Professional Documents
Culture Documents
Rail-Mounted Smart Temperature Transmitter Type Li-23: Electrical Diagrams
Rail-Mounted Smart Temperature Transmitter Type Li-23: Electrical Diagrams
12,5
99
114,5
[
?
?
?
?
?
?
?
Co
Electrical diagrams.
nw
er
ter
RS
RS232
GND
7
Rx
24
Tx
15
Rt
3
2
I
+
Tc
RS
_
+
7
Ucc
Ro
3
2
Rt
Tc
1
I
+
_
Hart
_
+
Conwerter
RS-Hart
RS232
Ucc
Ro > 250 W
52
Edition I/2007
Technical data
Input signal
Limit process
Min. measuring range
Output signal
Power supply
Load resistance
Alarm signal
Load current
Galvanic insulation
Accuracy
Thermal error
Voltage error
Ambient temperature
Degree of protection
Weight
K, J,S,B,N,T voltage
Pt100,Ni100 resistance
-10mV< E<90mV
20<R<380
2mV or 10
4 20mA
1036V DC
R0[k]<(Uz 11V)/25mA
23mA or 3,8 mA
<75mA
Optoelectrical
0,2%
0,1 %/10C
0,1%
-25+80C
IP20
0,1kg
Special version:
U input only for thermocouple
R input only for thermoresistance
V output signal 010V
Ordering procedure
LI-23/__/__/__/__/__:__C/___
Special version: U, R, V
Communication: RS
Measuring range
Sensor type
Example 1: Temperature transmitter LI-23, special version - only for thermocouple, RS communication,
sensor type K, measuring range from 400 to 800C, alarm signal 23mA.
LI-23/U/RS/K/400 - 800C/23mA
Example 2: Temperature transmitter LI-23, RS communication, no programmed.
LI-23/RS
53