Professional Documents
Culture Documents
Dr. N.G.P. Institute of Technology Coimbatore - 48.: Department OF Electronics and Communication Engineering
Dr. N.G.P. Institute of Technology Coimbatore - 48.: Department OF Electronics and Communication Engineering
0
Where D is size of the broad wall of horn aperture and
0
is free space wave length.
It is also very important to avoid reflection; antenna measurement is done at outdoor
ranges or in anechoic chambers made of absorbing materials.
Antenna measurements are mostly made with unknown antenna as receiver, therefore
several methods are there to measure the gain of antenna. One method is to compare the
unknown antenna with standard gain antenna with known gain. Another method is to use two
identical antennas. One as a transmitter and other as receiver and from following formula the
gain can be calculated.
P
t
0
G1 G2
P
r
= --------------------
(4S)
2
Where P
t
is transmitted power, P
r
is received power, G1, G2 is gain of transmitting and
receiving antenna, S is the radial distance between two antennas.
0
is free space wavelength.
If both transmitting and receiving antennas are identical having gain G then,
P
t
0
G
2
P
r
=--------------------;
(4S)
2
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
In the above equation and can be measured and gain can be computed .As from the above
equation it is not necessary to know the absolute value of P
r
and P
t
. Only ratio is required which
can be measured by VSWR meter.
PROCEDURE
A. Antenna Radiation Pattern
1. Setup the equipments as shown in the figure1. Keeping the axis of both antenna in same
line.
2. Energize the Gunn Oscillator for maximum output at desired frequency with square wave
modulation by tuning square wave amplitude and frequency of modulating signal of
Gunn Power Supply and by tuning the detector.
3. Also tune the S S tuner in the line for maximum output ( if S STunner is in the setup)
4. Obtain full scale deflection (0db) on normal db scale (0-1-db) at any convenient range
swith position of the VSWR meter by gain control knob of VSWR meter or by variable
attenuator.
5. Tune the receiving horn to the left in 2 or 5 steps upto 4-5 and note the corresponding
VSWR db reading in normal db range. When necessary , change the range switch to next
higher range and add 10 db to the observed value.
6. Repeat the above step but this time turn the receiving horn to the right and note down the
readings.
7. Plot a relative power pattern i.e. output vs. angle.
8. From diagram determine 3db width (beam width) of the horn antenna.
B. Gain Measurement
1. Setup the equipments as shown in the figure1. Both horns should be in line.
2. Keep the range db switch meter at 50db position with gain control full.
3. E Energize the Gunn Oscillator for maximum output at desired frequency with
modulating amplitude and frequency of Gunn Power Supply and by tuning of detector.
4. Obtain full scale deflection in VSWR meter with variable attenuator.
5. Replace the transmitting horn by detector mount and change the appropriate db position
to get the deflection. On scale (do not touch the gain control knob). Note and record the
range db position and deflection of VSWR meter,
6. Calculate the difference in db between the power measured in steps 4 and 5.
RESULT
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
Ex No: 6 DETERMINTAION OF GUIDE
WAVELENGTH, FREQUENCY
MEASUREMENT
Date:
AIM:
To determine the frequency and wavelength in a rectangular waveguide working in TE
mode.
EQUIPMENT:
SNo. EQUIPMENT SPECIFICATION QUANTITY
1 Klystron power supply SKPS-610 1
2 Klystron tube with Klystron Mount 2K-25, XM-251 1
3 Isolator XI-621 1
4 Frequency meter XF-710 1
5 Variable Attenuator XA-520 1
6 Slotted section XS 651 1
7 Tunable Probe XP-655 1
8 VSWR Meter SW-115 1
9 Movable Short XT-4813 1
10 Wave Guide Stand XU-535 2
11
Matched Termination XL-400 1
12 Oscilloscope -- 1
13 BNC Cable -- 1
14 Cooling Fan --- 1
THEORY
For dominant TE mode rectangular waveguide I
o,
I
g
and I
c
are related as below
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
(1/o
2
) =(1/g
2
) +(1/c
2
)
where I
o
is free space wavelength
I
g is
guide wavelength
I
c
is cutoff wavelength
For TE mode, I =2a where a is broad dimension of waveguide.
PROCEDURE
1. Setup the components and equipments as shown in figure.
2. Setup variable attenuator at minimum attenuation position.
3. Keep the control knobs of VSWR meter as below:
Range-50db
Input switch- crystal low impedance
Meter switch-normal position
Gain(Coarse & Fine)- Mid position
4. Keep the control knobs of Klystron power supply as below
Beam voltage- OFF
Mod-switch-AM
Beam Voltage knob-Fully anticlockwise
Reflector Voltage- Fully clockwise
AM- Amplitude knob- Around fully clockwise
AM- Frequency knob Around Mid position
5. Switch ON the Klystron power supply, VSWR Meter and cooling fan switch.
6. Switch ON the beam voltage switch and set beam voltage around 250V-300V with help
of beam voltage knob.
7. Adjust the reflector voltage to get some deflection in VSWR meter.
8. Maximize the deflection with AM amplitude and frequency control knob of power
supply.
9. Tune the plunger of Klystron Mount for maximum deflection.
10. Tune the reflector voltage knob for maximum deflection .
11. Tune the probe for maximum deflection in VSWR Meter.
12. Tune the frequency meter knob to get a dip on the VSWR scale and note down the
frequency directly from frequency meter.
13. Replace the Termination with movable short, and detune the frequency meter.
14. Move the probe along the slotted line. The deflection in VSWR meter will vary. Move
the probe to a minimum deflection position, to get accurate reading. If necessary increase
the VSWR Meter range dbswitch to higher position. Note and record the probe position.
15. Move the probe to next minimum position and record the probe position again.
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
16. Calculate the guide wavelength as twice the distance between two successive minimum
position obtained as above.
17.
18. Measure the waveguide inner broad dimension a which will be around 22.86 mm for X-
band.
19. Calculate the frequency by following equation
20. Verify the frequency obtained by frequency meter.
21. Above experiment can be verified at different frequencies.
RESULT:
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
Ex No: 7
IMPEDANCE MEASUREMENT BY
SLOTTED LINE METHOD
Date:
AIM:
To measure an unknown impedance using the smith chart.
EQUIPMENTS:
SNo. EQUIPMENT SPECIFICATION QUANTITY
1 Klystron power supply SKPS-610 1
2 Klystron tube with Klystron Mount 2K-25, XM-251 1
3 Isolator XI-621 1
4 Frequency meter XF-710 1
5 Variable Attenuator XA-520 1
6 Slotted section XS 651 1
7 Tunable Probe XP-655 1
8 VSWR Meter SW-115 1
9 Movable Short XT-4813 1
10 Wave Guide Stand XU-535 2
11 Matched Termination XL-400 1
12 S S Tuner XT 441
13 Oscilloscope -- 1
14 BNC Cable -- 1
15 Cooling Fan --- 1
THEORY:
The impedance at any point on a transmission line can be written in the form R+jX
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
For comparison SWR can be calculated
Where
Reflection co-efficient
Z
0
=characteristics impedance of w/g at operating frequency
Z=load impedance. The measurement is performed in following way.
The unknown device is connected to the slotted line and the position of one minima is
determined. The unknown device is replaced by movable short to the slotted line . Two
successive minima positions are noted.The twice of the difference between minima position will
be guidewave length. One of the minima is used as reference for impedance measurement .find
the difference of reference minima and minima position obtained from unknown load. Let it be
d. Take a smith chart , taking 1 as centre, draw a circle of radius equal to S. mark a point on
circumference of smith chart towards load side at a distance equal to d/g. join the centre with this
point . find the point where it cut the drawn circle.the co-ordinates of this point will show the
normalized impedance of load.
PROCEDURE:
1. Setup the components and equipments as shown in figure.
2. Setup variable attenuator at minimum attenuation position.
3. Keep the control knobs of VSWR meter as below:
Range - 50db position
Input switch - Crystal low impedance
Meter switch - Normal position
Gain(Coarse & Fine)- Mid position
4. Keep the control knobs of Klystron power supply as below
Beam voltage - OFF
Mod-switch -AM
Beam Voltage knob-Fully anticlockwise
www.Vidyarthiplus.com
www.Vidyarthiplus.com
Optical and MicrowaveLab Manual
Dr.NGP INSTITUTE OF TECHNOLOGY DEPT: ECE YEAR & SEM: IV & 07
Reflector Voltage- Fully clockwise
AM- Amplitude knob- Around fully clockwise
AM- Frequency knob Around Mid position
5. Switch ON the Klystron power supply, VSWR Meter and cooling fan switch.
6. Switch ON the beam voltage switch and set beam voltage around 250V-300V with help
of beam voltage knob.
7. Adjust the reflector voltage to get some deflection in VSWR meter.
8. Maximize the deflection with AM amplitude and frequency control knob of power
supply.
9. Tune the plunger of Klystron Mount for maximum deflection.
10. Tune the reflector voltage knob for maximum deflection .
11. Tune the probe for maximum deflection in VSWR Meter.
12. Tune the frequency meter knob to get a dip on the VSWR scale and note down the
frequency directly from frequency meter.
13. Keep the depth of pin S S. Tuner to around 3-4 mm and lock it.
14. Move the probe along the slotted line to get maximum deflection.
15. Adjust VSWR meter gain control knob and variable attenuator until the meter indicates
1.0 on the normal db SWR scale.
16. Move the probe to next minimum position and note down the SWR S
0
on the scale .also
note down the probe position. Let it be d.
17. Remove the SS tuner and matched termination and place movable short at slotted line.
The plunger of short should be at zero.
18. Note the position of two successive minima position .let it be as d
1
and d
2
.Hence
g
=
2(d
1
- d
2
).
19. Calculate
d
-----
g
20. Find out the normalized impedance as described in the theory section.
21. Repeat the same experiment for other frequency if required.
RESULT:
www.Vidyarthiplus.com
www.Vidyarthiplus.com