Acceptance criteria for various validation parameters

1) Validation Characteristics & Requirements Validation Parameter Specificity Linearity Accuracy Precision LOD LOQ Assay/active constituent range 2 100% Yes Yes Yes Yes No No Assay/Active constituent range 0.1 1% Yes Yes Yes Yes No Yes Contaminants/ Residues <0.1% Yes Yes Yes Yes Yes Yes

2) Precision Active ingredient content in sample 10.0 % 1.0 to 10.0% 0.1 to 1.0% < 0.1% RSD limit 2% 5% 10% 20%

3) Accuracy by Recovery Active ingredient content in sample 10.0 % 1.0 to 10.0% 0.1 to 1.0% < 0.1% Acceptable mean recovery 98 102% 90 110% 80 120% 75 125%

4) Linearity Parameter Assay Assay, Herbal matrix Content uniformity Contaminants/residues Linearity range 80 120% ( 5 points) 70 130% (7 points) 70 130% (5 points) Reporting level to 120% (5 points)

5) Possible Variations for carrying out robustness For HPLC analysis pH Mobile phase composition (change in organic phase ) Temperature Flow rate Wave length, 5nm For GC analysis Column (different supplier) Temperature Flow rate (gas)