ISOLTESTER-DIG-PLUS
2CSM341000R1501

ABB

2CSM445004D0202

ISOLTESTER-DIG-RZ

UK

Compatibility

The product is compatible with the following remote signalling


panel devices:
ABB Code
Type
2CSM310000R1521 QSD-DIG 230
2CSM273063R1521 QSD-DIG 230/24

ABB

Max. Num. QSD


4
4

ABB S.p.a ABB SACE Division


V.le dellIndustria, 18
20010 Vittuone (MI), Italy
Tel +39 02 9034 1
FAX +39 02 9034 7609

UK Main functions
ISOLTESTER-DIG is an insulation monitoring device for
medical locations supplied by IT-M networks. It monitors
thermal and electrical overloads in isolating transformers,
sending a warning when one of the monitored parameters
exceeds a set threshold value.
Error / Link-Fail is an auto-diagnostic system which checks if
the device is working correctly, monitoring the prescence of
voltage between terminals.
ISOLTESTER-DIG-PLUS is available with a serial RS485
output using ModubusRTU protocol which allows bidirectional
communication with control systems such as PLCs or LAN,
etc

UK Legend
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.

Display with 3 digits


Green LED, SET: program status
Yellow LED, Alarm: alarm due to one of the monitored
parameter threshold values being exceeded
Red LED, Output Relay: auxiliary relay status
Red LED, Error / Link Fail: alarm due to failure, open
circuit device, PT100 sensor failure
Red LED, R: insulation resistance (k)
Red LED, Z: insulation impedance ( only available for
ISOLTESTER-DIG-PLUS)
Red LED, T1: isolating transformer temperature
Red LED, T2: isolating transformer temperature
Red LED, I: rated current
+/- | UP/DOWN: parameter selection, settings adjustment
and max/min values visualisation (only available for
ISOLTESTER-DIG-PLUS)
RESET | SET: device program menu, alarms snooze
function and cancellation of max/min values
TEST | ENTER: device and panel periodic test and
SETUP settings acknowledgement

Signalling LEDs (points 6, 7, 8, 9, 10) flash when a given


parameter is over the set threshold value.

UK Programming

UK Installation
It is possible to install only one insulation monitoring device for
each independent supply line. Fasten the device onto DIN rail
(35 mm) using a proper fixing system on the rear.

Installation SETUP
Used to define the general functioning conditions
Button
12 + 11
13 / 12
13
13 + 12

Pressure
continuous
quick
quick
continuous

Function
Access installation setup
Adjust parameters
Confirm parameter / next parameter
Default values

Available settings from installation setup, (default values in


bold:)

UK
A.
B.
C.
D.
E.
F.
G.
H.
I.

Legend
Auxiliary supply (rated voltage 115/230 V 50-60Hz)
Error / Link Fail
Insulating resistance and impedance measurement
Output for remote signalling panels
Serial output RS485 (only for ISOLTESTER-DIG-PLUS)
Temperature sensor 2
Temperature sensor 1
Current transformer /5 A input
Programmable auxiliary relay output

Function

Adjustment

Delay for insulation alarm


Delay for impedance alarm
Config. temp. sensor T1
Config. temp. sensor T2
Delay for electrical overload
Transformation rate
Rated frequency
Fail safe (aux relay status)

14 sec
off / 14 sec
off/PTC/PT100
off/PTC/PT100
off / 160 sec
off/1 40
50/60 Hz
NOP (open)
NCL (closed)

Settings for serial output RS485 (only for ISOLTESTER-DIGPLUS)

Function
Network Address
Communication speed
Parity, bit data, bit stop

Adjustment

Indications
LED
Display
1255
SET
ID
2.40/4.80/9.60/19.2 SET
BDr
N81/N82/E81/O81 SET
PAr

Indications
LED
Display
SET+R DLY
SET+Z DLY
SET+T1 nOD
SET+T2 nOD
DLY
SET+I
Ct
SET+I
FrE
SET
SET+Ou SEt
tput
Relay

Adjustment SETUP:
To set threshold values and auxiliary relay activation mode
Button
12
13 /
12
13

Pressure Function
continuous Access adjustment setup
quick

Adjust parameter

quick

Confirm parameter / next parameter

Available settings from adjustment setup, (default values in


bold:)
Function

Adjustment

Resistance threshold
Aux relay activation for low
resistance
Impedance threshold

50500 k
off/on

Aux relay activation for low


impedance*
Aux relay activation for
Error / Link Fail*
Temp threshold T1/T2 (in
case of PT100)
Temp threshold T1/T2 (in
case of PTC)
Aux relay activation for
thermal overload T1/T2*
Current threshold
Aux relay activation for
electrical overload*

off/50500
k
off/on
off/on
off / 30200
C
off/on
off/on
off/099.9
off/on

Indications
LED
Disp.
R
SEt
R+ Output
rEL
Relay
Z
Set
Z+ Output
Relay
Output Relay
+Link Fail
T1/T2

rEL

T1/T2

rEL

T1/T2+
Output Relay
I
T2+ Output
Relay

rEL

* Available only for ISOLTESTER-DIG-PLUS

rEL
Set

Set
rEL

Alarm and error indications


NOTE:
A.

If temperature measurement is taken from a PTC sensor,


ISOLTESTER will display tLO for low temperature and
HOt temperature close to the threshold and tHI for
temperatures exceeding the threshold
B. If a given parameter exceeds the maximum value of the
measure scale, the display will show HI_ (hi_ HIGH)
C. The Output Relay LED will switch on only if the auxiliary
relay is on.
D. The parameters view can be substituted by an error
message and an alarm shown on the Error / Link Fail
LED.

Indication

Display

Low insulation
Low impedance
Thermal overload T1 (PT100)

0499
0499
30200 /
HI
Thermal overload T1 (PTC)
tHI
Thermal overload T2 (PT100)
30200 /
HI
Thermal overload T2 (PTC)
tHI
Electrical overload
199.9 /
HI
Error / Link Fail
LF
Short circuit temp sensor T1/T2 SHr
Open circuit temp sensor T1/T2 Ope

10

UK Wiring diagram

UK Wiring diagram

Single phase network and QSD-DIG 230/24 (or QSD-DIG 230)

Poly phase network

19

20

21

22

115V

Va ux: *

115V

RS485
RS485
(Only for
seriale
porta
ISOLTESTER(solo per
DIG-PLUS)
versione
P LUS )

AUX-L1 AUX-01 AUX-L2 AUX-02

alimentazione
ausiliaria
Vaux: 230Vca

PE

23

24
25
26

VC-L

PT2+A PT2+B

29

28

PT1+A PT1+B
PT1-

31

CT-S 2 CT-S 1

32

30

P T100

33

15

35

R-C

36

R-NC

QSD-DIG 230

34

18

FAULT

R-NO

17

uscita rel
ausiliario

V-P

16

ingresso
corrente
TA .../5A

13

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS

ACUS OVE R COM-P

14

ACK

12

RETE ISOLATA (IT-M)

QSD-DIG
230/24
pann
ello controllo
remoto

PT2-

11

ingresso 1
sonda
P T100
(o PTC)

10

27

VC-PE TES T+ TES T-

ingresso 2
sonda
P T100
(o PTC)

LF-L

controllo controllo
link fail isol ame nto

LF-PE

PC

P T100

LED
Alarm + R
Alarm + Z
Alarm + T1
Alarm + T1
Alarm + T2
Alarm + T2
Alarm + I
Link Fail+R/Z
T1/T2
T1/T2

F-S

V-P

ACUS

ACK

5
6
7
8
9
10
11

OVER FAULT COM-P TEST- TEST-S TEST+

pannello controllo remoto


QSD-DIG 230/24

COM-S

AGLI ALTRI
PANNELLI
QSD-DIG 230/24

CARICO

UK Wiring diagram

UK Wiring diagram

Measuring section detail for a single phase network:

Measuring section detail for a poly phase network:

Insulance
control

Insulance
control

13

14

UK Technical features
Signals
ISOLTESTER-DIG-RZ
Aux supply
Power
consumption
Rated voltage
Measuramentr
current
Measurement
voltage
Control signal
Internal
impedance
Insulation
measurement
Impedance
measurement
Temperature
measurement
Current
measurement
Capacity
measurement
Thresholds

ISOLTESTER-DIGPLUS
115 - 230 V 50-60Hz
5 VA max
6 VA max

Output

24230V 50-60Hz
24250V ac/dc
1 mA max
24V max
Continuous with digital
Composite codified
filter
200 k
- 0 999 k
- resolution 1 k
- accuracy 5% 1 digit
- 0 999 k
- resolution 1 k
- PT100 with 2 or 3 wires, PTC
- 0200 C,
- resolution 1 C,
- accuracy 2 % 1 digit
- TA /5 A external
- accuracy 5% 1 digit, (adjustable
transformation rate 140)
no
- 0 9,9 F
- resolution 0,1 F
Resistance: 50 500 k
Impedance: 50 500 k
Thermal overload: 30 200 C with PT100
Electrical overload: 1 99,9 A

- incorrect wiring (link fail)


- open/short circuit for temp sensor PT100
- internal error
QSD supply (max 2
QSD supply (max 4
QSD), max 24Vdc
QSD), max 24Vdc
Signals to QSD
aux relay for low
programmable aux
resistance, NO-C-NC relay, NO-C-NC 5A
5A 250Vac
250Vac
-

Modules
Weight
Mechanical
features
Terminals
IP degree
Operating
temperature
Storage
temperature
Insulation
Directives

Serial output RS485,


ModbusRTU protocol,
6
0,4 kg
0,5 kg
- fire resistant plastic case
- sealable transparent front cover
screw terminals 2,5 mm2
front IP50, elswhere IP20???
-10 60 C
-25 70 C, humidity < 95%
2,5 kV 60 sec.
CEI-EN 61010-1
CEI-EN 61557-8 / CEI 64.8/7-710 V2/ IEC
60364-7-710 / UNE 20615
CEI-EN 61326-1

UK

Overall dimensions
61

105

19 20 21 22 23 24 25 26 27

28 29 30 31 32 33 34 35 36

ISOLTESTER-DIG-PLUS

1 2 3

4 5 6 7

8 9

Alarm

Out put Error /


Relay Link Fail

45

SET
R [k W]
Z [kW] / C [ F]
T1 [C]
T2 [C]
I [A]

min/M ax

RESET TEST
SET

ENTER

10 11 12 13 14 15 16 17 18

44
ato 35mm
2
m

56

17

2CSM244000R1501

ISOLTESTER-DIG-PLUS
2CSM341000R1501

2CSM445004D0101

ISOLTESTER-DIG-RZ

ABB

Kompatibilitt

Die Gerte sind kompatibel mit den folgenden


Fernsignalisierungspanels:
ABB Identnummer Typ
2CSM310000R1521 QSD-DIG 230
2CSM273063R1521 QSD-DIG 230/24

ABB

Max. Anzahl QSD


4
4

ABB S.p.a ABB SACE Division


V.le dellIndustria, 18
20010 Vittuone (MI), Italy
Tel +39 02 9034 1
FAX +39 02 9034 7609

Funktionsprinzip

Der ISOLTESTER-DIG ist ein Isolationsberwachungsgert fr


IT-M Netzwerke in medizinischen Rumen. Es berwacht
thermische und elektrische berlastungen von Trenntransformatoren und meldet, wenn einer der zu berwachenden
Parameter seinen Grenzwert berschreitet.
Error / Link-Fail ist ein autodiagnostisches System, welches die
korrekte Funktion des Gertes berprft und anzeigt, wenn eine
Spannung an den Eingangsklemmen anliegt.
Der ISOLTESTER-DIG-PLUS ist mit einer seriellen Schnittstelle
RS485 lieferbar, die ein ModbusRTU Protokoll verwendet und
eine bidirektionale Kommunikation mit SPS-Steuerungen oder
LAN-Systemen etc. ermglicht.

Legende

1.
2.
3.

Digitalanzeige 3-stellig
Grne LED, SET: Programmier Status
Gelbe LED, Alarm: Alarm durch einen der berwachten
Parameter bei berschreitung des zulssigen Grenzwertes
Rote LED, Output Relay: Hilfskontakt Status
Rote LED, Error / Link Fail: Alarm wegen Ausfall eines
Stromkreises oder PT100-Sensor Fehler
Rote LED, R: Isolationswiderstand (k)
Rote LED, Z: Isolationsimpedanz ( nur verfgbar bei
ISOLTESTER-DIG-PLUS)
Rote LED, T1: Temperatur Trenntransformator
Rote LED, T2: Temperatur Trenntransformator
Rote LED, I: Bemessungsstrom
+/- | UP/DOWN: Parameter Auswahl, Einstellung und
Anzeige der max/min Werte (nur verfgbar fr
ISOLTESTER-DIG-PLUS)
RESET | SET: Gerte Programm Men, Alarm
Schlummerfunktion und Quittierung von max/min Werten
TEST | ENTER: Periodischer Gerte- und Anzeige Test
und Besttigung der Einstellwerte

4.
5.
6.
7.
8.
9.
10.
11.
12.
13.

Signal LEDs (siehe Punkte 6, 7, 8, 9, 10) blinken, wenn der


jeweilige Parameter ber dem eingestellten Grenzwert liegt.

Installation und Montage

Es ist mglich, fr jede eigenstndige Zuleitung nur ein


Isolationsberwachungsgert zu installieren.
Die Befestigung des Gertes auf der DIN-Schiene erfolgt mittels
eines dazugehrigen Befestigungssystems auf der
Gerterckseite.

Programmierung

Installations SETUP
Zur Einstellung der generellen Funktionen
Tasten
12+11
13 / 12
13
12+12

Bettigung
dauernd
kurz
kurz
dauernd

Funktion
Zugriff auf Installations SETUP
Einstellung der Parameter
Besttigung des Param. / Nchster Param.
Standardwerte

Verfgbare Einstellungen fr Installations Setup, (Standardwerte


fett:)

D
A.
B.
C.
D.
E.
F.
G.
H.
I.

Legende
Hilfsspannung (Bemessungsspannung:115/230 V 50 60Hz)
Fehler / Fehlerhafte Verbindung
Isolationswiderstands- und Impedanzmessung
Ausgnge fr Fernsignalisierungs Panels
Serielle Schnittstelle RS485 (nur fr ISOLTESTER-DIGPLUS)
Temperatursensor 2
Temperatursensor 1
Wandlereingang /5 A
Programmierbarer Hilfskontaktausgang

Funktion

Einstellung

Verz. Alarm Isolationsfehler


Verz. Alarm Impedanzfehler
Konfig. Temp. Sensor T1
Konfig. Temp. Sensor T2
Verz. Alarm el. berlast
bersetzungsverhltnis
Bemessungsfrequenz
Ausfallsicherung
(Hilfskontakt Status)

14 sec
off / 14 sec
off/PTC/PT100
off/PTC/PT100
off / 160 sec
off/1 40
50/60 Hz
NOP (open)
NCL (closed)

Einstellungen fr serielle Schnittstelle RS485 (nur fr


ISOLTESTER-DIG-PLUS
Funktion

Einstellung

Meldungen
LED
Anzeige
1255
Netzwerk Adresse
SET
ID
bertragungsgeschwindig- 2.40/4.80/9.60/19.2 SET
BDr
keit
Bit parity, data bit, stop bit N81/N82/E81/O81 SET
PAr

Meldungen
LED
Anzeige
SET+R DLY
SET+Z DLY
SET+T1 nOD
SET+T2 nOD
DLY
SET+I
Ct
SET+I
FrE
SET
SEt
SET +
Output
Relay

Einstellungs SETUP:
Einstellung der Grenzwerte und Hilfskontakt Aktivierungsmodus
Taste
12
13/12
13

Bettigung
dauernd
kurz
kurz

Funktion
Zugriff auf Einstellungs SETUP
Einstellparameter
Besttigung Parameter / Nchster Parameter

Verfgbare Einstellungen fr Einstellungs Setup, (Standardwerte


fett:)
Funktion

Einstellung

Meldungen
LED
Anz.
50500 k
Widerstandsgrenzwert
R
SEt
off/on
Hilfskontaktaktivierung fr
R+ Output
rEL
Geringer Widerstand
Relay
off/50500 k Z
Impedanzgrenzwert
Set
off/on
Hilfskontaktaktivierung fr
Z+ Output
rEL
Geringe Impedanz*
Relay
off/on
Hilfskontaktaktivierung fr
Output Relay rEL
Fehler / Fehlerh. Verbind.*
+Link Fail
Temperaturgrenzw. T1/T2 off / 30200 C T1/T2
Set
(bei Verwend. von PT100)
Temperaturgrenzw. T1/T2 off/on
T1/T2
rEL
(bei Verwend. von PTC)
off/on
Hilfskontaktaktivierung fr
T1/T2+
rEL
Therm. berlast T1/T2*
Output Relay
off/099.9
Stromgrenzwert
I
Set
off/on
Hilfskontaktaktivierung fr
T2+ Output rEL
Elektrische berlast*
Relay
* Nur bei ISOLTESTER-DIG-PLUS verfgbar

Alarm- und Fehler-Meldungen


Anmerkung:
A.

Wenn die Temperaturmessung mit einem PTC Sensor


durchgefhrt wird, zeigt der ISOLTESTER tLO fr niedrige
Temperatur und HOt bei Temperaturen nahe am Grenzwert
und tHI fr Temperaturen ber dem Grenzwert
B. Wenn ein bestimmter Parameter den maximalen Wert der
Mess-Skala berschreitet, erscheint in der Anzeige HI_ (hi_
HIGH)
C. Die Output Relay LED wird nur eingeschaltet, wenn der
Hilfskontakt aktiviert ist.
D. Die Parameter Ansicht kann ersetzt werden durch eine
Fehlermeldung und einen Alarm, angezeigt durch die Error /
Link Fail LED.

Meldung

Anzeige

Geringer Isolationswiderstand
Geringe Impedanz
Therm. berlast T1 (PT100)

0499
0499
30200 /
HI
Therm. berlast T1 (PTC)
tHI
Therm. berlast T2 (PT100)
30200 /
HI
Therm. berlast T2 (PTC)
tHI
Elektrische berlast
199.9 /
HI
Fehler / Fehlerhafte Verbindung LF
Kurzschluss Temp.Sensor T1/T2 SHr
Offener Stromkreis
Ope
Temp.Sensor T1/T2

10

Anschlussplan

1-phasiges IT-System und QSD-DIG 230/24 (oder QSD-DIG


230)

Anschlussplan

3-phasiges IT-System

19

20

21

22

115V

Va ux: *

115V

RS485
RS485
(Only for
porta
seriale
ISOLTESTER(solo per
DIG-PLUS)
versione
P LUS )

AUX-L1 AUX-01 AUX-L2 AUX-02

alimentazione
ausiliaria
Vaux: 230Vca

PE

23

24
25
26

VC-L

PT2+A PT2+B

29

28

PT1+A PT1+B
PT1-

31

CT-S 2 CT-S 1

32

30

P T100

33

15

35

R-C

36

R-NC

QSD-DIG 230

34

18

FAULT

R-NO

17

uscita rel
ausiliario

V-P

16

ingresso
corrente
TA .../5A

13

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS

ACUS OVE R COM-P

14

ACK

12

RETE ISOLATA (IT-M)

QSD-DIG
230/24
pann
remoto
ello controllo

PT2-

11

ingresso 1
sonda
P T100
(o PTC)

10

27

VC-PE TES T+ TES T-

ingresso 2
sonda
P T100
(o PTC)

LF-L

controllo controllo
link fail isol ame nto

LF-PE

PC

P T100

LED
Alarm + R
Alarm + Z
Alarm + T1
Alarm + T1
Alarm + T2
Alarm + T2
Alarm + I
Link Fail+R/Z
T1/T2
T1/T2

F-S

V-P

ACUS

ACK

5
6
7
8
9
10
11

OVER FAULT COM-P TEST- TEST-S TEST+

pannello controllo remoto


QSD-DIG 230/24

COM-S

AGLI ALTRI
PANNELLI
QSD-DIG 230/24

CARICO

Anschlussplan

Messstreckendetails fr 1-phasiges IT-System:

Anschlussplan

Messstreckendetails fr 3-phasiges IT-System:

Insulance
control

Hilfsspannung

Hilfsspannung

*Vaux: 230V AC

*Vaux: 230V AC

Trenntransformator
Trenntransformator

max 230V L-L


(IT-M)

13

14

Technische Daten
Meldungen
ISOLTESTER-DIG-RZ

Hilfsspannung
Leistungsaufnahme
Bemessungsspannung
Messstrom
Messspannung
Steuersignal

ISOLTESTER-DIGPLUS
115 - 230 V 50-60Hz
5 VA max
6 VA max

24230V 50-60Hz

24250V AC/DC

1 mA max
24V max
Konstant mit digitalem
Kodierte
Filter
Zusammensetzung
Innenwiderstand
200 k
Isolations- 0 999 k
messung
- Auflsung 1 k
- Genauigkeit 5% 1 digit
Widerstands- 0 999 k
messung
- Auflsung 1 k
Temperatur- PT100 mit 2 oder 3 Adern, PTC
messung
- 0200 C,
- Auflsung 1 C,
- Genauigkeit 2 % 1 digit
Strommessung - TA/5 A externer Wandler
- Genaigkeit 5% 1 digit, (einstellbares
bersetzungsverhltnis 140)
Kapazittsnein
- 0 9,9 F
messung
- Auflsung 0,1 F
Grenzwerte
Widerstand: 50 500 k
Impedanz : 50 500 k
Thermische berlast: 30 200 C mit PT100
Elektrische berlast: 1 99,9 A

Ausgnge

- Fehlerhafter Anschluss (fehlerh. Verbindung)


- Offener Stromkreis/Kurzschluss fr Temp.
Sensor PT100
- Interner Fehler
QSD Versorgung (max. QSD Versorgung (max.
2 QSD), max. 24V DC 4 QSD), max. 24V DC
Meldungen fr QSD
Hilfskontakt fr
Programmierbarer
geringen Widerstand, Hilfskontakt, NO-C-NC
NO-C-NC 5A 250V AC 5A 250V AC
-

Serielle Schnittstelle
RS485, ModbusRTU
Protokoll,
6
0,4 kg
0,5 kg
- Feuerbestndiges Kunststoffgehuse
- Plombierbare transparente Frontabdeckung
Schraubklemmen fr 2,5 mm2
Frontseite IP50, ansonsten IP20???
-10 60 C

Teilungseinheiten
Gewicht
Mechanische
Daten
Klemmen
IP Schutzart
Zulssige
Betriebstemperatur
Lagertemperatur
-25 70 C, Luftfeuchtigkeit < 95%
Isolation
2,5 kV 60 sec.
Standards/Normen VDE 0411 Teil 1 (DIN EN 61010-1)
VDE 0413 Teil 8 (DIN EN 61557-8)
VDE 0100 Teil 710 (DIN VDE 0100-710)
VDE 0843-20-1 (DIN EN 61326-1)

Gehuseabmessungen
61

105

19 20 21 22 23 24 25 26 27

28 29 30 31 32 33 34 35 36

ISOLTESTER-DIG-PLUS

1 2 3

4 5 6 7

8 9

Alarm

Out put Error /


Relay Link Fail

45

SET
R [k W]
Z [ kW] / C [ F]
T1 [C]
T2 [C]
I [ A]

min/M ax

RESET TEST
SET

ENTER

10 11 12 13 14 15 16 17 18

44
ato 35mm
2
m

56

17

2CSM244000R1501

ISOLTESTER-DIG-PLUS
2CSM341000R1501

2CSM445004D0701

ISOLTESTER-DIG-RZ

ES Compatibilidad
El producto es compatible con los siguientes paneles de
sealizacin a distancia:
ABB Cdigo tarifa

Tipo

2CSM310000R1521 QSD-DIG 230


2CSM273063R1521 QSD-DIG 230/24

ABB

ABB

Nmero
Mx.QSD
4
4

ABB S.p.a ABB SACE Division


V.le dellIndustria, 18
20010 Vittuone (MI), Italy
Tel +39 02 9034 1
FAX +39 02 9034 7609
2

ES Funciones principales

ES Leyenda

El ISOLTESTER-DIG es un monitor de aislamiento para


instalaciones hospitalarias alimentadas por redes IT.
Comprueba sobrecargas elctricas y trmicas en el
transformador de aislamiento, enviando un aviso cuando uno
de los parmetros monitorizados exceden el valor umbral
establecido.
Error / Link-Fail es un auto diagnstico que comprueba si el
aparato esta trabajando correctamente, monitorizando la
presencia de tensin entre los terminales.

1.
2.
3.

ISOLTESTER-DIG-PLUS incorpora salida serie RS-485 usando


el protocolo ModbusRTU que permite comunicacin
bidireccional con sistemas de control, como por ejemplo PLCs
o LAN, etc..

8.
9.
10.
11.

4.
5.
6.
7.

12.
13.

Display con 3 dgitos


LED Verde, SET: Estado del programa
LED Amarillo, Alarm: Alarma debido a un parmetro
monitorizado a excedido el valor umbral
LED Rojo, Output relay: Estado del rel auxiliar
LED Rojo, Error / Link fail: Alarma debido a un fallo, abre
el circuito, el sensor PT100 falla.
LED Rojo, R: Resistencia de aislamiento (K)
LED Rojo, Z: Impedancia de aislamiento (solo disponible
para ISOLTESTER-DIG_PLUS)
LED Rojo,T1: Temperatura transformador de aislamiento
LED Rojo,T2: Temperatura transformador de aislamiento
LED Rojo, I: Corriente nominal
+/- Subir/Bajar: Seleccin del parmetro, ajustes y
valores de visualizacin mx/min(solo disponible para
ISOLTESTER_DIG_PLUS)
RESET/ SET: Menu de programa, alarma con repeticin,
funcin y cancelacin de valores mx/mn
TEST/ENTER: Test peridico y confirmacin de ajustes
de configuracin

Los LEDs de sealizacin ( puntos 6, 7, 8, 9, 10) se iluminan


cuando el parmetro es superior al valor umbral establecido.

ES

Instalacin

ES Programacin

Es possible instalar solo un monitor de aislamiento por cada


lnea de alimentacin independiente.Montaje en carril DIN
(35mm).

Configuracin de instalacin
Para definir las condiciones generales de funcionamiento
Botn
12 + 11
13 / 12
13
13 + 12

Pulsacin
continuo
rpido
rpido
continuo

Funcin
Acceso a la configuracin de instalacin
Ajuste de parmetros
siguiente parmetro next parameter
Valores por defecto

Ajustes disponibles desde configuracin de instalacin (valores


de defecto en negrita:)

ES Leyenda
A.
B.
C.
D.
E.
F.
G.
H.
I.

Alimentacin 115/230 V 50-60Hz


Error / Link Fail
Medida de la resistencia y impedancia de
aislamiento
Salida para los paneles de sealizacin a
distancia QSD
Salida serie RS-485 (para ISOLTESTERDIG_PLUS)
Sensor de temperatura 2
Sensor de temperatura 1
transformador de corriente /5 A
Salida rel auxiliar programable

Funcin

Ajuste

Retardo en la alarma de
aislamiento
Retardo en la alarma de
impedancia
Configuracin temp. sensor
T1
Configuracin temp. sensor
T2
Retardo por sobrecarga
elctrica
ndice de transformacin
Rated frequency frecuencia
Fallo seguro (estado del
rel aux.)

14 seg

indicaciones
LED
Display
SET+R DLY

off / 14 seg

SET+Z

off/PTC/PT100

SET+T1 nOD

off/PTC/PT100

SET+T2 nOD

off / 160 seg

SET+I

off/1 40
50/60 Hz
NOP (abierto )
NCL (cerrado)

Ct
SET+I
FrE
SET
SET+Ou SEt
tput
Relay

Ajustes para la salida serie RS485 (solo para ISOLTESTERDIG-PLUS)


Funcin
Direccin red
Velocidad comunicacin
Paridad, bit de
informacin, bit stop

Ajuste

Indicaciones
LED
Display
1255
SET
ID
2.40/4.80/9.60/19.2 SET
BDr
N81/N82/E81/O81 SET
PAr

DLY

DLY

Configuracin de ajustes:
Para ajustar los valores de mbral y el modo de activacin del
rel auxiliar
Botn
12
13 / 12
13

Pulsacin
continuo
rpido
rpido

Funcin
Acceso a la configuracin de ajustes
Ajuste parmetro
Siguiente parmetro next parameter

Ajustes disponibles desde configuracin de ajustes (valores de


defecto en negrita)
Funcin

Ajuste

mbral de Resistencia
Activacin del rel auxiliar
por baja resistencia
mbral de impedancia

50500 k
off/on

off/50500
k
Activacin del rel aux. por off/on
baja impedancia*
Activacin del rel aux. por off/on
Error/Link fail*
mbral de temp. T1/T2 (en off / 30200
caso de PT100)
C
mbral de temp. T1/T2 (en off/on
caso de PTC)
Activacin del rel aux. por off/on
sobrecarga trmica T1/T2*
off/099.9
mbral de corriente
Activacin del rel aux. por off/on
sobrecarga elctrica*
* Disponible solo para ISOLTESTER-DIG-PLUS

Indicaciones
LED
Disp.
R
SEt
R+ Output
rEL
Relay
Z
Set
Z+ Output
Relay
Output Relay
+Link Fail
T1/T2

rEL

T1/T2

rEL

T1/T2+
Output Relay
I
T2+ Output
Relay

rEL

rEL
Set

Set
rEL

NOTA:
A.

Indicaciones de error y alarma

Si la temperatura medida es a travs de un sensor PTC ,


el ISOLTESTER indicar tLO para baja temperatura,

Hot para temperatura cercana al mbral y tHI para


cuando la temperatura excede del umbral.

B.

C.
D.

Si un parmetro excede del valor mximo de la


escala de medida, el display mostrar HI_ (hi_
HIGH)
El LED Output relay se encender si el rel auxiliar
esta en on.
Los parmetros vistos pueden ser sustituidos por un
mensaje de error y una alarma en el LED Error / link
Fail

Indicacin

Falta aislamiento
0499
Baja impedancia
0499
Sobrecarga trmica T1 (PT100) 30200 /

HI
tHI
199.9 /
HI
LF
SHr

Sobrecarga trmica T2 (PTC)


Sobrecarga elctrica
Error / Link Fail
Sensor de temepratura
cortocircuito T1/T2
Sensor de temperature circuito
abierto

Ope

10

ES

Esquema de conexiones

Red monofsica y QSD-DIG 230/24 ( QSD-DIG 230)

LED
Alarm + R
Alarm + Z
Alarm + T1

HI
Sobrecarga trmica T1 (PTC)
tHI
Sobrecarga trmica T2 (PT100) 30200 /

ES

Dispaly

Esquema de conexiones

Red trifsica+N

19

20

21

22

115V

Va ux: *

115V

RS485
RS485
(Only for
seriale
porta
ISOLTESTER(solo per
DIG-PLUS)
versione
P LUS )

AUX-L1 AUX-01 AUX-L2 AUX-02

alimentazione
ausiliaria
Vaux: 230Vca

PE

23

24
25
26

VC-L

PT2+A PT2+B

29

28

PT1+A PT1+B
PT1-

31

CT-S 2 CT-S 1

32

30

P T100

33

15

35

R-C

36

R-NC

QSD-DIG 230

34

18

FAULT

R-NO

17

uscita rel
ausiliario

V-P

16

ingresso
corrente
TA .../5A

13

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS

ACUS OVE R COM-P

14

ACK

12

RETE ISOLATA (IT-M)

QSD-DIG
230/24
pann
remoto
ello controllo

PT2-

11

ingresso 1
sonda
P T100
(o PTC)

10

27

VC-PE TES T+ TES T-

ingresso 2
sonda
P T100
(o PTC)

LF-L

controllo controllo
link fail isol ame nto

LF-PE

PC

P T100

Alarm + T1
Alarm + T2
Alarm + T2
Alarm + I
Link Fail+R/Z
T1/T2
T1/T2

F-S

V-P

ACUS

ACK

5
6
7
8
9
10
11

OVER FAULT COM-P TEST- TEST-S TEST+

pannello controllo remoto


QSD-DIG 230/24

COM-S

AGLI ALTRI
PANNELLI
QSD-DIG 230/24

CARICO

ES Esquema de conexiones

ES

Detalle de la seccin medida para un red monfsica

Esquema de conexiones

Detalle de la seccin medida para una red trifsica

Insulance
control

Insulance
control

13

14

ES Caractersticas tcnicas

Seales

ISOLTESTER-DIG-RZ
Tensin Aux.
Consumo de
potencia
Tensin nominal
Corriente
medida
Tensin medida
Seal de control
Impedancia
interna
Medida de
aislamiento
Medida de
impedancia
Medida de
temperatura
Medida de
corriente
Medida de
capacidad
mbral

ISOLTESTER-DIGPLUS
115 - 230 V 50-60Hz
5 VA mx
6 VA mx

24230V 50-60Hz
24250V ac/dc
1 mA mx
24V mx
Continuo con filtro
digital
200 k

codificado

- 0 999 k
- 1 k resolucin
- precisin 5% 1 digito
- 0 999 k
- resolucin1 k
- PT100 con 2 3 cables, PTC
- 0200 C,
- resolucin1 C,
- resolucin 2 % 1 digito
- TA /5 A externo
- precisin 5% 1 digito, (Rango de
transformacin 140)
no
- 0 9,9 F
- resolucin 0,1 F
Resistencia: 50 500 k
Impedancia: 50 500 k
Sobrecarga trmica: 30 200 C con PT100
Sobrecarga elctrica: 1 99,9 A

Salida

- Conexionado incorrecto (link fail)


- Cortocircuito o circuito abierto para sensor
temp.
- Error interno
Panel QSD (mx 2
Panel QSD (mx 4
QSD), mx 24Vdc
QSD), mx 24Vdc
Seales para QSD
Rel aux. para baja
Rel aux. programable,
resistencia,NO-C-NC NO-C-NC 5A 250Vac
5A 250Vac
-

Salida serie RS485


Protocolo ModbusRTU
6
0,4 kg
0,5 kg
- Caja de plstcico resitente al fuego
- Tapa frontal transparente
terminales tornillo 2,5 mm2
Frontal IP50, resto IP20

Mdulos
Peso
Caractersticas
mecnicas
Terminales
Grado
deproteccin IP
Temperatura de
-10 60 C
funcionamiento
Temperatura de
-25 70 C, humedad< 95%
almacenamiento
Aislamiento
2,5 kV 60 seg
Normativa
CEI-EN 61010-1
CEI-EN 61557-8 / CEI 64.8/7-710 V2/ IEC
60364-7-710 / UNE 20615
CEI-EN 61326-1

ES Dimensiones
61

105

19 20 21 22 23 24 25 26 27

28 29 30 31 32 33 34 35 36

ISOLTESTER-DIG-PLUS

1 2 3

4 5 6 7

8 9

Alarm

Out put Error /


Relay Link Fail

45

SET
R [k W]
Z [kW] / C [ F]
T1 [C]
T2 [C]
I [A]

min/M ax

RESET TEST
SET

ENTER

10 11 12 13 14 15 16 17 18

44
ato 35mm
2
m

56

17

2CSM244000R1501

ISOLTESTER-DIG-PLUS
2CSM341000R1501

2CSM445004D1601

ISOLTESTER-DIG-RZ

BR Compatibilidade
O produto compatvel com os seguintes painis de
sinalizao remota:
Cdigo ABB

Tipo

2CSM310000R1521 QSD-DIG 230


2CSM273063R1521 QSD-DIG 230/24

ABB

ABB

N. Max. de
QSD
4
4

ABB S.p.a ABB SACE Division


V.le dellIndustria, 18
20010 Vittuone (MI), Italy
Tel +39 02 9034 1
FAX +39 02 9034 7609
2

BR Funes Principais
ISOLTESTER-DIG um dispositivo de monitoramento de
isolao, para locais de uso mdico, providos de rede com
esquema de aterramento IT-Mdico. Monitora sobrecargas
trmicas e eltricas em transformadores de isolao, enviando
alerta quando um dos parmetros monitorados excede o valor
ajustado.
Error/ Link-Fail um sistema de auto-diagnstico que checa
se o dispositivo est operando corretamente, monitorando a
presena de tenso entre seus terminais.
ISOLTESTER-DIG-PLUS est disponvel com a sada serial
RS485 usando o protocolo Modbus RTU, que permite uma
comunicao bidirecional, com sistemas de controle como
PLCs ou LAN, etc

BR Legenda
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.

12.
13.

Display com 3 dgitos, para visualizao dos valores dos


parmetros sob controle e dos valores ajustados;
LED verde, SET: estado de programao do dispositivo;
LED amarelo, Alarm: alarme quando um dos valores
ajustados for excedido;
LED vermelho, Output Relay: indicao do estado de
sada do rel auxiliar;
LED vermelho, Error / Link Fail: alarme por falha interna,
falta de conexo com a linha, sensor PT100 aberto ou em
curto-circuito;
LED vermelho, R: visualizao do parmetro R (K);
LED vermelho, Z: visualizao do parmetro Z
(disponvel apenas para ISOLTESTER-DIG-PLUS);
LED vermelho, T1: visualizao do parmetro
temperatura no primrio do transformador;
LED vermelho, T2: visualizao do parmetro
temperatura no secundrio do transformador;
LED vermelho, I: visualizao do parmetro corrente;
Boto +/- | UP/DOWN: seleo de visualizao de
parmetro, seleo de ajustes e visualizao dos valores
mx/min (disponvel apenas para ISOLTESTER-DIGPLUS);
Boto RESET | SET: Menu de programao do
dispositivo, desligar o alarme e cancelamento dos valores
mx/min;
Boto TEST | ENTER: Teste peridico do dispositivo e do
painel e reconhecimento dos ajustes de SETUP.

Os LEDs de sinalizao (pontos 6, 7, 8, 9, 10) piscam quando


um dos parmetros est acima do valor pr-estabelecido.

BR Programao

BR Instalao
possvel instalar apenas um monitor de isolao para cada
linha de alimentao independente. Para fixao em trilho DIN
(35 mm) use um sistema de fixao apropriado na parte
traseira.

SETUP de Instalao
Usado para definir as condies gerais de operao
Boto
Pressionamento Funo
12 + 11 contnuo
Acesso ao setup de instalao
13 / 12 rpido
Ajuste de parmetros
Confirmao de parmetro /
13
rpido
prximo parmetro
13 + 12 contnuo
Valores padro
Ajustes disponveis para o setup de Instalao, (valores padro
em negrito:)

BR
A.
B.
C.
D.
E.
F.
G.
H.
I.

Legenda
Fonte de Alimentao Auxiliar (Tenso Nominal 115/230
V 50-60 hz)
Error / Link Fail
Resistncia de isolao e medio de impedncia
Sada para painis de sinalizao remota
Sada serial RS485 (apenas para ISOLTESTER-DIGPLUS)
Sensor de Temperatura do secundrio T2
Sensor de Temperatura do primrio T1
Transformador de corrente, entrada /5 A
Sada rel programvel auxiliar

Funo

Ajuste

Atraso para alarme de


isolao
Atraso para alarme de
impedncia
Config. temp. sensor T1
Config. temp. sensor T2
Atraso para sobrecorrente
Taxa de Transformao
Frequncia Nominal
Fail safe (status do rel
auxiliar)

14 seg

Indicaes
LED
Display
SET+R DLY

off / 14 seg

SET+Z

DLY

off/PTC/PT100
off/PTC/PT100
off / 160 seg
off/1 40
50/60 Hz
NOP (aberto)
NCL (fechado)

SET+T1
SET+T2
SET+I
SET+I
SET
SET+Ou
tput
Relay

nOD
nOD
DLY
Ct
FrE
SEt

Ajustes pra sada serial RS485 (apenas para ISOLTESTERDIG-PLUS)

Funo
Endereo de rede
Velocidade de
comunicao
Paridade, bit de dado, bit
de parada

Ajuste

Indicaes
LED
Display
1255
SET
ID
2.40/4.80/9.60/19.2 SET
BDr
N81/N82/E81/O81

SET

PAr

SETUP de Ajuste:
Para ajustar os valores e modo de ativao do rel auxiliar
Boto Pressionamento Funo
12
contnuo
Acesso ao setup de ajuste
13 / 12 rpido
Ajuste de parmetro
13
Confirmao de parmetro / prximo
rpido
parmetro
Ajustes disponveis para o setup de ajuste, (Valores padro em
negrito:)
Funo

Ajuste

50500 k
Ajuste de resistncia
Ativao do rel auxiliar por off/on
baixa resistncia
off/50500
Ajuste de impedncia
k
Ativao do rel auxiliar por off/on
baixa impedncia*
Ativao do rel auxiliar por off/on
Error / Link Fail*
off / 30200
Ajuste de temperatura
T1/T2 (em caso de PT100) C
off/on
Ajuste de temperatura
T1/T2 (em caso de PTC)
Ativao do rel auxiliar por off/on
sobretemperatura T1/T2*
off/099.9
Ajuste de corrente
Ativao do rel auxiliar por off/on
sobrecorrente*

Indicaes
LED
Disp.
R
SEt
R+ Output
rEL
Relay
Z
Set
Z+ Output
Relay
Output Relay
+Link Fail
T1/T2

rEL

T1/T2

rEL

T1/T2+
Output Relay
I
T2+ Output
Relay

rEL

* Disponvel apenas para ISOLTESTER-DIG-PLUS

rEL
Set

Set
rEL

Indicaes de alarme e erro


NOTAS:
A.

Se a medio de temperatura for tirada de um sensor PTC,


o display do ISOLTESTER ir mostrar tLO para uma
temperatura baixa, HOt para uma temperatura prxima da
ajustada e tHI para temperaturas que excedam o ajuste;
B. Se um dos parmetros dados exceder o fundo de escala, o
display ir mostrar HI_ (hi_ HIGH);
C. O LED Output Relay ir acender apenas se o rel auxiliar
estiver ligado;
D. A viso dos parmetros pode ser substituda por uma
mensagem de erro e um alarme mostrado no LED Error /
Link Fail.

Indicao

Display

LED

Baixa isolao
Baixa impedncia
Sobretemperatura T1 (PT100)

0499
0499
30200 /
HI
tHI
30200 /
HI
tHI
199.9 /
HI
LF
SHr

Alarm + R
Alarm + Z
Alarm + T1

Ope

T1/T2

Sobretemperatura T1 (PTC)
Sobretemperatura T2 (PT100)
Sobretemperatura T2 (PTC)
Sobrecorrente
Error / Link Fail
Curto circuito temp sensor
T1/T2
Abertura ou curto-circuito por
elevao de temperatura no
sensor PT100

10

BR Diagrama Eltrico

BR Diagrama Eltrico

Rede monofsica e QSD-DIG 230/24 (or QSD-DIG 230)

Rede trifsica

19

20

21

22

115V

Va ux: *

115V

RS485
RS485
(Only for
seriale
porta
ISOLTESTER(solo per
DIG-PLUS)
versione
P LUS )

AUX-L1 AUX-01 AUX-L2 AUX-02

alimentazione
ausiliaria
Vaux: 230Vca

PE

23

24
25
26

VC-L

PT2+A PT2+B

29

28

PT1+A PT1+B
PT1-

31

CT-S 2 CT-S 1

32

30

P T100

33

15

35

R-C

36

R-NC

QSD-DIG 230

34

18

FAULT

R-NO

17

uscita rel
ausiliario

V-P

16

ingresso
corrente
TA .../5A

13

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS

ACUS OVE R COM-P

14

ACK

12

RETE ISOLATA (IT-M)

QSD-DIG
230/24
pann
ello controllo
remoto

PT2-

11

ingresso 1
sonda
P T100
(o PTC)

10

27

VC-PE TES T+ TES T-

ingresso 2
sonda
P T100
(o PTC)

LF-L

controllo controllo
link fail isol ame nto

LF-PE

PC

P T100

Alarm + T1
Alarm + T2
Alarm + T2
Alarm + I
Link Fail+R/Z
T1/T2

F-S

V-P

ACUS

ACK

5
6
7
8
9
10
11

OVER FAULT COM-P TEST- TEST-S TEST+

pannello controllo remoto


QSD-DIG 230/24

COM-S

AGLI ALTRI
PANNELLI
QSD-DIG 230/24

CARICO

BR Diagrama Eltrico

BR Diagrama Eltrico

Detalhe da seo de medio de uma rede monofsica:

Detalhe da seo de medio de uma rede trifsica:

Controle de
Isolao

Controle de
Isolao

13

BR Caractersticas Tcnicas
ISOLTESTER-DIG-RZ

ISOLTESTER-DIGPLUS
115 - 230 V 50-60Hz

Alimentao
auxiliar
Consumo de
5 VA max
6 VA max
potncia
Tenso Nominal
24230V 50-60Hz
24250V ac/dc
Corrente
1 mA max
mxima de
medio
Tenso mxima
24V max
de medio
Tipo de sinal de Componente contnuo
Sinal composto
controle
com filtro digital
codificado
Impedncia
200 k
Interna
Medio de
- 0 999 k
isolao
- resoluo 1 k
- preciso 5% 1 dgito
Medio de
- 0 999 k
impedancia
- resoluo 1 k
Medio de
- PT100 com 2 ou 3 fios, PTC
Temperatura
- 0200 C,
- resoluo 1 C,
- preciso 2 % 1 dgito
Medio de
- TA /5 A externa
Corrente
- preciso 5% 1 dgito, (Taxa de
transformao 140)

14

Medio de
Capacitncia
Ajustes

Sinais

Sadas

No

- 0 9,9 F
- resoluo 0,1 F
Resistncia: 50 500 k
Impedncia: 50 500 k
Sobretemperatura: 30 200 C com PT100
Sobrecorrente: 1 99,9 A
- Ligaes incorretas (link fail)
- Abertura ou curto-circuito por elevao de
temperatura no sensor PT100
- Erro interno
Alimentao QSD (max Alimentao QSD (max
2 QSD), max 24Vdc
4 QSD), max 24Vdc
Sinais para o QSD
Rel auxiliar para baixa Rel auxiliar
resistncia, NO-C-NC programvel, NO-C-NC
5A, 250Vac
5A, 250Vac
-

Sada serial RS485,


protocolo ModbusRTU
Mdulos
6
Peso
0,4 kg
0,5 kg
Caractersticas - Caixa plstica resistente ao fogo
mecnicas
- Capa transparente vedada
Terminais
Bitola mxima 2,5 mm2
Grau IP
frontal IP50, outros IP20???
Temperatura de
-10 60 C
operao
Temperatura de
-25 70 C, umidade < 95%
armazenamento
Isolao
2,5 kV em 60 seg.
Normas
CEI-EN 61010-1
CEI-EN 61557-8 / CEI 64.8/7-710 V2/ IEC
60364-7-710 / UNE 20615
CEI-EN 61326-1

BR Dimenses
61

105

19 20 21 22 23 24 25 26 27

28 29 30 31 32 33 34 35 36

ISOLTESTER-DIG-PLUS

1 2 3

4 5 6 7

8 9

Alarm

Out put Error /


Relay Link Fail

45

SET
R [k W]
Z [kW] / C [ F]
T1 [C]
T2 [C]
I [A]

min/M ax

RESET TEST
SET

ENTER

10 11 12 13 14 15 16 17 18

44
ato 35mm
2
m

56

17

2CSM445004D0902

ISOLTESTER-DIG-PLUS
2CSM445004D0902

ABB

2CSM445004D0902

ISOLTESTER-DIG-RZ

Compatibilit

Il prodotto compatibile con i seguenti pannelli di segnalazione


a distanza:
Codice
Tipo
2CSM310000R1521 QSD-DIG 230
2CSM273063R1521 QSD-DIG 230/24

ABB

Num. Max. QSD


4
4

ABB S.p.a Divisione ABB SACE


V.le dellIndustria, 18
20010 Vittuone (MI)
Tel +39 02 9034 1
FAX +39 02 9034 7609
2

Principali funzionalit

ISOLTESTER-DIG un dispositivo per il controllo


dellisolamento in sistemi IT-M. Consente inoltre di effettuare il
monitoraggio del sovraccarico termico ed elettrico del
trasformatore di isolamento ad uso medicale, inviando una
segnalazione di allarme qualora uno dei parametri controllati
superi il valore soglia impostato.
Il sistema di auto-diagnostica Error / Link Fail supervisiona il
corretto funzionamento di ISOLTESTER-DIG, verificando la
presenza di tensione ai capi dei morsetti.
Nella versione ISOLTESTER-DIG-PLUS disponibile unuscita
seriale RS485 per la comunicazione bidirezionale con un
sistema di controllo tipo PLC o PC tramite protocollo
ModbusRTU.

Legenda

1.
2.
3.

Display a 3 digit
LED verde, SET: stato di programmazione strumento
LED giallo, Alarm: allarme per valore parametro fuori
soglia
LED rosso, Output Relay: stato rel ausiliario
LED rosso, Error / Link Fail: allarme per guasto interno,
mancato collegamento alla linea da controllare, sonda di
temperatura PT100 aperta o in corto circuito
LED rosso, R: resistenza di isolamento (k)
LED rosso, Z: impedenza di isolamento (solo per
ISOLTESTER-DIG-PLUS: impedenza di isolamento e
capacit della linea)
LED rosso, T1: temperatura trasformatore
LED rosso, T2: temperatura 2 sensore
LED rosso, I: corrente di linea
+/- | UP/DOWN: selezione del parametro da visualizzare,
regolazione delle impostazioni e visualizzazione dei valori
massimi e minimi memorizzati (solo per ISOLTESTERDIG-PLUS)
RESET | SET: accesso alla programmazione dello
strumento, tacitazione allarmi e cancellazione valori
memorizzati
TEST | ENTER: prova dello strumento e dei pannelli di
segnalazione a distanza e conferma impostazioni SETUP

4.
5.
6.
7.
8.
9.
10.
11.

12.
13.

I LED di segnalazione (punti 6, 7, 8, 9, 10) lampeggiano qualora


il valore del parametro controllato sia al di fuori del valore di
soglia impostato

Installazione

Per ogni linea indipendente possibile utilizzare un solo


dispositivo per il controllo dellisolamento. Fissare il dispositivo
su barra DIN 35mm utilizzando lattacco rapido posto sul retro.

Programmazione

SETUP di installazione
Permette di impostare le condizioni di funzionamento di base
Pulsante Pressione Funzione
12 + 11 continua
ingresso nel setup di installazione
13 / 12 breve
modifica paramentri
conferma impostazione / parametro
13
breve
successivo
13 + 12 continua
ripristino dei valori di fabbrica (default)
Parametri impostabili nel setup di installazione, in grassetto i
valori di fabbrica:

I
A.
B.
C.
D.
E.
F.
G.
H.
I.

Legenda
Alimentazione ausiliaria (tensioni nominali 115/230 V 5060Hz)
Ingresso controllo Error / Link Fail
Misura resistenza ed impedenza isolamento
Connessioni ai pannelli di segnalazione a distanza
Uscita seriale RS485 (solo per ISOLTESTER-DIG-PLUS)
Sonda di temperatura 2
Sonda di temperatura 1
Ingresso trasformatore amperometrico di misura /5 A
Uscita rel ausiliario programmabile

Funzione

Regolazione

Intervento isolamento
Intervento impedenza
Config. sensore di temp. T1
Config. sensore di temp. T2
Ritardo intervento corrente
Rapporto trasformazione
Frequenza nominale rete
Fail safe (stato rel aux)

14 sec
off / 14 sec
off/PTC/PT100
off/PTC/PT100
off / 160 sec
off/1 40
50/60 Hz
NOP (aperto)
NCL (chiuso)

Impostazioni porta seriale RS485 (solo per versione PLUS)

Funzione
Indirizzo di rete
Velocit comunicazione
Parit, bit dati, bit stop

Regolazione

Indicazioni
LED
Display
1255
SET
ID
2.40/4.80/9.60/19.2 SET
BDr
N81/N82/E81/O81 SET
PAr

Indicazioni
LED
Display
SET+R DLY
SET+Z DLY
SET+T1 nOD
SET+T2 nOD
DLY
SET+I
Ct
SET+I
FrE
SET
SET+Ou SEt
tput
Relay

SETUP di regolazione
Permette di regolare le soglie di intervento e le modalit di
attivazione del rel ausiliario
Pulsante Pressione Funzione
12
continua ingresso nel setup di regolazione
13 / 12 breve
modifica parametri
13
conferma impostazione / parametro
breve
successivo
Parametri impostabili setup di regolazione, in grassetto i valori
di fabbrica
Funzione

Regolazione

Soglia intervento resistenza


Attivazione rel aux per
bassa resistenza
Soglia intervento
impedenza
Attivazione rel aux per
bassa impedenza*
Attivazione rel aux per
Error / Link Fail*
Soglia sensore T1/T2 (in
caso di PT100)
Soglia sensore T1/T2 (in
caso di PTC)
Attivazione rel aux per alta
temperatura T1/T2*
Soglia intervento corrente
Attivazione rel aux per alta
corrente*

50500 k
off/on
off/50500
k
off/on
off/on
off / 30200
C
off/on
off/on
off/099.9
off/on

Indicazioni
LED
Disp.
R
SEt
R+ Output
rEL
Relay
Z
Set
Z+ Output
Relay
Output Relay
+Link Fail
T1/T2

rEL

T1/T2

rEL

T1/T2+
Output Relay
I
T2+ Output
Relay

rEL

* Disponibile solo per ISOLTESTER-DIG-PLUS

rEL
Set

Set
rEL

NOTE:

Segnalazioni di allarme ed errore

A.

Segnalazione

Display

LED

Basso isolamento
Bassa impedenza
Sovratemperatura T1 (PT100)

0499
0499
30200 /
HI
tHI
30200 /
HI
tHI
199.9 /
HI
LF
SHr
Ope

Alarm + R
Alarm + Z
Alarm + T1

Se la misura della temperatura avviene attraverso sonde


tipo PTC i valori indicati possono essere solo tLO (bassa
temperatura), HOt (temperatura vicina alla soglia di
intervento) oppure tHI (alta temperatura)
B. Se un parametro di misura supera il valore massimo della
scala il display visualizza la scritta HI_ (HIGH)
C. Il LED Output Relay si accende solo se il rel ausiliario
attivato.
D. La visualizzazione dei parametri di misura abilitati potrebbe
essere sostituita dalleventuale messaggio di errore
associato allaccensione del LED Error / Link Fail.

Sovratemperatura T1 (PTC)
Sovratemperatura T2 (PT100)
Sovratemperatura T2 (PTC)
Sovracorrente
Error / Link Fail
Corto circuito sensore T1/T2
Circuito aperto sensore T1/T2

10

Schema di collegamento

Rete monofose e QSD-DIG 230/24 (o QSD-DIG 230)

Schema di collegamento

Rete trifase

19

20

21

22

115V

Va ux: *

115V

RS485
RS485
(Only for
seriale
porta
ISOLTESTER(solo per
DIG-PLUS)
versione
P LUS )

AUX-L1 AUX-01 AUX-L2 AUX-02

alimentazione
ausiliaria
Vaux: 230Vca

PE

23

24
25
26

VC-L

PT2+A PT2+B

29

28

PT1+A PT1+B
PT1-

31

CT-S 2 CT-S 1

32

30

P T100

33

15

35

R-C

36

R-NC

F-S

V-P

ACUS

ACK

5
6
7
8
9
10
11

OVER FAULT COM-P TEST- TEST-S TEST+

AGLI ALTRI
PANNELLI
QSD-DIG 230/24

CARICO

pannello controllo remoto


QSD-DIG 230/24

COM-S

QSD-DIG 230

34

18

FAULT

R-NO

17

uscita rel
ausiliario

V-P

16

ingresso
corrente
TA .../5A

13

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS

ACUS OVE R COM-P

14

ACK

12

RETE ISOLATA (IT-M)

QSD-DIG
230/24
pann
ello controllo
remoto

PT2-

11

ingresso 1
sonda
P T100
(o PTC)

10

27

VC-PE TES T+ TES T-

ingresso 2
sonda
P T100
(o PTC)

LF-L

controllo controllo
link fail isol ame nto

LF-PE

PC

P T100

Alarm + T1
Alarm + T2
Alarm + T2
Alarm + I
Link Fail+R/Z
T1/T2
T1/T2

Schema di collegamento

Connessione sezione misura su linea IT-M monofase con


trasformatore di isolamento senza presa centrale:

Schema di collegamento

Connessione sezione misura su linea IT-M trifase senza neutro:

13

Caratteristiche tecniche
ISOLTESTER-DIG-RZ

ISOLTESTER-DIGPLUS
115 - 230 V 50-60Hz
5 VA max
6 VA max
24230V 50-60Hz
24250V ca/cc
1 mA max
24V max

Alimentaz. aux
Autoconsumo
Tensione di rete
Corrente misura
Tensione
misura
Segnale di
continuo con filtro
composto codificato
controllo
digitale
Impedenza
200 k
interna
Misura
- 0 999 k
isolamento
- risoluzione 1 k
- precisione 5% 1 digit
Misura
- 0 999 k
impedenza
- risoluzione 1 k
Misure
- PT100 a 2 o 3 fili, PTC
temperatura
- 0200 C,
- risoluzione 1 C,
- precisione 2 % 1 digit
Misura corrente - TA /5 A esterno
- precisione 5% 1 digit, (rapporto
trasformazione selezionabile 140)
Misura capacit
no
- 0 9,9 F
- risoluzione 0,1 F
Soglie
Resistenza: 50 500 k
intervento
Impedenza: 50 500 k
Sovratemperatura: 30 200 C con PT100
Sovraccarico elettrico: 1 99,9 A

14

Segnalazioni
diagnostiche
Uscite

- dispositivo non connesso alla linea (Link Fail)


- sonda termica PT100 aperta o in corto circuito
- errore interno
alimentazione QSD
alimentazione QSD
(max 2 QSD), max
(max 4 QSD), max
24Vcc
24Vcc
Segnalazioni verso QSD
rel ausiliario attivabile
rel ausiliario con
per bassa resistenza, funzioni programmabili,
NO-C-NC 5A 250Vca NO-C-NC 5A 250Vca
-

Moduli
Peso
Caratteristiche
meccaniche
Connessioni
Grado IP
Temperatura di
funzionamento
Temperatura di
stoccaggio
Isolamento
Norme di
riferimento

seriale RS485,
protocollo ModbusRTU

6
0,4 kg
0,5 kg
- contenitore plastico autoestinguente
- calotta frontale trasparente piombabile
morsetti a vite 2,5 mm2
frontale IP50, contenitore IP20
-10 60 C
-25 70 C, umidit < 95%
2,5 kV 60 sec.
- CEI-EN 61010-1
- CEI-EN 61557-8
- UNE 20615
- CEI-EN 61326-1

Dimensioni di ingombro
61

105

19 20 21 22 23 24 25 26 27

28 29 30 31 32 33 34 35 36

ISOLTESTER-DIG-PLUS

1 2 3

4 5 6 7

8 9

Alarm

Out put Error /


Relay Link Fail

45

SET
R [k W]
Z [kW] / C [ F]
T1 [C]
T2 [C]
I [A]

min/M ax

RESET TEST
SET

ENTER

10 11 12 13 14 15 16 17 18

44
ato 35mm
2
m

56

17

2CSM244000R1501

ISOLTESTER-DIG-PLUS
2CSM341000R1501

2CSM445004D2301

ISOLTESTER-DIG-RZ

GR

:
ABB

2CSM310000R1521 QSD-DIG 230


2CSM273063R1521 QSD-DIG 230/24

ABB
ABB
GR

. . QSD
4
4

ABB S.p.a ABB SACE Division


V.le dellIndustria, 18
20010 Vittuone (MI), Italy
Tel +39 02 9034 1
FAX +39 02 9034 7609

ABB SA
13th km Athens Lamia National Rd
GR 14452 Metamorphossi, Greece
Tel: +30 210 2891900
Fax: +30 210 2891999
email: abb@gr.abb.com

GR

ISOLTESTER-DIG
IT-M .

,

.
/ (Error/Link Fall)

,
.

1.
2.
3.

ISOLTESTER-DIG-PLUS RS485
ModbusRTU
PLCs LAN,
..

8.
9.
10.
11.

4.
5.
6.
7.

12.
13.

3
LED, (SET):
LED, (ALARM):

LED, :
LED, / :
, ,
PT100
LED, R: (k)
LED, Z: (
ISOLTESTER-DIG-PLUS)
LED, 1:
LED, 2:
LED, :
+/- | (UP/DOWN): ,
./. (
ISOLTESTER-DIG-PLUS)
/ (RESET | SET):
,
./.
(TEST | ENTER):
SETUP

LED (6, 7, 8, 9, 10)


.

GR

GR

.
DIN (35mm)
.

(SETUP)

12 + 11

13 / 12
13
13 + 12

(setup) ,(
)



. 1
. 2



(
. )

GR
A.
B.
C.
D.
E.
F.
G.
H.
I.

LED

14sec

SET+R

DLY

off / 14sec

SET+Z

DLY

off/PTC/PT100
off/PTC/PT100
off / 160sec

SET+T1
SET+T2
SET+I

nOD
nOD
DLY

off/1 40
50/60 Hz
NOP () NCL
()

SET+I
SET
SET+

Ct
FrE
SEt

RS485 ( ISOLTESTER-DIG-PLUS)

( 115/230V 50-60Hz)
/


RS485 ( ISOLTESTERDIG-PLUS)
2
1
/5A



Parity, bit data, bit stop

ID
BDr
PAr

:
A.

12

13 / 12
13

B.
C.

(setup) , (
)

LED


.


.
*
. /
*
. 1/2 (
PT100)
. 1/2 (
PTC)
.
T1/T2*

50500k
off/on

R
R+

SEt
rEL

off/50500k
off/on

Z
Z+

Set
rEL

off/on

+

T1/T2

rEL

off /30200C
off/on

T1/T2

rEL

off/on

T1/T2+

rEL

off/099.9

Set

.
*

off/on

T2+

rEL

* ISOLTESTER-DIG-PLUS

1255
2.40/4.80/9.60/19.2
N81/N82/E81/O81

LED
SET
SET
SET

(SETUP) :

Set

D.

PTC,
ISOLTESTER tLO HOt
tHI .
, HI_
(hi_ HIGH)
LED
( on).
LED
Error / Link Fail .

GR

19

20

21

22
23
24
25
26

VC-L

PT2+A PT2+B

29

28

PT1+A PT1+B
PT1-

31

CT-S 2 CT-S 1

32

30

P T100

P T100

33

15

34
35

R-C

36

R-NC

18

FAULT

R-NO

17

uscita rel
ausiliario

V-P

16

ingresso
corrente
TA .../5A

13

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS

ACUS OVE R COM-P

14

ACK

12

QSD-DIG
230/24
pann
remoto
ello controllo

PT2-

11

ingresso 1
sonda
P T100
(o PTC)

10

27

VC-PE TES T+ TES T-

F-S

V-P

ACUS

ACK

5
6
7
8
9
10
11

OVER FAULT COM-P TEST- TEST-S TEST+

pannello controllo remoto


QSD-DIG 230/24

COM-S

GR

ingresso 2
sonda
P T100
(o PTC)

LF-L

AGLI ALTRI
PANNELLI
QSD-DIG 230/24

CARICO

RS485
RS485
(Only for
seriale
porta
ISOLTESTER(solo per
DIG-PLUS)
versione
P LUS )

RETE ISOLATA (IT-M)

GR

controllo controllo
link fail isol ame nto

LF-PE

T1/T2
T1/T2

Va ux: *

(Link Fail)+R/Z
SHr

T1/T2
Ope
T1/T2

LF

115V

199.9 / HI

tHI

115V

T2 (PTC)

30200 / HI

AUX-L1 AUX-01 AUX-L2 AUX-02

tHI

T2 (PT100)

alimentazione
ausiliaria
Vaux: 230Vca

T1 (PTC)

PE

30200 / HI

0499

T1 (PT100)

QSD-DIG 230/24 ( QSD-DIG 230)

(Alarm) +
R
(Alarm) +
Z
(Alarm) +
T1
(Alarm) +
T1
(Alarm) +
T2
(Alarm) +
T2
(Alarm) +
I

LED

0499

PC

10

Insulance
control

QSD-DIG 230

GR

GR

Insulance
control

13

- ( )
- / PT100
-
QSD (. 2 QSD),
QSD (. 4 QSD),
. 24V dc
. 24V dc
QSD
.
. ,
, NO-C-NC 5A 250V ac NO-C-NC 5A 250V ac
RS485,
ModbusRTU
6 (17,5mm)
0,4kg
0,5kg
-
-
2,5 mm2

ISOLTESTER-DIG-RZ
ISOLTESTER-DIG-PLUS
115 - 230V 50-60Hz
. 5VA.
. 6VA
24230V 50-60Hz
24250V ac/dc
. 1mA
. 24V


200k

- 0 999k
- (resolution) 1k
- 5% 1

- 0 999k

- (resolution) 1k
- PT100 2 3 , PTC
- 0200 C,
- (resolution) 1C,
- 2% 1

- TA /5A
- 5% 1 , ( 140)

- 0 9,9F
- (resolution)
0,1F

: 50 500k
: 50 500k
: 30 200C PT100
: 1 99,9A

GR

14

61

105

19 20 21 22 23 24 25 26 27

28 29 30 31 32 33 34 35 36

ISOLTESTER-DIG-PLUS

SET
R [k W]
Z [ kW] / C [ F]
T1 [C]
T2 [C]
I [ A]

IP 50, IP 20
-10 60 C
1 2 3

4 5 6 7

8 9

Alarm

Out put Error /


Relay Link Fail

45

min/M ax

RESET TEST
SET

ENTER

10 11 12 13 14 15 16 17 18

-25 70C, < 95%


2,5kV 60sec.
CEI-EN 61010-1
CEI-EN 61557-8 / CEI 64.8/7-710 V2/ IEC 60364-7-710 / UNE 20615
CEI-EN 61326-1