Techfors S

sterilisation

Step 1
0

Air in
Steam in

Water out

Stop
Wait for launching
sterilisation

0
0

0
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 2
0

sterilisatio
n

Air in
Steam in

Water out

Sterilisation start
Wait for T >=121C
or abort sterilisation

0
0
I

0
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 3
0

sterilisatio
n

Air in
Steam in

Water out

Sterilisation
plateau
Wait for plateau time
>= Sterillisation time
or abort sterillisation

0
0
I

0
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 4
0

sterilisatio
n

Air in
Steam in

Water out

Cooling to 103
Wait for T <= 103C

0
0

0
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 5
0

sterilisati
on

Air in
Steam in

Water out

Cooling to 98C
Wait for T <= 98C

I
0

I
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 6
0

sterilisatio
n

Air in
Steam in

Water out

Cooling to setpoint +10


Wait for T <= Set point T +10C
or Stop sequence

I
0

I
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change
0

Step 7
0

operation
Air in
Steam in

Water out

End of cooling
Wait for T <= Set point T
or stop
or launch an other
sequence

I
0

I
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 8
0

operation
Air in
Steam in

Water out

Waiting
Wait for operator stop
or launch an other
sequence

I
I
0

I
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change

Step 9
0

operation
Air in
Steam in

Water out

Heating

I
0

I
Water in

Valves:
0= closed
I= open
C= controlled
Orange= Status
change