Abstract
There are various kinds of temperature sensors used in the chemical industries such as thermistors, RTDs, pyrometers and many more. All of them have a unique temperature range and sensitivity. Economic considerations are always very important in installing these sensors. Their accuracy, repeatability, reproducibility and ability to eliminate errors should also be analysed closely. The following assignment will give u an idea about thermocouples and its calibration mainly. As thermocouples is a very vast topic so I have included the important topics such as the three important laws of thermocouples which are Law of intermediate materials, Law of homogeneous material and Law of successive or intermediate temperatures. It also includes types of thermocouples such as refractory metal thermocouples, base metal thermocouples and noble metal thermocouples and their sub types too. Its applications and different temperature ranges for the given types. The pros and cons of using a thermocouple are also included. Then comes the calibration procedure of a thermocouple. Basically in a thermocouple we have a hot junction and a cold junction so ways to calibrate it are shown and illustrated by diagrams. Thermocouples have various applications in heating appliance safety, steel industries, thermopile radiation sensors, process plants manufacturing, Radioisotope thermoelectric generators and many more.
www.engineering-resource.com
Table of Contents
4 4
4 4 5
3. Types of Thermocouple
Base metal Thermocouples Noble Metal Thermocouples Refractory Metal Thermocouples
5
5 6 7
4. 5. 6. 7.
Facts About Thermocouple Advantages and Disadvantages Aging of Thermocouple Calibration procedure
Controlled Temperature Sources Reference Junction Reference Thermometers
8 8 9 10
10 11 13
8. Applications
14
www.engineering-resource.com
THERMOCOUPLE
A thermocouple is formed when two dissimilar metals are joined and a temperature difference occurs between the two ends. This temperature difference creates a current and is called the SeeBeck Effect. This was named after the scientist Thomas Seebeck who discovered this in 1821. A temperature difference between the two ends of the circuit, voltage is generated within the circuit. The voltage is measured in the 0.001 th of a volt. Higher the temperature difference creates higher voltage. Right pair of metals should be used to measure the temperature. The junction put into the process in which temperature is measured is known as the Hot Junction. The other junction which is always at some kind of measuring instrument is called the Cold Junction. For example if we want to measure the temperature inside a furnace; inside the furnace its 1000C and outside temperature is measured to be 100C. Thermocouple would thus indicate a difference of 900C. The problem with the thermocouple is if we want to know the temperature inside the furnace and not the temperature difference between inside and outside. For this we apply the Cold Junction Compensation. For this, we need to know the cold junction temperature and normally the measuring instrument normally does it. It measures the point at which the thermocouple attaches then adds the temperature back into the equation. The result is then displayed. Thermocouple material should be used throughout the circuit. Specially coded extension wire is normally used. If the extension wire is not used in the circuit an error may occur due to incorrect cold junction compensation. (1)
LAWS OF THERMOCOUPLES
Law of intermediate materials
Algebraic sum of the thermoelectric forces in a circuit comprising of any number of dissimilar metals is zero if all of the junctions are at a uniform temperature. If a third metal is inserted and if the two new junctions are at the same temperature, there will be no net voltage generated by the new metal.
www.engineering-resource.com
Type J
Type J thermocouples have an iron positive leg and a constantan negative leg. They have a temperature range of 32 to 1400F (0 to 750C) and can be used in vacuum, oxidizing, reducing and inert atmospheres. Due to the oxidation (rusting) problems associated with the iron leg, care must be used when choosing this type for use in oxidizing environments above 1000F. The color code for type J is white for positive and red for negative.
Type E
The type E thermocouple is composed of a positive leg of chromel (nickel/10% chromium) and a negative leg of constantan (nickel/45% copper). The temperature range is 330 to 1600F (-200 to 900C). The type E thermocouple has the highest millivolt output. Type E sensors can be used in sub-zero, oxidizing or inert applications but should not be used in sulfurous, vacuum or low oxygen atmospheres. The color code for type E is purple for positive and red for negative.
www.engineering-resource.com
Type K
The type K thermocouple has a Chromel positive leg and an Alumel (nickel/ 5% aluminum and silicon) negative leg. Temperature range for type K alloys is 328 to 2282F (-200 to 1250C). Type K sensors are recommended for use in oxidizing or inert environments. Type K should also not be used in sulfurous environments. Because type K has better oxidation resistance than types E, J and T, its main area of usage is at temperatures above 1000F but vacuum and low oxygen conditions should be avoided.
Type T
Type T thermocouples are made with a copper positive leg and a constantan negative leg. The temperature range for type T is 328 662F (-200 to 350C) and color code is blue for positive and red for negative. Type T sensors can be used in oxidizing (below 700F), reducing or inert applications.
Type N
Type N thermocouples are made with a Nicrosil (nickel 14% chromium 1.5 % silicon) positive leg and a Nisil (nickel 4.5% silicon - .1% magnesium) negative leg. The temperature range for Type N is 450 to 2372F (-270 to 1300C) and the color code is orange for positive and red for negative. Type N is very similar to Type K except that it is less susceptible to selective oxidation. Type N should not be used in vacuum and or reducing environments in an unsheathed design.
Type B
Type B thermocouples are made with a platinum/30% rhodium positive leg and a platinum/6% Rhodium negative leg. The temperature range for this type is 32 3092F (0 to 1700C) and the color code is gray for positive and red for negative.
Type R
Type R thermocouples are made with a platinum/13% rhodium positive leg
www.engineering-resource.com
7 and a pure platinum negative leg. The temperature range for type R is 32 2642F (0 to 1450C) and the color code is black for positive and red for negative.
Type S
Type S thermocouples are made with a platinum/10% rhodium positive leg and a pure platinum negative leg. The temperature range for type S is 32 2642F (0 to 1450C) and the color code is black for positive and red for negative.
Type C
Type C thermocouple is made with a tungsten/5% rhenium positive leg and tungsten 26% rhenium negative leg and has a temperature range of 32 4208F (0 2320C). The color code for this type is white with red for positive leg and red for the negative leg. (3)
ERROR LIMITS
Accuracy of temperature sensors is referred to as limits of error and apply only to brand new, temperature sensors. Once a sensor is exposed to elevated temperatures, there is no guaranteed accuracy.
www.engineering-resource.com
FACTS
According to ASTM color code guidelines, which apply to most North American sensor manufacturers, the Red leg is always negative.
The hot junction of a thermocouple requires a good, constant contact between the two wires. Material may be any.
Two types of thermocouples (types J and K) have one leg, which is magnetic. With these two types, you can use a magnet to determine polarity.
Non-thermocouple materials can be used in thermocouple circuits under the right conditions. Non-thermocouple connectors, terminals and slices can be used where there is no temperature gradient present at the areas where these items are used.
Special limits of error thermocouple sensors do not have to have special limits of error extension wire.
We can get an average temperature reading using multiple thermocouples as long as the sensors are wired in parallel and the resistance of different sensors is the same. Extension wire does not have to be a large gauge to work in an application where the sensor is placed a long way from the measuring instrument. Most modern temperature monitoring instruments are current based so lead wire resistance is not critical. (4)
www.engineering-resource.com
9 Thermocouples have the ability to be brought into direct contact with the material they are measuring. Since they use voltage readings to measure temperatures, the only requirement when they are brought in direct contact with the material is that the thermocouple is safely grounded. Thermocouples are made up of two dissimilar metals so they are vulnerable to corrosion, which can sometimes be difficult to detect. It can result in misreading by the thermocouple; therefore, proper care and maintenance of thermocouples is necessary. Thermocouples are complex and have many sources of error. This is mostly due to their operation. Thermocouples use two dissimilar metals that have two different voltage readings produced from a change in temperatures on alternative ends of the metals. They also have a cold junction, which consists of copper wires to allow the metal to meet the instrumentation. They use various types of electronic compensation at the cold junction to allow for more accurate readings; this account for marginal errors, depending upon the type of thermocouple used The correct calibration of thermocouples can be tedious and difficult. It is always essential thermocouple calibration be done next to another--already calibrated--thermocouple. During the calibration bath, the output is not reproduced exactly as it was, making correct calibration essential. The relationship between the process temperature and the thermocouple signal (millivolt) is not linear.
(5)
AGING OF THERMOCOUPLES
Thermocouples are often used at high temperatures and in reactive furnace atmospheres. The practical lifetime is determined by aging. The thermoelectric coefficients of the wires in the area of high temperature change with time and the measurement voltage drops. The simple relationship between the temperature difference of the joints and the measurement voltage is only correct if each wire is homogeneous, unlike the aged thermocouple. Relevant for the generation of the measurement voltage are the properties of the metals at a temperature gradient. If an aged thermocouple is pulled partly out of the furnace, the aged parts from the region previously at high temperature enter the area of temperature gradient and the measurement error is significantly increased. However an aged thermocouple that is pushed deeper into a surface gives a more accurate reading.
(6)
www.engineering-resource.com
10
Fixed Point: For highly accurate measurements, fixed point cells are used. It consists of a metal sample inside a graphite crucible with a graphite thermometer sufficiently submerged into the metal sample. When the metal sample is heated to the freezing point, it will produce a stable and constant temperature. For better comprehension of the operation of fixed point cells, the following terms are useful.
Freezing Point: The fixed point between the solid and liquid phases of a material.
REFERENCE JUNCTIONS:
A thermocouple's output is based on the difference in temperature between the measuring junction (hot junction) and the reference junction (cold junction). See Figure A.
www.engineering-resource.com
11
ICE BATH:
One of the most common reference junctions is the ice bath. The ice bath is made up of a mixture of melting shaved ice and water. The ice bath is a convenient and inexpensive way to achieve an ice point, it can be easily reproduced and with high accuracy. Junctions formed between the thermocouple materials and instrument leads can be simply immersed into the slush mixture, or alternatively glass "U" tubes containing a quantity of mercury approximately 3/4" to 1" depth can be placed into the slush mixture. Quick electrical connection can then be made between thermocouple and instrument leads through the mercury. Figure B
www.engineering-resource.com
12
Note: An improperly used ice bath can result in serious errors. The largest error arises due to melting of the ice at the bottom of the bath until the reference junctions are below the ice level and surrounded by water alone. This water may be as much as 7F above the ice point.
www.engineering-resource.com
13
ELECTRONIC COMPENSATION:
This method employs a compensation circuit containing a source of current and a combination of fixed resistors and a temperature sensitive resistor (TSR). This device can be designed to produce similar EMF to that of the thermocouple being calibrated. The Electronic Compensator will make EMF compensations to the thermocouple circuit based in the difference in EMF from 32 to ambient temperature.
MEASURING INSTRUMENTS:
The choice of a specific instrument for measuring the thermocouple output will depend on the accuracy required of the calibration being performed. Fluke 702 calibrator or Altek 422 is sufficient for most thermocouple calibrations.
REFERENCE THERMOMETERS:
The reference thermometer used for the comparison calibration of a thermocouple depends on the temperature range covered, the accuracy desired, the capabilities, or the preference of the calibration laboratory. The following are different examples of reference thermometers.
LIQUID-IN-GLASS THERMOMETERS:
Liquid-in-glass thermometers cover the range from -300 to 950Fahrenheit with an accuracy of 1- 3 Fahrenheit depending on the thermometer type and the width of the range covered. They are inexpensive but fragile, and if the highest degree of accuracy is to be achieved, an individual thermometer must cover a very narrow temperature range so that the graduation intervals can be as large as possible. A disadvantage of the liquid-in-glass thermometer is distinct reading errors because of fine graduations. Taylor Instruments offers Superior Grade Certified Secondary Reference Thermometers individually or in matched Celsius or Fahrenheit sets, which Branom stocks.
www.engineering-resource.com
14
APPLICATIONS
Thermocouples are suitable for measuring a large temperature range, up to 2300 C. They are less suitable for applications where smaller temperature differences need to be measured with high accuracy, for example the range 0100 C with 0.1 C accuracy. Thermistors and resistance temperature detectors are more suitable for these applications. Applications include temperature measurement for kilns, gas turbine exhaust, diesel engines, and other industrial processes.
www.engineering-resource.com
15 thermocouple temperature falls, causing the voltage across the thermocouple to drop and the valve to close. Some systems, known as millivolt control systems, extend this concept to the main gas valve as well. Not only does the voltage created by the pilot thermocouple activate the pilot gas valve, it is also routed through a thermostat to power the main gas valve as well. Here, a larger voltage is needed than in a pilot flame safety system described above, and a thermopile is used rather than a single thermocouple. Such a system requires no external source of electricity for its operation and so can operate during a power failure, provided all the related system components allow for this. Note that this excludes common forced air furnaces because external power is required to operate the blower motor, but this feature is especially useful for un-powered convection heaters. (8)
Steel industry
Steel and Iron industries make extensive use of type B, S, R and K thermocouples for temperature monitoring and chemistry of the steel making process. In an electric arc furnace, disposable, immersible, type S thermocouples are regularly used to accurately measure the temperature of steel before tapping. The cooling curve of a small steel sample can be analyzed and used to estimate the carbon content of molten steel.
Manufacturing
Thermocouples can be used in the testing of prototype electrical and mechanical apparatus. For example, switchgear under test for its current carrying capacity may have thermocouples installed and monitored during a heat run test, to confirm that the temperature rise at rated current does not exceed limits.
www.engineering-resource.com
16
Process plants
Chemical production and petroleum refineries employ computers for logging and limit testing various temperatures associated with a process. For such cases a number of thermocouple leads will be brought to a common reference block (a large block of copper) containing the second thermocouple of each circuit. The temperature of the block is then measured by a thermistor. Simple computations are used to determine the temperature at each measured location. (9)
www.engineering-resource.com
17
References
1. http://www.sensortecinc.com/docs/technical_resources/Thermocouple_Theory.pdf (Retrieved on 12/10/10) 2. http://en.wikipedia.org/wiki/Thermocouple#Laws_for_thermocouples (Retrieved on 12/10/10) 3. Daniel D. Pollock: The Theory and Properties of Thermocouple Elements; American Society for Testing and Materials; Hardcover; Pg15-20; 1971.
4. David R Keyser., Naval Ship Engineering Center Philadelphia PA Philadelphia Div; Computer Optimization Of Thermocouple Calibration; Defense Technical Information Center ; Pg 50-59; 1970 5. http://www.engineeringtoolbox.com/thermocouples-d_496.html (Retreived on 13/10/10) 6. http://www.branom.com/literature/thermocouple.html (Retreived on 13/10/10) 7. http://en.wikipedia.org/wiki/Thermocouple#Applications (Retreived on 14/10/10) 8. John G. Webster: The measurement, instrumentation, and sensors handbook; Springer; Volume 1; Pg 23-36; 1999 9. Thomas D. McGee, Thomas Donald McGee: Principles and methods of temperature measurement; Wiley-IEEE; Hardcover; Pg 278-285; 1988