The following Procedure should be carried out at least twice a year at 1000 hour and 2000
hour Preventative Maintenance visits or as required to remove Bacterial or Fungal buildup from the RX Imola Analyser.
C1 Wash solution is available from Randox order number RX3973.
1.
2.
3.
4.
5.
6.
7.
8.
9.
9.
Loosen bottle tops and remove all tubes and the floating sensor.
Empty the sub tank and fill it with C1 solution up to the high level mark.
Insert all tubes back into the sub Tank.
Repeat steps from 3 to 6 for the R-sub tank.
Disconnect water supply tubing.
Switch on the analyser and the PC.
Press cancel on the pop-up screen appears. See on picture below
Composed VY
Checked RD
May 25th 2006
ver. 1.0