SMART CHARGE II MANUAL - INDUSTRIAL LEAK DETECTION DIVISION

V. ALARM/ FAULT CONDITION DEFINITION

VENT FAULT: If the part pressure remains above the VENT LEVEL setpoint value,
in excess of the VENT WATCHDOG time, the part is rejected and a
VENT FAULT is displayed.

EVACUATION FAULT: After the EVACUATION LEVEL setpoint has been reached; if the
pressure level rises above the EVACUATION FAULT LEVEL after
the duration of the EVACUATION TIME, the part is rejected and an
EVACUATION FAULT is displayed.

EVAC. HIGH PRESSURE: If the pressure rises above 45 PSIG at any time, the part is rejected
and an EVAC HIGH PRESSURE fault is displayed.

HELIUM CHARGE FAULT: If the HELIUM CHARGE LEVEL setpoint is not met within the
HELIUM CHARGE WATCHDOG time, the part is rejected and a
HELIUM CHARGE FAULT is displayed.

LARGE LEAK DETECTED: If an increase in the pressure level of the part occurs, and this change
value exceeds the programmed REJECT LEVEL setpoint before the
LEAK TEST TIME is elapsed, the part is rejected and a LARGE
LEAK DETECTED fault prompt is displayed.

LEAK DETECTED FAULT: If an increase in the pressure level of the part occurs, and this
differential change value exceeds the programmed REJECT LEVEL
setpoint before the LEAK TEST TIME is elapsed, the part is rejected
and a LEAK DETECTED FAULT prompt is displayed.

LOW CHARGE FAULT: If a decrease in the pressure level of the part occurs, and this pressure
change value exceeds the programmed LOW He. CHARGE
DIFFERENTIAL setpoint before the completion of the SNIFF TEST
TIME, the part is rejected and a LOW CHARGE FAULT prompt is
displayed.

861-018 Rev. E 29