Liquid flow performance

Mass
lb/min

kg/hr

Volume (1)
gal/min

l/hr

Nominal flow range (2)

CMF010
CMF025
CMF050
CMF100
CMF200
CMF300
CMF400

0 to 3
0 to 40
0 to 125
0 to 500
0 to 1600
0 to 5000
0 to 15,000

0 to 82
0 to 1090
0 to 3400
0 to 13,600
0 to 43,550
0 to 136,080
0 to 409,000

0 to 0.4
0 to 5
0 to 15
0 to 60
0 to 192
0 to 600
0 to 1800

0 to 82
0 to 1090
0 to 3400
0 to 13,600
0 to 43,550
0 to 136,080
0 to 409,000

Maximum flow rate

CMF010
CMF025
CMF050
CMF100
CMF200
CMF300
CMF400

4
80
250
1000
3200
10,000
20,000

108
2180
6800
27,200
87,100
272,160
545,500

0.4
10
30
120
385
1200
2400

108
2180
6800
27,200
87,100
272,160
545,500

Mass flow accuracy (3)

Transmitters with
MVD Technology

0.10% of rate (4)

All other transmitters

0.10% --------------------------------- 100 % of rate


flow rate

Mass flow repeatability (3) Transmitters with


MVD Technology
All other transmitters

Zero stability

CMF010
CMF010P
CMF025
CMF050
CMF100
CMF200
CMF300
CMF400

zero stability

0.05% of rate (4)


zero stability

0.05% --------------------------------- 100 % of rate


flow rate
lb/min
0.000075
0.00015
0.001
0.006
0.025
0.08
0.25
1.50

kg/hr
0.002
0.004
0.027
0.163
0.680
2.18
6.80
40.91

(1) Specifications for volume measurement are based on a process-fluid density of 1 g/cc (1000 kg/m3). For fluids with density other
than 1 g/cc (1000 kg/m3), the volumetric flow rate equals the mass flow rate divided by the fluids density.
(2) Micro Motion has adopted the terminology nominal flow range. The upper limit of this range is the flow rate at which water at
reference conditions causes approximately 15 psid (1 bar) of pressure drop for ELITE sensors.
(3) Accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based on
reference conditions of water at 68 to 77 F (20 to 25 C) and 15 to 30 psig (1 to 2 bar), unless otherwise noted.
zero stability
zero stability
(4) When flow rate is less than --------------------------------- , accuracy equals --------------------------------- 100 % of rate and
flow rate
0.001
zero stability
repeatability equals --------------------------------- 100 % of rate.
flow rate

Micro Motion ELITE Mass Flow and Density Meters