This document summarizes an experiment measuring the volume of water using different instruments to determine the most accurate and precise. A group of students used a 50mL beaker, 10mL graduated cylinder, 10mL pipette, and 1mL pipette to measure 30mL of water three times each. The 50mL beaker was the least accurate and precise, while the 10mL graduated cylinder was the most precise and the 10mL pipette was the most accurate based on the averages and standard deviations calculated. Sources of error were likely human errors in using the instruments and properties of water.
This document summarizes an experiment measuring the volume of water using different instruments to determine the most accurate and precise. A group of students used a 50mL beaker, 10mL graduated cylinder, 10mL pipette, and 1mL pipette to measure 30mL of water three times each. The 50mL beaker was the least accurate and precise, while the 10mL graduated cylinder was the most precise and the 10mL pipette was the most accurate based on the averages and standard deviations calculated. Sources of error were likely human errors in using the instruments and properties of water.
This document summarizes an experiment measuring the volume of water using different instruments to determine the most accurate and precise. A group of students used a 50mL beaker, 10mL graduated cylinder, 10mL pipette, and 1mL pipette to measure 30mL of water three times each. The 50mL beaker was the least accurate and precise, while the 10mL graduated cylinder was the most precise and the 10mL pipette was the most accurate based on the averages and standard deviations calculated. Sources of error were likely human errors in using the instruments and properties of water.
ArminaCasildo measured volume of water using 1 mL pipette, computed for the average and standard deviation, contributed in conclusion Ralph Dela Cruz measured volume of water using 1mL pipette, computed for the average and standard deviation, contributed in introduction and recommendation Jason Lumabas measured volume of water using 10mL pipette, contributed in results and discussion & conclusion, created the graphical representation Gary Tomwong measured volume of water using beaker and graduated cylinder, took pictures of the experiment, contributed in methods and conclusion
Introduction
Two important concepts that would tell that the result of an experiment is correct and reliable are accuracy and precision. Measurements are never identical even if measuring the same individual since the results will somehow be dispersed. It will always be a subject to error until making enough trials to come up with an average or mean of all the measurements. According to Bell (1999), accuracy describes the closeness of agreement between the measurements to the true value. Basically, accuracy is how close the average of the measurement is to the true value. Precision, on the other hand, is the degree of exactness of a measurement (Zitzewits, 2013). Exactness meaning how close the measurements is to each other after several trials. Using the concept of accuracy and precision can also be limited because of several instruments and methods that is why the concept itself can be judged using simple techniques. Comparing an instrument with another instrument is a useful technique in determining how accurate is an instrument while repeating the measurement for several trials can support the precision of a measurement. Basic knowledge about several laboratory equipment like beakers, graduated cylinder and pipettes is needed before doing any thesis work or laboratory experiment to know which instrument to use to get the most reliable results. The concept of accuracy and precision must be observed to back up or support any experiment to lessen the subject for error.
The experiment is done in order to:
Objectives 1. To determine which instrument is best used for getting the most accurate and most precise measurement 2. To verify if the true value agrees with the average of the individual measurements of the most accurate and most precise instrument
Expected Outcomes 1. The 1mL pipette would yield the most precise and most accurate measurement 2. The 50mL beaker would yield the least precise and least accurate measurement
II. Methodology
The group first checked the equipment given for cracks or any damages to the instruments. The 100ml beaker was then weighed once and the reading obtained was initial weight of the beaker. The group wiped the 100ml beaker with a laboratory gown for each trial. The other beaker was used for obtaining tap water. Four instruments were tested for precision and accuracy. The first one was the 50ml beaker. 30ml tap water was obtained using the 50ml beaker and then transferred to the 100ml beaker. The beaker is weighed and recorded after. This procedure was done three times. The next instrument tested was the 10ml graduated cylinder. 10ml of tap water was measured using the graduated cylinder and then transferred to the 50ml beaker. This was done two more times to get the target volume of 30ml. The beaker is weighed and recorded after. The whole procedure is done three times. The 10ml pipette was the next instrument tested. It was used to measure 10ml of tap water, transferred the sample to the beaker and repeated two more times to get a total volume of 30ml. The beaker is weighed and recorded after. The whole procedure was done three times. The last instrument tested was the 1ml pipette. 1ml of tap water was obtained, transferred the water to the beaker and was repeated twenty nine more times to get a total volume of 30ml. The beaker is weighed and recorded after. The whole procedure was done three times.
III. Results
The following data were obtained from the experiment: Based on Table 1, the 50mL beaker showed the average (27.85) farthest to the true value which is 30mL. Beaker was then followed by 1mL pipette (28.75), 10mL graduated cylinder (29.95) and 10mL pipette (29.95). The instrument that presented the average closest to the true value is the most accurate instrument. In this experiment, the most accurate instrument is the 10mL pipette while the least accurate instrument is the 50mL beaker. It can be seen that basing at the standard deviation in Table 1, the instrument that showed the least precision is the 50mL because it has the highest standard deviation (1.01). It is followed by the 1mL pipette with the standard deviation of 0.40, then the 10mL pipette with 0.01 and the instrument that presented the most precise measurement is 10mL graduated cylinder with the lowest standard deviation of 0.0071. This data is also presented graphically in Figure 1.
Figure 1. Graphical representation of average volume of instruments used
The standard deviation of each measurement is presented in Figure 1 by the error bars. It can be noticed that the 50mL beaker has a relatively long error bars compared to the others which means the measurements of the beaker is very dispersed.
IV. Discussion
Getting precise and accurate results is important in scientific studies in order to get reliable results and be able to defend those results properly. This experiment showed students the concept and more understanding on how to get accurate and precise results while also teaching them how to limit errors. Based on Table 1, the results of each trial was recorded in 2 decimal places because the group applied the rule of uncertainty since the pointer was not directly pointing at the graduation in the scale. The group estimated the last decimal place since the pointer was pointing to the space between one graduation to another. The results also showed that the expected outcome for the least precise and least accurate instrument of the group is congruent but it showed that the most accurate and most precise instrument is not congruent with each other. To understand why the expected outcome was not congruent with the final results the group researched more about the instruments that was used.According to Cady (n.d.), a person using a beaker must allot a room for plus or minus 5% error since it is not designed for accuracy or precision. The instrument is commonly used to store liquids during experiments and not to get accurate and precise results. Since precision is the degree of exactness of a measurement, an instrument with the finest division on its scale usually produces the most precise measurement. (Zitzewits, 2013). This is another reason why the beaker can not guarantee precise results because its graduation is too far from each other.
Both the graduated cylinder and the pipette possess divisions that is closer to each other that is why it gathered more precise results than the beaker. According to Cady (n.d.), pipettes are made for accuracy and precision because it is commonly used in quantity sensitive studies and experiments. There were many errors that lead to the groups results and most of them were human errors since the 1mL pipette should have presented the most accurate and most precise results. Parallax errors, spillage of the tap water and wrong scale reading could have affected the results and the people conducting the experiment caused this. The tiredness from too much repetition of getting the target volume of tap water also increased the risk of committing human errors. The procedure to the get the target volume which is 30mL is tiresome since the instrument limits its maximum capacity to a certain measurement. For example, to obtain the target amount of water using the 1mL pipette, the person conducting the experiment must repeatedly use the aspirator 90 times in total to get 30mL. Another possible source of error is the use of tap water itself. Water is known to have a property of adhesion.Adhesion is defined as the process of attachment of a substance to the surface of another substance (Vert et al., 2012). The tap water may not be transferred completely to the pre weighed beaker and there was still residue in the instrumentthat caused errors in the measurement.
Based on the possible sources of error, the result was most likely affected by errors committed by the group members since most groups in the class presented results that are congruent to our expected outcome.
V. Conclusion The group concludes that the least precise and least accurate instrument is the 50mL beaker. The most precise instrument, on the other hand, is the 10mL graduated cylinder and the most accurate is the 10mL pipette. Theoretically, the most accurate and precise results would be the 1mL pipette because its commonly use to measure volumes in small scales that is why it possess fine graduations but because of errors committed by the group the measurements were affected greatly. The group recommends taking several samples while randomizing the order of the instruments that will be used to eliminate bias of being under the same condition while measuring.
VI. Literature Cited Bell, S. (1999). A beginners guide to uncertainty of measurement. Retrieved September 13, 2014, from https://www.wmo.int/pages/prog/gcos/ documents/gruanmanual/UK_NPL/mgpg11.pdf
Cady, S (n.d). Measurements. Retrieved September 13, 2014 at http://www.esu.edu/~scady/measurement.shtml
Vert, M., Doi, V. Hellwich, K., Hess, M., Hodge P., Kubisa, P., Rinaudo, M., Schue, F. (2012). Terminology for biorelated polymers and applications. Pure and Applied Chemistry Journal, 84(2), 377-410.doi:10.1351/PAC-REC-10-12-04
Zitzewitz, P., Haase, D., Harper, K. (2012). Physics principles and problems. Retrieved September 13, 2014, from https://archive.org/details/PhysicsPrinciplesAndProblemsByAGlencoeProgram