Professional Documents
Culture Documents
2.10 Diagnostics and Service: Damping With A Field Communicator
2.10 Diagnostics and Service: Damping With A Field Communicator
Section 2: Configuration
00809-0100-4100, Rev AA
March 2013
3051 Wireless pressure transmitter, damping only takes effect when the device is placed in high
power refresh mode and during calibration. In normal power mode, the effective damping is 0.
Note that when the device is in high power refresh mode, battery power will be depleted rapidly.
Determine the appropriate damp setting based on the necessary response time, signal stability,
and other requirements of the loop dynamics of your system. The damping value of your device
is user selectable from 0 to 60 seconds.
2, 2, 2, 5
2.9.3
1.
2.
Within the Pressure Setup box, enter desired damping value and click Send.
3.
Carefully read the warning and click Yes if it is safe to apply the changes.
Write protect
The Rosemount 3051 Wireless pressure transmitter has a software write protect security
feature.
2, 2, 6, 3
2.10
1.
2.
3.
24
Configuration