STRAIGHT CONNECTOR CHANGE-OUT- NO LUER LOCK

Established: 1998

Revised: Nov. 2013

PURPOSE
To safely change-out a faulty or clotted connector

SITE APPLICABILITY
Applicable to all areas within BCCH and all patients requiring ECLS.

PRACTICE LEVEL/COMPETENCIES
ECLS is considered an advanced skill and only those with advanced training in ECLS are to perform this
task.

Perfusionist

In the event of an emergency or directed by the Perfusionist

ECLS Team leader

ECLS Specialist

EQUIPMENT
1.
2.
3.
4.
5.
6.
7.
8.

8 clamps.
Sterile towels
Sterile scissors
2 x 60 ml syringes
Sterile connector
Sterile gloves (staff protection)
2% chlorhexidine gluconate with 70% alcohol swab (CHG/alcohol)
Blunt needle

PROCEDURE
1. Prepare for separation from bypass as per protocol.
2. Cool the patient if applicable (see cooling guidelines).
3. Separate as per protocol
4. Place clamp #1 at connector inlet tubing place clamp #2
5. Place clamp #3 at connector outlet, tubing place clamp #4
6. CHG/alcohol swab between clamp #1 and clamp # 2 and between clamp # 3 and clamp #4.
7. Use sterile scissors to cut the tubing next to clamp #1 and clamp #3.
8. Insert new connector into inlet tubing
9. Fill and de-air the new connector with crystalloid using a 60ml syringe and blunt needle.
10. Fill the outlet tubing with crystalloid and connect the outlet tubing to the new connector. Have an
assistant drip into connector point during the maneuver.
Policy #

BCCH Child & Youth Health Policy and Procedure Manual

Page 1 of 2

STRAIGHT CONNECTOR CHANGE-OUT- NO LUER LOCK


Established: 1998

Revised: Nov. 2013

11. Remove clamps # 3 and # 4


12. Remove clamp from bridge
13. Turn the pump RPMs to 1000 and re-circulate via the bridge.
14. Check for air at the oxygenator and the blood pump.
15. If circuit is free of air clamp the bridge and resume ECLS as per routine.
16. Re-establish bridge flow via laminar bridge clamp or thumb screw clamp.
17. Dispose of old equipment appropriately

DOCUMENTATION
Chart all ECLS parameters on ECLS flow sheet.