ApplicationsNote:38
Technique:Spectrophotometry
ApplicationAreas:Clinical
Determinationofglucoseinclinicalsamples
Introduction
Glucoselevelsinphysiologicalfluidsareincreasedindiabetesmellitusandin
conditionsofthyroid,pituitaryoradrenalhyperactivity.Lowlevelsarecaused
byinsulinoverdosage,insulinsecretingtmours,hypoadrenalismandglucose
maladsorption.Thenormalrangeinplasma(fasting)is75-115mg/100ml.
Principle
Thephysiologicalfluidsampleisaddeddirectlytoacuvettecontainingthe
reagentatroomtemperature(20-25C),andtheglucoseconcentrationisread
directlyfromthedisplayorprintedout.Thismakesitsuitabletechniquefor
singleassays.
Useofaspectrophotometermakesthisanalysisstraightforwardandfeasible
foruseinnon-laboratoryenvironment.
Method
Thereagentcanbepreparedreadyforuseorobtainedasakit(RandoxLabs.
UK)
Pipette1mlofreagentcontaining:
GlucoseoxidasekU/l
Peroxidase1.55kU/l
4-aminophenazone0.25mmol/l
Phenol11mmol/l
Phosphatebuffer0.1mmol/l,pH7.0
intoeachof2disposablecells(80-2004-53).
.
Spectrophotometry Page 2 of 3 Application Note: 38
LibraS21/S22operation
SelectBasicModes(1)
SelectConcentration
SelectWavelength=500pressOK(F3)
EnterFactor324.1pressOK(F3)
Add0.01mlofsampleand0.01mlwater,fortheblankanalysis,respectively
andmix.
After25minutesat20-25
o
C:
Insertblankandpressgreenrunkey.
o Asingleblankasabovesufficesforsubsequentanalysesinthe
sameseries.
Insertsampleandpressgreenrunkey.
Theanalysisislinearovertherange2-500mg/100ml.
(Forassayswherethereisnoestablishedconcentrationfactor,calibration
shouldbecarriedoutusingpreparedstandards.)
PrintoutsmaybeobtainedbysettinguptheprinteroptionsinSystemUtilities
andPreferences(3)printsresults.Thisisautomaticwithautoprint.
Additionallypress.toprintresultifauto-printisoff,ortore-printresultifauto-
printison
TheaboveprocedurecanbeeasilyusedwithotherinstrumentsintheLibra
range
Spectrophotometry Page 3 of 3 Application Note: 38
OrderingDetails
LibraS5 80-2115-00
LibraS11 80-2115-15
LibraS12 80-2115-10
LibraS21 80-2115-25
LibraS22 80-2115-20
Libras32 80-2115-30
Thereactioncanbeacceleratedforincreasedsensitivityifwarmed.Forthis
purposetheLibraS21/S22havethefollowingaccessories:
8positionwaterheatedcellchanger(80-2109-70)usedwithanexternal
heatingbath
6positionPeltierheatedcellchanger(80-2106-04)andTemperature
ControlUnit(80-2112-49)
Singlepositionwaterheatedcellholder(80-2106-08)usedwithanexternal
heatingbath
Singlepositionelectricalcellholder(80-2106-12),temperaturesselectable
from25,30and37
o
C
SinglepositionPeltiercellholder,temperaturesselectableoverthewhole
rangefrom20-49
o
C(80-2106-13).
TheSipper(80-2112-25)enablessomeautomationoftheanalyses,andcan
beusedtogetherwithaheated(notwaterheated)ornon-heatedsinglecell
holder.