Date
29/09/2014
Time
14:05
printed by winspiroPRO 5.3.0 - 29/09/2014 14:10:44 - Mod.D10
Calibration test SUCCESSFULLY COMPLETED. The Device is now using the following correction values :
(Expiration: 0,71% ,Inspiration: -1,05%)
Device
SPIROBANK II
Serial Number
Version
Turbine
003885
3.6
Reusable
BTPS
Test Target
1,063
3L
Expiration Correction
0,71 %
Inspiration Correction
-1,05 %
3,04
2,70
3,09
3,09
3,08
3,08
2,92
2,78
3,06
3,06
3,01
2,96
3,03
Expiration (L)
Inspiration (L)
IMPORTANT: Flow and volume values measured during calibration are at ATP (Ambient Temperature and
Pressure) conditions. All spirometry parameters instead are converted to BTPS conditions (Body Temperature and
Pressure, Saturated). Eg a 3L syringe (ATP) measured during calibration will produce an FVC of 3.08 L (BTPS).