SA-XT Firmware Release Notes

Version 1.2 20110603

30303 Aurora Road, Cleveland (Solon) Ohio 44139-2794 Phone: (440) 248-1200 Fax: (440) 248-5426 www.bird-technolgies.com

Part Number Filename File Size

7002A726-2 7002A726-2.saxt 589,312 bytes

Includes support for USB power sensors. The following sensors are supported: 5012, 5014, 5015, 5016, 5017, 5018, and 5019.

Prerequisites
This firmware update requires that the SA-XT has firmware version 1.1 20110517 or later installed. Firmware version information is displayed in the Utility screen. Displaying Firmware 1. Turn on the SA-XT. 2. Press the Mode button. 3. Press the Utility softkey. NOTE: The firmware version is displayed in the Utility screen. NOTE: A version of Firmware v1.1 17 May 2011 or later is acceptable for this release.

Updating Firmware
1. Turn the SA-XT off. 2. Attach the AC external power supply. 3. Connect the SA-XT to a PC via type A/B USB cable. 4. Hold down the power button for 6 seconds or more. NOTE: The SA-XT will enumerate as a Mass Storage Device. It will appear as a drive in Windows explorer with a drive label Removable Disk. The SA-XT LCD display will indicate that it is in firmware update mode. 5. Go to the PC. 6. Copy the firmware image file to the SA-XT drive. 7. Disconnect the SA-XT from the PC. NOTE: The SA-XT will now apply the firmware update. 8. Turn the SA-XT on. NOTE: The SA-XT should power off automatically after the update has been applied. If this does not happen, disconnect the external power supply. 9. Verify the firmware has been updated. See Displaying the Firmware above.

Please Refer to the Units Original Operator Manual for Additional Operation & Maintenance Information