SYRINGE PUMP TE-331 S/332 S

INSPECTION RECORD

No

Item

Visual Inspection

Battery Power

Self Check

AC power

AC power

Syringe Detection

Nearly Empty
Alarm

Plunger/Clutch
Disengagement

Stop Switch

10

11

Occlusion
Detection

Flow rate

Check Point
Check the pump body and pole clamp for
damage, etc
Check that unit can be operated on
battery power
Check that self check is made normally
when power is switched on
1. Check the [ Battery ] lamps light on
when AC power is connected
Check [AC/DC]/[Battery] lamp light on
as far as the pump is operated on AC
power
Check that 10, 20, 30, 50 mL syringes
can be detected normally
Check that the [ Nearly Empty ] lamp
blinks when the slider without syringe is
moved to the end of its travel
Plunger/Clutch Disengagement is
detected properly by pushing
in/releasing the contact pin
Check that pressing the [ STOP ] switch
can stop the pump operation
Check that occurance of [ OCCLUSION ]
alarm is within spesification.
50 mL Syringe: 41.4 52.2 N ( 4.23
6.04 kgf ) at flow rate of 100 mL/h and
occlusion limit of level II, or
50 mL Syringe: 53.4 80.0 kPa ( 0.54
0.82 kgf/cm2 ) at flow rate of 50 mL/h
and occlusion limit of level II
Check flow rate accuracy:
Syringe Size (50 mL Syringe )
Test Time (1 Minute )
Flow rate (300 mL/h )

Result
Yes N
o

Value:
..............
...

Value:
..............
...

Remarks:
.....................................................................................

..........................................................................................................

................................................
............................................................................................................................
...........................................................
Serial
Number:

Date:

Checked by: