Characterization of Ge-Doped Optical Fibres For MV Radiotherapy Dosimetry

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 1

Characterization of Ge-doped optical fibres for MV radiotherapy

dosimetry
ABSTRACT
Ge-doped optical fibres offer promising thermoluminescence (TL) properties
together with small physical size and modest cost.
Their use as dosimeters for postal radiotherapy dose audits of megavoltage photon
beams has been investigated.
Key dosimetric characteristics including reproducibility, linearity, dose rate,
temperature and angular dependence have been established.
A methodology of measuring absorbed dose under reference conditions was
developed. The Ge-doped optical fibres offer linearity between TL yield and
dose,with are producibility of better than 5%, following repeated measurements
(n5) for doses from 5 cGy to 1000cGy.
The fibres also offer dose rate, angular and temperature independence, while an
energy- dependent response of 7% was found over the energy range 6MV to
15MV (TPR20,10 of 0.660,0.723 and 0.774 for 6,10 and 15MV respectively). The
audit methodology has been developed with an expanded uncertainty of 4.22% at
95% confidence interval for the photon beams studied.

RESUMEN
Las fibras pticas dopadas de germanio (Ge) ofrecen prometedora
termoluminiscencia (TL), as como su pequeo tamao y modesto costo.
Se ha investigado su uso como dosmetro, establecindose caractersticas
dosimtricas claves como: reproducibilidad, linealidad, tasa de dosis, temperatura
y dependencia angular.
La fibra ptica dopada de germanio (Ge) ofrece linealidad entre el rendimiento TL
(termoluminiscencia) y dosis.

You might also like