Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

CALIBRATION CERTIFICATE

ISSUED BY
:ABD CALIBRATION LABORATORY
DATE OF ISSUE : 23 August 2015
CERTIFICATE NO: T4539

ARAB BRITISH DYNAMICS


P O. BOX: Sakr Factory No (3) Egypt. Post Code: 11361
Tel.: 22692124 22694759 Fax: 22688726

REFERENCE:
CUSTOMER & ADDRESS:
CALIBRATED DATE:

TAKI/005/2015
TAKI, Egypt
23 August 2015

DESCRIBTION:

DIGITAL THERMOMETER

Page 1 of 2 pages
Approved Signature
Signature:
Name: S.AHMED
Title: T.M

DUE DATE: 23 August 2016

CAL. TEMP.

: 23 3 oC

RESOLUTION

: 0.1 & 1 oC

CODE No.

: SC01463

MANUFACTURER

: KM

TEMP SCALE

: ITS-90

BASIS OF THE TEST:


Calibration 3 points.
CALIBRATION METHOD:
The instrument was calibrated by comparison with calibrated thermocouple type R
(Code: HQ/18/140/0/01).
All measurement scale in use at this laboratory is the international temperature scale of 1990 and SI units.

Calibrated By:

The laboratory follows the international standards for competence of testing and calibration laboratories (ISO/IEC 17025) for
quality system and technical requirements.

CALIBRATION CERTIFICATE

CERTIFICATE No. T4539


Page 2 of 2 Pages

DESCRIPTION: DIGITAL THERMOMETER


CODE No.
: SC01463
CALIBRATION RESULTS

REFERENCE
READINGS
oC

UNIT UNDER
TEST
READINGS oC

ERROR

99.80

101.60

1.80

199.40

200.00

0.60

298.80

299.00

0.20

398.10

397.00

-1.10

497.30

498.00

0.70

oC

REF.
UNCERTAINTY

oC

1.10

You might also like