3051 Wireless pressure transmitter, damping only takes effect when the device is placed in high
power refresh mode and during calibration. In normal power mode, the effective damping is 0.
Note that when the device is in high power refresh mode, battery power will be depleted rapidly.
Determine the appropriate damp setting based on the necessary response time, signal stability,
and other requirements of the loop dynamics of your system. The damping value of your device
is user selectable from 0 to 60 seconds.

Damping with a Field Communicator


From the HOME screen, enter the fast key sequence

Device Dashboard Fast Keys

2, 2, 2, 5

Enter desired Damping Value and select APPLY.

Damping with AMS Device Manager


Right click on the device and select Configure.

2.9.3

1.

Select Manual Setup.

2.

Within the Pressure Setup box, enter desired damping value and click Send.

3.

Carefully read the warning and click Yes if it is safe to apply the changes.

Write protect
The Rosemount 3051 Wireless pressure transmitter has a software write protect security
feature.

Enabling write protect with a Field Communicator


From the HOME screen, enter the fast key sequence

Device Dashboard Fast Keys

2, 2, 6, 3

Select Write Protect to enable.

Enabling write protect with AMS Device Manager


Right click on device and select Configure.

2.10

1.

Select Manual Setup.

2.

Select the tab labeled Device Information.

3.

Select Write Protect to enable this feature.

Diagnostics and service


Diagnostics and service functions listed below are primarily for use after field installation. The
Transmitter Test feature is designed to verify that the transmitter is operating properly, and can