Professional Documents
Culture Documents
Kyoritsu 2009 Full Line Test Equipment Catalogue
Kyoritsu 2009 Full Line Test Equipment Catalogue
Catalogue 2009
Japan
Italy
U.K.
U.S.A.
China
Thailand
Headquarters
Affiliates
Please read the "Safety Warnings" in the instruction manual supplied with the instrument thoroughly and completely
Safety Warnings : for correct use. Failure to follow the safety rules can cause fire, trouble, electrical shock, etc. Therefore, make sure
to operate the instrument on a correct power supply and voltage rating marked on each instrument.
In consideration of the environment, soy ink and recycled paper were used in this publication. The contents of this leaflet are subject to change without notice. GeneCata09-1E Sep. 08 AD
Greetings
The electrical instrumentation industry has been playing a key role as a basis for
technological development to sustain the growth of the world.
With the passage of time there is an ever increasing need for us to meet the more
diversified and sophisticated demand from all walks of our industrial society.
Since its foundation in 1940 Kyoritsu Electrical Instruments Works, Ltd. has been pursuing
a solid business philosophy of "Customer First" in every area of our production and
servicing activities, thus establishing a relationship of mutual trust with our customers.
Here in our company all of us are expected to fully develop our individual personality
and creativity into a valuable corporate asset for our future growth.
We are united together by the guiding principle of always being diligent, honest and
appreciative of all those around us.
We sincerely wish that with this corporate culture we can make some contribution
towards building a more affluent human society.
Thank you in advance for all your continuous support and patronage.
ISO9001
QUALITY STANDARDS APPROVAL
(Headquarters・Ehime Factory)
Ehime Factory of Kyoritsu Electrical Instruments Works, Ltd. obtained ISO 9001 certification in
December, 1992.
The certification was given by Bureau Veritas which is one of the globally recognized
certification bodies.
C O N T E N T S
ANALOGUE MULTIMETERS 1106, 1109, 1110...................................................... 6
1009, 1011, 1012...................................................... 7
1018, 1018H, 1030................................................... 8
DIGITAL MULTIMETERS 1061, 1062................................................................ 9
2000, 2001..............................................................11
VOLTAGE TESTER 1700,1710............................................................... 10
FORK CURRENT TESTERS 2300R......................................................................11
ANALOGUE CLAMP METERS 2608A, 2805.......................................................... 12
2002PA, 2002R, 2006, 2007A.............................. 13
2017, 2027, 2031.................................................... 14
2040........................................................................ 15
DIGITAL CLAMP METERS 2003A, 2004, 2009A.............................................16
2010, 2033, 2037...................................................17
2046R, 2055, 2056R............................................. 18
2412, 2413F, 2413R................................................ 19
LEAKAGE CLAMP METERS 2417, 2431, 2432................................................... 20
2433, 2433R, 2434................................................ 21
MULTI-TRANS, CLAMP ADAPTOR 8004, 8008, 8006.................................................. 24
CLAMP ADAPTORS 8112, 8112BNC, 8113............................................. 25
3001B, 3005A, 3007A........................................... 26
DIGITAL INSULATION/CONTINUITY TESTERS 3021, 3022, 3023.................................................. 27
3111V, 3131A.......................................................... 28
ANALOGUE INSULATION/CONTINUITY TESTERS 3132A...................................................................... 29
3144A, 3145A, 3146A, 3161A.............................. 29
ANALOGUE INSULATION TESTERS 3165, 3166, 3315, 3316......................................... 30
3321A, 3322A, 3323A...........................................31
3121, 3122, 3123, 3124......................................... 33
HIGH VOLTAGE INSULATION TESTERS 3125........................................................................ 34
3128........................................................................ 35
4102A..................................................................... 37
EARTH TESTERS 4105A..................................................................... 38
4106........................................................................ 39
EARTH CLAMP TESTER 4200........................................................................ 40
4116A, 4118A......................................................... 41
LOOP/PSC TESTERS 4120A...................................................................... 42
5402D, 5406A........................................................ 43
RCD TESTERS 5410........................................................................ 44
5000, 5001............................................................. 45
LOGGERS 5010, 5020............................................................. 47
6010A..................................................................... 50
6010B...................................................................... 51
6011A...................................................................... 52
MULTI FUNCTION TESTERS 6016........................................................................ 53
6017, 6018.............................................................. 55
6020, 6030............................................................. 56
6050........................................................................ 57
PORTABLE APPLIANCE TESTERS 6200, 6201A, 6202............................................... 58
6300........................................................................ 59
POWER METERS 6310.........................................................................61
5201, 5202............................................................. 65
5500, 5510............................................................. 66
OTHERS 8030, 8031,8131F.................................................. 67
Demonstration Panels............................................... 68
8124, 8125, 8126, 8127, 8128, 8129..................... 69
SENSORS 8121, 8122, 8123, 8146, 8147, 8148..................... 70
8141, 8142, 8143, 8309, 8325F............................ 71
ACCESSORIES Test Leads................................................................. 73
Instrument Glossary................................................. 78
Kyoritsu Quality Control Concept............................ 80
PRODUCT INDEX..................................................... 81
Safety Warnings
Please read the “Safety Warnings” in the instruction manual supplied with the instrument thoroughly and completely for
safety use. Failure to follow the safety rules can cause fire, trouble, electrical shock, etc. Therefore, make sure to operate the
instrument on a correct power supply and voltage rating marked on each instrument.
K YO R I T S U L I N E U P
ANALOGUE MULTIMETERS DIGITAL MULTIMETERS
MODEL 1106 MODEL 1109 MODEL 1110 MODEL 1009 KEW 1011/1012 KEW 1018/1018H KEW 1030 KEW 1061/1062
P6 P6 P6 P7 P7 P8 P8 P9
UÊ £ä£Ó
MODEL 2006 MODEL 2007A MODEL 2017 MODEL 2027 MODEL 2031 KEW 2040
P13 P13 P14 P14 P14 P15
φ33 MAX
AC 600A
φ19 MAX
AC 200A AC V DC V
AC V DC V φ33 MAX
AC 600A
φ33 MAX
AC 600A
φ33 MAX
AC 600A Hz
RMS
Continuity AC V Continuity AC V Continuity AC V Continuity φ24 MAX
AC 200A Continuity
#!4Ⅳ
6
φ33 UÊ ÓäxÈ,
-!8
AC V c
UÊ ÓäxÈ, φ40 Ó{£Î,
UÊ
!#$# !
RMS
DC V φ40 MAX
AC/DC 1000A AC V MAX
AC 500A
Resolution
0.01mA
φ68
Hz °C DC V Hz AC V MAX
AC1000A
Resolution
0.1mA
φ40 MAX
AC 500A
φ24 MAX
AC 200A
Continuity
RMS
#!4Ⅳ OUT External OUT PEAK Resolution Resolution
#!4Ⅳ
Continuity 6 PUT Filter Power Supply PUT 10/100ms Filter 0.1mA Filter WP 0.01mA Filter
6
3021(125V/250V/500V/
1000V)3022(50V/100V/250V/ 3144A(250V/500V) 3146A(50V/125V) 3165(500V)
500V)3023(100V/250V/500V/1000V) 3111V(250V/500V/1000V) 3131A(250V/500V/1000V) 3132A(250V/500V/1000V) 3145A(125V/250V) 3161A(15V/500V) 3166(1000V)
RMS
VOLTAGE TESTERS DIGITAL MULTIMETERS FORK CURRENT TESTER ANALOGUE CLAMP METERS DIGITAL CLAMP METERS
KEW 1700 MODEL 2000 φ6 MAX
AC/DC 60A MODEL 2300R MODEL 2608A MODEL 2805 MODEL 2002PA
KEW 1710 MODEL 2001 φ10 MAX
AC/DC 100A MODEL 2002R
P11 P11 P12 P12 P13
2002R
P10
•
DC V MAX
AC 300A °C φ35 MAX
AC 600A AC V DC V
2433R
•
φ40 MAX
AC 100A
φ40 MAX
AC 400A
φ28 MAX
AC 100A
RMS
KEW ANALOGUE MULTIMETERS
MODEL 1106
MODEL 1106 DC V
AC V
0.5/5/25/100/250/500V(5kΩ/V)
±3% of FS
10/50/250/500V(2.5kΩ/V)
±3% of FS
DC A 200µA/2.5/25/250mA
±3% of FS
±5% of FS(250mA)
Ω 3/30/300kΩ
±3% of scale length
Battery test 2V(20Ω load)
Temperature -20°C~+150°C
±5°C(0°C~+100°C) ±10°C(other ranges)
(with the use of Temperature probe 7060)
Withstand voltage 3000V AC for 1 minute
Power source R6P(AA)(1.5V) × 1
Dimensions 130(L) × 85(W) × 38(D)mm
Weight 175g approx.
• Fuse and diode protect circuit.
Accessories 7066(Test leads) 8901(Fuse[0.5A/250V]) × 1
• Safety designed input terminals and
R6P(AA) × 1 Instruction manual
test leads.
Optional 9059(Carrying case)
7060(Temperature probe)
1109
MODEL 1109
DC V 0.1/0.5/2.5/10/50/250/1000V
MODEL ±3% of FS (20kΩ/V)
AC V 10/50/250/1000V(9kΩ/V)
±3% of FS
DC A 50µA/2.5/25/250mA
±3% of FS
AC A 15A
±3% of FS
Ω 2/20kΩ/2/20MΩ
±3% of scale length
Decibel -10~+62dB
hFE 0~1000(Ω× 10)
• Mirrored scale for easy and ac- ±3% of scale length
curate reading. Withstand voltage 6000V AC for 1 minute
• Fuse and diode protect circuit Power source R6P(AA)(1.5V) × 2 6F22(9V) × 1
(except 15A AC range). Dimensions 150(L) × 100(W) × 47(D)mm
• Output terminal to cut off DC Weight 330g approx.
MODEL
1110
MODEL 1110
DC V 0.3V(16.7kΩ/V) ±3% of FS
MODEL 3/12/30/120/300/600V(20kΩ/V)
±3% of FS
AC V 12V(9kΩ/V) ±4% of FS
30/120/300/600V(9kΩ/V)
±3% of FS
MODEL
DC A 60µA/30/300mA
±3% of FS
Ω 3/30/300kΩ
1109
• 1m drop-proof heavy duty designed taut-band movement. standard CAT.2 600V Pollution degree 2
• Can measure line voltage up to AC 600V. IEC 61010-031, IEC 61326-1
(Voltage to ground MAX AC300V) (Protected by 600V ceramic fuse Power source R6P(AA)(1.5V) × 2
against accidental overload) Dimensions 94(L) × 140(W) × 39(D)mm
• Continuity buzzer, battery check, LED check, temperature measurement Weight 280g approx.
function. Accessories 7066(Test leads) 8923(Fuse[F500mA/600V]) × 2
• Skeleton type robust and clear case with carrying handle furnished as R6P(AA) × 2 9103(Carrying case) Instruction manual
standard accessory. Optional 7060(Temperature probe)
6 FS : full scale
KEW DIGITAL MULTIMETERS
1009
MODEL 1009
KEW 1011/1012
KEW 1012 KEW 1011
£ä£Ó
UÊ
DC V 600.0mV/6.000/60.00/600.0/600V
(Input impedance :10MΩ, 100MΩ only 600mV)
±0.5%±2dgt(600.0mV/6.000/60.00/600.0V) ±0.8%±3dgt(600V)
AC V 6.000/60.00/600.0/600V 6.000/60.00/600.0/600V
(Input impedance :10MΩ) (Input impedance :10MΩ)
±1.5%±5dgt(6.000V) ±1.0%±3dgt(6.000/60.00/600.0V)
±1.2%±3dgt(60.00/600.0V) ±1.5%±3dgt(600V)
±1.5%±5dgt(600V)
DC A 600/6000μA/60/600mA/6/10A
±1.2%±3dgt(600/6000μA/60/600mA) ±2.0%±5dgt(6/10A)
AC A 600/6000μA/60/600mA/6/10A
1009
±1.5%±4dgt(600/6000μA/60/600mA) ±2.2%±5dgt(6/10A)
Ω 600Ω/6/60/600kΩ/6/60MΩ
±1.0%±2dgt(600Ω/6/60/600kΩ/6MΩ) ±2.0%±3dgt(60MΩ)
Continuity buzzer 0~600Ω(Buzzer sounds below 100Ω)
MODEL
Diode test 2.8V release voltage : Approx. 0.4mA test current
photo : 1012 Capacitance test 40/400nF/4/40/400/4000μF
Frequency 10/100/1000Hz/10/100/1000kHz/10MHz
(Input sensitivity
Current:more than 800μA[~10kHz] Voltage:more than 1.5V[~10kHz]
•6040 counts with Bar Graph display Hz:more than 0.4V[~10kHz])
•MIN/MAX function enables to record min & max value DUTY 0.1~99.9%(Pulse width/Pulse period) ±2.0%±2dgt(~10kHz)
•REL(relative value) function to indicate the measurement variation Temperature -50~300°C(-58~572°F)
Saving the initial value at the start of measurement as a reference value(= zero) (with the use of Temperature
1011/1012
The difference between the later measured values and the reference value is indi- probe 8216)
cated on the display Withstand voltage AC3700V / 1min.
• Temperature measurement, selectable for °C and °F (KEW 1011) Applicable IEC 61010-1 CAT.3 300V Pollution degree 2
supplied with K-type temperature probe (8216): -50~300 °C (-58~572°F) standards CAT.2 600V Pollution degree 2
• True RMS can measure and indicate distorted waveforms (KEW 1012) IEC 61010-031 IEC 61326
•DUTY function Power source R6P(1.5V) × 2 (Auto power off : approx. 15 minutes)
(It is possible to measure Pulse width / Pulse period) Dimensions 161(L) × 82(W) × 50(D)mm
•Data Hold function, Auto Power Off function
KEW
Accessories
7
KEW DIGITAL MULTIMETERS
KEW 1018 (Soft case type)/1018H (Hard case type)
KEW 1018/1018H
KEW 1030
KEW
KEW 1030
1018/1018H
Frequency 5/50/500/5k/50k/200kHz
• Double moulding provides comfortable and good feeling in hand ±0.1%rdg±5dgt
1030
• Penlight illuminates brightly the point to be measured, even in dark place (Input sensitivity
• Backlight LCD is highly visible, even in darkness Voltage:more than 1.5V[~50kHz] Voltage:more than 10V[>200kHz])
• Auto power-off function saves the battery life Duty 0.1~99.9% ±2.5%rdg±5dgt (Pulse width / Pulse cycle)
• Unique wrapping mechanism for test lead in the rear side compartment Data hold The measured value can be hold by pressing Data hold button
• Unique protection mechanism for the test prod for safety Battery voltage warning When the battery voltage drops to 2.4V±0.2V or less
• All ranges including Ohm range are protected against overload voltage of 600V Applicable standards IEC 61010-1 CAT.3 600V
• Designed in confor mity wit h Int er national safety s t andards IEC 61010-031 IEC 61326-1(EMC)
IEC61010-1:2001 CAT.3 600V Power source Button type battery LR44(SR44)(1.5V) × 2 (Auto power off : approx. 30 minutes)
Dimensions 190(L) × 39(W) × 31(D)mm
Weight Approx. 100g (including batteries)
Accessories 9130(Carrying case) LR44(1.5V) × 2 Instruction manual
8
KEW DIGITAL MULTIMETERS
KEW 1061 (Professional model)/1062 (Top of the range Professional model)
High Accuracy, High Performance and Reliable Measurements.
• 0.02% basic DC accuracy
• Large display with 50,000 counts
• Dual display for double indication for AC and DC, V and Hz, etc.
RMS • True-RMS Measurements AC and AC+DC
• Wide AC Frequency bandwidth from 10Hz~100kHz.*
• True-RMS or MEAN value detection mode can be selected.*
• Fast Peak Hold response time of 250μs.*
• Low-pass filter for motor drive measurements.*
• Low Power W for measuring the resistance of parts on PCB under low
measurement current.*
• User calibration function.
Safety design for industrial use.
• Complies with IEC 61010-1 CAT.31000V, CAT.4 600V
• Terminal shutter to prevent incorrect test leads’ insertion in current terminals.
• Very wide operating temperature range from -20~+55°C
Reliable support for data management.
• Large data logging memory
• KEW 1062 : 10,000 data
• KEW 1061 : 1,000 data
• Download data and Live Monitoring on a PC via the USB interface.
photo : 1062
(Option for USB Communication set) *Only for KEW 1062
1061
500.00/5000.0μA/50/500mA/5/10A
DC A [MEAN]* -
±1.5%rdg±20dgt (Basic accuracy)
500.00/5000.0μA/50.000/500.00mA/5.0000/10.000A
DCA+ACA
±1.5%rdg±10dgt (Basic accuracy) ±1%rdg±10dgt (Basic accuracy)
KEW
500.00W/5.0000/50.000/500.00kW/5.0000/50.000MW
W
±0.1%rdg±2dgt (Basic accuracy) ±0.05%rdg±2dgt (Basic accuracy)
5.000/50.00/500.0kW/5.000MW
LowPower-W* -
±0.2%rdg±3dgt (Basic accuracy)
Continuity buzzer 500.0W (The buzzer turns on for resistances lower than 100±50W.)
Diode test 2.4000V ±1%rdg±2dgt Open-loop voltage:<1.5V(Approx. 0.5mA Measuring Current)
5.000/50.00/500.0nF/5.000/50.00/500.0μF/5.000/50.00mF
Capacitance
±1%rdg±5dgt
2.000~9.999/9.00~99.99/90.0~999.9/0.900~9.999/9.00~99.99Hz
Frequency
±0.02%rdg±1dgt (Basic accuracy) 1062
DUTY 10~90% ±1%rdg
Temperature -200~1372°C ±1%rdg±1.5°C(with the use of K-type Temperature probe)
DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel (dBV, dBm),
Measurement functions
Continuity Check, Diode Test, LowPower-W*
KEW
Data Hold (D•H), Auto Hold (A•H), Peak Hold* (P•H), Range Hold (R•H), Maximum value (MAX), Minimum value (MIN), Average value (AVG),
Other functions
Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight
Withstand voltage 6,880V/5 sec.
Application standards IEC 61010-1 CAT.4 600V, CAT.3 1000V Pollution degree 2, IEC 61010-031, IEC 61326-1(EMC)
Power source R6(1.5V)×4 (Auto power off: approx. 20 minutes)
Dimensions/Weight 192(L)×90(W)×49(D)mm / Approx. 560g (including batteries)
Accessories 7220(Test leads), R6×4, Instruction manual, 8926(Fuse[440mA/1000V])×1 (included), 8927(Fuse[10A/1000V])×1 (included)
*Only for KEW 1062
Optional
Description MODEL Specification Description MODEL Specification
USB Communication set 8241 USB adaptor+USB cable+DMM Software Max. 500°C (Surface type, Point material:
8405
DMM Printer full set 8249 8243+8246+8248 Ceramic)
Printer Communication set 8243 Printer Adapter +RS232 cable Temperature probes 8406 Max. 500°C (Surface type)
Printer 8246 Printer + Thermal paper(1 roll) 8407 Max. 700°C (Liquid, Semi-solid)
AC adapter for printer[EU] 8248 AC230V±10% 8408 Max. 600°C (Air, Gas)
Thermal paper for printer 8247 10 rolls For the main unit with test leads and
Carrying case 9150
communication cable
9
KEW VOLTAGE TESTERS
1700/1710
KEW 1700/1710
Voltage test
KEW Voltage range 12~690V AC/DC
LED
Nominal voltage 12/24/50/120/230/400/690V
Probe Protection AC(45~400Hz), DC (±)
Cover Tolerance Light on at more than :
(Threshold voltage) 7±3V(12V LED)
18±3V(24V LED)
37.5±4V(50V LED)
75%±5% of nominal voltage
(120/230/400/690V LED)
Response time < 0.5s at 100% of each nominal voltage
LCD(1710 only)
Range / Resolution 300V(7.0~299.9V)/ 0.1V
(Auto-range) 690V(270~759)/ 1V
Accuracy(23±5˚C) ±1.5V(7~100V)
±1%±5dgt(100~690V)
AC(45~400Hz), DC(±)
Overrange indication “OL”
Also available Responce time < 2s at 90% of each voltage
non LCD model KEW 1700. Peak current Is<3.5mA(at 690V)
Measurement duty 30s ON(operation time)
240s OFF(recovery time)
• Designed to meet international safety Internal battery consumption Approx. 33mA(battery 3V,measuring 690V AC)
standards IEC 61243-3 / IEC 61010-1 Battery life Approx. 2500 operations(30s ON/240s OFF duty)
Measurement Category (CAT.) 4 600V Single-pole phase test
• Self-Diagnostic test Voltage range 100~690V AC (45~100Hz)
• AC and DC voltage tests up to 690V 180~690V AC (100~400Hz)
with LEDs and LCD Phase rotation test
• Polarity indication System Three-phase 4-wire system
• Single-pole phase test 200~690V phase-to-phase
• Phase rotation test (100~400V earth-to-phase)
• Double molding gives comfortable grip AC 50/60Hz
• Continuity test Phase range 120±5 degree
• Auto-power ON / OFF Continuity test
• Pen light for illuminating measurement Detection range 0~400kΩ+50%
points Test current Approx. 1.5µ A (battery 3V, 0Ω)
• Variable distance in socket, 16.7mm or Internal battery consumption Approx. 30mA (battery 3V, 0Ω)
19.0mm Operating temperature -10∼55°C, max 85% RH (no condensation)
KEW
AC +DC -DC
Unscrew Tighten
16.7mm 19mm
/100A(MODEL 2001) with OPEN CLAMP SENSOR. Dimensions 128(L) × 87(W) × 21(D)mm 128(L) × 91(W) × 27(D)mm
• 3400 counts with bargraph display. Weight 210g approx. 220g approx.
2000/2001
• Pocket size and heavy duty design. Accessories R03(1.5V) × 2
• Sleep function to save battery consumption. Instruction manual
• Designed to international safety standard IEC61010-1 CAT.3 300V Optional 9107(Carrying case[Soft])
MODEL
2300R
•Set the DC current range to zero in one touch with the Zero Adjust function.
•Auto Power Off.
•Data Hold (AC A / DC A only).
•Designed to international safety standard IEC61010-1 CAT.3 300V.
KEW FORK 2300R can be used in crowded connection boxes, where cables are very
short, and space is too limited to clamp cables using with a traditional clamp meter.
11
KEW ANALOGUE CLAMP METERS
2608A
MODEL 2608A
AC A 6/15/60/150/300A ±3% of FS
MODEL AC V 150/300/600V ±3% of FS
DC V 60V ±3% of FS
Ω
φ33 MAX
AC 300A °C
1/10kΩ(25/250Ω mid-scale)
±2% of scale length
Temperature -20°C~+150°C(with the use of Temperature probe 7060)
AC V DC V ±5°C(0°C~+100°C) ±10°C(other ranges)
Conductor size φ33mm max.
Frequency response 50/60Hz
Withstand voltage 3700V AC for 1 minute
Applicable IEC 61010-1 CAT.3 300V Pollution degree 2
standards IEC 61010-031 IEC 61010-2-032
Power source R6P(AA)(1.5V) × 1
• DC voltage range is also available Dimensions 193(L) × 78(W) × 39(D)mm
especially for checking emergency Weight 275g approx.
battery operated power supply. Accessories 7066(Test leads) 8901(Fuse [0.5A/250V]) × 2
• Can measure temperature using op- 9097(Carrying case) R6P(AA) × 1
tional probe. Instruction manual
• Tear drop shaped transformer jaws Optional 7060(Temperature probe)
for ease of use. 8004/8008(Multi-tran)
2805
MODEL 2805
AC A 6/20/60/200/600A ±3% of FS
MODEL AC V 150/300/600V ±3% of FS
Ω 2kΩ(25Ω mid-scale)
φ35 MAX
AC 600A AC V
±3% of scale length
Conductor size φ35mm max.
Frequency response 50/60Hz
Withstand voltage 2200V AC for 1 minute
Power source R6P(AA)(1.5V) × 1
Dimensions 220(L) × 83(W) × 40(D)mm
• Range switch selects only one rotary Weight 390g approx.
scale at a time, making it easy to take Accessories 7067(Test leads) 8901(Fuse [0.5A/250V]) × 2
readings correctly. 9144(Carrying case) R6P(AA) × 1
• A long time seller with proven reputation Instruction manual
world wide for its easy-to-use functions. Optional 8004/8008(Multi-tran)
MODEL
Selection Guide
MODEL 2608A MODEL 2805
2608A
AC A 300A 600A
AC V 600V 600V
DC V 60V
Ω 10kΩ 2kΩ
Temperature
Clamp meters are a very convenient testing instrument that Using the clamp meter, however, we can measure current
permits current measurements on a live conductor without by simply clamping on a conductor as illustrated in Fig.2.
circuit interruption. When making current measurements with One of the advantages of this method is that we can even
2805
the ordinary multimeter, we need to cut wiring and connect measure a large current without shutting off the circuit being
the instrument to the circuit under test as shown in Fig.1. tested.
Current Current
Power
Power Supply Load
Supply Cut wiring and Load
connect multimeter
Clamp
to current circuit
Meter
under test.
2006
MODEL 2006
AC A 2/20/200A
MODEL ±2%rdg±5dgt[50Hz~1kHz](2/20A)
±2%rdg±5dgt[50/60Hz](200A)
2002PA/2002R
φ19 MAX
AC 200A AC V
±3%rdg10±dgt[40Hz~1kHz](200A)
AC V 2/20/200/500V
±1.5%rdg±2dgt[50/60Hz] ±1.5%rdg±5dgt[40Hz~1kHz]
DC V Continuity DC V 2/20/200/500V ±1%rdg±2dgt
Ω 2kΩ ±1.5%rdg±2dgt
Continuity buzzer buzzer sounds below 300~400Ω
Conductor size φ19mm max.
• Pocket size, handy clamp meter. Frequency response 40Hz~1kHz
• UL listed fire retardant plastic housing Withstand voltage 2200V AC for 1 minute
case. Power source R6P(AA)(1.5V) × 2 ※Continuous measuring time : approx. 300 hours
MODEL
• Multifunction instrument having voltage Dimensions 184(L) × 54(W) × 32(D)mm
and resistance measurement functions Weight 160g approx.
in addition to all the ranges incorpo- Accessories 7066(Test leads)
rated. 9066(Carrying case)
R6P(AA) × 2
Instruction manual
Optional 8004/8008(Multi-tran) 2006
2007A
MODEL 2007A
AC A 400/600A
MODEL
φ33 MAX
AC 600A AC V
±1.2%rdg±3dgt[50/60Hz] ±1.5%rdg±4dgt[40~400Hz]
Ω 400/4000Ω
±1.5%rdg±2dgt
Continuity Continuity buzzer buzzer sounds below 50±35Ω
Conductor size φ33mm max.
Frequency response 40Hz~400Hz
2007A
AC A 200/600A
MODEL ±1.5%rdg±4dgt[50/60Hz](200A)
±1%rdg±3dgt[50/60Hz](600A)
φ33 MAX
AC 600A AC V
±2%rdg±5dgt[45Hz~1kHz]
AC V 200/600V
±1%rdg±2dgt[50/60Hz]
Continuity ±1.5%rdg±4dgt[45Hz~1kHz]
Ω 200Ω
±1.2%rdg±2dgt
• Tear drop shaped jaws for ease of use in Continuity buzzer buzzer sounds below 30±20Ω
tight places and crowded cable areas. Conductor size φ33mm max.
• Designed to meet internationally proven Frequency response 45Hz~1kHz
safety standards such as IEC61010-1, Withstand voltage 5550V AC for 1 minute
UL3111 and VDE0411. Applicable IEC 61010-1 CAT.3 600V Pollution degree 2
• Three functions in one unit; AC current, standards IEC 61010-031
AC voltage and resistance. Power source 6F22(9V) × 1 ※Continuous measuring time : approx. 200 hours
• Resistance range provides audible con- Dimensions 208(L) × 91(W) × 40(D)mm
tinuity test. Weight 400g approx.
• Frequency response from 40Hz to 1kHz Accessories 7066(Test leads)
on AC current and voltage ranges. 9079(Carrying case)
6F22 × 1
Instruction manual
Optional 8004/8008(Multi-tran)
2027
MODEL 2027
AC A 200/600A(True RMS)
MODEL ±1.5%rdg±4dgt[50/60Hz] (CF<3)
±2%rdg±5dgt[40Hz~1kHz](Sine wave)
2031
MODEL 2031
AC A 20A
MODEL ±2%rdg±5dgt[50Hz~1kHz]
2027
200A
φ24 MAX
AC 200A
±2%rdg±5dgt[50/60Hz]
±3%rdg±10dgt[40Hz~1kHz]
Conductor size φ24mm max.
Frequency response 40Hz~1kHz
Withstand voltage 3700V AC for 1 minute
Applicable IEC 61010-1 CAT.3 300V
standard
Power source LR-44(1.5V) × 2
※Continuous measuring time : approx. 100 hours (Auto power off : approx. 10 minutes)
Dimensions 147(L) × 58.5(W) × 26(D)mm
MODEL
Instruction manual
Optional 8004/8008 (Multi-tran)
14
KEW DIGITAL CLAMP METERS
2040
KEW 2040
AC A 0~600.0A
KEW ±1.5%rdg±5dgt(50/60Hz)
±3.5%rdg±8dgt(40~400Hz)
φ33 MAX
AC 600A AC V
AC V 6/60/600V
Auto Ranging
±1.3%rdg±4dgt(50/60Hz)
DC V Hz ±3.0%rdg±5dgt40~400Hz)
DC V 600m/6/60/600V
#!4Ⅳ Auto Ranging
Continuity 6
±1.0%rdg±3dgt
Ω 600/6k/60k/600k/6M/60MΩ (Auto Ranging)
±1%rdg±5dgt(600~6M) / ±5%rdg±8dgt(60M)
Continuity buzzer Buzzer Sounds at 100Ω
Hz 10/100/1k/10kHz(Auto Ranging)
(Input sensitivity
Current:more than 50A[~1kHz] Voltage:more than 1V[~10kHz]
DUTY 0.1~99.9%
(Pulse width / Pulse cycle) ±2.5%rdg±5dgt
Conductor size φ33
Withstand voltage 6880V AC for 5 seconds
Applicable standards IEC 61010-1 CAT.4 600V IEC 61010-031
IEC 61010-2-032 IEC 61326
Power source R03 (1.5V)(AAA) × 2
※Continuous measuring time : approx. 30 hours (Auto sleep function : approx. 15 minutes)
Dimensions 243(L) × 77(W) × 36(D) mm
Weight 300g
Accessories 7066(Test leads) 9094(Carrying case)
R03 × 2
Instruction manual
Optional 8004/8008(Multi-tran)(AC only)
2040
How Do Clamp Meters Operate?
In general AC clamp meters operate on the principle of
KEW
current transformer(CT) used to pick up magnetic flux gen- l(A)
proportional to the product of bias current and magnetic Magnetic flux at this gap is proportional
to primary current under test l(A)
field on the output terminal when bias current is applied to
the input terminal.
15
KEW DIGITAL CLAMP METERS
2003A
KEW 2003A
AC A 400A/2000A(0~1000A)
KEW ±1.5%rdg±2dgt[50/60Hz] ±3%rdg±4dgt[40~500Hz]
±5%rdg±4dgt[500Hz~1kHz]
φ55 MAX
AC/DC 2000A AC V
2000A(1001~2000A)
±3%rdg±2dgt[50/60Hz]
DC A 400/2000A ±1.5%rdg±2dgt
DC V Continuity AC V 400/750V
±1.5%rdg±2dgt[50/60Hz] ±1.5%rdg±4dgt[40Hz~1kHz]
OUT MAX #!4Ⅳ DC V 400/1000V ±1%rdg±2dgt
PUT HOLD 6
Ω 400/4000Ω ±1.5%rdg±2dgt
Continuity buzzer buzzer sounds below 50±35Ω
Conductor size φ55mm max.
Frequency response 40Hz~1kHz
Output voltage 400/2000A DC400mV/200mV
Withstand voltage 6880V AC for 5 seconds
Applicable IEC 61010-1 CAT.4 600V, CAT.3 1000V
standards IEC 61010-031 IEC 61010-2-032
• Equipped to measure both AC and DC cur- Power source R6P(AA)(1.5V) ×2
※Continuous measuring time : approx. 100 hours (Auto sleep function : approx. 10 minutes)
rent with transformer jaws of large diameter.
• Can measure AC and DC currents up to Dimensions 250(L) × 105(W) × 49(D)mm
2000A. Weight 530g approx.
• Output terminal for connection to recorders. Accessories 7107(Test leads) 8201(Output plug)
• AC/DC voltage, resistance measurement 9094(Carrying case) R6P(AA) × 2 Instruction manual
and continuity functions also available. Optional 8008(Multi-tran)(AC only) 7014(Output cord)
2004
MODEL 2004
AC A 20/200A
MODEL ±1%rdg±2dgt[50/60Hz](20A)
±1.5%rdg±4dgt[40Hz~1kHz](20A)
φ19 MAX
AC/DC 200A AC V
±1.5%rdg±4dgt[50/60Hz](0~150A)
±2%rdg±5dgt[40Hz~1kHz](0~150A)
±3.5%rdg[40Hz~1kHz](>150A)
DC V DC A 20/200A ±1%rdg±2dgt(20A)
±1.5%rdg±2dgt(0~150A) ±3%rdg(>150A)
AC V 500V
• Smallest clamp meter capable of AC and ±1.5%rdg±2dgt[50/60Hz] ±2%rdg±4dgt[40Hz~1kHz]
DC current measurements. DC V 200V ±1%rdg±2dgt
• 20A range has a minimum resolution of Ω 200Ω ±1.5%rdg±2dgt
0.01A AC/DC. Conductor size φ19mm max.
• AC/DC voltage and resistance measure- Frequency response DC, 40Hz ~ 1kHz
KEW
2009A
MODEL 2009A
AC A 400A(0~400A)(True RMS)
MODEL ±1.5%rdg±3dgt[50/60Hz] ±3%rdg±4dgt[30Hz~1kHz]
MODEL
2000A(150~1700A)(True RMS)
φ55 MAX
AC/DC 2000A
±1.5%rdg±2dgt[50/60Hz] ±3%rdg±4dgt[30Hz~1kHz]
2000A(1701~2000A)(True RMS)
2004
±3.5%rdg±3dgt[50/60Hz]
AC V DC V DC A 400/2000A ±1.5%rdg±2dgt
RMS
AC V 40/400/750V(True RMS)
OUT
Hz Continuity PUT
±1.5%rdg±3dgt[50/60Hz] ±2%rdg±4dgt[30Hz~1kHz]
DC V 40/400/1000V ±1%rdg±2dgt
PEAK Ω 400/4000Ω ±1.5%rdg±2dgt
10ms Continuity buzzer buzzer sounds below 20Ω
Hz 10 ~ 4000Hz ±1.5%rdg±5dgt
(Input sensitivity Current:more than 40A Voltage:more than 10V)
AC A 200mA/2/20A
MODEL ±1%rdg±2dgt[50/60Hz](200mA)
±1.5%rdg±4dgt[40Hz~2kHz](200mA)
φ7.5 MAX
AC/DC 20A
OUT
PUT
±1%rdg±2dgt[50/60Hz](2A)
±2.5%rdg±5dgt[40Hz~2kHz](2/20A)
DC A 2/20A
External
Power Supply ±1%rdg±2dgt(2A) ±1.5%rdg±4dgt(20A)
Conductor size φ7.5mm max.
Frequency response DC 40Hz~2kHz
Output voltage 200mV DC for 200mA AC/DC and 2A/20A AC/DC full scale
Withstand voltage 750V AC for 1 minute
Power source 6LF22(9V Alkalline battery) × 1 or AC adaptor
※Continuous measuring time : approx. 20 hours (DC)/approx. 40 hours (AC)
2033
MODEL 2033
AC A 40/300A
MODEL ±1%rdg±4dgt[50/60Hz](0~40A)
±2.5%rdg±4dgt[20Hz~1kHz](0~40A)
φ24 MAX
AC/DC 300A
±1.5%rdg±4dgt[50/60Hz](20~200A)
±2.5%rdg±4dgt[20Hz~1kHz](20~200A)
±3.5%rdg[50/60Hz](200~300A)
±4%rdg[20Hz~1kHz](200~300A)
• Smallest clamp meter capable of AC and DC A 40/300A ±1%rdg±4dgt(0~±40A)
DC current measurements. ±1.5%rdg±4dgt(±20~±200A) ±3%rdg(±200~±300A)
• 300A auto ranging has minimum resolu- Conductor size φ24mm max.
tion of 0.01A AC/DC. Frequency response DC 20Hz~1kHz
• Auto-zero function to allow one touch Withstand voltage 3700V AC for 1 minute
zero adjustment. Applicable IEC 61010-1 CAT.3 300V
standards IEC 61010-2-032
Power source LR-44(1.5V) × 2
2010
※Continuous measuring time : approx. 10 hours (Auto sleep function : approx. 5 minutes)
MODEL
LR-44 × 2
Instruction manual
Optional 8004/8008(Multi-tran)(AC only)
2037
MODEL 2037
AC A 400/600A(True RMS)
MODEL ±1.5%rdg±5dgt[50/60Hz](CF<3)
±3.5%rdg±5dgt[40Hz~1kHz]
PEAK
DC200ms
DC A 400/1000A 2033
AC500ms Continuity ±1%rdg±5dgt
AC V 40/400/600V(True RMS)
φ33 MAX
AC 600A
MAX
DC1000A ±1.5%rdg±5dgt[50/60Hz](CF<3)
RMS ±3.5%rdg±5dgt[40Hz~1kHz]
MODEL
AC V DC V DC V 40/400/600V
±1%rdg±5dgt
Ω 400/4000Ω
Hz AVG LoHz
±1%rdg±5dgt
Continuity buzzer buzzer sounds below 20Ω
Hz 3000Hz ±1.5%rdg±5dgt
(Input sensitivity Current:more than 40A Voltage:more than 8V)
※Continuous measuring time : approx. 15 hours (Auto sleep function : approx. 30 minutes)
• True RMS reading instrument that permits
precise measurements of non-sinusoidal Dimensions 208(L) × 91(W) × 40(D)mm
waveform AC current and voltage. Ac- Weight 450g approx.
curate measurements are also possible at Accessories 7066(Test leads) 9079(Carrying case)
low frequencies from 10Hz upwards. 6F22 × 1
• Peak, average and frequency measure- Instruction manual
ment functions. Optional 8004/8008(Multi-tran)(AC only)
17
KEW DIGITAL CLAMP METERS
2046R
KEW 2046R
φ33 -!8
!#$# !
±1.5%rdg±4dgt(50/60Hz) ±3.5%rdg±5dgt40~400Hz)
DC V 600m/6/60/600V(Auto Ranging) ±1.0%rdg±3dgt
Ω 600/6k/60k/600k/6M/60MΩ(Auto Ranging)
AC V DC V ±1%rdg±5dgt(600~6M) / ±5%rdg±8dgt(60M)
RMS
Continuity buzzer Buzzer Sounds at 100Ω
Continuity Hz Hz 10/100/1k/10kHz(Auto Ranging)
(Input sensitivity Current:more than 50A[~1kHz] Voltage:more than 1V[~10kHz])
DUTY 0.1~99.9%
#!4Ⅳ
°C 6
±2.5%rdg ±5dgt (Pulse width/Pulse cycle)
Capacitance test 400n/4µ/40µF(Auto Ranging)
Temperature -50°C~+300°C(with the use of Temperature probe 8216)
Conductor size φ33
Withstand voltage 6880V AC for 5 seconds
Applicable standards IEC 61010-1 CAT.4 600V
IEC 61010-031, IEC 61010-2-032, IEC 61326
Power source R03 (1.5V)(AAA) × 2
※Continuous measuring time : approx. 10 hours (Auto power off : approx. 15 minutes)
2055/2056R
KEW 2055 KEW 2056R
AC A 0~600.0/1000A 0~600.0/1000A
KEW ±1.5%rdg±5dgt(50/60Hz) ±2.0%rdg±5dgt(50/60Hz)
±3.0%rdg±5dgt(40~400Hz) ±3.5%rdg±5dgt(40~500Hz)
φ40 MAX
AC/DC 1000A AC V
DC A 0~600.0/1000A ±1.5%rdg±5dgt
AC V 6/60/600V(Auto Ranging) 6/60/600V(Auto Ranging)
±1.3%rdg±4dgt(50/60Hz) ±1.5%rdg±4dgt(50/60Hz)
DC V Continuity ±3.0%rdg±5dgt40~400Hz) ±3.5%rdg±5dgt40~400Hz)
DC V 600m/6/60/600V(Auto Ranging) ±1.0%rdg±3dgt
#!4Ⅳ
Hz 6
Ω 600/6k/60k/600k/6M/60MΩ (Auto Ranging)
±1%rdg±5dgt(600~6M) / ±5%rdg±8dgt(60M)
Continuity buzzer Buzzer Sounds at 100Ω
ÓäxÈ,
UÊ
Capacitance test 400n/4µ/40µF(Auto Ranging)
Temperature -50°C~+300°C
UÊ ÓäxÈ, (with the use of Temperature probe 8216)
KEW
Hz 10/100/1k/10kHz(Auto Ranging)
c
UÊ ÓäxÈ, (Input sensitivity Current:more than 50A[~1kHz] Voltage:more than 1V[~10kHz])
2046R
• Red LED, as “Non Contact Voltage” function, gives warning to the user Weight 310g
on the presence of AC voltage. Accessories 7066(Test leads) 9094(Carrying case) R03 × 2
2055/2056R
AC A 20/200mA/2/20/200/500A ±1.5%rdg±5dgt(20/200mA/2A)
MODEL (50/60Hz) ±2%rdg±5dgt(20/200A) ±2.5%rdg±5dgt(500A)
AC A 20/200mA/2/20/200/500A
±1%rdg±3dgt[50/60Hz] ±5%rdg±5dgt[40~400Hz](20/200mA/2A)
φ40 MAX
AC 500A
Resolution
0.01mA
(WIDE)
±1.5%rdg±3dgt[50/60Hz] ±5%rdg±5dgt[40~400Hz](20/200A)
±2%rdg±3dgt[50/60Hz] ±5%rdg±5dgt[40~400Hz](500A)
OUT
AC V PUT AC V 600V ±2%rdg±5dgt[50/60Hz] ±5%rdg±5dgt[40~400Hz]
Ω 200Ω ±1.5%rdg±5dgt
External
Filter Power Supply
Conductor size φ40mm max.
Frequency response 40~400Hz
• Digital clamp meter with tear drop Effect of external stray10mA AC max.
shaped, medium size transformer magnetic field φ15mm 100A
jaws specially designed for leak- Output DC200mV against 2000 count
age current measurement. Withstand voltage 3700V AC for 1 minute
• Frequency filter switch to eliminate Power source 6F22(9V) × 1 or AC adaptor
※Continuous measuring time : approx. 100 hours (Auto power off : approx. 60 minutes)
the effect of harmonics.
• Output terminal for connection to Dimensions 209(L) × 96(W) × 45(D)mm
recorders and facility to operate Weight 450g approx.
from external power supply permit Accessories 7066(Test leads) 9072(Carrying case)
continuous leakage current moni- 8025(Output plug) 6F22 × 1 Instruction manual
toring. Optional 8004/8008(Multi-tran) 8022(AC adaptor)(110V)
8023(AC adaptor)(220V) 7014(Output cord)
These Multi-trans can not be used for leakage current measurement.
2412
Effect of external stray 10mA AC max.
photo : 2413R
magnetic field
φ15mm 100A
• Large tranceformer jaws of 68mm diameter makes it possible to clamp on all Output AC/DC 200mV against 2000 count
KEW
three or four wires (3 phases) together for leakage current measurement. Crest factor 3.0 or Less
• Frequency filter switch to eliminate the effect of the harmonics. Withstand voltage 3700V AC for 1 minute
• 2 way analogue output terminal Applicable standards IEC 61010-1 CAT.3 300V IEC 61010-2-032
• Measurement from mA up to AC 1000A Power source 6F22(9V) × 1 ※Continuous measuring time : approx. 35 hours (2413F)
※Continuous measuring time : approx. 60 hours (2413R)
• True RMS enables an accurate measurement for distorted waveforms.(2413R)
• Peak hold function Dimensions 250(L) × 130(W) × 50(D)mm
• Data hold function: Allows for easy reading in dimly light or hard-to-reach locations Weight 570g approx. 600g approx.
• Back light in the LCD (2413R) Accessories 9094(Carrying case) 6F22 × 1 Instruction manual
• International safety standard IEC61010-1 CAT. 3 300V Optional 7073(2WAY Output cord)
2413F
i>ÃÕÀiiÌÊLÞÊV>«}Ê
ÌÜÜ>ÞÊÜÀiÃÊÌ}iÌ
iÀ
of current flowing between the incoming wire and out- i>>}iÊVÕÀÀiÌÊ}£Ó®ÜÊLi
`ë>Þi`®
going wire and display the result. If there is no leakage
on the load side, the instrument display reads zero. If i>>}iÊ
ÕÀÀiÌÊ}
the leakage occurs on the load side, the leakage cur-
KEW
i>ÃÕÀiiÌÊÊi>ÀÌ
}ÊÜÀi
rent will flow back into the power supply through earth, i>>}iÊVÕÀÀiÌÊ}ÊÜÊLiÊ`ë>Þi`®
resulting in the difference of current flowing between
the two-way wires which will then be displayed on the
instrument as a value of leakage current. Fig. Method of leakage current measurement.
19
KEW LEAKAGE CLAMP METERS
MODEL 2417
MODEL 2417 AC A
(50/60Hz)
200/2000mA/20/200/500A(True RMS)
±1.5%rdg±6dgt(200/2000mA)
±2%rdg±6dgt(20/200A)
±2.5%rdg±6dgt(500A)
φ40 MAX
AC 500A AC A 200/2000mA/20/200/500A(True RMS)
(WIDE) ±1%rdg±4dgt[50/60Hz]
Resolution
0.1mA Filter WP ±3%rdg±4dgt[40Hz~1kHz](200/2000mA)
RMS ±1.5%rdg±4dgt[50/60Hz]
±3.5%rdg±4dgt[40Hz~1kHz](20/200A)
±2%rdg±4dgt[50/60Hz]
±4%rdg±4dgt[40Hz~1kHz](500A)
Conductor size φ40mm max.
Frequency response40~400Hz
• Water and dust proof construction. 10mA AC max.
Effect of external stray
The instrument is protected against magnetic field φ15mm 100A
water and dust. Withstand voltage 3700V AC for 1 minute
• True RMS for accurate measure- Applicable standards IEC 61010-1 CAT.3 300V IEC 61010-2-032
ment of non-sinusoidal waveform Power source 6F22(9V) × 1
※Continuous measuring time : approx. 50 hours (Auto power off : approx. 30 minutes)
current.
• Selectable frequency response of Dimensions 209(L) × 96(W) × 45(D)mm
50/60Hz only or up to 1kHz. Weight 450g approx.
• Automatically turns power off in Accessories 9079(Carrying case) 6F22 × 1
about 30 minutes to conserve bat- Instruction manual
tery life. Optional 8004/8008(Multi-tran)
These Multi-trans can not be used for leakage current measurement.
2431
MODEL 2431
AC A 20/200mA/200A
MODEL (50/60Hz) ±3%rdg±5dgt(20/200mA/100A)
±5%rdg±5dgt(200A)
20/200mA/200A
φ24 MAX
AC 200A
Resolution
0.01mA
AC A
±2%rdg±4dgt[50/60Hz](20/200mA/0~100A)
(WIDE)
±5%rdg±6dgt[40~400Hz](20/200mA/0~100A)
Filter ±5%rdg±4dgt[50/60Hz](100.1~200A)
Conductor size φ24mm max.
Frequency response40~400Hz
Effect of external10mA AC max.
stray magnetic field
• Frequency Selector Switch to φ15mm 100A
eliminate the effect of harmonics. Withstand voltage 3700V AC for 1 minute
MODEL
• Three AC current ranges 20mA/ Applicable standards IEC 61010-1 CAT.3 300V IEC 61010-2-032
200mA/200A. Power source LR-44(1.5V)× 2
※Continuous measuring time : approx. 15 hours (Auto power off : approx. 10 minutes)
• 20mA range with a minimum
2417
AC A 4/40mA/100A
MODEL (50/60Hz) ±1%rdg±5dgt(4/40mA)
±1%rdg±5dgt(0~80A)
2431
±5%rdg(80.1~100A)
φ40 MAX
AC 100A
Resolution
0.001mA 4/40mA/100A
AC A
(WIDE) ±1%rdg±5dgt[50/60Hz] ±2.5%rdg±10dgt[20Hz~1kHz](4/40mA)
PEAK
Filter 10ms ±1%rdg±5dgt[50/60Hz] ±2.5%rdg±10dgt[40Hz~1kHz](0~80A)
±5%rdg[50/60Hz] ±10%rdg[40Hz~1kHz](80.1~100A)
Maximum 600V AC/DC (between line/neutral)
circuit voltage 300V AC/DC (against earth)
Conductor size φ40mm max.
Frequency response 20Hz~1kHz(40Hz~1kHz:100A)
Effect of external 2mA AC approx. in proximity to a 15mm-dia
stray magnetic field conductor carrying 100A AC
MODEL
IEC 61010-2-032
Power source R03(DC1.5V) × 2
• Frequency Selector Switch to eliminate the effect of harmonics. ※Continuous measuring time : approx. 40 hours (Auto power off : approx. 10 minutes)
2434
MODEL 2434
2433/2433R
stray magnetic field φ15mm 100A
Withstand voltage 3700V AC for 1 minute
Applicable standards IEC 61010-1 CAT.3 300V IEC 61010-2-032
Power source R03(AAA) (1.5V) × 2
※Continuous measuring time : approx. 150 hours(Auto sleep function : approx. 10 minutes)
MODEL
These Multi-trans can not be used for leakage current measurement.
21
KEW LEAKAGE CLAMP METERS
The KEW leakage clamp meters way to use
Let’s go back to the customer’s problem. If the RCD trips, it current the fault will be found.
should be temporarily “bridged out”. The contractor now simply In the fig. below there is a practical example how to trace the
clamps the Leakage Clamp Meter around both phase and neu- fault measuring the leakage current.
tral conductors on the supply side of the RCD(for 3-phase sys-
tems, all three live conductors and the neutral conductor should
be enclosed). Normally, using this tracing system the fault will be found but sometimes
The instrument display will then directly read the leakage current the earth leakage current will not be caused exclusively by low insulation
resistance.
to earth in the installation with a high resolution. In fact could happen that performing an insulation test there is not a low
Suppose the display reads 43.5 mA, simply tracing the leakage value of insulation resistance even if the RCD still trips!
Meter 0.02 =The leakage clamp does not read the leakage current
Conductor
with bad
To PE insulation
RCD Temporarily
bridge out
43 0.
.5 02 43
.5
43 43
.5 .5
0.
02 0.
02
0.
02
PE
22
KEW CLAMP METERS
Selection Guide
Analogue
Digital Clamp Meter AC/DC Digital Clamp Meter
Clamp Meter
MODEL 2608A 2805 2002PA 2006 2007A 2017 2027 2031 2040 2003A 2004 2009A 2010
2002R
Conductor size φ33 φ35 φ55 φ19 φ33 φ33 φ33 φ24 φ33 φ55 φ19 φ55 φ7.5
(MAX mm)
ACA MAX 300A 600A 2000A 200A 600A 600A 600A 200A 600A 2000A 200A 2000A 20A
MIN 6A 6A 400A 2A 400A 200A 200A 20A 400A 20A 400A 200mA
DCA MAX 2000A 200A 2000A 20A
MIN 400A 20A 400A 2A
ACV MAX 600V 600V 750V 500V 750V 600V 600V 600V 750V 500V 750V
MIN 150V 150V 40V 2V 400V 200V 200V 6V 400V 40V
DCV MAX 60V 1000V 500V 600V 1000V 200V 1000V
MIN 40V 2V 600mV 400V 40V
NCV ●
Frequency 10Hz~10kHz 10~4000Hz
Ω 1k/10k 2k 400/4k/ 2k 400/4k 200 200 600/6k/ 400/4k 200 400/4k
40k/400k 60k/600k/
6M/60M
Continuity buzzer ● ● ● ● ● ● ● ●
Sample rate Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx.
2.5 2 2.5 2.5 2.5 2 3 2.5 3 3 3
counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec
Data hold ● ● ● ● ● ● ● ● ● ● ● ● ●
Peak hold ● ● MAX ●
Analogue output ● ● ● ●
(FS) 400mV 400mV 400mV 200mV
Frequency 50/60 50/60 40~1k 40~1k 40~400 45~1k 40~1k 40~1k 40~400 DC DC DC DC
response (Hz) 40~1k 40~1k 30~1k 40~2k
True RMS ●(2002R) ● ●
Frequency filter
Temperature ●
Water proof
Power source R6P × 1 R6P × 1 R6P × 2 R6P × 2 R03 × 2 6F22 × 1 6F22 × 1 LR44 × 2 R03 × 2 R6P × 2 R6P × 2 6F22 × 1 6F22 × 1
Weight 275g 390g 470g 160g 260g 400g 400g 100g 300g 530g 170g 540g 220g
Page 11 11 12 12 12 13 13 13 14 15 15 15 16
Fork
AC/DC Digital Clamp Meter Digital Leakage Clamp Meter Current
Tester
MODEL 2033 2037 2046R 2055 2412 2413F 2417 2431 2432 2433 2434 2300R
2056R 2413R 2433R
Conductor size φ24 φ33 φ33 φ40 φ40 φ68 φ40 φ24 φ40 φ40 φ28 φ10
(MAX mm)
ACA MAX 300A 600A 600A 1000A 500A 1000A 500A 200A 100A 400A 100A 100A
MIN 40A 400A 600A 20mA 200mA 200mA 20mA 4mA 40mA 400mA
DCA MAX 300A 1000A 600A 1000A 100A
MIN 40A 400A 600A
ACV MAX 600V 600V 600V 600V
MIN 40V 6V 6V
DCV MAX 600V 600V 600V
MIN 40V 600mV 600mV
NCV ● ● ●
Frequency 10~3000Hz 10Hz~10kHz 10Hz~10kHz
Ω 4k 600/6k/ 600/6k/ 200
60k/600k/ 60k/600k/
6M/60M 6M/60M
Continuity buzzer ● ● ●
Sample rate Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx. Approx.
2.5 3 3 3 3 3 3 2 2.5 2.5 2.5 2
counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec counts / sec
Data hold ● ● ● ● ● ● ● ● ● ● ● ●
Peak hold ● ● ●(2056R) ● ● ●
Analogue output ● ●
(FS) 200mV 200mV
Frequency DC DC DC DC 40~400 40~1k 40~1k 40~400 40~1k 40~1k 40~400 50/60
response (Hz) 20~1k 10~1k 40~400 40~400
True RMS ● ● ●(2056R) ●(2413R) ● ●(2433R) ●
Frequency filter ● ● ● ● ● ● ●
Temperature ● ●(2056R)
Water proof ●
Power source LR-44 × 2 6F22 × 1 R03 × 2 R03 × 2 6F22 × 1 6F22 × 1 6F22 × 1 LR-44 × 2 R03 × 2 R03 × 2 R03 × 2 R03 × 2
Weight 100g 450g 300g 310g 450g 570g 450g 120g 290g 270g 210g 110g
Page 16 16 17 17 18 18 19 19 19 20 20 9
23
KEW MULTI-TRANS, CLAMP ADAPTOR
Adaptor designed to increase the measuring capability of your clamp MODEL 8004 MODEL 8008
8004 MULTI-TRAN
3000A(30 seconds max.)
Conductor size φ60mm max. (50 × 75mm) φ100mm max.(100 × 150mm)
MODEL Frequency response 50Hz/60Hz
Safety standard IEC 61010-1 CAT.3300V
Pollution degree 2
φ60 MAX
AC1000A
Input to Output Ratio
10:1 Withstand voltage 3000V AC for 1 minute
Dimensions 240(L) × 85(W) × 30(D)mm 317(L) × 150(W) × 30(D)mm
45(L) × 40(W) × 10(D)mm Output coil 45(L) × 40(W) × 10(D)mm Output coil
Weight 490g approx. 750g approx.
Accessories 9022(Carrying case) 9056(Carrying case)
Applicable models 2002PA,2002R,2003A(AC only),2004(AC only),2006,
2007A(AC only),2009A(AC only),2608A,2805,2017,2027,
2037(AC only),2040(AC only),2046R(AC only),
2055(AC only),2056R(AC only),2412,2417,8113(AC only),
2031,2033(AC only),2431,2432,2433,2433R,2434
These Multi-trans can not be used for leakage current measurement.
How to Use
Clamp your instrument around the pick-up coil of the multi-
tran. Then, clamp the transformer jaws of the multi-tran on
MODEL 8008 MULTI-TRAN the conductor, as shown below. Multiply the reading of your
clamp meter by 10 to obtain the current value measured.
Pickup Coll
Clamp Meter
8006
MODEL
φ60 MAX
1500A
1A➞1mV
MODEL 8006
24
KEW CLAMP ADAPTORS
8112
MODEL 8112/8112BNC
MODEL AC CLAMP ADAPTOR Range Measuring ranges Output voltage Accuracy Frequency response
200mA AC 0~500mA AC1V/A ±1.5%rdg±0.2mA 50Hz~1kHz
AC 0~1000mA (1000mA/1V) ±3%rdg±0.4mA 40Hz~10kHz
±1%rdg±1mA 40Hz~1kHz
φ8 MAX
AC 120A
2A AC 0~20A AC100mV/A
(20A/2V) ±1.5%rdg±2mA 1k~10kHz
20A AC 0~20A AC10mV/A ±1%rdg±0.01A 40Hz~1kHz
200mA 2A 20A
1A➞1V 1A➞100mV 1A➞10mV AC 20~60A (120A/1.2V) ±2.5%rdg 50Hz~10kHz
AC 60~120A ±2.5%rdg 100Hz~10kHz
Conductor size φ8mm max.
Frequency response 30Hz~100kHz(-3dB)
Applicable standard IEC 61010-1 CAT. 2
100V Pollution degree 2(8112 Only)
Withstand voltage 500V AC for 1 minute
Dimensions 153(L) × 18(W) × 23(D)mm
Weight 100g approx.
Accessories 9057(Carrying case)
Instruction manual
MODEL 8112BNC
8112
Model 8112BNC is an AC clamp adaptor designed for use with oscillo-
scopes. Output cord has a BNC connector which enables direct observa-
MODEL 8113
tion of current waveform on oscilloscope. Refer to a diagram shown at left
Range Measuring ranges Output voltage Accuracy Frequency response
for application. Specifications are same as those for Model 8112.
MODEL
AC 2A AC5A AC100mV/A ±2%rdg±5mA DC~1kHz
(5A/500mV)
20A AC50A AC10mV/A ±1.5%rdg±0.04A DC~1kHz
(50A/500mV)
8113
200A AC150A ±2%rdg±0.5A DC~1kHz
AC/DC AC1mV/A (0~150A)
MODEL CLAMP ADAPTOR AC200A (200A/200mV) ±3.5%rdg DC~1kHz
(150~200A)
φ19 MAX DC 2A DC5A DC100mV/A ±1.5%rdg±5mA
AC/DC 200A (5A/500mV) 8112BNC
2A 20A 200A 20A DC50A DC10mV/A ±1%rdg±0.02A
1A➞100mV 1A➞10mV 1A➞1mV (50A/500mV) (0~50A)
200A DC150A ±1.5%rdg±0.5A
DC1mV/A (0~150A)
DC200A (200A/200mV) ±3%rdg
MODEL
(150~200A)
Conductor size φ19mm max.
Frequency response DC~1kHz
Withstand voltage 1000V AC for 1 minute
Power source R6P(AA)(1.5V) × 2
Dimensions 180(L) × 54(W) × 31(D)mm
Weight 170g approx.
Accessories 7061(Output cords)
8113
Model 8113 is a clamp-on AC/DC current/voltage converter that permits AC/ R6P(AA) × 2
DC current measurements from 1mA to 200A in conjunction with a voltmeter 9055(Carrying case)
having 0.1mV resolution. Instruction manual
MODEL
25
KEW DIGITAL INSULATION/CONTINUITY TESTERS
3001B
MODEL 3001B
Insulation resistance
MODEL Test voltage 500V/1000V
Measuring ranges 200MΩ(2/20/200MΩ) autoranging
Compact and lightweight digital (Mid-scale value)
Insulation/Continuity tester. Output voltage Rated test voltage +12%
on open circuit
Rated current 1~1.1mA DC approx.
Output short 1.5mA DC approx.
circuit current
Accuracy ±2%rdg±1dgt
Continuity test
Measuring ranges 200Ω(20/200Ω) autoranging
(Mid-scale value)
Output voltage 600mV DC approx.
on open circuit
3005A
Output short 6mA approx.
circuit current
MODEL Accuracy ±2%rdg±0.1Ω±1dgt
General
Withstand voltage 2200V AC for 1 minute
Power source R6P(AA)(1.5V) × 8
Dimensions 144(L) × 93(W) × 61(D)mm
Weight 460g approx.
Accessories 7025(Test leads)
9050(Carrying case)
R6P(AA) × 8
Instruction manual
MODEL 3005A/3007A
Insulation resistance
Test voltage 250V/500V/1000V
Measuring ranges 20MΩ/200MΩ/2000MΩ
(Mid-scale value)
3007A
Output voltage Rated test voltage+20%
on open circuit -0%
MODEL Nominal current 1mA DC min.
Output short 1.5 mA DC approx.
circuit current
Accuracy ±1.5%rdg±5dgt(20MΩ/200MΩ) ±10%rdg±3dgt(2000MΩ)
MODEL
Continuity test
Measuring ranges 20Ω/200Ω/2000Ω
(Mid-scale value)
3007A
• Displays the value of external AC voltage along with flashing symbol. Weight 990g approx.(3007A)
• Auto null function to automatically subtract the test lead resistance before 970g approx.(3005A)
displaying the real continuity resistance value. Accessories 7122A(Test leads) 9074(Cord case)
• Trac-Lok mode to conserve battery life on insulation and continuity tests 8923(Fuse[F500mA/600V] ) × 2
(Model 3007A only ). R6P(AA) × 8 9121(Shoulder strap)
• Live circuit warning beeper. Instruction manual
• Releasing the test button automatically discharges the charges stored in
the circuit under test.
• Backlight function to view the test results in dimly lit areas (Model 3007A only).
• 200mA continuity measuring current to IEC 61557.
MODEL
photo : 3021