Your signature indicates that you have completely read the entire report and agree with everything here in. Failure to sign will result in a zero for your personal grade unless a formal exception is filed with your TA. Please Print and Sign Full Name Principal investigator: Skeptic: Researcher TA Ryan Samons_________________________________________ Morgan Culver_________________________________________ Logan Murphy_________________________________________ Justin Woods__________________________________________
Role
DC
AD
RC
Q1
Q2
PI
PG
Legend:
I introduction DC data and calculation AD analysis and discussion RC results and conclusion Q1/Q2 quiz/prelab PI principal investigator points PG personal grade (average of individual score and Lab Report Score)
Page 2
Introduction:
After familiarizing ourselves with equipment during lab one, lab two focuses on learning how to use and becoming familiar with the oscilloscope. We learned how to look at the data displayed on the oscilloscopes screen and interpret that data to find other measurements. From this experiment will verify ohms law, and confirm that there are different ways to calculate voltages but ohms law works for all the methods and the values calculated are not significantly different between the methods used. Equipment: 1 Digital Multimeter 91910044 1 Oscilloscope C050155 1 DC power supply C1010444 2 AC power supply (function generator) FG24243,FG24241 1 10-k resistor 5 Banana clip wires 2 BNC cables
Post Lab Meetings: Monday February 24th there was a meeting in the lab to finish and finalize the data from the lab. Morgan and Logan were in attendance. Ryan had a class at the time of the office hours were the lab was completed therefore he was unfortunately unable to attend. The meeting lasted from approximately 11:00 am to 11:40 am. Thursday February 27th there was a meeting during the lab time. We discussed how we were going to proceed on the lab and discussed the results and comments given by the TA. This meeting lasted approximately 30 minutes. References: Ellis, Steven L. Laboratory Manual Spring 2013-2014 Appendix Fa - Tektronix_PWS4305 Appendix Fb - Tektronix_PWS4305 Appendix H- Tenma function generator 72-7210 Appendix K - Digital Multimeter Appendix S - Tenma Function Generator72 7210 Manual Appendix J Error propagation Appendix C TDS 100B Oscilloscopes users manual Standard Cover Sheet
Page 3
Procedure:
The research team began the experiment by reviewing the lab manual and trying to learn the general procedures of the lab before moving any further. Next, Ryan Samons and Morgan Culver retrieved the required equipment from the filing cabinet and Logan Murphy recorded the serial numbers of the equipment. Table 1 The PI Ryan Samons referenced the lab manual for the proper setup of the equipment for this part of the experiment. By placing cables between the oscilloscope and the DC power supply we were able to complete the circuit. These banana clips were ran through the resistor. The Digital Multimeter was then placed where it would read the voltage running through the system. At this point the DMM Voltage heading of Table 1 was completed by reading the DMM. Next the oscilloscope was used to record the scale and divisions of the data given in order to find the voltage reading on that device. These steps were repeated for .5V,1V,2V,4V, and 8V. The percent difference were then calculated by Skeptic Morgan Culver and recorded in Table 1. The diagram below shows the setup for this part of the experiment.
Table 2 For this part of the experiment the DC power supply was replaced with the function generator in order to find the same measurements using an AC power supply. All group members helped in switching these devices. The connection between the oscilloscope and function generator was made using BNC to banana clip cables. The TA Mr. Wood was asked by Logan Murphy to check the setup before continuing.
Page 4
The DMM readings were taken in the same manner as before, with the DMM being placed on top of the banana wires in the experiment in order to accurately record the data. Next the oscilloscope was used to find the divisions and scale of the data given in order to calculate the voltage readings for this part of the experiment. The percent difference was then calculated by the Skeptic Morgan Culver and recorded along with the other data The table below shows the setup for this part of the experiment.
Table 3 The next part of the experiment focused on using the function generator and seeing the effect the shape of the wave and the frequency strength had on the voltage given off. The same setup from table 2 was used in this part of the experiment also. Mr. Wood was asked by Morgan Culver to check the setup before preceding. First the data was recorded for all of the sine, triangle, and square waves at the 400Hz values, then the 4000Hz value, and the 40000Hz value. The divisions and scale factor data were gathered by Ryan Samons and Logan Murphy from the screen of the Oscilloscope itself. The divisions were the number of lines in between the two points needed. All the data was then recorded by the Skeptic Morgan Culver. The table below shows the setup for this part of the experiment.
Table 4
Page 5
The last part of the experiment involved 2 function generators. The two function generators were connected to the oscilloscope. One in channel one and the other in channel two. Mr. Wood was asked by Ryan Samons to check the setup before preceding. Both were turned on and the frequency values set to about equal, one double the other, and one triple the other, with the data being recorded along the way. The data recorded were the number of loops along the Lissajous pattern. After finding the display button we were able to find the proper display to show the data in this form. The number of loops were counted in the horizontal and vertical directions. The diagram below shows the setup for this part of the experiment.
Although the percent difference equation does not apply here because the data was exactly the same the equation for percent difference will be used throughout the experiment and thus will be referenced in this section.
Page 6
Equation 1
Below is a sample equation for Equation 1. Since the values from the table above all give the % to be zero, the sample calculation will be if the values were 4V and 4.2V.
This next equation applies to all the numbers underneath the oscilloscope heading in the charts throughout the experiment. In order to find the DC voltage you must multiply the scale by the divisions. Equation 2
Table 2 This part of the lab used the function generator to produce current using a sine wave and was measured both with the DMM and the oscilloscope. Before comparing the data the DMM voltage reading had to be converted to the peak voltage readings and the divisions and Scale had to be found on the oscilloscope. The data for this part of the experiment is shown below.
Function Generator min 0 db mid 0 db max 0 db mid -20 db max -20 db Oscilloscope Divisions 5 6.4 4.6 6.4 4.6 Scale .1V 1V 5V .1V .5V V(pp) .5V 6.4V 23V .64V 2.3V Digital Multimeter V(rms) .1738V 2.21V 7.68V .22V .795V Digital Multimeter V(pp) .491V 6.25V 21.73V .62V 2.25V %difference 1.81 2.37 5.68 3.17 2.2
There a few equations needed for the completion of the table above. The first is the conversion of the Digital Multimeter V(rms) into the V(pp). This equation is listed below. Equation 3
Page 7
Sample equation for equation 2 will be used at the mid 0 db row of calculations above.
The percent difference for this part of the experiment was found using Equation 1 from earlier in the experiment. The V(pp) underneath the oscilloscope was calculated by simply multiplying the divisions and the scale together. Also under the oscilloscope in the table there is a divisions and scale column. Using Equation 2 you find the values for the next column in the table. Equation 4
Table 3 In this part of the experiment the function generator was used to find the changes sine, triangular, and square waves affected the frequency readings. The table below shows the data gathered from this section of the lab.
Table 3
Wave Function form generato type r frequenc y
Sine Sine Sine Triangl e Triangl e Triangl e Square Square Square 400 Hz 4000 Hz 40000 Hz 400 Hz 4000 Hz 40000 Hz 400 Hz 4000 Hz 40000 Hz
Volt s
.32 .33 .31 .37 .365 .375 .55 .54 .45
Sec.
% diff in freq.
The percent difference is found in this table by once again using Equation 1. The two volts categories under amplitude and peak to peak were calculated by taking the divisions and multiplying them by the scale factor. Another equation used in this chart is Equation 2. This is
Page 8
used by multiplying the divisions and scales in order to get the values. This just leaves the calculated frequency. Calculating frequency has the following equation. Equation 5
The sample calculation for this equation will be using the information in the square 40000Hz row.
Table 4 The last part of the experiment involved using two function generators changing the frequency of one of them and creating Lissajous Pattern. We had to set up the oscilloscope in the XY mode and looked at the ratio of vertical loops to horizontal loops. The chart below was created after looking at different ratios of frequencies given off by the function generator.
The picture below shows the oscilloscope at the time when both function generators were set at 400Hz. As you can see there is one vertical and one horizontal loop.
Page 9
The next picture shows the oscilloscope when one of the function generators is at double the frequency of the other. As you can see there is two horizontal loops and one vertical loop.
The last ratio tested was when one function generator was 3 times the other. As you can see there is three horizontal loops and one vertical loop.
Page 10
Page 11
where: Vps is the error in the power supply output measured in volts : Vps is the voltage setting on the power supply : [2] is the number added to the smallest significant digit of the voltage of the power supply. Below is a sample equation of the error for the Power Supply using the voltage at 0V.
By using this equation, the error regarding the power supply can be found for each individual measurement taken. These errors can be found in Errors in Table 1. The systematic error in the digital multi-meter can also be found using a similar equation show below,
where: V is the digital multi-meter voltage measurement. : is the total error in the power supply
Page 12
Because the digital multi-meter is measuring the power supply voltage, the error in the digital multi-meter will also include the error in the power supply as seen above. Below is a sample equation of the error for the DMM device using the voltage at 0V.
The error in the oscilloscope will also affect the measurements found in Data Table 1. The error regarding this piece of equipment is more complex than the previous two. The total error in the DC voltage measurements taken has three different components. The first two are the error in the number of divisions and the error in the scale factor. The error in the divisions was given as 0.1 because it is designated as half of the smallest possible measurement. The error in the scale factor will be as 0.3% of the scale factor. In order to find the total error in DC measurement, these two values are multiplied together. Because of this, the total error must be propagated using the following equation:
This equation can be simplified, however, to an even simpler equation show below. The partial derivative of the voltage with respect to the divisions is just S. The partial derivative of the voltage with respect to the scale will be D. These can be substituted into the equation and then it be further simplified,
where:V is the error in the voltage : S is the scale factor : D is the error in the divisions : D is the number of divisions : S is the error in the scale
Similar to the digital multi-meter, the error in the oscilloscope will also inherit the error in the power supply. Because of this, the total error in the voltage measured by the oscilloscope can be found using the equation below.
Below is a sample calculation where S is .2V with an uncertainty of .006, a division of 2.5 and an uncertainty of .1, and the error in the power supply is .045V.
Page 13
Error in Table 1 Error in DMM Error in Voltage Divisions Reading .07 .1 .11 .1 .19 .1 .35 .1 .67 .1
Oscilloscope Error in Scale .0006 .0015 .003 .006 .015 Error in DC Voltage .065 .12 .22 .42 .92
Page 14
: Vrms is the original measurement : [2] is the number added to the smallest significant figure. Below is a sample calculation where the peak to peak voltage is .1738V.
The other measurable error for the digital multi-meter is the error in measuring peak to peak voltage. This error can be found using a simple process. Because the peak to peak voltage is found by multiplying the root mean square measurement by 22, the error in peak to peak voltage is simply: = Below is a sample calculation with Vrms is *2 . =.0037* 2 =.0105 The measurements taken using the oscilloscope will also have a certain amount of measurable error. These can be found by solving the following equation for peak to peak voltage.
In this equation the variables represent the same values that they did for the error in the oscilloscope measurement in the previous section. A sample calculation using this equation can be seen below where the function generator is set to min 0 db, the scale is .1V, the uncertainty in divisions is .1, the number of divisions is 5 and the uncertainty in the scale is .0003V.
Page 15
Error in Divisions .1 .1 .1 .1 .1
Error in Table 2 Oscilloscope Error in Error in Vpp Scale .0003 .01 .003 .1 .015 .5 .0003 .01 .0015 .05
where: S is the seconds : D is the error in the divisions : D is the number of divisions : S is the error in the seconds A sample calculation can be seen below where the scale factor is .001s, the error in the divisions is .1, the number of divisions is 2.46, and
Page 16
The error in frequency can also be calculated using a similar equation. The main difference is now the derivative of frequency is taken with respect to the divisions and the scale.
Because seconds is equal to the divisions multiplied by the scale, this equation can be rewritten by replacing the two partial derivatives. The equation for error in frequency is now:
A sample calculation using this equation is shown below where the number of divisions in 2.46, the scale is .001, the error in divisions in .1, and the error in the scale is 3.0E-6.
There will also be a measurable error in volts in amplitude for this section of the lab. This error can be calculated using the equation below.
Page 17
This equation is the same one used to calculate the total error in voltage in part 2 of the lab. A sample equation can be seen below.
The error in peak to peak amplitude can be found using the same equation seen in part 3 of the lab. = *2
A sample calculation regarding this section of the lab can be seen below. = *2
=.01* 2 Vpp=.028
Error in Table 3
Waveform type Function generator frequency Amplitude Error in Div Error in Scale factor r (V)
.0003 .0003 .0003 .0003
Error in Volts
Period Error in Volts Error in Div Error in Scale factor r (ms) .003 .0003 .0000 3 .003 Error in Sec.
.1 .1 .1 .1
.1 .1 .1 .1
Page 18
Triangle Triangle Square Square Square 4000 Hz 40000 Hz 400 Hz 4000 Hz 40000 Hz .1 .1 .1 .1 .1 .0003 .0003 .0006 .0006 .0006
.1 .1 .1 .1 .1
.1 .1 .1 .1 .1
Page 19
DMM vs Oscilloscope (2 V)
DMM 2 0.19V Oscilloscope 2 0.22V 0 0.5 1 1.5 2 2.5
We can see from the number line that the DMM and Oscilloscope voltage overlap (they were stacked to aid viewing). This shows that these instruments can be trusted to produce an accurate value for the voltage. Now the AC voltage is measured. The AC voltage is provided by the function generator. Below is a number line that compares the oscilloscope, and the DMM measurements for the voltage. The number line is based on the 400 Hz and 0 db setting for the function generator. The errors are taken from the Error in Table 2 table from the skeptic. Number Line 2:
Page 20
Here we can see that the DMM and the Oscilloscope measurements for voltage overlap. The percent difference between the two measurements was 1.81%. This, plus the overlapping error bars, shows that the oscilloscope and the DMM can be used to accurately determine the voltage. We can conclude because these values overlap, as well as the overlap from Number Line 1, that the adjusting of the DC power supply and the Function generator using only the front control knobs can yield a precise output from either unit. This next section compares the different frequencies of the function generator and that of the oscilloscope. The uncertainty in the frequency could be calculated by defining the uncertainty to be that of the error of the inverse of the period. The first number line in this section will be that of the sin at 400 Hz. The frequency measured by the oscilloscope is the one to trust in this situation because setting the frequency on the function generator is difficult because of the sensitivity of the knob on the front of the unit. It is also important to note that one should try to fit the entire period onto the screen of the oscilloscope to allow for a more accurate measurement. If the entire period was not on the screen, it would be difficult to determine where the period began and ended. The errors came from the Error in Table 3 table in from the skeptic. Below is the comparison for the sin wave at 400 Hz vs. the function generator at 400 Hz. Number Line 3:
We can see that the error bounds on the oscilloscope's measurements overlapped the frequency of the function generator. This shows that the oscilloscope can produce an accurate measurement for the frequency from the function generator; however, it may not be a precise measurement. Below is the comparison for the sin wave at 4000 Hz vs. the function generator at 4000 Hz.
Page 21
Number Line 4:
3600
3700
3800
3900
4000
4100
Because the oscilloscope's measurement and errors overlapped, we can draw the same conclusion as we did from number line 4, that is, that the oscilloscope can return an accurate measurement of the frequency. One thing to notice is that the measured frequency of 3846.15 Hz is lower than the frequency of 4000 Hz we were supposed to compare the oscilloscope to. This is probably due to a random error or a mistake in setting the function generator's frequency. The random errors and systematic errors were discussed in the Skeptic's portion of this paper. Below is the comparison for the sin wave at 40000 Hz vs. the function generator at 40000 Hz. Number Line 5:
38000
39000
40000
41000
42000
We can see again that the oscilloscope's measurement and errors overlap the frequency of the function generator. Based on this, we can again conclude that the oscilloscope will produce an accurate measurement of the frequency.
Conclusion:
In this lab, we learned how to use the oscilloscope and accompanying software, although not required, to determine voltage and frequency. We learned about using the oscilloscope to measure AC and DC voltages. We learned how to measure the period and frequency of a signal using the oscilloscope, and we learned how to make a Lissajous figure using two AC voltage input signals. This lab also introduced us to the oscilloscope, which allows us to make a more indepth analysis of electrical signals for future labs.