Professional Documents
Culture Documents
A93a01239ben Magnesium Rtu
A93a01239ben Magnesium Rtu
2010/03/30 A93A01239BEN
A11A01646
2 x 25 mL
Intended Use: Diagnostic reagent for quantitative in vitro determination of Magnesium in serum and plasma by colorimetry.
Handling
1. Identify the cassette by using dedicated reagent stickers with barcode (605). 2. Transfer the reagent into compartment 1 (30 ml capacity) of the cassette 30/10 provided (see diagram below). The compartment 2 of the cassette will be unused.
Method (3)
Photometric test using xylidyl blue. Magnesium ions form a purple colored complex with xylidyl blue in alkaline solution. In presence of GEDTA, which complexes calcium ions, the reaction is specific. The intensity of the purple color is proportional to the magnesium concentration.
3. If present, remove foam by using a plastic pipette. 4. Place the reagent cassette at an available position on the reagent tray in the refrigerated Pentra C200.
Calibrator
For calibration, use: ABX Pentra Multical, Ref. A11A01652 (not included) 10 x 3 mL (lyophilisate)
Reagents
ABX Pentra Magnesium RTU is ready-to-use.
Control
For internal quality control, use: ABX Pentra N Control, Ref. A11A01653 (not included) 10 x 5 mL (lyophilisate) ABX Pentra P Control, Ref. A11A01654 (not included) 10 x 5 mL (lyophilisate) Each control should be assayed daily and/or after a calibration.
Reagent:
Ethanolamine pH 11.0 GEDTA (Glycoletherdiamine acid) Xylidyl blue Detergents
Form-0846 Rev.4
ABX Pentra Magnesium RTU should be used according to this reagent notice. The manufacturer cannot guarantee its performance if used otherwise.
S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B
Clinical Chemistry
Packaging spoiling
In case of protective packaging spoiling, do not use the reagent if the damages might have an effect on the product performance.
Waste Management
Please refer to local legal requirements.
General Precautionsb
This reagent is for professional in vitro diagnostic use only. S25: avoid contact with eyes. Observe the standard laboratory precautions for use. The reagent vials should be discarded after use. Please refer to the MSDS associated with the reagent. Do not use the product if there is visible evidence of biological, chemical or physical deterioration.
Specimena
Serum. Plasma in lithium heparin. Do not use EDTA plasma. Stability (4): 7 days at 20-25C 7 days at 4-8C 1 year at -20C
Assay Procedure
Test instructions for other automated systems than Pentra C 200 are available on request (not available in the USA).
a Modification b Modification
from index A to B: modification of specimen stability. from index A to B: addition of safety phrases.
S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B
Clinical Chemistry
Triglycerides:
No significant influence is observed up to 7 mmol/L (612.5 mg/dL) (as Intralipid, representative of lipemia). Total Bilirubin: No significant influence is observed up to 375 mol/L (21.9 mg/dL). Direct Bilirubin: No significant influence is observed up to 375 mol/L (21.9 mg/dL). Other limitations are given by Young as a list of drugs and preanalytical variables known to affect this methodology (12, 13). Calibration Stability: The reagent is calibrated on Day 0. The calibration stability is checked by testing 2 control specimens. The calibration stability is 19 days. Note: A recalibration is recommended when reagent lots change, and when quality control results fall outside the range established. Application releasec : 01.xx Conversion factor: mmol/L x 24.3 = mg/L mmol/L x 2.43 = mg/dL
Reproducibility (total precision) 3 specimens of low, medium and high levels and 2 controls are tested in duplicate for 20 days (2 series per day) according to the recommendations found in the CLSI (NCCLS), EP5-A2 protocol (8).
Mean value mmol/L Control specimen 1 Control specimen 2 Specimen 1 Specimen 2 Specimen 3 0.89 1.59 0.66 0.99 1.58 mg/dL 2.2 3.9 1.6 2.4 3.8 CV % 2.91 2.85 3.23 3.15 2.77
Warning
It is the user's responsibility to verify that this document is applicable to the reagent used.
Measuring Range: The assay confirmed a measuring range from 0.05 to 1.9 mmol/L (0.12 to 4.61 mg/dL), with an automatic postdilution up to 5.7 mmol/L (13.83 mg/dL). The reagent linearity has been assessed up to 1.9 mmol/L (4.61 mg/dL) according to the recommendations found in the CLSI (NCCLS), EP6-A protocol (9). Correlation: 93 patient samples (serum) are correlated with a commercial reagent taken as reference according to the recommendations found in the CLSI (NCCLS), EP9-A2 protocol (10). Values ranged from 0.09 to 1.80 mmol/L (0.22 to 4.37 mg/dL). The equation for the allometric line obtained using PassingBablock regression procedure (11) is: Y = 1.02 X - 0.01 (mmol/L) Y = 1.02 X - 0.04 (mg/dL) with a correlation coefficient r2 = 0.9683. Interferences: Haemoglobin: No significant influence is observed up to 100 mol/L (172 mg/dL).
c Modification
Reference
1. Thomas L. Clinical Laboratory Diagnostics. 1st ed. Frankfurt: THBooks Verlagsgesellschaft (1998): 231-241. Endres DB, Rude RK. Mineral and bone metabolism. In: Burtis CA, Ashwood ER, editors. Tietz Textbook of Clinical Chemistry. 3rd ed. Philadelphia: WB Saunders Company (1999): 1395-1457. Burcar PJ, Boyle AJ. Spectrophotometric determination of magnesium in blood Serum Using Magon. Clin. Chem. (1964) 10: 1028-1038. Guder WG, Zawta B et al. The Quality of Diagnostic Samples. 1st ed. Darmstadt: GIT Verlag (2001): 38-39, 50-51. Sitzmann FC. Normalwerte. Mnchen: Hans Marseille Verlag GmbH (1986): 166. Protocols for determination of limits of detection and limits of quantitation. Approved Guideline, CLSI (NCCLS) document EP17-A (2004) 24 (34) .
2.
3. 4.
5. 6.
S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B
Clinical Chemistry
11.
12.
13.
S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B