Professional Documents
Culture Documents
Article Comparison
Article Comparison
Article Comparison
Ryan Clark
2/27/2016
Dos 711 Research Methods in Medical Dosimetry I
The field of medical dosimetry is heavily dependent on the constant advances of medical
technology that are currently being implemented in todays medical field. Advances in the
quality and safety of imaging devices greatly benefit 3D treatment plans. PET/CT scans are an
important part of the treatment planning process as they can be fused with simulation CT scans
to allow for better contouring and to easily define target volumes. This article shows that
manufacturers are working to improve the slice thickness and reduction of noise in future PET
scans. PET/CT scanners are being utilized more often in treatment clinics and may be utilized
even more if manufactures are able to further minimize the radiation exposure for each exam
while still improving the quality of the images.
When analyzing this article, there are several key points that I felt greatly benefited this
piece. The problem of why PET scans are having difficulties in reducing the dose of radiation
given to patients is clearly stated and is then followed by what the manufactures of the PET/CT
scanners are doing to help address the issue. The word choice used was appropriate and did not
use medical jargon. If there was a term or phrase that might confuse the average reader,
Orenstein would quickly follow up with a definition. The facts provided on how the industry was
improving were thought provoking and came from the leading manufactures in the industry such
as Siemens, GE, and Toshiba. The article guides readers through simple to more complex
solutions on solving the problem of helping reduce the patient dose of PET/CT scans. These
solutions are organized by explaining what each of these manufactures has done and will do to
help make reducing patient dose a top priority.
Despite these positive points, I feel that there are still a few flaws that can be found
within this trade publication article. Since this is a trade publication, it is not reviewed by any
peer editors and shows a biased view towards only the positive things the manufacturers are
providing. There is never a simple counter statement or differing view on PET/CT scans from
anyone outside of the industry that manufactures the scanners. There is only a small excerpt at
the end by Richard Morin where he is quoted that PET/CT scans should only be used for certain
clinical applications.2 Theoretically, the best way to reduce the radiation dose to a patient would
be to not have a PET/CT scan performed at all. Instead, I feel that this article could have better
highlighted ways in which standards could be created by those in the field of nuclear medicine to
determine when a patient truly would need a PET/CT scan. This article seems biased towards
only focusing on the benefits of what the scanners are going to be capable of in the future.
References:
1. Lenards N, Weege M. Readings in Radiation Therapy & Medical Dosimetry.
[Powerpoint]. La Crosse, WI: UW-L Medical Dosimetry Program; 2016.
2. Orenstein B. Reducing PET Dose. Radiology Today. January 2015; 17(1):22. Retrieved
from http://www.radiologytoday.net/archive/rt0115p22.shtml
from the scans with the manufacturer reported HU values.2 The difference of PDD for when the
couch was present or not was illustrated in a table for both 4-MV and 10-MV. The data is
accurately shown for the PDDs acquired by using measurement square points and calculations
lines. The results are then related to different IMRT treatment plans in a table to show the 1, 2,
and 3 percent difference marks. These treatment plans include how the accuracy of the table
density can affect brain, pelvis, and abdomen treatment plans.2
The discussion takes the data and concisely answers the question that was asked in the
introduction portion of the article. The results show that there is a significant difference between
the HU units that are provided by the table manufacturer and by what is actually scanned by the
CT.2 Despite this, there was only a negligible 0.8% difference in dose given between the two
couch structures. The significant find of the study was that there is a large difference in the dose
delivered between if a table is included or not included within the treatment plan. It was found
that the dose delivered dose can be as large as 9.5% and 7.3% for 4-MV and 10-MV beams
respectively.2 The discussion does a great job evaluating how the results found in the study
confirm what other prior studies have found relating to couch structures and PDD. There is no
evidence of bias that could possibly hinder the data that was found from being accurate in this
article.
The conclusion accurately summarized the research process and highlighted the most
important findings from the tests that were performed. The authors stressed the importance that
future treatment plans should always include the couch structures for IMRT and RapidArc.2 A
wide variety of different articles and journals were used from reputable sources within the
reference section. Though, some of these references seem a little dated and go back to the late
1990s and early 2000s.
I felt that the peer-reviewed article did a much better job of walking the reader through
the scientific research process to answer the question at hand. It was clear that more extensive
research and thought was put into the work. The author seemed more unbiased and even came to
the conclusion that some of the research that was conducted did not have any significant result
for the field of medical dosimetry. Overall, I felt that the article was very well written and
presented the data that was found in an informative manner for the reader. I feel that this article
comes as an important reminder for current medical dosimetrists to always include the treatment
couch for IMRT and RapidArc plans.
References:
1. Lenards N, Weege M. Readings in Radiation Therapy & Medical Dosimetry.
[Powerpoint]. La Crosse, WI: UW-L Medical Dosimetry Program; 2016.
2. Aldosary G, Nobah A, Al-Zorkani F, et al. A practical method of modeling a treatment
couch using cone-beam computed tomography for intensity-modulated radiation therapy
and RapidArc treatment delivery. Medical Dosimetry. 2015; 40(4):304-313.
http://dx.doi.org/10.1016/j.meddos.2015.03.008. Published April 29 2015. Accessed
February 27, 2016.