Professional Documents
Culture Documents
Pressure Overview of Measurement and Calibration
Pressure Overview of Measurement and Calibration
Measurement and
Calibration
Course Objectives
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Gauge
Vacuum
Compound
Differential
Absolute
26
Gauge Pressure
27
Vacuum
Vacuum gauges measure negative pressures, i.e.
the removal of atmospheric pressure. Using
"gauge pressure zero" as the starting point, the
gauge will indicate the "vacuum level" in positive
numbers as more pressure is removed. Vacuum
gauges usually position zero at 5:00 on the dial
and rotate counter clockwise. Most vacuum
gauges are rated to a full scale of 30 In Hg or 15
psi. Remember that since "vacuum" is simply the
removal of atmospheric pressure, the highest level
of vacuum that can be achieved on a given day is
equal to the barometric pressure.
28
Vacuum (Cont.)
(You can only remove what's there to remove in
the first place!). Users often comment that they
cannot get the gauge to read "all the way down
to 30 In Hg, therefore the gauge is not working
properly ." In most cases, the reason that they
cannot achieve a 30 In Hg vacuum reading is
that the existing barometric pressure is less
than 30 In Hg, thus a 30 In Hg vacuum reading
is unachievable. Vacuum gauges will not tell
you how far you are from a complete vacuum;
they will only tell you how far you are from
gauge pressure zero.
29
Compound Pressure
Starting at "gauge pressure zero," a
compound gauge simply combines the
vacuum indication of the straight vacuum
gauge with the gauge pressure indication of
a gauge pressure gauge. The position of
the zero is dependent upon the full scale
rating of the pressure side. The pointer will
travel in the counterclockwise direction for
vacuum indications, and clockwise for
pressure indications.
30
Differential Pressure
31
Absolute Pressure
32
33
34
ACCURACY:
PROCEDURES/DEFINITIONS
Accuracy inspection Readings at
approximately five points equally
spaced over the dial should be taken,
both upscale and downscale, before
and after lightly rapping the gauge to
remove friction. A pressure standard
with accuracy at least 4 times
greater than the accuracy of the
gauge being tested is recommended.
Biometrix Corporation (80
35
36
37
38
39
40
41
42
43
44
45
NOTES:
1 For operation of test gauge external zero
reset, refer to Figure 3
2 For test gauge calibration procedure,refer
to Figure 4
6.3 Positive Pressure Ranges
a. Remove dial and lightly press pointer onto
pinion at 9:00 oclock position.
b. Apply full scale pressure and rotate span
block as shown in Figure 5a until pointer
rests at 6:00 oclock position.
46
47
Vacuum range
a. Using a pencil, refer to dial and mark the 0 and 25
inch of Hg positions on the case flange.
b. Remove the dial.
c. Apply 25 inches of Hg vacuum.
d. Lightly press pointer onto pinion carefully aligning it
with the 25 inch of Hg vacuum mark on case flange.
e. Release vacuum fully.
f. Note agreement of pointer to zero mark on case
flange.
g. If span is high or low, turn span block as shown in
Figure 5b
h. Repeat steps 4 through 8 until span is correct.
i. Remove the pointer.
48
49
50
51
52
53
54
55
56
57
58
Pressure
Transducer/Transmitter
Pressure transducers are generally available with three types of electrical output;
millivolt, volt and 4-20mA. Below is a summary of the outputs and when they are
best used.
Millivolt Output Pressure Transducers
Transducers with millivolt output are normally the most economical pressure
transducers. The output of the millivolt transducer is nominally around 30mV. The
actual output is directly proportional to the pressure transducer input power or
excitation. If the excitation fluctuates, the output will change also. Because of this
dependence on the excitation level, regulated power supplies are suggested for use
with millivolt transducers. Because the output signal is so low, the transducer
should not be located in an electrically noisy environment. The distances between
the transducer and the readout instrument should also be kept relatively short.
59
60
61
Transducer/Transmitter
Calibration
Calibrations of Transducers/Transmitters
62
In most cases the desired output will be 420mA. Calibration adjustments will be
made at zero and span.
Knowing the fullscale pressure of the
device under test will come in handy when
adjusting for 20mA. Zero of course is just
zero psi although the output in this case
will have a value of 4mA.
63
64
Calibrators
Prior to performing any calibration the selecting
the correct standards is necessary.
In selecting the standards it is important to
consider
1) the range of pressure over which the test will be
performed.
2) the range of temperature over which the test
will be performed
3) The Duration of test.
4) Degree of accuracy of the Unit under test.
65
Documentation
66