Industrial & Precision Flowmeters

CRYOGENIC FLUIDS

Proven Performance for Over 50 Years


A DIVISION OF

POWER GENERATION

For over 50 years, COX Flow Measurement has met the test of endurance, which can only be achieved by manufacturing a quality flowmeter with unsurpassed repeatability and linearity. The COX turbine flowmeter is embedded in more military applications than any other manufacturer. From this platform, COX launched products into every liquid and gas application in the industry. Today, we offer the highest quality turbine meters available with unmatched calibration uncertainty. COX products can be found in applications ranging from fuels, demineralized water and chemicals, to cryogenic flow measurement of CO2. With a variety of end-fittings to mate with piping and an array of electronics for signal processing, COX Flow Measurement provides the right solution for demanding flow applications. Our sales engineers stand ready to help you with your applications and our customer service team provides quality after-sales product support.

BULK LOADING

FOOD PROCESSING COX Flow Measurement maintains an accredited NVLAP Primary Standard Calibration Laboratory

FM78587 AS9100 and ISO 9001

REFINING

The Company
COX Flow Measurement is a division of Badger Meter, Inc. USA, based in Milwaukee, Wisconsin. With sales of more than $250 million, combined with the dedication of more than 1,200 employees around the world, Badger Meter has been a leading marketer and manufacturer of flow measurement and control devices since 1905. COX Flow Measurement consists of COX Precision and Industrial flowmeters, Exact Flow high-accuracy dual rotor flowmeters, and the prestigious Flow Dynamics NVLAP (Lab Code 200668) accredited Primary Standard Calibration Laboratory. These companies complement the best-in-class technology of Badger Meter.

COX Flow Measurement offers a variety of flow metering products, signal conditioners, indicators and flow computers, which provide compensation algorithms to achieve the ultimate accuracy in flow. COX Flow Measurement maintains a competent NVLAP accredited calibration facility, in accordance with the U.S. Code of Federal Regulations (CFR, Title 15, Part 285) and encompassing the requirements of ISO/IEC 17025. The ongoing NVLAP audits demonstrate COX Flow Measurements excellence in maintaining quality systems, test and calibration methods, and measurement traceability. In addition, all products are manufactured with care and conformity to AS9100 and ISO 9001:2008 directives.

Badger Meter Milwaukee, Wisconsin

COX Flow Measurement Scottsdale, Arizona

COX Flow Measurement Provides: 1. Precision Turbine Meters 2. Primary Standard Calibration Service 3. Flow Transfer Standard Calibration Systems 4. Primary Standard Calibrators 5. OEM Manufacturing
FM78587 AS9100 and ISO 9001

Table of Contents

FLOWMETER CONFIGURATIONS ......................................................................................4 COX HIGH-ACCURACY PRECISION AND INDUSTRIAL FLOWMETERS ...........................6 CX FLOW RANGE TABLE ................................................................................................7 CX STANDARD METER MATERIALS ................................................................................7 CX HOUSING PRESSURE RATINGS .................................................................................8 FLANGE PRESSURE RATINGS.........................................................................................8 CX MODEL NUMBERS ...................................................................................................9 TURBINE FLOWMETER DIMENSIONS ...........................................................................10 CX AN/NPT TURBINE FLOWMETER DIMENSIONS..........................................................10 CX SA TURBINE FLOWMETER DIMENSIONS .................................................................10 FLANGED TURBINE FLOWMETER DIMENSIONS............................................................11 CLF LOFLO TURBINE METER DIMENSIONS ...............................................................11 LOFLO SERIES .............................................................................................................12 LOFLO MODEL NUMBER AND FLOW RANGE TABLE......................................................13 EXACT SERIES DUAL ROTOR FLOWMETERS...............................................................14 EXACT SIZE AND FLOW RANGE....................................................................................15 EXACT MODEL NUMBERS............................................................................................16 EXACT AN TURBINE FLOWMETER DIMENSIONS ...........................................................17 EXACT DUAL ROTOR TURBINE CONCEPT .....................................................................18 FLOWMETER CALIBRATION CODES .............................................................................18 CALIBRATION CAPABILITY ............................................................................................19 ELECTRONIC PRODUCTS ...............................................................................................20 SIGNAL CONDITIONERS...............................................................................................20 INDICATOR DISPLAYS ..................................................................................................20 PANEL MOUNT FLOW COMPUTERS .............................................................................21 FLOW COMPUTERS, METER MOUNTED & REMOTE......................................................21 FLOW STRAIGHTENERS...............................................................................................22 INSTALLATION.............................................................................................................22 FLOW TRANSFER STANDARDS ....................................................................................23 FLOW DYNAMICS PRIMARY STANDARD PD CALIBRATORS ......................................24 FLOW CALIBRATORS ...................................................................................................24 OEM SEMI-AUTOMATIC CALIBRATORS ........................................................................26 CALIBRATOR CHECK STANDARDS ...............................................................................26 MANUFACTURING CELLS............................................................................................27

FLOWMETER CONFIGURATIONS

CX SERIES END-FITTING OPTIONS SINGLE ROTOR

EXACT SERIES END-FITTING OPTIONS DUAL ROTOR

Wire Leads Amplified Pickoff Flanged Series Pipe Fittings 150# to 600# AN Series Pipe Fittings 37" Flare From 1/4" to 2" 0.25 to 310 GPM

Sine Wave RF Carrier

NPT Series Pipe Fittings From 1/4" to 2" 0.25 to 310 GPM

Flanged Pipe Fittings From 1/4" to 4" 0.025 to 1,500 GPM

AN Series Pipe Fittings 37" Flare From 1/4" to 2" 0.25 to 310 GPM

Tri-Clover Pipe Fittings From 1/4" to 2" 0.25 to 310 GPM

MS Amplified Pickoff

Amplified Square Wave

Housing Wire Termination

Wire Leads Magnetic Pickoff LoFlo Series AN Pipe Fittings 1/2", 37" Flare 0.007 to 1.25 GPM
4

Sine Wave MV Signal

FLOWMETER CONFIGURATIONS

METER MOUNTED ELECTRONICS

PANEL MOUNTED ELECTRONICS

CUSTOMERS CONTROL SYSTEM

Meter Mounted Model: EC80 Flow Computer Enclosure: Explosion-proof cast aluminum Power: 24 VDC Nominal Inputs: Single/Dual Rotor freq., RTD Outputs: 2 freq., 2 analog (4-20 mA, 0-10 VDC) Digital Outputs: 0-5 V TTL, RS485

Panel Mounted Indicator Model: CUB5 Rate or Total indicator up to 8 digits Programmable rate scaling and decimal point placement Power: 9-28 VDC Enclosure: NEMA 4X/IP65 Input: 4-20 mA and RTD

Customers Data Acquisition System Not sold by COX Flow Measurement

Remote Mounted Model: 4080 Flow Computer Enclosure: Aluminum with mounting flange NEMA 4 Power: 24 VDC Nominal Inputs: Single/Dual Rotor freq., RTD Outputs: 2 freq., 2 analog (4-20 mA, 0-10 VDC) Digital Outputs: 0-5 V TTL, RS485

Customers Control System

Model: RT112 Rate & Total Indicator 15-point linearization Power: Loop or battery-powered, 8-24 V AC/DC or 115-230 V AC power supply Input: Sine wave, pulse, 4-20 mA or 0-10 VDC Output: Scaled 4-20 mA or 0-10 VDC

Panel Mounted Flow Computer Model: FC30 Flow Computer Liquid and gas applications Batching and temperature viscosity corrections Power: 110/220 VAC and 12/24 VDC Input: Single/Dual Rotor freq., RTD Output: Pulse, RS232 and 485

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 5

PROVEN PERFORMANCE FOR OVER 50 YEARS

COX HIGH-ACCURACY PRECISION AND INDUSTRIAL FLOWMETERS


COX Flow Measurement continues to improve meter performance by offering long-life ceramic bearings and unsurpassed primary standard calibration uncertainties of 0.05% of reading. Coupled with an advanced EC80 flow computer, these flowmeter systems provide fully compensated precision flow measurements.

COX Flow Measurement has been the icon of turbine flowmeters since its inception. The installation base is a testimony to customers who have discovered the performance and longevity of COX turbine meters and associated electronics. DESCRIPTION COXs superior performance is credited to the axial helical rotor design, which enhances linearity, repeatability and speed-of-response and reduces pressure drop. A single-bearing design minimizes friction and delivers linearity beyond that of the standard 10:1 range. This bearing system also extends the repeatable range to 60:1 on small meters and up to 150:1 on large meters.

In addition, COX Flow Measurement offers the EC80 flow computer, which will linearize the meter within 0.1% linearity over the entire repeatable range. This flow computer offers temperature and pressure compensation with multiple outputs for frequency and analog in engineering units. COX precision flowmeters utilize an improved, extended 6-vane support to reduce fluid swirl from entering the rotor. Ceramic ball bearings are standard for improved durability. The 316 SS housing exceeds the pressure ratings of the AN (MS) fittings. Hex or square housings make it easy for tightening the threaded fittings during installation and removal.

ADVANTAGES
Superior linearity Exceptional repeatability Wide flow range Fast response to rate changes 316 SS housings standard Ceramic bearings 4-week delivery or less on standard models Primary Standard Calibrations traceable to NIST & NVLAP (Lab Code 200668) accredited
Repeatability Linearity

PERFORMANCE SPECIFICATIONS
Calibrator Uncertainty 0.05% of reading 0.02% of reading 0.5% of reading 0.1% of reading 500 to 1500 Hz standard See charts 0-10 V (square wave pulse) 2-3 mS or better

With Electronic Linearizer Frequency Output (maximum) Pressure Rating Output Signal Response Time

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 6

CX FLOW RANGE TABLE

Model CX Meter Size Inches Meter Size in 1/16"

Std. Linear Range GPM

Std. Linear Flow Ratio

P @ 10:1 Flow Ratio

Repeatable Range GPM Repeatable Flow Ratio

Freq. @ Max. Flow

P @ Repeatable Flow Ratio

1/4" 3/8" 1/2" 5/8" 3/4" 1" 1" 1" 2" 3" 4"

8-4 8-6 8 10 12 16 20 24 32 48 64

0.25 - 3.0 0.50 - 6.0 0.75 - 9.5 1.25 - 15.5 2.50 - 30 5.0 - 65 7.5 - 95 12.5 - 155 25 - 310 53 - 800 100 - 1500

12:1 12:1 12.6:1 12.5:1 12:1 13:1 12.6:1 12.4:1 12.4:1 15:1 15:1

3.5 4.5 6.0 5.0 5.0 5.0 5.5 6.0 6.5 10.0 10.0

0.05 - 3.0 0.06 - 6.0 0.09 - 9.5

60:1

6.5 8.5 8.5

100:1 0.16 - 15.5 0.30 - 30 0.45 - 65 0.65 - 95 150:1 1.05 - 155 2.10 - 310 20 - 800 50 - 1250 25:1 25:1 45 Hz 20 Hz 9.0 9.0 10.0 10.0 1500 Hz 9.0 9.0 9.0 8.5

Standard ranges = 10-point calibration and extended ranges = 20-point calibration Meter linearity was developed using MIL-C-7024C, Type 2 at 80F

CX STANDARD METER MATERIALS


Helical Rotor Material: 17-4 SS Ball Bearing: Ceramic Meter Body Material: 316 SS Note: Other materials available upon request. Lightweight helical rotors provide 3 mS, or less, speed-of-response and extended linear ranges. The helical blade profile transfers energy more efficiently, reducing bearing wear, pressure drop and their associated cost. Industrial applications will benefit from using ceramic flow-through ball bearings. This ceramic bearing system is extremely versatile and the recommended bearing for most applications, including cryogenics. This superior lowfriction bearing outperforms journal bearings and extends the usable range of the turbine flowmeter. Whatever the fluid media, COX Flow Measurement has a proven solution.
Versatile ceramic bearing for water, cryogenics and hydrocarbons Lightweight helical rotor

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 7

CX HOUSING PRESSURE RATINGS

The type of end-fitting used may have a lower pressure rating than the AN flowmeter housing

FLANGE PRESSURE RATINGS

Maximum non-shock allowable working pressure in psig at 100F or less Alloy 150 Flg Stainless Steel 316 A-181 275 300 Flg 720 Pressure Class 600 Flg 900 Flg 1440 2160 1500 Flg 3600 2500 Flg 6000

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 8

CX MODEL NUMBERS

Model Number Format for AN, NPT, SA and Flange End-Fittings

CX

NP

C2

CB

XXX

COX Model Size Number (inches) CX 1/4" 3/8" 1/2" 5/8" 3/4" 1" 1" 1" 2"
Only Available in Flange Sizes

AN Sizes 8-4 8-6 8 10 12 16 20 24 32 40 48 64

EndFittings AN=37Flare NP=NPT Male SA=Sanitary F1=150# Flng SS F3=300# Flng SS F6=600# Flng SS C1=F150C C3=F300C C6=F600C

Pickoff Description C1=Carrier MS 3 Pin C2=Carrier NPT threaded with leads C3=Car MS RTD C4=Car NPT RTD C5=Car MS RTD IS C6=Car NPT RTD IS M1=Mag MS M2=Mag NPT M3=Mag MS RTD M4=Mag NPT RTD M5=Mag MS RTD IS M6=Mag NPT RTD IS

Bearing CB=Ceramic Ball SB=SST Ball

3-Digit Special

Used to identify special configurations

2" 3" 4"

Refer to Electronic Products Section for a complete review of all electronic signal conditioners, indicators and flow computers. MODEL NUMBER EXAMPLES: CX1/2AN-C3CB-XXX CX2F1-C2CB-XXX

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 9

TURBINE FLOWMETER DIMENSIONS

CX AN/NPT TURBINE FLOWMETER DIMENSIONS

COX Model CX-1/4 CX-3/8 CX-1/2 CX-5/8 CX-3/4 CX-1 CX-1 CX-1 CX-2

A NPT 2.7 (69) 2.7 (69) 2.7 (69) N/A 3.29 (84) 3.78 (96) 4.23 (107) 4.67 (119) 5.89 (150)

A AN 2.45 (62) 2.45 (62) 2.45 (62) 2.72 (69) 3.25 (83) 3.56 (90) 4.06 (103) 4.59 (117) 6.06 (154)

B 1.12 (28) n 1.12 (28) n 1.12 (28) n 1.25 (32) n 1.25 (32) n 1.63 (41) 1.88 (48) 2.25 (57) 2.75 (70)

C Max 3.43 (87) 3.46 (88) 3.52 (89) 3.59 (91) 3.65 (93) 3.77 (96) 3.89 (99) 4.01 (102) 4.24 (108)

Tube Size 1/2" (127) 1/2" (127) 1/2" (127) 5/8" (15.9) 3/4" (19) 1" (25.4) 1" (31.7) 1" (38) 2" (51)

AN EndFitting 8 8 8 10 12 16 20 24 32

n Square Body Hex Body

Note: Housing dimensions are in inches (mm). All dimensions are for reference only. Actual dimensions should be obtained from the factory.

CX SA TURBINE FLOWMETER DIMENSIONS

COX Model CX-1/4 SA CX-3/8 SA CX-1/2 SA CX-5/8 SA CX-3/4 SA CX-1 SA CX-1 SA CX-1 SA CX-2 SA

A 3.56 (90) 3.56 (90) 3.56 (90) 3.56 (90) 3.56 (90) 3.56 (90) 4.59 (117) 4.59 (117) 6.06 (154)

B 0.984 (24.99) 0.984 (24.99) 0.984 (24.99) 1.984 (50.39) 1.984 (50.39) 1.984 (50.39) 1.984 (50.39) 1.984 (50.39) 2.516 (63.91)

C Max 3.43 (87) 3.46 (88) 3.52 (89) 3.59 (91) 3.65 (93) 3.77 (96) 3.89 (99) 4.01 (102) 4.24 (108)

Tube Size 1/2" (127) 3/4" (19) 1" (25.4)

Clamp Size

Code

3/4" (19)

T1

1" (38) 1" (38) 2" (51) 2" (51)

T2

T3

Note: Housing dimensions are in inches (mm). All dimensions are for reference only. Actual dimensions should be obtained from the factory.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 10

TURBINE FLOWMETER DIMENSIONS

FLANGED TURBINE FLOWMETER DIMENSIONS

COX Model CX-1/4 F_ CX-3/8 F_ CX-1/2 F_ CX-5/8 F_ CX-3/4 F_ CX-1 F_ CX-1 F_ CX-1 F_ CX-2 F_ CX-2 F_ CX-3 F_ CX-4 F_

A 5 (127) 5 (127) 5 (127) 5.5 (140) 5.5 (140) 5.5 (140) 6 (152) 6 (152) 6.5 (165) 7 (178) 10 (254) 12 (305)

B F1=150 lb Flange 3.5 (89) 3.5 (89) 3.5 (89) 3.5 (89) 3.87 (98) 4.25 (108) 4.62 (117) 5 (127) 6 (152) 7 (178) 7.5 (191) 9 (229)

B F2=300 lb Flange 3.75 (95) 3.75 (95) 3.75 (95) 3.75 (95) 4.62 (117) 4.87 (124) 5.25 (133) 6.12 (156) 6.5 (165) 7.5 (191) 8.25 (210) 10 (254)

B F3=600 lb Flange 3.75 (95) 3.75 (95) 3.75 (95) 3.75 (95) 4.62 (117) 4.87 (124) 5.25 (133) 6.12 (156) 6.5 (165) 7.5 (191) 8.25 (210) 10.75 (273)

ANSI raised-face integral flanges

Note: Housing dimensions are in inches (mm). All dimensions are for reference only. Actual dimensions should be obtained from the factory.

CLF - LOFLO TURBINE METER DIMENSIONS

COX Model CXLF-8-6

Tube Size 3/8" (9.5)

A 2.22 (56.4)

B 1 (25.4)

C Max 4.25 (108)

AN Size 6

Note: Housing dimensions are in inches (mm). All dimensions are for reference only. Actual dimensions should be obtained from the factory.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 11

PROVEN PERFORMANCE FOR OVER 50 YEARS

LOFLO SERIES
Bearings are the key factor in performance when measuring low flow. That is why COX Flow Measurement uses ceramic ball bearings, which are near perfect in roundness within 10 millionths of an inch variation. This means lower friction and less vibration, which equals superior repeatability.

The COX LoFlo Series is a robust axial turbine flowmeter with exceptional repeatability of 0.25% of reading. When paired with the EC80 Flow Computer, the calibration data is linearized to within 0.1% of reading, allowing for precision flow measurement over the entire usable range of the meter. COX LoFlo Series flowmeters are not position-sensitive and can be mounted in any piping orientation. The custom long-life, ceramic low-friction ball bearing system used in the COX LoFlo Series meets the demands of water, hydrocarbons and cryogenic fluids. Capable of measuring flows as low as 0.006 GPM (0.024 LPM), this meter is an ideal solution for a multitude of applications. Typical uses include fuel injection production systems, blending of costly chemical additives, onboard fuel consumption, refrigerants, dye injections, and more. Ceramic ball bearings have much less fiction than journal bearings and provide a wider measureable flow range with superior repeatability. The COX LoFlo Series is designed to be rugged and repeatable with the assurance of the most accurate calibration provided on our positive displacement primary standard calibrators.

ADVANTAGES
Low-friction ceramic ball bearings Suitable for water, hydrocarbons and cryogenic service 17-4 helical rotor Compact size (2.22" face dimension)
Repeatability

PERFORMANCE SPECIFICATIONS
0.25% 1500-1800 Hz 6000 psi 20-30 mS 316 SST -6 AN (MS) Frequency Output Pressure Rating Response Time Body Construction End Fittings

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 12

PROVEN PERFORMANCE FOR OVER 50 YEARS

LOFLO MODEL NUMBER AND FLOW RANGE TABLE English Units


COX Model Number CLF6-000 CLF6-00 CLF6-0 CLF6-1 CLF6-2 CLF6-3 Std 10:1 Range Magnetic Pickoff GPM 0.007 - 0.075 0.012 - 0.125 0.025 - 0.25 0.050 - 0.50 0.075 - 0.75 0.125 - 1.25 Lbs. / Hour 2.85 - 28.5 4.75 - 47.5 9.50 - 95.0 19.0 - 190 28.5 - 285 47.5 - 475 Extended Range Carrier Pickoff GPM 0.006 - 0.075 0.009 -0.125 0.014 - 0.25 0.023 - 0.50 0.037 - 0.75 0.06 - 1.25 Lbs. / Hour 2.2 - 28.5 3.4 - 47.5 5.4 - 95.0 8.8 - 190 13.2 - 285 22 - 475 3/8" 3/8" 3/8" 3/8" 3/8" 3/8" Tube Size (inches) Max. Frequency (Hz) 1800 1800 1800 1500 1500 1500 *P @ Max Flow Range (PSI) 20.0 12.0 4.5 4.0 4.5 7.0

Standard ranges = 10-point calibration and extended ranges = 20-point calibration * Pressure drop calculation based on liquid with a specific gravity of 0.76.

Metric Units
COX Model Number CLF6-000 CLF6-00 CLF6-0 CLF6-1 CLF6-2 CLF6-3 Std 10:1 Range Magnetic Pickoff Lit. / Min. 0.27 - 0.284 0.045 - 0.473 0.095 - 0.946 0.189 - 1.89 0.284 - 2.84 0.473 - 4.73 Kg. / Hour 1.29 - 12.9 2.15 - 21.5 4.31 - 43.1 8.62 - 86.2 12.9 - 129 21.5 - 215 Extended Range Carrier Pickoff Lit. / Min. 0.024 - 0.284 0.034 - 0.473 0.053 - 0.946 0.087 - 1.89 0.140 - 2.89 0.227 - 4.73 Kg. / Hour 0.998 - 12.9 1.54 - 21.5 2.45 - 43.1 3.99 - 86.2 5.99 - 129 9.98 - 215 3/8" 3/8" 3/8" 3/8" 3/8" 3/8" Tube Size (inches) Max. Frequency (Hz) 1800 1800 1800 1500 1500 1500 *P @ Max Flow Range (Kg/cm2) 1.40 0.84 0.32 0.28 0.32 0.49

Standard ranges = 10-point calibration and extended ranges = 20-point calibration * Pressure drop calculation based on liquid with a specific gravity of 0.76.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 13

PROVEN PERFORMANCE FOR OVER 50 YEARS

EXACT SERIES DUAL ROTOR FLOWMETERS

WORLDS MOST PRECISE STANDARD & INDUSTRIAL TURBINE FLOWMETERS


COX Flow Measurement offers proven flow metering solutions for the most demanding aerospace, automotive, industrial and OEM applications. The Exact Series Dual Rotor Turbine Flowmeter is the worlds most precise meter of its kind, providing extended range performance not obtainable with traditional single rotor designs. DESCRIPTION The Exact Series Dual Rotor Turbine Flowmeter provides an extended flow range capability, which often eliminates the need for manifold systems and thus simplifies installation and lowers costs. The meters exceptional performance, enabled by the innovative dual rotor configuration, enhances UVC curves and extends the usable measurement flow range. With the Exact Series Dual Rotor Turbine Flowmeter (Standard Model CDX/CDL), flow straighteners are not required to control process fluid swirl, as the dual rotor system cancels out the rotor acceleration effect. With flow straighteners, bearing diagnostics can be determined by monitoring the ratio of the rotors to detect wear or cleanliness. The Exact Series meters also utilize a unique pickoff system, which is impervious to vibration and minimizes space. This allows for direct coupling of electronics on a robust mount having a lower profile. Both integral and remote electronics are available to process the signal output. Exact Series Dual Rotor technology is ideal for a wide range of applications. The meter is designed for bi-directional flow and high-shock environments. It utilizes a robust bearing system having dual ceramic bearings on each rotor with the internals securely locked in place, resulting in excellent repeatability. The dual rotor meter allows for installation in applications that do not have space for flow straighteners without loss of measurement accuracy. Pressures of up to 30,000 psig can be contained, while compensating for viscosity changes using integrated pressure sensors.

DX explosion-proof with flow computer

COX Flow Measurement offers the pinnacle of turbine flowmeter performance. The Dual Rotor is NISTs golden meter for round robin testing. Its use around the world for primary standard calibrator correlation tests is proof of performance.

DL MS connector for remote flow computer

ADVANTAGES
No need for flow straighteners Enhanced performance due to helical rotors Superior absolute accuracy Excellent repeatability Wider operating flow range Extended UVC flow range Integral pickoff impervious to vibration NVLAP calibration uncertainty
Repeatability

PERFORMANCE SPECIFICATIONS
Calibration Uncertainty 0.05% of reading 0.02% of reading 0.01% of reading -450F (-270C) to 300F (150C) standard Up to 30,000 PSIG depending on size and end-fittings 14 PSID at max. flow rate @ 1.2 cStk Ceramic standard (water & hydrocarbons)

Linearity (with linearizer) Process Temperature Operating Pressure Pressures Drop Bearings

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 14

PROVEN PERFORMANCE FOR OVER 50 YEARS

STANDARD METER MATERIALS


EXACT SIZE AND FLOW RANGE TABLE

Model CDX/CDL Meter Size (inches) 1/4" 3/8" 1/2" 5/8" 3/4" 1" 1" 1" 2" 2" 3" 4" Meter Size in 1/16" 8-4 8-6 8 10 12 16 20 24 32 40 48 64

Flow Range @ 1 cStk S.G. = 1 GPM 0.025 - 3.0 0.035 - 6.0 0.040 - 10 0.045 - 16 0.08 - 30 0.15 - 65 0.25 - 95 0.35 - 155 0.60 - 310 1.0 - 500 2.0 - 800 5 - 1500 LPM 0.095 - 11.4 0.132 - 22.7 0.151 - 37.8 0.170 - 60.5 0.302 - 113.5 0.378 - 246 0.946 - 359.6 1.324 - 586.7 2.27 - 1173.4 3.78 - 1892.7 7.57 - 3028 18.9 - 5678

Turndown range 1 cStk 120:1 171:1 286:1 355:1 375:1 433:1 380:1 443:1 517:1 500:1 400:1 300:1

UVC Turndown Range 1 cStk 30:1 35:1 40:1 45:1 50:1 60:1 60:1 60:1 60:1 60:1 60:1 60:1

Typical K-Factor Pulses/Gallon Rotor A 30000 15000 9480 6000 3640 1680 1149 705 242 132 45 20 Rotor B 33000 16500 10440 6375 4000 1846 1263 774 266 145 50 22

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 15

PROVEN PERFORMANCE FOR OVER 50 YEARS

EXACT MODEL NUMBERS

CDL

8-4

AN

CB

XXX

Exact Model Number CDL=Low Profile MS Connector CDX= Explosion-Proof Spud

Size (inches) 1/4"

AN Sizes 8-4

End-Fittings AN

Bearing CB-Ceramic Ball

Temperature Sensor R = RTD

3-Digit Special

3/8" 1/2" 5/8" 3/4" 1" 1 1 2" 2 3" 4"

8-6 08 10 12 16 20 24 32 40 48 64

F1=F150S F3=F300S F6=F600S

Used to identify special configurations

Flange sizes only

MODEL NUMBER EXAMPLES: CDX3/4-F1-CB-R-XXX and CDL1AN-CB-R-XXX Bearings are the most critical component of a turbine flowmeter in terms of both repeatability of the meter output and the security of the calibration over time. This is why all COX Flow Measurement turbine meters use ceramic ball bearings. There are several advantages in using ceramic material: Ceramic can be used in non-lubricating fluids Ceramic is non-corrosive and heat-tolerant, making it ideal for water or hydrocarbons Ceramic offers near perfect roundness, minimizing vibration Ceramic is harder and measures RC 75-80 as compared to SST at RC58-65 Ceramic is 50% lighter than SST, extending the life of the bearing

All the above attributes reduce heat, providing five times more wear resistance. Ultimately, this preserves the calibration of the meter, while being more tolerant to particulates.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 16

PROVEN PERFORMANCE FOR OVER 50 YEARS

EXACT AN TURBINE FLOWMETER DIMENSIONS

Exact Series Model Number

Tube Size (inches) 1/4" 3/8" 1/2"

A AN 2.75 (70) 2.75 (70) 2.75 (70) 2.92 (74) 3.25 (83) 3.56 (91) 4.06 (103) 4.59 (117) 6.06 (154) N/A N/A N/A

B AN 0.87 (20.1) 0.87 (20.1) 0.87 (20.1) 1.00 (25.4) 1.37 (34.8) 1.50 (38.1) 1.88 (47.7) 2.00 (50.8) 2.62 (66.5)

A Flange

AN End-Fitting 8 8 8 10 12

CXL Dual Rotor, MS Connector CDX Dual Rotor, ExplosionProof

5/8" 3/4" 1" 1" 1" 2" 2" 3" 4"

5.50 (140) 6.00 (152) 6.00 (152) 6.50 (165) 7.00 (178) 10.00 (254) 12.00 (305)

16 20 24 32 N/A N/A N/A

Note: Housing dimensions are in inches (mm). All dimensions are for reference only. Actual dimensions should be obtained from the factory.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 17

PROVEN PERFORMANCE FOR OVER 50 YEARS

EXACT DUAL ROTOR TURBINE CONCEPT

1. Unconditioned flow enters flowmeter. 2. Straightening vanes smooth the flow as it enters the first rotor. 3. Flow transfers momentum to the first rotor, making it spin counterclockwise. Flow then exits rotor, with a clockwise spin. 4. Flow enters second rotor with a nearly perpendicular angle of attack, thereby transferring additional momentum to the second rotor. This additional momentum results in greatly extended turndown. 5. Flow exits flowmeter. 6. Dual pickoffs transmit the rotor frequency signal to remote instrumentation. NOTE: The ratio of both pickoff frequencies is used for bearing diagnostics.

Blue is the uid ow Orange is the rotor rotation

FLOWMETER CALIBRATION CODES


Standard Calibration Codes Model # CAL __ __ __ (Meter Size) # of Cal Points 10 = 10 cal points 20 = 20 cal points Fluids S = Solvent B = Oil Blend W = Water 1 cStk Viscosity Qty. 1 = 1 viscosity 2 = 2 viscosities 3 = 3 viscosities Comments All dual rotor meters must use the EC80 Flow Computer to sum rotor frequencies and linearize the output to 0.1% of reading.

NOTE: 1. Calibrations are performed on primary standard calibration equipment operating in a NVLAP-accredited facility with an uncertainty of 0.5% of reading. 2. Solvent ranges from 1 to 1.14 cStk = MIL-C-7024C, Type 2 @ 80F. 3. Blends are a mixture of oils and solvent to obtain a specified viscosity.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 18

PROVEN PERFORMANCE FOR OVER 50 YEARS

CALIBRATION CAPABILITY
Calibrations are performed at COX Flow Measurement, a NVLAP (Lab Code 200668) accredited facility using Primary Standard Calibrators producing 0.05% of reading uncertainty.

Customers can select from a variety of calibrations to enhance the accuracy of the flow measurement reading. Often, manufacturers perform water calibrations and then use a correction factor. This can be misleading, as the calibration curve will vary in more viscous fluids. COX Flow Measurement avoids calibration shifts by using the actual fluid viscosity in the calibrator to develop the calibration curves. Multiple calibrations can be provided to form a single Universal Viscosity Curve, which adjusts the frequency/viscosity relationship to match the actual temperature of the known fluid. CALIBRATION CAPABILITY Liquid Flow Measurement: 0.0025 to 1,500 GPM hydrocarbon 0.75 to 1,000 centistokes viscosity 0.01 to 300 GPM water Gas Flow Measurement: 0.001 to 3,500 SCFM in air 0.001 to 400 SCFM inert or flammable gases Gas correlations for toxic gases Partial vacuum to 500 PSIA Primary Gas Calibrators: Uncertainty: 0.20% to 0.25% of reading Repeatability: 0.10% of reading CALIBRATOR UNCERTAINTY Primary Liquid Calibrators: Uncertainty: 0.025% to 0.05% of reading Repeatability: 0.012% of reading

COX Flow Measurement is a participant in the NIST round robin proficiency tests intended to eliminate calibrator bias. This goes beyond traditional NIST traceability and provides correlation data using NIST flowmeter artifacts to compare directly with NIST calibrations.

PDCL-400 Liquid Primary Standard Calibrator

100 Cubic Foot Gas Primary Standard Bell Prover

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 19

PROVEN PERFORMANCE FOR OVER 50 YEARS

ELECTRONIC PRODUCTS
SIGNAL CONDITIONERS Carrier Pickoff/Amplifier
Description: COX Carrier Pickoff/Amplifier extends the usable range of the flowmeter by eliminating magnetic drag on the rotor. This enhances not only the range, but also the linearity at low flow. The modulated carrier frequency is amplified and conditioned to provide a square wave pulse output, capable of transmitting up to 100 ft. Specifications: Input Power: Min 11.0 VDC; Max 30.0 VDC Input Current Standard: 15 mA Output Voltage: 0 to 10 VDC Temperature Range: -40F to 248F (-40C to 120C)

INDICATOR DISPLAYS Digital Display, Panel Mount


Description: The indicator toggles manually or automatically, to view either rate or total and displays up to 8 digits with separate scaling and decimal point placement. Programmable update times, front panel reset and dual counters make this miniature digital indicator ideal for flowmeters with pulse outputs. The LCD display, with letter height of 0.46", is provided in positive reflective or negative transmissive with yellow/green or red LED backlighting, to meet your visual requirements. Specifications: Input Power: 9 to 28 VDC Non-volatile memory NEMA 4X/IP65 front bezel Operating Temperature: -31F to 167F (-35C to 75C) depending on intensity level and display color Display response rate indicator has low or high update times Frequency: Min 0.01Hz Max 20 KHz

RT112 Flow Rate Indicator/Totalizer, with Linearization


Description: The RT112 has a large, UV-resistant display of 0.67" (17 mm) with digits of 0.31" (8 mm), which can be set to show flow rates or totals in selectable engineering units. Linearization is standard with up to 15 data points of calibration and selectable output signal options. The linearized accumulated total can register up to 11 digits and is backed up in EEPROM memory every minute. All settings are safely stored in EEPROM memory in the event of a sudden power failure. The display enclosures are water-tight, conforming to NEMA 4X/ IP67, and available in aluminum and molded configurations. An intrinsically safe option is available conforming to ATEX, IECEx, FM and CSA. The unit has a maximum operating temperature of 158F (70C). The RT112 can be mounted directly on the flowmeter, wall mount, exterior panel mount or pipe stand. Specifications: Input Power: Long-life lithium battery - 110/230 VAC 50/60 Hz - 8/24 VAC/DC, 10 watts max. 4-20 mA loop-powered Auto Backup: All settings and running totals Operating Temperature: -22F to 178F (30C to 80C) Update Rate: User-definable from 10 times/sec 30 secs Linearization: 15 data points Signal Output: Rate = 4-20 mA, 0-10 VDC, Total = scaled pulse Display: Rate & Total with selectable on-screen engineering units Configuration Settings: Front panel input with simple operator menu. Pass-code protected. Modbus Communication: RS232/RS485

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 20

PROVEN PERFORMANCE FOR OVER 50 YEARS

PANEL MOUNT FLOW COMPUTERS Series FC30 Flow Computer, Panel Mount
Description: COX FC30 Flow Computer services both liquid and gas applications. The FC30 has features such as: batching, 40-point flowmeter linearization tables, temperature viscosity correction using Strouhal/Roshko advanced linearization methods, and multiple flow and control inputs. Relay outputs are independently assignable for alarms and general purpose warnings. The FC30 has serial communications, both RS232 and RS485, and isolated analog and pulse outputs. The display has two lines of 20 characters (0.3" in height), with blue VFD or backlit LCD. During start-up, there is a service or test mode to check out the system by monitoring inputs and exercising outputs, and printing system setup. The unit is CEcompliant, UL/CUL listed, and housed in a DIN enclosure made of flame-retardant plastic. Specifications: Input Power: 110/220 VAC 50/60 Hz, 12/24 VDC, 300 mA max. Operating Temperature: 32F to 122F (0C to 50C) Accuracy: 0.01% FS at 20C Resolution: 16-bit Update Rate: 1 update/sec min. Linearization: 40-point table, UVC table or Strouhal/Roshko computations Temperature Input: 100 ohm RTD, resolution 0.01C Serial Communication: RS232 or RS485 (printing, data recording and communications with a computer) Flow Inputs: Pulse, single rotor, dual rotor, single rotor with dual pickups or quadrature Control Inputs: Menu-selectable for Start, Stop, Reset, Lock, Inhibit, Alarm Acknowledge, Print or Not Used Relay Outputs: Menu-assignable Low Rate Alarm, Hi Rate Alarm, Prewarn Alarm, Preset Alarm, or General Purpose Warning

FLOW COMPUTERS, METER MOUNTED & REMOTE Series EC80 ExactComp, Flowmeter Interface
Description: COX Series EC80 ExactComp Flowmeter Interface provides total flow compensation to enhance flowmeter accuracy, while extending the linear flow range. The Series EC80 tracks all variables to compensate for viscous and inertial effects due to fluid temperature and pressure variations. Enhanced DSP technology allows exceptional signal characterization and fast response outputting dynamic data in engineering units. Meeting the demanding requirements of the aerospace, automotive and test & measurement industries, the Series EC80 provides significant improvements in flowmeter accuracy under extreme temperature conditions. It can be mounted directly on the flowmeter or in a remote box.

Meter-mounted Flow Computer

Remote-mounted Flow Computer

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 21

FLOW COMPUTERS, METER MOUNTED & REMOTE Series EC80 ExactComp, Flowmeter Interface Specifications:
Performance Linearized Frequency: 0.1% of reading Linearized Analog Output: 0.1% of full scale RTI (>100K load impedance) Process Latency: 100 microseconds Input Power 24 VDC nominal, 3W max. 9-32 VDC (digital output only) 12-32 VDC (analog output voltage) 15-32 VDC (analog output, 4-20 mA) Temperature Environment Operating: -40F to 185F (-40C to 85C) Storage: -67F to 257 F (-55C to 125C) Humidity: 0-80% RH non-condensing Temperature Input (2-, 3- or 4-wire) RTD 100 Ohm Platinum, 0.00385 Alpha IEC751 Class B Standard Usable Range: -65F to 365F (-54C to 185C) Voltage: 0-10 VDC or 0-5 VDC Analog Outputs 2 Independent Channels Ch 1, 4-20 mA or 0-10 VDC Ch 2, 0-10 VDC only 16-Bit Resolution, 4-20 mA, 0-10 Volts 15-Bit Resolution, 0-5 Volts Load Impedance, <100K Ohm (voltage) Digital Outputs 2 Independent Channels 0-5 Volt, TTL, 0 to 5 Hz step response Minimum Load Impedance, 1K Ohm Remote Cable Length 100 feet environment dependent RS485 Serial Data (calibration interface only) BAUD 56.4K Raw Frequency Output 3 Independent Channels 0-5 Volt, TTL, 5 to 3500 HZ, square wave, 5K Ohm minimum load Enclosure Options Remote aluminum enclosure with mounting flange weatherproof Explosion-proof conforming to NEMA 4, CLI GR DC CL II, GR EFG CL III WET LOC cast aluminum Software Software conforms to SAE ARP 4990 calculations

FLOW STRAIGHTENERS
Required on single rotor turbine meters to condition the flow. Flow conditioners larger than 1/2" have internal straightening vanes and are furnished standard with pressure taps. Temperature taps are also provided. AN End-Fittings Flange End-Fittings NPT End-Fittings Tri-Clover End-Fittings

INSTALLATION
COX Turbine Flowmeters can be installed in any position, without affecting performance. To reduce flow turbulence, COX recommends using flow straighteners, which have 10 diameters in length upstream and 5 diameters in length downstream, with vanes. Where impractical due to space limitations, careful attention to the location of valves, bends, etc., is suggested. A 10-micron
22

filter on the small meters is recommended to prevent bearing contamination. Electrical installation requires a 3-conductor shielded cable to the readout device.

PROVEN PERFORMANCE FOR OVER 50 YEARS

FLOW TRANSFER STANDARDS


COX Flow Measurement offers a field solution for calibrating all types of flowmeters. Instead of removing equipment from service for recalibration, the innovative Flow Gator Flow Transfer Standard (FTS) allows users to bring the calibrator to the flowmeter. This affordable documenting field flow calibrator is intended for in-line calibration and validation of meters using the actual process liquid. The Flow Gator FTS is compatible with any flowmeter with an electronic output or visual indicator, and is capable of measuring and correcting the influences of line pressure and temperature effects on flow. The calibrator reads signals from a master flowmeter, a unit under test (UUT) and fluid temperature. With this data, it calculates and compares the precise flow rate from the master meter and UUT. Rugged, lightweight and compact, the Flow Gator incorporates hand-held electronics, thus eliminating the need to carry a laptop computer into the field. The master meter can be any meter of choicepreferably an Exact Series Dual Rotor Turbine Flowmeter, which provides exceptional repeatability of 0.02%, enabling total system uncertainty of 0.25% of reading.

Flow Gator Handheld DAQ System

Flow Gator Master Meter System

The Flow Gator is capable of controlling an extended flow range manifold and will automatically select the proper master meter to compare with the UUT.

The Flow Gator is a fully contained data acquisition system, which includes signal conditioning. Use by the U.S. Armed Forces to calibrate flowmeters onboard ships and in aircraft test facilities gives credence to its capability and ease of use. Meets SAE ARP4990 Calculation Standards.

ADVANTAGES
NIST traceable SAE ARP 4990 conformance Works with all types of master meters Performs automatic viscosity and density correction Handheld data acquisition system, including all sensor interface amplifiers No need for a PC in the field Stores up to 5 master meters and 5 fluids User-friendly, easily programmable Includes optional manifold control capability

PERFORMANCE SPECIFICATIONS
System Uncertainty Frequency Meter Inputs Analog Meter Inputs Display Linearity (with linearizer) Operating Temperature Engineering Units Temperature Units 0.15% of reading (Depending on Master Meter type and calibration) 0.001% of reading Accurate to 25 nanoseconds 0.02% of reading Accurate to 16-bit resolution 4 Lines x 20 Characters 0.01% of reading -40F to 140F (-40C to 60C) GPM, LPM, PPM, PPH F,C, R, K
23

PROVEN PERFORMANCE FOR OVER 50 YEARS PERFORMANCE SPECIFICATIONS

FLOW DYNAMICS PRIMARY STANDARD POSITIVE DISPLACEMENT FLOW CALIBRATORS


Flow Dynamics, part of the COX Flow Measurement family, is known for the industrys most advanced primary standard positive displacement flow calibrators. These liquid flow calibrators are the equipment-of-choice across a wide range of industries, and are in use throughout the world in the most prestigious government primary standard laboratories. The Flow Dynamics Positive Displacement Calibrator, Liquid (PDCL) 10, 60 and 400 Series calibrators perform NIST traceable calibrations ranging from 0.01 GPM (0.04 LPM) to 400 GPM (1514 LPM) with an uncertainty of 0.05% and a repeatability of 0.01% of reading. They are compatible with most common flowmeters providing a visual, pulse, or analog output. The PDCL Series is a precision volumetric device displacing an exact known volume of fluid passing through the flowmeter under test, while compensating for temperature effects on viscosity. The PDCLs low fluid usage supports the need to reduce hazardous fluids, while keeping fluid costs to a minimum. Blending fluids is made easy with an automatic cycling feature for mixing fluid while in the machine. In addition, the calibrators piping loop is designed to assist draining fluids out of the system and facilitate flushing to clean residual fluid out of the piping. Flo-Cal User Interface: The PDCL Series employs Flo-Cal software, which is seamlessly compatible with Windows operating systems, to graph and produce calibration data reports. The software graphics display visual aids calibration data/plots, and the data reports are user-configurable. Calibration files are automatically stored for flowmeter performance history and can be recalled for plot comparison to the current calibration. Field Service: In addition to manufacturing new primary standard calibration systems, COX Flow Measurement offers field certification and service worldwide. All calibrators are 100% documented to ensure service support for the life of the calibrator.

The PDCL Series Primary Standard Calibrators are used daily in our prestigious NVLAP calibration lab. As a user, we continually implement time-saving features such as fluid filling, mixing and viscosity temperature control, while improving on operator efficiency and ensuring minimal maintenance.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 24

PRIMARY STANDARD POSITIVE DISPLACEMENT FLOW CALIBRATORS ADVANTAGES


High precision measurement capability Heater/Chiller to control fluid viscosity Calibrations traceable to national standards Same calibrator design as used by NIST User-friendly software allows easy operation Provides automatic data acquisition Low fluid usage reduces costs User features to reduce calibration time

PERFORMANCE SPECIFICATIONS
Measurement Range Uncertainty Repeatability Temperature Reading Viscosity Range Ambient Operating Temperature Fluid Operating Temperature Operating Pressure 0.01 to 400 GPM 0.05% of reading 0.01% of reading 0.02F of reading 0.5 to 1000 Centistokes 60F to 90F (15.5C to 32.2C) 40F to 140F (4.4C to 60C) Up to 120 psig

Operator enhancements incorporated in the PDCL Series calibrators

Integral Data Acquisition System

3 Valve Manifold Flow Adjustments

Heater & Chiller Fluid Viscosity Control

Servo Driven Telescope Tube

Auto Pneumatic Air Bleed Valves

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 25

OEM SEMI-AUTOMATIC CALIBRATORS

Semi-automatic calibrators are available for OEMs with high-volume calibration requirements. These systems incorporate flowmeter quick-load fixturing. CALIBRATOR CHECK STANDARDS
Check Standards are used to verify the repeatability of a primary standard calibrator. They ensure that nothing has changed in the ability of the calibrator to repeat its former calibration data. Calibration data curves should lay over each other within <0.1% over the entire range of the Check Standard (30:1 turndown). Check Standard sets are offered with every PDCL Series calibrator.

COX Flow Measurement Check Standards are used by NIST in their intra-laboratory test to determine bias between primary standard calibration labs for the armed forces and major labs.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 26

PROVEN PERFORMANCE FOR OVER 50 YEARS

MANUFACTURING CELLS
COX Flow Measurement recognizes the need for 100% product performance when manufacturing flow measurement instruments for OEM applications. Our world-class flowmeter production line is compliant with the ISO/TS-16949 automotive industry standard. We have implemented a comprehensive quality documentation system addressing every process required to economically produce the highest quality flow measurement products. COX Flow Measurements OEM manufacturing cell capability provides material and assembly traceability on each flowmeter delivered, encompassing total quality from raw materials to finished products. The end-result drives product confidence, eliminating liability and potential costly replacements, while minimizing product cost. ADVANTAGES Compliant with ISO/TS16949 APQP includes: Production Part Approval Process (PPAP) Process Failure Modes Effects Analysis (PFMEA) Design Failure Mode Effects Analysis (DFMEA) Six Sigma-level quality standards

COX Flow Measurement invites you to evaluate the cost of manufacturing your flowmeter in one of our world-class cells. High volume calibrations are also provided for flowmeter manufacturers with timely deliveries you can count on.

COX Flow Measurement application engineers will help you configure the proper system for your application. They can be reached at 480-922-7446 or by email at sales@cox-instruments.com. Application technical tips are also available at www.coxflow.com
Proven Performance for Over 50 Years 27

COX Flow Measurement, based in Scottsdale, Arizona USA, offers one of the industrys largest selections of flow measurement products and services. Customers are also encouraged to take advantage of our world-class applications engineering assistance. We have the expertise to help you solve the most difficult liquid and gas flow metering applications! COX precision flowmeters are recognized for their uncompromised accuracy and reliability. All of our products are manufactured in compliance with AS9100, ISO 9001 and other strict industry quality standards. Most COX flowmeters are delivered in 4 weeks or less, and many models are in stock. We also offer expedited deliveries, depending on the complexity of the calibration. If you demand superior flowmeter accuracy and reliability, choose COX!

Proven Performance for Over 50 Years

(480) 922-7446
www.coxflow.com

15555 North 79th Place Scottsdale, AZ 85260 Tel: (480) 922-7446 Fax: (480) 948-3610 sales@cox-instruments.com www.coxflow.com

Manufactured under an ISO 9001 and AS9100 registered quality management system.

FM78587 AS9100 and ISO 9001

2011 COX Flow Measurement, Inc.