Basic Principles of HPLC: Office of Quality Assurance

You might also like

Download as ppt, pdf, or txt
Download as ppt, pdf, or txt
You are on page 1of 17

Basic Principles of HPLC

Martin R. Hackman
NJ- DEP
Office of Quality Assurance
H igh
Performance
L iquid
C hromatography
H igh
Pressure
L iquid
C hromatography
H igh
Priced
L iquid
C hromatography
HPLC Methods
• Parameter Group Method Compounds

• SDW05.23000’s EPA 555 Cl-PhenoxyAcids


• WPP05.06000’s EPA 605 Benzidines

• WPP05.13000’s EPA 610 PAHs


• SHW06.26000’s SW-846 8316 Acrylics

• SHW06.28000’s SW-846 8330’s Explosives


• SHW07.06000’s SW-846 8325 Benzidines and
N- Pesticides
Compounds

2,4,5-T

Benzidine

Fluorene
Compounds

TNT (2,4,6-Trinitrotoluene)

H2C=CH-CN Acrylonitrile

Carbaryl
Partitioning
• Separation is based on the analyte’s relative
solubility between two liquid phases

Mobile Phase Stationary Phase

Solvent Bonded Phase


HPLC - Modes
• Normal Phase.
- Polar stationary phase and non-polar
solvent.

• Reverse Phase.
- Non-polar stationary phase and a polar
solvent.
Common Reverse Phase Solvents
• Methanol
CH3OH
• Acetonitrile CH3CN

• Tetrahydrofuran

• Water H2O
Columns
• Solid Support - Backbone for bonded phases.
– Usually 10µ, 5µ or 3µ silica or polymeric particles.
• Bonded Phases - Functional groups firmly linked
(chemically bound) to the solid support.
– Extremely stable
– Reproducible
• Guard - Protects the analytical column:
– Particles
– Interferences
– Prolongs the life of the analytical column
• Analytical - Performs the separation.
Bonded Phases

• C-2 Ethyl Silyl -Si-CH2-CH3

•C-8 Octyl Silyl -Si-(CH2)7-CH3

• C-18 Octadecyl Silyl -Si-(CH2)17-CH3

•CN Cyanopropyl Silyl -Si-(CH2)3-CN


Instrumentation
Gradient
Controller

Pump Column
Detector
Injector
Mobile Phases
Detectors
• UV
– Single wavelength (filter) [610, 8330]
– Variable wavelength (monochromator) [8316, 8325]
– Multiple wavelengths (PDA) [555]
• Fluorescence [610]
• Electrochemical [605]
• Mass Spectrometric [8325]
Chromatograms

Restek® ULTRA C-18 and CN Columns (250mm x 4.6mm, 5µ),


Mobile Phase: (1:1 Methanol:Water), 1.5 mL/min.
Chromatograms

A Supelcosil LC-PAH Columns. B


Conditions: A: 150mm x 4.6mm, 5µ. Conditions: B: 50mm x 4.6mm, 3µ.
Flow Rate: 1.5 mL/min Flow Rate: 3.0 mL/min

You might also like