Vicat Device Callibration
Vicat Device Callibration
Instrument name: ID: Referent norm: Date of calibration: Calibration tools: Timer device: Length device: Weight device: ID*: ID*: ID*: EN196-3
R011/1
APPROVED
Table 1. Timer inspection: Units: Time (Vicat): Time device: Difference: Range test: s 30 1 15 60 minute 120 180 240 Test passed
30
0 OK
1
0 OK
15
0 OK
60
0 OK
120
0 OK
181
1 OK
240
0 OK
OK
d1
d2
d3
1.11
1.11
1.09 1.21
1.11
Test pas
OK
OK
OK
OK
47
45 OK
Test pass
OK
YES
OK
299.2
297.1
295
298.3
Test pass
OK
Table 6. Mold inspection: Dimension (mm) Mold Dimension (mm) Required (mm) max min
R011/1
L1
L2
L3
69.0
65 85 OK
70.0
65 85 OK
40.0
39.5 40.5 OK Test pass
Range test: Table 7. Software inspection: Needle ZERO position, auto calibration Table 8. Software inspection: Software parameters Check in the last version of LI001 Table 9. Base plate dimension: All base plates used for setting time test has to be same thickness Table 10 Visual inspection:
OK
Test passed:
YES
No , needle ZERO position is not set
OK
Test passed:
YES
No , parameters are not set good
OK
Test passed:
YES
Yes , all base plates are same thickens No , If not, remove all which are different
OK
Device is clean Any visible damage (if YES, give description in annex) Does the room condition suitable for this device Electricity connector is proper and safe
Final Conclusion:
APPROVED
OK
If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this document)
2)
R011/1
--
Table 1. Timer inspection: Units: Time (Vicat): Time device: Difference: Range test: --------------s 30 1 15 60 minute 120 180 240 Test passed
--
d1
d2
d3
Test pas
--
--
--
--
Table 6. Mold inspection: Dimension (mm) Mold Dimension (mm) Required (mm) max min
R011/1
L1
L2
L3
Test pass
65 85 --
65 85 --
39.5 40.5 --
Range test: Table 7. Software inspection: Needle ZERO position, auto calibration Table 8. Software inspection: Software parameters Check in the last version of LI001 Table 9. Base plate dimension: All base plates used for setting time test has to be same thickness Table 10 Visual inspection: Device is clean
OK
Test passed:
Yes , needle ZERO position is set No , needle ZERO position is not set
-Test passed:
Yes , parameters are set good No , parameters are not set good
-Test passed:
Yes , all base plates are same thickens No , If not, remove all which are different
--
Any visible damage (if YES, give description in annex) Does the room condition suitable for this device Electricity connector is proper and safe
Final Conclusion:
--------
--
If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this document)
1)
2)
R011/1
Table 7
Table 9
Table 2
d1 d2 d3
Table 3 L = 45mm
Table 4
Table 6 L1 = 65-85mm
L3 = 40.0 2mm
L2 = 65-85mm NOT OK