DD+DIS378.

05E

Controls, Connections, and Setup Procedures

2.1

Checking Installation Site Prerequisites


Check that all site prerequisites shown in the Installation Planning-Checklist (can be found at the end of Chapter 11, CR 85-X Service Manual) are fulfilled before visiting the site.

2.2
2.2.1

Required Service Programs


CCM-Tool The CCM-Tool is needed to create and modify the configuration file (adc.cpf) for the CR 85-X and the other CR system components. The latest version of the CCM-Tool is installed by default on the NX processing station (Go to Start Agfa Service CCM Tools). If not yet done create and/or adapt the configuration file adc.cpf together with the application specialist. NOTE: When integrating the digitizer into an existing network, it is strongly recommended to create a CPF file beforehand (see CR 85-X Service Manual, Chapter 11, Installation Planning, checklist). For detailed information on the configuration file, see: ADC Application Manual - Chapter 11, DD+DIS181.03E.

2.2.2

Error Message Program The "SHOW ERROR PROGRAM" with the "ERROR.MSG file" translates the CR 85-X error codes into clear text. The "SHOW ERROR PROGRAM" is a part of IMOS, the "ERROR.MSG file" is loaded in the digitizer.

DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 2, Revision 1 06-2008

CR 85-X Type 5148/100 Chapter 1 / 7 CR 75.0 Type 5146/105 (as of SN 6000) Agfa Company Confidential