Professional Documents
Culture Documents
Instant Micro
Instant Micro
Goals of Presentation
Optical instrument (IMD) for instantaneous detection of microbes in the environment Intrinsic fluorescence from certain metabolites differentiates microbes from inert particles Process analytical tool to monitor and control an aseptic environment
RMM Defined
Rapid Microbiological Method (RMMs) instrument/procedure designed to obtain test results faster than the traditional culture-based approach
FDA encourages RMMs through the Process Analytic Testing and Quality by Design (QbD) initiatives for continuous, online monitoring RMMs have the potential to enable a paradigm shift from batch product release to parametric release
Spectroscopy
Chemical Byproduct
Intrinsic Fluorescence
Extrinsic Fluorescence
Extrinsic Fluorescence
Intrinsic Fluorescence
Intrinsic Raman
IMD (BioVigilant)
Bactometer
Impedance
Pallchek (Pall)
Advance (Celsis)
(BioMerieux)
Test Locations:
General Environment
Fill line Sterility isolator Raw materials Bioreactors
Media fills
Environmental Monitoring Investigational testing Process refinement/continuous improvement
Test Substrates:
Test Methods:
At a fixed location Mobile test Air Liquid Surface
IMD-A Technology
Viable Monitoring
Limitations of conventional air sampling
a periodic test operator can cause biological contamination labor intensive retrospective sample collection efficiency (impaction or impingement), selectivity of media VBNC discrepancies
IMD Monitoring
IMD techniques offer incremental benefits, :
Ability to detect VBNC microbes Ability to give immediate results
10
11
AIR INLET
FAN ASSEMBLY
12
13
Customers
Follow PAT guidance IQ/OQ/PQ Side by side testing with IMD-A and conventional methods (i.e., currently used air samplers)
Accuracy
Precision Linearity Limit of Detection Limit of Quantification Ruggedness Robustness Range Specificity
14
15
16
Summary Test Matrix for USP <1223> Test Fuctions Performed at AlburtyLab IMD-A-200-1 vs. Andersen 6-Stage Viable Air Sampler
T1 13% T2 Concentrations T3 20% T4 17% T5 17%
Legend:
Did not meet acceptance criteria Test Fuctions Accuracy Bacterial Species Wet Bacillus atrophaeus Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans Wet Bacillus atrophaeus Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans Wet Bacillus atrophaeus Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans IMD: Wet Bacillus atrophaeus And: Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans IMD: Wet Bacillus atrophaeus And: Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans Wet Bacillus atrophaeus Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans Range IMD: 5.217-11.304 Wet Bacillus atrophaeus And: 20.563-107.730 Dry Bacillus atrophaeus Escherichia coli Micrococcus lylae Staphylococcus epidermidis Corynebacterium afermentans
2
IMD-A-200-1 Passing Rate All # of tests # of passing 138 120 87% Exclude wet Bg 115 114 99%
pass z-test
Precision
RSD 37%
RSD 38%
% passed
Specificity 1
42%
6.140-8.643 39.245-76.010
73.913 576.276
One instance involves <60 sec sampling time for Andersen 6-stage
Linearity
17
APS
IMD-A-200-1
Andersen 6-Stage
1.0E+03 1.0E+02 1.0E+01 1.0E+00 1.0E-01 0 Titer 1 10 Titer 2 20 Titer 3 30 Titer 4 40 Titer 5 50
Runs
Both IMD-A-200-1 and Andersen trend well with APS particle counter Variability seen in all three instruments within concentration runs In general, 200-1 counts higher than Andersen over all runs
18
19
Example
Description of the location Personnel passing Criteria High Average Low Close Median Far High Average Low Short Median Long Rating point 3 2 1 3 2 1 3 2 1 3 2 1
Proximity to exterior
20
(3) Grid-pattern survey of a room (see graphs below): divide room into grid, use IMD to test each point; then generate a two-dimensional bio-burden map (lower left)
14 12 10 8 6 4 2 0 1 2 3 4 5 S1 6 7 S3
Door
S4
workstation
S3 S2
1 2 3 4 5 6
S1
Company X Fill Room IMD Biologic Counts 09/09/2008: two workstations at (2, S1) and (6, S1); entrance door at (7, S4)
21
Applications
Clean Room/Critical Area monitoring RABS monitoring (Restricted Access Barrier System) Aseptic isolator monitoring Bacterial challenge chambers (ingress testing vs. experimental testing) Investigation of excursions
22
Summary
Quality by design is a goal for pharmaceutical manufacturing
QbD requires control of the production process
23
Contact Information
Thank you for your attention
BIOVIGILANT : PMT FRANCE :
jengle@biovigilant.com jean-sebastien.bonin@pmtfrance.fr
http://www.biovigilant.com/
24