Strata X 30mg/3mL
(Part # 8B-S100-TBJ)
Specimen Pretreatment:
Mix sample 250L in 2.5-mL of water and 2.5-mL of dichloromethane/hexane (2:1 v/v),
vortex for 3 minutes and filter. Separate the (organic) lipid- and hydro-soluble layers in a
separatory funnel. Add 50mg NaCl to the aqueous layer and mix.
Take the hydro-layer for sample loading. Do no discard the organic layer until you have
completed the method.
Load:
Apply 5-10 mL of sample at a flow rate of 1-2 mL/ min. Slow is
Safe.
Wash:
Pass through the cartridge to waste,
2x 300 L D.I. water
Dry column 30 60 seconds at full vacuum to remove all traces of wash solvents.
Elution:
Turn off vacuum, remove manifold lid, and place collection tubes in rack.
Place lid back on manifold.
With the vacuum OFF apply,
2x 600 L Methanol
Allow to sit on the cartridge for 30 seconds then slowly aspirate through the
cartridge under very low vacuum, at a rate not to exceed 2 mL/ min. Slow is
safe.