASTM E74 OR E2428 LLF % at the lowest calibration force or torque to be used Combined Type A Uncertainty
0.10417%
normal
32
1.04E-3 1.04E-3
1.09E-6 1.09E-6
99.91% 999.07E-3
36.79E-15 36.79E-15
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % AT THE LOWEST FORCE TO BE APPLIED) 0.10422%
1 0.00201 0.00201 0.00200 0.00200 0.00202 0.00200 0.00199 0.00200 0.00200 0.00199 0.002002 9.19E-6 84.44E-12 11.6762E-6 18.827E-6
Repeatability and Reproducibility PLUS ANOVA 2 3 4 5 6 0.00201 0.00200 0.00199 0.00199 0.00195 0.00203 0.00201 0.00200 0.00200 0.00201 0.00195 0.00200 0.00200 0.00201 0.00196 0.00199 0.00198 0.00203 0.00202 0.00204 0.001987 0.00201 #DIV/0! #DIV/0! #DIV/0! 25.84E-6 17.64E-6 #DIV/0! #DIV/0! #DIV/0! 667.78E-12 311.11E-12 #DIV/0! #DIV/0! #DIV/0! F calc Maximum Std. Dev. 25.84E-6 3.8464 F crit Minimum Std. Dev. 9.19E-6 3.3541
PLUS ANOVA 7 8 9 10
ity data
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 10000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00208%
normal
32
20.83E-6 20.83E-6
434.03E-12 434.03E-12
30.09% 300.90E-3
5.89E-21 5.89E-21
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 37.98E-6 212 1.97 74.87E-6
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00380%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 9000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00231%
normal
32
23.15E-6 23.15E-6
535.84E-12 535.84E-12
34.70% 346.99E-3
8.97E-21 8.97E-21
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 39.30E-6 185 1.97 77.53E-6
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00393%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 8000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00260%
normal
32
26.04E-6 26.04E-6
678.17E-12 678.17E-12
40.21% 402.10E-3
14.37E-21 14.37E-21
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 41.07E-6 155 1.98 81.12E-6
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00411%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 7000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00298%
normal
32 9
29.76E-6 29.76E-6
885.77E-12 885.77E-12
46.76% 467.63E-3
24.52E-21 24.52E-21
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 43.52E-6 126 1.98 86.13E-6
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00435%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 6000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00347%
normal
32 9
34.72E-6 34.72E-6
1.21E-9 1.21E-9
54.45% 544.54E-3
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00471%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 5000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00417%
normal
32 9
41.67E-6 41.67E-6
1.74E-9 1.74E-9
63.26% 632.58E-3
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00524%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 4000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00521%
normal
32 9
52.08E-6 52.08E-6
2.71E-9 2.71E-9
72.90% 729.01E-3
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00610%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 3000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.00694%
normal
32 9
69.44E-6 69.44E-6
4.82E-9 4.82E-9
82.71% 827.06E-3
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00764%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 2000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.01042%
normal
32 9
104.17E-6 104.17E-6
10.85E-9 10.85E-9
91.50% 914.97E-3
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.01089%
MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 1000 LBF 0.001% 0.000% 9/26/2013
CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1
Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
0.02083%
normal
32
208.33E-6 208.33E-6
43.40E-9 43.40E-9
97.73% 977.29E-3
Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df
normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty
FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.02107%