2 calibration ST 400 O2
The following describes how to calibrate the O2-detector:
Open the detector cover and connect an ampere-meter to the
test points MP1 (RED +) and MP2 (BLACK -). Go to the
zero-adjustment menu (13) and choose the O2-detector
(normally Detector no 2). Press F1. Connect the
nitrogen-gas.Press F1. Adjust the flow on flow meter FM2 to
20 l/h. Adjust POT3 (ZERO) on the detector to 4.0 mA.
Press F1 (ADJUST) in the system. Close the gas. Press F2
followed by F1. Go to the calibration menu (12) and choose the
O2-detector (normally Detector no 2). Adjust POT1 (SPAN) on
the detector to set the calibration level. Normally the system
should show 20,9 % O2 and the ampere-meter 17.4mA.