Download as pdf or txt
Download as pdf or txt
You are on page 1of 76

CewePrometer

User Manual
Ver. 1.0

Contents
Introduction........................................................................................................................5
About this user manual.................................................................................................5
Contacting us................................................................................................................5
Product description ...........................................................................................................6
Sealing of CewePrometer-W........................................................................................9
Sealing of CewePrometer-R.......................................................................................10
Connections ................................................................................................................10
Mechanical design......................................................................................................13
Function modules .......................................................................................................16
Measuring principles ..................................................................................................17
Configuration, reading and maintenance ......................................................................18
Connecting to CewePrometer.....................................................................................19
Basic configuration.....................................................................................................20
Overview of functions ................................................................................................22
Changing configuration..............................................................................................24
Working with configurations......................................................................................25
Reading.......................................................................................................................25
Information about the meter .......................................................................................26
Versions and version conflicts ...................................................................................27
Updating firmware .....................................................................................................27
Language ....................................................................................................................28
Resetting registers and logs........................................................................................28
Functions..........................................................................................................................29
Meter clock.................................................................................................................29
Energy registers..........................................................................................................30
Instant values..............................................................................................................31
Digital inputs and outputs...........................................................................................33
Display sequences ......................................................................................................36
Communications and security ....................................................................................37
Loggers.......................................................................................................................39
Alarms ........................................................................................................................41
Event log.....................................................................................................................43
Maximum demand......................................................................................................44
Historical registers......................................................................................................45
Time of use.................................................................................................................46
Transformer compensation.........................................................................................48
3

Contents
Power quality..............................................................................................................50
Miscellaneous.............................................................................................................52
Using the display .............................................................................................................53
Voltage order display layout ......................................................................................53
Navigating in display sequences ................................................................................54
Display layouts with active choices ...........................................................................54
Using the display for communication diagnostics .....................................................56
Appendix A Display Layouts .......................................................................................57
Appendix B - Events ........................................................................................................61
Appendix C Serial Communication Port RS232/422 ..................................................63
Appendix D - Module block diagram ..............................................................................65
Appendix E Frequently Asked Questions...................................................................66
Appendix F Calculation Principles ..............................................................................67
Appendix G Material Declaration.................................................................................72
Appendix H Connection Data ......................................................................................73

Introduction
Introduction
Thank you for choosing the CewePrometer.
The CewesPrometer is a multi-functional, electronic, electricity energy meter. It provides very high
accuracy in measuring electrical energy and also in measuring instantaneous quantities such as
current, voltage, power, frequency etc. CewePrometers extensive configurable functional features
together with the high accuracy enable application areas more numerous than for traditional electric
meters. Besides having well-designed traditional features such as meter registers, data logging and
tariff capability, there are also extensive functions for monitoring alarm conditions, harmonics,
measuring sags and swells and other power quality quantities.

About this user manual


This user manual describes the CewePrometers functions and provides the information needed to
configure and use the meter. The manual covers all versions of the CewePrometer-R (rackmounted) and CewePrometer-W (wall-mounted). Some of the described functional properties can
be missing in certain meter versions. The CewePrometer is complemented with CeweConfig, a PC
program for configuring, manual reading and maintenance.

Contacting us
For more information and technical support, please contact Cewe Instrument.
Internet

www.ceweinstrument.se

E-mail technical support


E-mail order and product information

support@ceweinstrument.se
marknad@ceweinstrument.se

Telephone

+46 (0)155 77500

Address

Cewe Instrument AB
Box 1006
SE-611 29 Nykping
Sweden

Product description
Product description
15

14
13
12

11
10
3

Figure 1 CewePrometer-W

10
15
14

13

12

16

Figure 2 CewePrometer-R

11

Product description
1

Display

Optical communication port


The optical port communicates in accordance with IEC62056-21/IEC1107.

Alarm LED
The alarm LED flashes when an alarm state has occurred. For alarms, such as
for high or low voltage, phase balance or similar external errors, the LED stops
flashing when the alarm state has passed. If the meter has had an internal error,
the alarm must be acknowledged and manually reset.

Calibration LED for active energy


The LED is the source of the energy signal for accuracy testing of active energy.
The light pulses have a 50% pulse width and a frequency proportional to the
measured active power. A constant that specifies the total number of pulses per
kWh can be configured. The same constant applies to active and reactive energy.

Auxiliary power LED


When the CewePrometer is receiving auxiliary power, the green LED turns on to
indicate that the meter's power supply is working.

Sealing points for terminal cover


The terminal cover can be sealed with lead seals (or comparable seals) and
sealing wire through the sealing points.

Terminal cover

Sealing points for front panel window


The front panel exterior window can be sealed with lead seals (or comparable
seals) and sealing wire through the sealing points. The seal prevents access to
the fourth button (sealed button) and the optional backup battery for the realtime clock.

Meter cover

10

Front panel window


The front panel window has hooks for attaching the meter label. In addition, the
window is provided with a pocket for inserting an extra label. The extra label
can be used when transformer ratios are changed.

11

Calibration LED for reactive energy


The LED is the source of the energy signal for accuracy testing of reactive
energy. The light pulses have a 50% pulse width and a frequency proportional to
the measured reactive power. A constant that specifies the total number of pulses
per kvarh can be configured. The same constant applies to active and reactive
energy.

12

Sealed button
The button is hidden under the front panel window and provides access to a fifth
configurable display sequence. Holding down the button for about two seconds
opens a system menu.

Product description
13

Step backward
The button is used to step backward to a previous display layout in the selected
display sequence, or to step downwards in a menu.

14

Menu/Apply
The Menu/Apply button is used to open the menu where one of four
configurable display sequences can be chosen. The button confirms choices
made or initiates other activities depending on the current display layout.

15

Step forward
The button is used to step forward to the following display layout in the selected
display sequence, or to step upwards in a menu.

16

Sealing points for rack mounting


The rack mounting bracket for the CewePrometer-R can be sealed with lead
seals (or comparable seals) and sealing wire through the sealing points.

Product description
Sealing of CewePrometer-W
See Figure 1 CewePrometer-W for location of the sealing points. The front panel window can be
sealed with lead seals (or comparable seals) and sealing wire through the sealing points. The seal
prevents access to the fourth sealed button and the optional backup battery for the real-time clock.

Battery cover

Sealed button

Figure 3 Under the front panel window is a fourth button and backup battery for the real-time clock.

The terminal cover can be sealed with lead seals (or comparable seals) and sealing wire through the
sealing points.
The meter cover can be sealed with lead seals (or comparable seals) and sealing wire through the
two sealing points under the terminal cover.
Sealing point

Figure 4 The meter cover can be sealed at the sealing rings on both sides under the terminal cover.

Product description
Sealing of CewePrometer-R
See Figure 2 CewePrometer-R for location of sealing points. The enclosure permits the meter to be
sealed at three points. The top and bottom covers are normally each sealed with a lead seal and with
sealing wire through a sealing screw and a fixed hole on the cover at the rear of the meter. The
exterior window on the meter front can be sealed on a sealing point with a lead seal and sealing
wire. In this way, the meter can be sealed so that it cannot be opened without breaking all the seals.
One of the buttons on the front is protected from access behind the sealed exterior window.

Battery
Sealed button

Under the front panel window is a fourth button and backup battery for the real-time clock.

Connections
Connections to the CewePrometer-W are made on the meter terminal under the terminal cover.
Connections to the CewePrometer-R are made on the rear panel of the meter subrack. The
connections are: measuring voltages, measuring currents, relay outputs, optically isolated digital
inputs, separate auxiliary power and connections to communication modules, if any.
Digital outputs

Digital inputs

Output #8

Output #1

I'L1 UL1

IL1

I'L2 UL2

IL2

I'L3 UL3

Input #1

IL3

Input #4

N
Separate aux. power
(Uaux)

Connections for CewePrometer-W

10

Product description

Connections

Connections for CewePrometer-R

UL1
UL2
UL3
N

B1
B2
B3
B0

Uaux1
Uaux2

B5
B6

IL1
I'L1
IL2
I'L2
IL3
I'L3

A01
A1
A02
A2
A03
A3

Current, voltage and auxiliary power connections


for CewePrometer-R.

Digital I/O variants


6 outputs, 4 inputs
OUT1 C0, C5
OUT2 C1, C6
OUT3 C2, C7
OUT4 D0, D5
OUT5 D1, D6
OUT6 D2, D7
10 outputs, 2 inputs
OUT1 C0, C5
OUT2 C1, C6
OUT3 C2, C7
OUT4 D0, D5
OUT5 D1, D6
OUT6 D2, D7
OUT7 D3, D8
OUT8 D4, D9
OUT9 B7, B8
OUT10 B4, B9

IN1
IN2
IN3
IN4

C4, C9
C3, C8
D4, D9
D3, D8

IN1
IN2

C4, C9
C3, C8

12 outputs, 0 inputs
OUT1 C0, C5
OUT2 C1, C6
OUT3 C2, C7
OUT4 D0, D5
OUT5 D1, D6
OUT6 D2, D7
OUT7 D3, D8
OUT8 D4, D9
OUT9 B7, B8
OUT10 B4, B9
OUT11 C3, C8
OUT12 C4, C9

Connector for CewePrometer-R

A standardised Entrelec Essailec plug-in connector is used, of the same type as for other meters
according to standard DIN 43862. The cable connection is made in the subrack. The meter is
connected via the contact pins after insertion. The connector has a short-circuit protection feature
for electrical circuits, permitting the meter to be withdrawn and inserted into the subrack during
operation.
Auxiliary power

The CewePrometer is supplied with separate auxiliary power. Auxiliary power can be supplied both
with alternating current and polarity-independent direct current within a specified range. There are
two versions that have either a high or a low auxiliary voltage range.
Digital outputs

The CewePrometer's outputs are solid-state MOS-FET bipolar semiconductor relays, with normally
open contact function. Internal current limit protects the relay from being damaged by excessively
high current.
11

Product description
Digital inputs

The optically isolated digital inputs consist of opto couplers, where the LED is powered from an
external voltage signal through series resistance to limit the current. This means that an external
voltage must be connected between the passive pulse and level transmitter and the opto coupler's
LED, or the transmitter must deliver an active voltage signal to send pulses that are detected by the
optical input. The optically isolated inputs are unaffected by reverse polarity. Both DC and 50/60
Hz AC can be used.
Serial communication ports

The CewePrometer is always equipped with one optical port and is available in versions with or
without additional communication ports. A meter can be equipped with up to two extra ports, either
with the RS232 or RS422 interface.

Communications through the infrared IEC62056-21/IEC1107 optical port are transferred


via an optical read/write adapter, an optical head. The optical head is positioned and
secured over the optical port by a magnet on the head and a steel plate around the opening
of the optical port.

The RS232 standard communication port is used to connect communication equipment


(computers, modems) through a serial cable with a standard D-sub, 9-pin connector.

The RS422 serial communication port is intended to be used to connect the CewePrometer
to a network for multi-drop communications. The CewePrometer is connected to the
communication equipment (RS422 converter) with two twisted-pair wires connected to a
5-pole connector.

For information on pin configuration, see Appendix C Serial Communication Port RS232/422
(pg. 63).
Communication port #2

Communication port #1

Numbering of the communication ports from right to left on the CewePrometer-W.

Communication port #1

Communication port #2

Numbering of the communication ports on the CewePrometer-R

12

Product description
Mechanical design

Alternative
mounting

Dimensions for CewePrometer-W.

13

Product description

Dimensions for CewePrometer-R.

14

Product description

Dimensions for CewePrometer-R subrack.

The subrack is also available in alternative connector versions that will fit other brands of meters
with the same type of connector. Contact Cewe Instrument for more information.
Enclosure and protective earth

CewePrometer-W
The enclosure consists of a meter base, meter cover, terminal cover and a plastic exterior window.
The meter is not connected with protective earth. There is also an optional 19 inch rack/panel
mounting kit available for to CewPrometer-W to be mounted side by side.
CewePrometer-R
Enclosure fits for installation of two meters in a 19" subrack in compliance with DIN438862. The
enclosure consists of a box made of surface-finished steel sheet and a front with a plastic window.
The meter is designed for protective earthing through the 19" subrack.
Isolation and personal safety

The electronic circuits are isolated from all connections to the supply mains with transformers.
Digital I/O connections are isolated with opto couplers and solid state relays. The communication
ports are also isolated but for a lower voltage.

15

Product description
Function modules
Modulisation

The CewePrometer is made up of function modules. Modulisation is primarily used to make it easy
to produce different versions with the same basic design, and to enable development of new
modules for the basic design to add or replace functions. Modulisation is not primarily intended for
being able to switch plug-in modules in the field the meter must be opened and the seals broken to
switch modules.
Modules

The CewePrometer consists of the following basic modules, which are necessary for a functioning
meter in its simplest version:

Measurement module

Power module

Display and register module

Moreover, the meter is normally equipped with the following modules for additional functionality:

Digital I/O module

Communication module

There is also space for additional function modules for expansion of function properties to meet
future market needs.
For more information, see Appendix D - Module block diagram (pg. 65).

16

Product description
Measuring principles
The measuring circuit in CewePrometer consists of current and voltage transformers that provide
signals to six parallel AD converters (analogue to digital converters) that are synchronised by a
common clock signal. The digital signals are thereafter processed by a DSP. Using voltage
transformers makes the electronics in the meter galvanically isolated from the measurement voltage,
which provides good personal safety and protection for connected equipment, such as modems.
Calculation flow

All values are calculated in the CewePrometer based on calibrated current and voltage values.
Current and voltage amplitudes and phase angles are fully compensated in regards to accuracy,
harmonics, frequency and temperature. Adjustments for accuracy are made throughout the dynamic
range for both voltage and current. Based on these individually compensated current and voltage
signals, power, energy, power factor and all other quantities that the CewePrometer can present, are
subsequently calculated. This means that accuracy for instant values is good and that active and
reactive power are correctly calculated, including harmonic power. For more information on
calculation principles and the formulas used, see Appendix F Calculation Principles (pg. 67).
3-element meter

On the 3-element CewePrometer, phase voltages and neutral wires are connected to the meter. The
voltages measured are phase voltages. Power and energy are calculated from three phase voltages
and three currents. Harmonic measurement, alarm monitoring and transformer corrections are made
on phase voltages. The phase to phase voltage is calculated from the phase voltages.
2-element meter

On the 2-element CewePrometer, the neutral conductor is not connected to the meter and the three
voltages are internally D-connected. The voltages measured are subsequently phase to phase
voltages. Power and energy are calculated based on two phase to phase voltages (U12 and U23) and
two currents (I1 and I3) according to the 2-watt meter method. The 2-element meter is primarily
used for D-connected systems (3-wire). Harmonic measurement, alarm monitoring and transformer
corrections are made on phase to phase voltages.

17

Configuration, reading and maintenance


Configuration, reading and maintenance
CeweConfig is a PC program that makes all CewePrometer functions available. With CeweConfig,
you can:

Configure
Configuring means that parameters affecting meter function can be set. Examples of
parameters that can be configured are: transformer ratios, logging interval and limits for
alarms.

Reading
Examples of information that can be read are: registers, logged values, alarms and
harmonics. The information can be printed out or saved to a file.

Maintenance
Examples of maintenance tasks are: resetting the event log and updating the firmware in the
meter.
Menu

Structure tree

18

Toolbar

Status bar

Configuration form

Configuration, reading and maintenance


Connecting to CewePrometer
To be able to configure or read values in the CewePrometer, CeweConfig must be connected and
have authorisation to access the meter. The meter has five different authorisation levels that can be
configured with passwords. See the section Communications and security (pg. 37). With the meter's
factory settings, no password is configured, and subsequently no password is necessary when you
connect.
To communicate with a meter, the PC must be physically connected to the CewePrometer in one of
the following ways:

PC Optical head Meter

PC Crossed serial cable (null modem) Meter

PC Straight serial cable RS422 converter Daisy chained meters

PC Modem Modem Meter

How to connect to the meter


1. Connect to Meter
Choose Connect to Meter from the File menu or by clicking the toolbar button.
2. Communication channel
Click the Settings tab and choose either Optical Port, Serial Port or Modem. If the serial port or a
modem is used, the baud rate must be selected. For meters with factory settings, the baud rate is
9600. When the optical port is used, CeweConfig adapts itself to the baud rate the meter is
configured for, making selection of the baud rate unnecessary.
Note: The selectable communication ports can be changed by changing the XML file Settings.xml
that is in CeweConfig's installation directory.
3. Modem
If a modem connection is used, a modem must be chosen. The modems available are those installed
in Windows. Click the Common tab and enter the telephone number to be called.
19

Configuration, reading and maintenance


4. Password
Click the Common tab. A password is entered if one is configured in the meter. With the meter's
factory settings, no password is configured, and subsequently no password is necessary when you
connect.
5. Serial number
Serial number is only required if a special meter is to be addressed when several meters are
connected together with RS422.
6. Trace to file
Under the Settings tab, a box can be checked to activate tracing to a file. A file named
SerialTrace.log is then created in CeweConfig's installation directory. All communications between
CeweConfig and the meter are presented in the file. The function can be used to analyse
communication problems.
7. Click the Connect button.
Problems with connecting

If the meter cannot be connected, an error message is displayed. Depending on the reason, the
message can suggest corrective actions, such as changing the port or port baud rate.
Tip: When communication problems occur, the trace feature on the meter's display can be used to
see if that which has been sent has been registered by the meter. See the section Using the display
for communication diagnostics (pg. 56).

Basic configuration
Some basic settings may be required before the CewePrometer will be able to measure and operate
correctly in a system.
Note: Settings are only necessary if they have not been made at the factory prior to delivery.
Meters supplied without backup batteries retain the time setting for a maximum of three days. For
time-dependent functions such as logging and event logging to work, the clock may need to be
adjusted.
To change the configuration for a meter, you must be connected to it. Click the Configuration
folder in the structure tree to the left in CeweConfig to display the various functions that can be
configured. For more information see the section Changing configuration (pg. 24).
Tip: You can save a configuration from a meter to a file. A summary of the configuration can also
be printed out. You can also create a configuration without being connected to a meter. For more
information, see Working with configurations (pg. 25).

20

Configuration, reading and maintenance

Meter clock: To configure the meter clock, choose the node Meter Configuration Clock in the structure tree. Find out how the meter clock works and how it can be
configured in the section Meter clock (pg. 29).

Transformer ratios: For the meter to measure accurately, the ratios must be correct. To
configure the transformer ratio in CeweConfig, choose the node Meter Configuration
Measuring in the structure tree. Click the General tab in the window that opens. Fill in
the primary and secondary values for current and voltage.
Note: The values you choose as primary and secondary values will be considered as the
nominal values. These nominal values are used by several functions in the meter: alarms,
voltage monitoring and transformer compensation.

Presentation format for energy registers: To configure the presentation format for
energy registers, choose the node Meter Configuration Measuring in the structure
tree. Choose the prefix and number of decimals for energy registers. Based on the nominal
values, a capacity is calculated for how long the meter can measure without registers
resetting. Read about energy registers in the section Energy registers (pg. 30).

Pulse constants for pulse outputs (Digital I/O): To configure pulse constants for pulse
outputs in CeweConfig, choose the node Meter Configuration Digital I/O in the
structure tree. Find out how Digital I/O works and how it can be configured in the section
Digital inputs and outputs (pg. 33).

Display sequences: To configure display sequences in CeweConfig, choose the node


Meter Configuration Display sequences in the structure tree. Find out how display
sequences work and how they can be configured in the section Display sequences (pg. 36).

21

Configuration, reading and maintenance


Overview of functions
The following is a brief overview of the functions available in CewePrometer. All functions in the
meter can be both configured and read in CeweConfig. In many cases, CeweConfig can also export
data to a file or print out data.
Function
Communication speed
Set the baud rate for the meter's
optical or serial port.
Passwords
Regulate access to data in 5
levels with passwords or
hardware strap.
Data logging
Log energy or instant values.
Alarms
Activate alarms for events such
as voltage unbalance. Determine
how alarms will be presented.
Maximum demand
Determine values that are to be
stored as maximum average
values.
Historical registers
Choose how a historical period
will be finished.
Time of use
Determine how rates shift over
the course of a day and which
registers will be divided by rates.
Transformer compensation
Compensate errors and losses for
instrument and power
transformers.
Power quality
Activate monitoring of voltage
sags, swells and interrupts.
Language on display
Choose between several
available languages for the
display.
Information texts
Enter information texts that can
be read on the display and via
communications.
Calibration LED
Set a pulse constant for the
calibration LED. This is used to
test meter accuracy.
Accumulate incoming pulses
Configure registers to accumulate
pulses representing a selectable
quantity.

22

Configuration location in
CeweConfig
Meter Configuration
Communication

Section in handbook describing


the function.
Communications and security (pg.
37)

Meter Configuration
Communication

Communications and security (pg.


37)

Meter Configuration Loggers

Loggers (pg. 39)

Meter Configuration Alarm

Alarms (pg. 41)

Meter Configuration
Maximum Demand

Maximum demand (pg. 44)

Meter Configuration
Historical Registers

Historical registers (pg. 45)

Meter Configuration Time of


use

Time of use (pg. 46)

Meter Configuration
Measuring, Transformer
Compensation tab

Transformer compensation (pg.


48)

Meter Configuration
Measuring, Power Quality tab

Power quality (pg. 50)

Meter Configuration Misc.

Miscellaneous (pg. 52)

Meter Configuration Misc.

Miscellaneous (pg. 52)

Meter Configuration
Measuring, General tab

Product description (pg. 6)

Meter Configuration Digital


I/O, Inputs tab

Digital inputs and outputs (pg. 33)

Configuration, reading and maintenance


Indicate alarms via pulse
output
Choose a pulse output to be
activated in case of alarms.
Synchronise clock via
incoming pulse
Set an interval that the clock shall
be synchronised to upon
incoming pulses.
Adjusting the clock for daylight
saving time
Set date and time for daylight
saving time.

Meter Configuration Digital


I/O, Outputs tab

Digital inputs and outputs (pg. 33)

Meter Configuration Clock


Meter Configuration Digital
I/O, Inputs tab

Meter clock (pg. 29)


Digital inputs and outputs (pg. 33)

Meter Configuration Clock

Meter clock (pg. 29)

23

Configuration, reading and maintenance


Changing configuration

To open a configuration form, click the folder Configuration in the structure tree and then click
one of the nodes: Display Sequences, Measuring, Clock, etc. Configuration changes can be made
in all configuration form. In the lower right corner, there is an Apply button. If a meter is connected
and you click Apply, changes to the configuration will be immediately transferred to the meter. If
the changes affects registers or measurement, a warning dialog will be displayed. You can choose
not to send a portion of the configuration by clearing checkboxes in the dialog.

Warning dialog that sometimes is displayed when configurations are sent to the meter.

24

Configuration, reading and maintenance


Working with configurations
In CeweConfig, you can work with configurations as a collection of CewePrometer's settings and
save them in a file. On CeweConfig's File menu, are the commands Save configuration, Open
configuration and New configuration. A configuration can either be for a 2- or 3-element meter.
You cannot transfer a 2-element configuration to a 3-element meter or vice versa. Transformer
corrections or passwords are not saved in configuration files. On the File menu, there is a command
for creating a configuration summary. Below is a list of how you can use CeweConfig's functions to
work with configurations.

Creating a configuration file without being connected to a meter


Choose New configuration when CeweConfig is not connected to create a configuration
file offline. The configuration file can be either of the 2- or 3-element type. Make all settings
that are to be included in the configuration file and save the file. The file's configuration can
later be transferred to a meter.

Saving a meter's configuration to a file


Choose Save configuration when CeweConfig is connected to a meter to save the meter's
configuration to a file. The configuration file can later be used as a backup or be transferred
to another meter.

Transferring a configuration file to a meter


Choose Open configuration when CeweConfig is connected to a meter to transfer a
configuration file to the meter. The changes in configuration cause a warning dialog box to
open. The configuration file can either be created without being connected to a meter, or be
saved from a meter; see the points above.

Printing out a summary of a meter's configuration


Choose View configuration Configuration summary when CeweConfig is connected to
a meter to create a summary of the meter's entire configuration. Now choose Print.

Printing out a summary of a configuration file


Open a configuration file and choose View Configuration summary to create a summary
of the configuration that is in the file. Now choose Print.

Note: Choosing New configuration or Open configuration when CeweConfig is connected to a


meter opens a warning dialog box with the message that the configuration in the meter will be
written over if you continue.

Reading

If you are connected to a CewePrometer with an authorisation level between 1 and 5, several
alternatives will be available in the structure tree under the node Reading. For the alternatives
25

Configuration, reading and maintenance


Instant Values and Time, displayed values are constantly updated. The others are only updated
when the window is opened or when the Update button is clicked. Read values can be printed out
and often also saved to files, this can be done by with buttons Print and Save.

Information about the meter

Information about the connected meter can be obtained by choosing View Information about the
meter. Some of the information displayed is the same as what appears on the meter plate.
Additionally, there is information on when the meter was manufactured and when the most recent
configuration change was made.
Under the Modules heading, the firmware currently used in the meter is listed. See the section
Updating firmware (pg. 27).

26

Configuration, reading and maintenance


Versions and version conflicts
The latest version of CeweConfig can be used with all firmware versions of CewePrometer-R/W.
The version number for CeweConfig is displayed on the application's title bar or under About on
the Help menu. The version number for the meter's firmware can be viewed under View
information about the meter on the View menu.
CewePrometer and CeweConfig have three-digit version numbers according to the format main
version.sub-version.build number. As long as the main version and sub-version are the same,
CeweConfig and the meter are compatible. If the meter is of a newer version than CeweConfig and
the main version and/or sub-version are different, CeweConfig will display a message that
connection is not possible. CeweConfig must be updated.

Updating firmware
CewePrometer is designed with a number of modules that each have their own processors and their
own firmware. The module firmware is distributed in a package file (.pkg) which is transferred to
the meter using CeweConfig. Newly developed and improved functions can thus be added in a
meter that lacked these functions when delivered.
Note: Be sure to update CeweConfig to the latest version before updating the meter. There is
otherwise the risk that CeweConfig will no longer be version-compliant after firmware updating.

On the Tools menu, there is an Update firmware command when the meter is connected with
authorisation level 4 or higher. Begin by choosing the file that contains the update. The file name
and version number will then be displayed, and sometimes a message. Click the Advanced button
to view CeweConfig's version number by module for both the file and the meter. Additionally, you
can force modules to be updated even if the file has the same version number. Click Update to
begin updating.

27

Configuration, reading and maintenance


During the time the update is being installed, the meter stops measuring and registering energy.
Depending on the size of the file to be transferred and the baud rate, the time for updating can vary
from a few minutes to a half hour at the highest baud rate. If possible, connect at the highest baud
rate (19200 bps) to speed updating. After updating, the meter is restarted to complete installation of
the meter's new firmware.
Updating can be performed via a modem but this should be avoided since reliability for this method
of communication is too low. An interrupted update can leave a meter with incomplete firmware
that cannot measure at all. If an update via a modem has failed, the next update must be made via
RS232/422 or the optical head.

Language
CeweConfig can be set to different languages. The available languages can be seen under
Language on the View menu.

Resetting registers and logs


CewePrometer's various registers and log data can be reset with CeweConfig if a meter is connected
and you are connected with a sufficiently high authorisation level. You will find a number of
choices under Reset on the Tools menu. Some options may be marked in grey if your authorisation
level is not high enough. To reset energy registers, authorisation level 5 is required, and level 3 for
other options. A warning dialog box opens before resetting.

28

Functions
Functions
Meter clock
The CewePrometer has an integrated real-time clock for time-dependent functions. Information
about time adjustment and daylight saving time status is recorded with tags on the logged values;
see Loggers (pg.39). Time adjustments are also noted as an event in the event log; see Appendix B Events (pg. 61).
Instantaneous adjustment

The meters date and time can be set to an absolute point in time. Instantaneous adjustment of the
meters time can influence logged values. To avoid this, the meter's log memory is reset when doing
instantaneous adjustment. For this reason, instantaneous adjustment of the meter clock is primarily
intended for use at initial configuration of the meter.
Sliding adjustment

The meters time can be adjusted successively. Instead of the meter adjusting the clock
instantaneously, the total adjustment is spread out over a longer period. A speed of adjustment can
be chosen in the 140% range. With 40%-adjustment, the clock is adjusted 24 seconds for each
minute until the total adjustment has been made.
Daylight savings time

CewePrometer offers the alternative of letting the meter clock follow daylight savings time. At a
specified date, the meter clock is adjusted forward, and at another, adjusted backward.
Example: On 28 March the clock is to be adjusted forward, from 02:00 to 03:00. The adjustment
back to standard time is to occur on 31 October at 3:00 (daylight savings time) when the clock is to
be set back to 02:00. The following is set in the meters: Begin March, 28, 02:00. End October, 31,
03:00 and the standard time is to be adjusted by 60 minutes.
External synchronisation

The meter time can be adjusted by a pulse on one of the meters digital inputs. When a pulse is
registred, the clock is adjusted to the closest multiple of a specified synchronisation interval. If the
synchronisation interval is, for example, one hour and the time is 13:29, a pulse will adjust the clock
to 13:00. If the time had instead been 13:31, the clock would have been adjusted to 14:00.
Available synchronisation intervals are:
10, 15, 20 and 30 minutes
1 hour
Times 12:00 and 00:00
Time 00:00
Besides specifying that the meter clock will synchronise via external pulses, a digital input must
also be configured for clock synchronisation; see the section Digital inputs and outputs (pg. 33).

29

Functions
Energy registers
Energy registers in the CewePrometer are electronic counters that accumulate energy. Prefixes and
number of decimals are configured which apply to all energy registers. Energy registers have a
width of nine digits, regardless of the total number of decimals and the prefix configured. When an
energy register has reached its maximum figure (for example, 999999.999 kWh), it resets to zero.
Fewer decimals and larger prefixes give the energy registers space for more energy without
resetting. Read more about how energy is calculated and defined by quadrants in Appendix F
Calculation Principles (pg. 67).
Overview

The table shows the energy registers available in the meter.


Energy type
Active energy

Energy direction
Import, export

Total values
Yes

Reactive energy

Import, export,
quadrants I-IV,
capacitive, inductive
Import, export

Apparent energy

Active power export


(-)

Reactive power
import (+)

Total number
8

Yes

Per phase values


Yes (3-element
meter only)
No

Yes

No

Active power import


(+)

II

I
S

Capacitive
Q

P
Inductive
Reactive power
export (-)

III

IV

Quadrants
I

Phase angle
0 to 90

Current relative to voltage


Lagging

II

90 to 180

Lagging

III

-180 to -90

Leading

IV

-180 to 0

Leading

Storage in non-volatile memory

All registers are saved in non-volatile memory once per second. Loss of auxiliary voltage therefore
results in no more than one second of lost historical energy measurement data. For optimal
safekeeping, all registers are saved in three alternating memory areas.

30

Functions
Instant values
Besides energy, the CewePrometer can also measure instant values. Instant values are constantly
changing values such as current, voltage, power and harmonics. The formulas and definitions used
to calculate the values are presented in Appendix F Calculation Principles (pg. 67).
Overview

This table provides an overview of the instant values that can be read on the meter. Readings can be
viewed with CeweConfig, on the display and with other software that has implemented
CewePrometers communication protocol. Most instant values can be logged as average values; for
more information, see section Loggers (pg. 39).
Instant value
Frequency
Power factor total
Power factor per phase
Active power total
Active power per phase
Reactive power total
Reactive power per phase
Apparent power total
Apparent power per phase
Phase to phase voltage L12, L23, L31
Phase voltage
THD voltage
Amplitude voltage harmonics (231)4
Phase symmetry voltage
Current
THD current
Amplitude current harmonics (231)4
Phase symmetry current
Phase angle total
Phase angle per phase

Available on 3-element
meter
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes (calculated) 1
Yes
Yes (phase volt.)
Yes (phase volt.)
Yes (phase volt.)
Yes (L1, L2, L3)
Yes (L1, L2, L3)
Yes (L1, L2, L3)
Yes
Yes
Yes

Available on 2-element
meter
Yes
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes (phase to phase volt.) 2
Yes (phase to phase volt.)2
Yes (phase to phase volt.)
Yes (L1, L3) 3
Yes (L1, L3)
Yes (L1, L3)
Yes 4
Yes
No

Update frequency

The update frequency for instant values is proportional to the frequency of the measuring voltage.
With a 50 Hz measuring voltage, updating occurs 12.5 times per second, and at 60 Hz, 15 times per
second. Instant values can be read up to twice per second via the serial port when it is set to 19200
baud.

Calculated from the voltages fundamental vectors.


The harmonics for voltage on the 2-element meter are to be seen from the standpoint of the voltages being Dconnected internally.
3
Current L2 is calculated internally in the 2-element meter for monitoring.
4
The value can be read in CeweConfig and via the communication protocol but cannot currently be presented on the
display (meter firmware 1.2.0).
2

31

Functions
Accuracy

The table shows typical accuracy for a CewePrometer with accuracy class 0.2 for a selection of
instant values.
Instant value
Voltage amplitude
Current amplitude
Current amplitude
Frequency
Power

Range
Umin - Umax
5 % of Imax - Imax
15% of Imax
4763 Hz
1 % of Imax - Imax

Power factor

At PF=0.5 and current 5% of Imax


- Imax

Accuracy better than:


0.1% of reading
0.1 % of reading
0.05 % of Imax
0.02 % of reading
Corresponds to meters class
according to energy measurement
standard
0.1 % of reading

Harmonics measurement

Harmonics numbers 2 to 31 are measured for all currents and voltages. At a fundamental frequency
of 50 Hz, the second harmonic is 100 Hz, the third harmonic is 150 Hz, etc. Both the harmonics
amplitude and phase angle are measured and included in the calculation of power and energy, and
can be read via the meters communication protocols. In CeweConfig, harmonic amplitudes are
presented with a diagram.
THD

THD stands for Total Harmonics Distortion and is a measurement of the amount of harmonics
present in a signal. Voltages and currents THD can be read via CeweConfig and on the display.
Harmonics on 2-element meter

The harmonics measurements of voltage on the 2-element meters should be seen from the
standpoint of the voltages being D-connected internally in the meter. The harmonics measured are
the vector sums of the phases harmonics.

32

Functions
Digital inputs and outputs
The CewePrometer has several inputs and outputs that can be configured to perform various tasks.
Both inputs and outputs are protected against overvoltages by varistors. They also have an isolated
interface between the electronics and the surroundings to ensure personal safety. For electrical data
on the meters inputs and outputs, see Appendix H Connection Data (pg. 73).
Inputs

The inputs can be configured as follows:

Not used
The input is not used.

Finish historical period


An incoming pulse will result in the present period ending and registers being copied to
historical registers. For a pulse to finish historical period, it is also necessary that the
historical registers be configured to allow this. For more information, see the section
Historical registers (pg. 45).

Pulse input
To register pulses from pulse-producing units, such as energy meters or water meters, pulse
inputs are used. Incoming pulses are accumulated in registers called external registers. There
is an external register connected to each input on the meter. For external registers, a factor is
configured by which the number of incoming pulses is multiplied. Prefixes and the number
of decimals can also be configured for the registers. Moreover, the registers can be
configured with descriptive texts.

Time synchronisation
When incoming pulses are received, the meters clock is synchronised at a specific interval
For available synchronisation intervals and more detailed information on time
synchronisation, see the section Meter clock (pg. 29).

Registration of pulses
The meter registers pulses on positive or negative flanks, depending on if the input is set to inverted
or not. A pulse must be at least 16 ms long to be guaranteed of being detected by the meter. The
maximum pulse width that the meter can handle is 60 s. By setting limits for maximum and
minimum pulse lengths, the meter can be limited as to what it detects as a valid pulse. Pulses with
lengths beyond the established limits are ignored. The occurrence of to long or to short pulses can
also be configured to be recorded in the event log.
VCC

GND

Pulse
length
Pulse length
(inverted input)

The figure shows pulse lengths when an input is inverted or non-inverted, respectively.

33

Functions
Outputs

The outputs can be configured as follows:

Not used
The output is not used.

Pulse output
The output is used to pulse an energy type that the meter is measuring. A pulse constant is
specified for the output as pulses/unit, primary or secondary, and the pulse length is
specified for all pulse outputs. The shortest possible pulse length is 40 ms. Pulses are not
allowed not come too often, and because of this, there is a relationship between the pulse
length and the specified pulse constant that maximises the pulse frequency to 1000/(pulse
length(ms) * 2).

Pulse

Gap

Maximum pulse frequency at outputs limited so that the gap is at least as long as the pulse
length.

Remote control
With this function, the output can be made active or inactive by sending commands to the
meter via the IEC62056-21/IEC1107 protocol. This functions could be used control
anything that can be controlled with a digital relay output.

Alarm output
When an output is set to functions as an alarm output, one or more of the user-defined
alarms can be chosen to indicate at the output. When an alarm occurs, the output switches to
active, and when the alarm state ceases, the output returns to inactive. In the section Alarms
(pg. 41), user-defined alarms are described and how they can be configured.
In contrast to the other selectable functions for outputs, alarm outputs can be inverted. Note
that outputs are inverted via firmware. If the meter loses its auxiliary power, the relay will
open, regardless of it is inverted or not.

End of MD period (maximum demand period)


At the end of a set MD period, the output will go active for one second before returning to
the inactive state. See the section Maximum demand (pg. 44).

Output states
An active output means a closed relay when the output is not inverted. When the output is inverted,
the active relay is open. Only when the output is used as an alarm output can it be inverted. The
CewePrometer outputs are of the solid-state type and when the meter is turned off, they are open.

34

Functions
Meter variants
Meter
CewePrometer-R
CewePrometer-R
CewePrometer-R
CewePrometer-R
CewePrometer-W
CewePrometer-W

No of inputs
4
2
0
0
4
0

No of outputs
6
10
12
0
8
0

35

Functions
Display sequences
The following is a description of the function properties of CewePrometers display sequences. The
section Using the display (pg. 53) describes how to navigate in the display sequences via the
meters buttons, and Appendix A Display Layouts (pg. 57) lists all of the CewePrometers
available display layouts.

Menu for choosing one of the four display sequences. The names of the sequences can be configured.

The CewePrometer is equipped with a graphic display that can show the information available in
the meter. To organise the information, there can be up to five display sequences with a maximum
of 40 selectable display layouts in each. Each display sequence can be given a name that will be
presented on the display.
Display sequence five is only accessible via the sealed button under the front panel window. The
front panel window can be sealed to provide additional security. For this reason, it is appropriate
that display layouts with sensitive information and sensitive functions be placed in display sequence
five.
Automatic display cycling

Automatic display cycling means that the display layouts in one or more display sequences are
automatically cycled by the meter. If, for example, display sequences 1 and 3 are configured for
automatic display layout cycling, the meter will first cycle the display layouts in sequence 1 from
the first to last layout, and then continue with sequence 3. After the last layout in sequence 3,
cycling restarts with the first layout in sequence 1. The display layouts cycle with a configurable
delay of 1 to 30 seconds. If a user manually navigates in the display sequences via the meter's
buttons, cycling automatically stops and continues again after one minute.
Passwords and security

The display sequences can be configured to be password protected to limit access for various users.
When a user chooses a password-protected display sequence via the meters buttons, a password
entry field is displayed. Only a correctly entered password gives access to the display sequence and
to the display layouts that are included in it. A password can be up to six characters long and
include the characters A-Z, 0-9 and :;<>=?@. Passwords are not case-sensitive.
Note: If a password-protected display sequence is configured for automatic display layout cycling,
password protection is deactivated.
Automatic return to first display layout

One minute after a user has stopped navigating in the display sequences, the CewePrometer returns
to the first display layout in display sequence 1 if automatic display layout cycling is not activated
for any display sequence.

36

Functions
Communications and security
All CewePrometers are equipped with an optical port for communication. The meter can be
optionally equipped with one or two additional communication ports with RS232 or RS422
interfaces. The communication protocol used is IEC62056-21, or as the older version of the
standard is called, IEC1107. For more information on how the protocol is implemented, see the
document CewePrometer IEC1107/IEC62056-21 meter reading.
Communication speed

The meters optical port always starts with a baud rate of 300 bps, regardless of what is configured,
before shifting over to the specified communication speed. This means that software (for example,
CeweConfig) that communicates with the meter via the optical port does not need to know the
speed that the meters optical port is set to. The optional ports designated as communication ports
#1 and #2 differ in this respect. They start at the specified baud rate from the beginning, which
means that connected software must be aware of the speed to be able to communicate.
Communication port #1 can be set at a speed of between 300 bps and 19200 bps, and
communication port #2 can be set at a speed of between 1200 and 19200 bps. The optical port can
be set at a speed of between 300 and 9600 bps.
Modem initiation

If a modem is connected to one of the optional ports, a modem initiation string can be configured
that the meter sends via the serial port. The initiation string can be used to set the correct
communication settings for the modem and can consist of AT-Commands. The meter sends the
initiation string when the meter is turned on and thereafter every tenth minute. During ongoing
communications, the meter waits to send the initiation string until communications has stopped.
Security

The meter has five authorisation levels that can limit access to the meter during communication via
any of the meters communication ports. Authorisation levels 1 and 2 are password-protected.
Authorisation levels 3 to 5 may be protected by passwords or by a circuit board jumper.
Authorisation levels
1 Provides access to reading.
2 Provides access to everything in level 1 plus
access to set the clock and finish historical
periods (also resets maximum demand
values).
3 Provides access to everything in level 2 plus
access to configure the meter.
4 Provides access to everything in level 3 plus
access to transfer new firmware to the meter.
5 Provides access to everything in level 4 plus
access to calibrate the meter and reset
registers.

37

Functions
Limitation of total access attempts
The meter limits the total number of access attempts to six when incorrect passwords are entered.
At the seventh attempt, the meter blocks access whether the password is correct or not. The block is
in effect until the next hour shift. After that, new password attempts may be made.
Passwords
A password consists of up to 12 case insensitive alfa-numerical characters. The authorisation check
may be deactivated for a level by deleting the password. When connecting to the meter, access is
granted to the highest level that is lacking password regardless of the password given by the user.
Access restriction via circuit board jumper
Inside the CewePrometer, there is space for a circuit board jumper that provides access to
authorisation level 5 when fitted. If the meter is connected with authorisation at level 5, the function
of the jumper can be modified. Security can be heightened by requiring the jumper even for lower
authorisation levels down to level 3. Security can also be relaxed by making authorisation level 5
accessible via a password. Cewe Instrument can provide a technical fact sheet, tfcr0004, that
describes the procedure for opening the meter and fitting the jumper.

38

Functions
Loggers
The CewePrometer has two identical, parallel and individually configurable loggers. That which is
described in this section applies both to logger 1 and logger 2.
Overview

A logger in a CewePrometer can log average values for instant quantities, energy registers and
external registers. Some quantities can be logged both by phase and as total values for all three
phases, others only as total values. The table provides an overview of quantities that can be logged.
Certain instant values in the table are not available in 2-element meters and thus cannot be logged;
see the section Instant values (pg. 31).
Quantity
Active energy import
Active energy export
Reactive energy import
Reactive energy export
Reactive energy inductive
Reactive energy capacitive
Reactive energy QI
Reactive energy QII
Reactive energy QIII
Reactive energy QIV
Apparent energy import
Apparent energy export
Phase voltage
Phase to phase voltage
Current
Active power
Reactive power
Apparent power
Frequency
Phase angle
Power factor
THD voltage
THD current
External registers 18

By phase
Yes
Yes
No
No
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Not applicable
Yes
Yes
Yes
Yes
Not applicable

Total
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Energy is logged as register values, i.e. at each logging occasion the current registers values is
logged. Based on these register values, the reading software can present the energy as register
values, periodic energy or as average power. CeweConfig provides all three alternatives.
Logging interval and total channels

A logger can store data in 1 to 10 channels. Common for all channels in a logger is that that the
logging interval that can be configured from one minute up to one hour. A loggers capacity is
dependent on number of channels and logging interval. When the logger is full, the oldest values
will be written over. The table shows the capacity in number of days before the oldest value is
written over.

39

Functions
Capacity in days
Number of logging channels
Logging interval
1
2
3
4
(min)
1 21
14
10
8.6
2 43

28

21

17

10

7.2

6.2

5.4

4.8

4.3

3.9

14

12

10

9.6

8.6

7.9

5 108

72

54

43

36

31

27

24

21

19

10 217

144

108

86

72

62

54

48

43

39

15 326

217

163

130

108

93

81

72

65

59

30 652

434

326

260

217

186

163

144

130

118

60 1304

869

652

521

434

372

326

289

260

237

Storage of logged values

Logged values are saved with time stamps, tariff information and flags that indicate events that have
occurred during the logging interval.

The time stamp indicates the end-time. If the logging interval is configured to one hour, a
value with the time stamp 15:00 refers to the period 14:00 to 15:00.

Tariff information indicates active rates for energy and power during the past interval.

To indicate events or states during an interval, a logged value can be stored with one or
more flags.

Event or state
Time adjusted

Name of flag
T

Disturbed

Alarm

Parameter changed

Incorrect energy
direction

Daylight savings
time
Voltage
loss/missing
Invalid value

Explanation
During the past interval, the meter clock has been adjusted either instantaneous or a sliding
adjustment is in progress.
The past interval is incomplete. For example, an interval shortened by the meter being
without auxiliary power or if the logging memory has been reset. The first value after the
logging memory having been configured will thus always be indicated with "Faulty value
(the logging memory is reset in conjunction with reconfiguration).
In conjunction with user-defined alarms being configured, it may be specified that an alarm
will also be indicated with logged values. When a user-defined alarm has triggered during
the past interval, this is indicated with the flag Alarm.
The CewePrometers configuration, calibration or initiation has changed. Which of these
three the flag refers to can be seen in the event log.
The CewePrometer can be configured for a normal energy direction, import or export. If the
meters energy direction deviates from this, this is indicated with the flag Incorrect energy
direction.
Daylight saving time has been in effect during the past interval.

During the past interval all measuring voltages have been lost or missing.

To be able to guarantee that a logged value is correct, it is logged with a control sum. If the
check sum is incorrect when the value is read, this is indicated with Invalid value. This is a
rare event, but could occur if the meter is subjected to powerful, external electromagnetic
interference.

For several of the flags, additional information can be viewed in the event log. A more exact
time for events is specified in the log.
Note: When power is calculated from logged energy values, the resulting values will be somewhat
more precise than when power is logged directly. This is because power is logged as an average
value of instant values. The instant values are read twice per second, while energy is accumulated
continuously.

40

Functions
Alarms
The CewePrometer is equipped with alarms to be able to indicate when measured quantities are
over or under a configurable limit value. The meter enters the alarm state when the limit value is
reached. An alarm is generated only after the alarm state has continued for a configurable time
(delay). Alarms are configurable by the user and are therefore called user-defined alarms. For the
CewePrometer 1.2.0 or earlier, the maximum delay is 64 seconds.
Note: There are alarms and events that cannot be configured, but instead, are always active.
Examples of such alarms are indication that the clock has been changed or that an auxiliary power
loss has occurred. For more information, see the section Event log (pg. 43).
Overview

For most user-defined alarms, the limit value is specified as a percentage of the nominal value,
which is the configured, nominal primary value (current, voltage or power). For 3-element meters,
the limit value corresponds to phase voltage, and for 2-element meters, phase to phase voltage. The
following table provides an overview of available alarms.
Alarm
Low voltage
High voltage
Low power factor
Voltage unbalance

Current unbalance

Low active power


High active power
High THD voltage
High THD current
High harmonic voltage

High harmonic current

Reverse energy direction

Alarm state applies when


The average value of the voltages is beneath the
nominal voltage limit value.
The average value of the voltages is above the
nominal voltage limit value.
The power factor for the 3-phase system is
beneath the limit value.
A voltage deviates percentage-wise from the
average value of all voltages more than the limit
value.
A current deviates percentage-wise from the
average value of all currents more than the limit
value.
The 3-phase systems total power is beneath the
nominal power limit value.
The 3-phase systems total power is above the
nominal power limit value.
The average value of THD for all voltage phases
exceeds the limit value.
The average value of THD for all current phases
exceeds the limit value.
An overtone (2nd to 31st) relative to the basic
tone on same voltage phase exceeds the limit
value.
An overtone (2nd to 31st) relative to the basic
tone on same current phase exceeds the limit
value.
The phase angle for a phase deviates by more
than 90 degrees from another phase.

Voltage phase missing

One or more voltage phases is missing

Internal error

See the section Event log (pg. 43).

Comments

The event log identifies the


phase and harmonic no.
The event log identifies the
phase and harmonic no.
The event log identifies the
phase with the deviating energy
direction. This alarm is not
available on 2-element meters.
The event log identifies the
missing phase or phases.

41

Functions
Indication

Alarms are always stored in the event log. Alarms can also be configured to indicate in one or more
of the following ways:

Alarm LED on meter front

Changed digital output level

Indication of a logged value with a flag

The alarm LED stops flashing and the digital output returns to inactive low after the alarm state
passes.

42

Functions
Event log
The CewePrometer has a number of defined events and alarms that are stored in the meters event
log when they occur. The event log can contain a maximum of 40 events. Thereafter, the oldest will
be removed to make room for the new. The meter has a counter for the total number of events that
have occurred since the latest reset.
An event is represented by a time stamp and a code that indicates what has occurred. See Appendix
B - Events (pg. 61) for a complete list of all events. For certain events, data is also stored for the
events, for example, which phase the event concerns.
Events are divided into three categories:

User-defined alarms
The meter has a number of alarms that when they occur, generate an event in the event log.
The alarms each have an individual configuration that determines the conditions for when an
alarm is to be activated. See the section Alarms (pg. 41).

Meter events
The meter is also equipped with a number of alarms that are not configurable by the user,
so-called meter events. These have set conditions and are stored only in the event log.
Examples of meter events are auxiliary power loss and clock resetting. A complete list of
meter events is in Appendix B - Events (pg. 61).

Internal errors
The contents of the internal data structures, such as configuration and energy registers, are
checked by the meter. If an error is detected in one of the data structures, it is designated as
an internal error. If an internal error is indicated, the event log should be checked for related
information, and suitable actions should be taken. Internal errors are listed in Appendix B Events (pg. 61).

Display of events

The event log can be viewed on the meters display if the display layout Events is configured to be
included in one of the meters display sequences. Only the time stamp and event code are shown on
the display. A more detailed description of the meter's events can be viewed in CeweConfig, which
converts the code into plain text and displays any related data. Via CeweConfig, the event log can
also be printed out or saved to a file.

43

Functions
Maximum demand
Maximum demand is the highest average value of, for example, active power during a historical
period. The average value is calculated for an interval and compared with a previously stored value.
If the new average value exceeds the previously stored value, it becomes the new value for
maximum demand. When a historical period ends, the current maximum demand values are copied
to historical registers and then reset. The CewePrometer always stores the three highest average
values from three different days.
Overview

Up to eight quantities can be saved in the maximum demand registers, which makes a total of 24
values with time stamps for when they occurred. The table shows the types of energy that can
produce maximum demand values.
Active energy import
Active energy export
Reactive energy import
Reactive energy export
Reactive energy inductive
Reactive energy capacitive
Reactive energy QI
Reactive energy QII
Reactive energy QIII
Reactive energy QIV
Apparent energy import
Apparent energy export

The calculation interval for average values determines the period during which average values are
calculated. At the end of the interval, the new average value is compared with the previous and
possibly stored as a new maximum demand value. The calculation interval can be chosen between 1
and 60 minutes in the following predefined steps: 1, 2, 3, 4, 5, 10, 15, 30 and 60 minutes.
Example: A historical period has just been finished and all maximum demand values have been
reset. The meter is configured with a calculation interval of 30 minutes and active power as the
quantity for which to calculate the average value. The table is read from top to bottom and
illustrates what will be stored in the maximum demand registers at the various times.
Point in time

Average
power

Highest value

Second highest

Third highest

12:59, 4 July
13:00, 4 July
13:30
14:00

90 kW
89 kW
91 kW

90 kW 07-04 13:00
90 kW 07-04 13:00
91 kW 07-04 14:00

00:00, 5 July
00:30

75 kW
95 kW

91 kW 07-04 14:00
95 kW 07-05 00:30

75 kW 07-05 00:00
91 kW 07-04 14:00

00:00, 6 July

93 kW

95 kW 07-05 00:30

93 kW 07-06 00:00

91 kW 07-04 14:00

16:00, 7 July

94 kW

95 kW 07-05 00:30

94 kW 07-07 16:00

93 kW 07-06 00:00

Note: There is never more than one maximum demand value for one day.
44

Functions
Historical registers
Historical registers are used by the CewePrometer to store current register values at defined points
so as to be able to read them later. Stored in historical registers are all maximum demand values,
external registers, TOU registers and energy registers, with the exception of energy registers by
phase. The historical registers are time stamped to indicate when storage occurred. The
CewePrometer can store up to 14 historical registers.
Date and time
Energy registers

Maximum demand
values

Time of use registers

External registers

Historical register includes a number of stored registers, and dates/times when they were stored.
Finish historical periods
By finishing a historical period, the current registers values are stored in historical registers and the
maximum demand values are reset. When a period is finished, an event is stored in the meters
event log. Periods can be finished in various ways:

Via meter button

The period is finished when the meters Menu/Apply button


is held down for more than two seconds when the display
layout Finish period is active.
This requires both that the historical registers are configured
to permit finish via the meter button and that a display
layout for this is entered in a display sequence.

Via CeweConfig

The period is finished when a command is given from


CeweConfig or third-party software.

At the end of the


month

The period is finished when the meter clock reaches the end
of the month.

Via digital input.

The period ends when a pulse is received at a digital input


on the meter.
This requires both that the historical registers are configured
to permit ending via a digital input and that an input is
configured for this purpose.

Note: The display layout Finish period will only work if the meter is configured to permit ending a
period via a button.

45

Functions
Time of use
Time of use is a function that enables energy to be divided up into various registers depending on
the rate that applied when the energy was measured. In the CewePrometer, tariff structures can be
stored that switch rates at predetermined times according to a configurable pattern. A tariff structure
consists of seasons, day types and special days. The maximum number of rates is eight.

Day types specify how rates change during a 24-hour day.

Seasons specify the day types that apply during the days of the week, Monday to Sunday.

Special days specify the day type that applies on a certain date.

Tariff structures

In the CewePrometer, there are two separate tariff structures. Both tariff structures have a set of day
types, seasons and special days. By setting a starting date for a tariff structure, it can be configured
before it goes into force.
Day type

In the CewePrometer, there are 16 day types (AP). A day type specifies witch rate, from a
maximum of eight, should apply when during the day. 16 changing points per day may be specified.
A day type can be connected to a day of the week in a season or to a special day.
Season

A season refers to a period. During this period, the season defines the day types that will apply
during the days of the week. In the CewePrometer, there are 16 seasons per tariff structure. The
seasons are arranged in a sequence where one season replaces the previous at a predetermined date.
Special days

Days that change rates according to a pattern that is not covered by linking day types to seasons
(such as holidays) are called special days. A special day specifies the day type that applies on a
certain date. Special days can be configured to apply every year on the same date or for a single
year. In the CewePrometer, up to 30 special days per tariff structure can be configured.
TOU registers

An energy register or external register that is chosen to be divided into the rates is a TOU register.
There are eight TOU registers and each TOU register has separate registers for eight rates.
Maximum demand values and tariffs

Maximum demand values are stored together with information on which rate applied when
registered. For one and the same day, there can be separate day types for energy and maximum
demand values.

46

Functions
Example: The conditions are Monday to Saturday, rate 2 applies from 7:00 in the morning until
17:00 in the afternoon. For the rest of the day, rate 1 applies. On Sundays, rate 1 applies around
the clock. During the summer months, 1June to 31 August, rate 1 applies around the clock. On
Christmas Eve, 24 December, rate 1 applies around the clock. For maximum demand values, rate 3
always applies. It is active energy import that is divided in to different rates.
Day types
Three day types must be configured. Day type A to apply to energy during the autumn, winter and
spring. Day type B to apply during the summer months, and day type C for maximum demand
values throughout the year.
Day type A
00:00
Rate 1
07:00
Rate 2
17:00
Rate 1

Day type B
00:00
Rate 1

Day type C
00:00
Rate 3

Seasons
Two seasons must be configured. The first season will apply during the summer period and the
second the rest of the year. The season that applies from 1 September will also apply from 1
January to 31 May.
06-01

Energy
MD

A
C

A
C

A
C

A
C

A
C

A
C

B
C

09-01

Energy
MD

B
C

B
C

B
C

B
C

B
C

B
C

B
C

Special day
For rate 1 to apply all of Christmas Eve, a special day must be configured. Because no year is
specified, the special day will apply every year.
12-24

Energy
MD

C
C

TOU register
A TOU register is configured to accumulate active energy in several registers for the different
rates.

47

Functions
Transformer compensation
Transformer compensation is a function for compensating for measurement errors in instrument
transformers and for losses in power transformers. The function enables the CewePrometer to
present measurement values for which errors and losses have been compensated. The formulas used
in the meter are presented in Appendix F Calculation Principles (pg. 67).
Overview

The tables present an overview of the transformer compensations in CewePrometer.


Instrument transformer compensations
Name
Value to entered
Voltage error L1, L2, L3
Amplitude error as percent
Phase angle in minutes
Current error L1, L2, L3
Amplitude error as percent
Phase angle in minutes
Power transformer compensations
Name
Copper losses, Total values
Iron losses, Total values

Value to entered
Active loss as percent of nominal power
Reactive loss as percent of nominal power
Active loss as percent of nominal power
Reactive loss as percent of nominal power

Instrument transformer compensations

To compensate for errors in instrument transformers, their amplitude error in percent and phase
angle error in minutes are configured in the meter. One minute is equal to the angle 1/60. The
errors can be specified separately for all voltages and currents. When instrument transformer
compensations are used, current and voltage are affected, as well as all quantities that arise from
these: power, energy, etc.
2-element meter

When voltage errors are compensated on a 2-element meter, this is done on phase to phase voltages
L12, L23 and L31 instead of on phase voltages. Only L12 and L23 are included in the calculation of
power and energy. Compensation of L31 has no significance in this respect. In compensation of
current errors for 2-element meters, only L1 and L3 can be compensated for because they are the
only currents measured.
Power transformer losses

Power transformer losses consist of copper and iron losses. They are expressed as percentages of
nominal power. One value is specified for active loss and one for reactive. When compensation of
losses is configured, power, energy, power factor, etc. are affected but not current and voltage.

48

Functions
Calculating loss values
Based on the nominal total power and the measured loss value in watts, a loss value can be
calculated as a percentage of nominal power. It is the loss value that is configured in the meter.
Nominal power is calculated with configured nominal current and voltage.
Nominal power: Powernom = Current nom Line voltagenom 3
Loss value: Loss value = Loss / Powernom 100

49

Functions
Power quality
Power quality encompasses voltage monitoring and harmonics measurement. Harmonics
measurement is described in the section Instant values (pg. 31).
Voltage monitoring

Voltage monitoring monitors three states: swells (overvoltage), sags (undervoltage) and interrupts.
Monitoring is activated by configuring the limit values for sags and swells, and is deactivated by
setting the limits to zero. The limits are expressed as percentages of configured nominal voltage.
Example: For an upper limit of 110% and a lower limit of 90%, and the configured nominal
primary voltage of 10 kV, the limits attained are 9 kV and 11 kV primary.
For 3-element meters, the average value for phase voltage is monitored, and for 2-element meters,
the average value for phase to phase voltage is monitored.
Sags and swells
Sags and swells shorter than one second are registered by accumulating registers. If the state lasts
longer than one second, it is instead registered in the event log with a time stamp and duration. The
accumulating counters and the event log can be read in CeweConfig.
Interrupts
When the average value for voltage falls to under 10% of the configured nominal voltage, this is
registered as an interrupt in the event log with a time stamp and duration. Duration for an interrupt
of less than three seconds, and down to 10 ms, is presented as "< 3 s".
State
Sags and swells

Duration
160 ms 5 1 s

Registered in
Accumulating counter

Duration presented as
Duration not specified

Sags and swells


Interrupt

> 1s
10 ms 6 3 s

Event log
Event log

Duration of state
<3s

Interrupt

>3s

Event log

Duration of state

5
6

Applies to basic frequency 50 Hz. Corresponding time at 60 Hz is 130 ms.


Applies to basic frequency 50 Hz. Corresponding time at 60 Hz is 8.3 ms.

50

Functions
Hysteresis

Hysteresis means that the limit for a state and the limit for when it is restored are different. This to
avoid several states being registered when the voltage level varies around a limit. The hysteresis
limit is always midway between the low or high limit and nominal voltage.
Primary voltage (example)

Percent

11 kV

110%

10,5 kV

105%

10 kV

100%

9,5 kV

95%

9 kV

90%

Limit for swells


Hysteresis limit
Nominal voltage
Hysteresis limit
Limit for sags

Duration sag

1 kV

10%

Limit for interrupt

Duration
interrupt

51

Functions
Miscellaneous
General information texts

General information texts are used in the CewePrometer for all types of information, such as the
name of the station where the meter is installed. The information texts can be shown on the meter's
display and read with CeweConfig. There are four information texts that consist of a label and an
associated text. Both the label and text can be up to 17 characters long. To present information texts
on the display, display layouts for this can be entered in a display sequence.

Example of general information texts where the label is "Owner" and the text "Cewe Instrument".

Language on display

The CewePrometers display can present texts in a number of languages. The following languages
are available in meter version 1.2.0.

52

Swedish

English

German

French

Italian

Using the display


Using the display
This section describes how the CewePrometers display can be used to read measurement values
and also to control certain functions in the meter. The display layouts are arranged in display
sequences. For details on display sequences, see Display sequences (pg. 36). A complete listing of
display layouts is presented in Appendix A Display Layouts (pg. 57).
Quadrant indicator
In most display layouts, a quadrant indicator is shown in the lower-right corner. The quadrant
indicator shows the quadrant in which energy is currently being measured. See Appendix F
Calculation Principles (pg. 67) for a definition of quadrants.
Quantity

Unit
Quadrant I

Phase

Quadrant II
Quadrant III
Quadrant IV

Measurement value Quadrant indicator

Display layout

Voltage order display layout


When the CewePrometer starts, the display layout Voltage Order is always shown for about 15
seconds. The display layout shows the order of the connected measurement voltages. If the voltage
order is correct, the current direction is indicated with a + or a -. If not all voltages are connected,
Phase missing is shown on the display. The current direction is indicated with a + if the currents
phase angle is following its voltage with a maximum deviation of 90; if it is not, a - is displayed.
Correct voltage order and current direction.

Reversed current direction on phase 1.

Reversed voltage order (no current direction shown).

53

Using the display


Navigating in display sequences
Use the buttons on the front of the CewePrometer to step through the display sequences. For more
information, see the Product description (pg. 6) section for button placement.
Step forward
Step to next layout in the display sequence or moves upward in a menu.
Menu/Apply
The Menu/Apply button is used to open the menu where one of four configurable display sequences
can be chosen. The button is used to confirm choices or initiate other activities depending on the
current display layout.
Step backward
Step to previous layout in the display sequence or moves downward in a menu.
Sealed button
A Short press to the sealed button provides access to the fifth display sequence. Press the button for
about two seconds to open the system menu. From the system menu, you can view program
versions for all modules and activate the trace function for the communication ports. For more
information, see Using the display for communication diagnostics (pg. 56)

Display layouts with choices


In certain CewePrometer display layouts, active choices can be made.
Time of Use

TOU register 2 of 8, rate 1

To choose between rates, press the Menu/Apply button for about two seconds. Use the step buttons
to change rates and Menu/Apply to confirm your choice. The rate will continue to apply for other
display layouts that show TOU registers.
Finish historical periods

Layout for finishing historical periods. #5 indicates total number of previously finished periods.

Finish a historical period means that all registers are saved and maximum demand values are reset.
The meter may be configured so that periods may be finished by pressing a button on the meter. For
more information, see the section Historical registers (pg. 45). Additionally, the display layout
shown above must be included in a display sequence. To finish a period, press the Menu/Apply
button for about two seconds.
54

Using the display


Historical registers

Historical register period 02, finished 12 July 2005, 14:14.

There are display layouts for viewing historical energy registers, external registers and maximum
demand registers. The title line with the date indicates that the display layout is showing a historical
register. To step between periods, the Menu/Apply button must first be pressed for about two
seconds. The step buttons are then used to change periods. Press Menu/Apply to return to stepping
between display layouts. The period will continue to apply for other display layouts. See Historical
registers (pg. 45).
Historical TOU registers

Historical TOU register

Switching historical periods

Switching rates

Historical registers also contain TOU registers, and for these, rates also can be changed. Choose
historical period and then press Menu/Apply to be able to change rate with the step buttons. Press
Menu/Apply to return to stepping between display layouts.
Events

No of events.

Press Apply for about two seconds


to browse through the events.

The meter registers different types of events; see the section Event log (pg. 43). This display layout
shows how many events are saved in the meter. Press Menu/Apply for about two seconds to view a
list of events in chronological order with ID numbers. Explanations of these IDs are in Appendix B
- Events (pg. 61).
Confirming alarms

By pressing Apply for about two seconds, alarms of the type internal error in the meter can be
confirmed/acknowledged. For more information on internal errors, see section Event log (pg. 43).
55

Using the display


Using the display for communication diagnostics
To facilitate troubleshooting of communications with the meter, there is a trace function integrated
into the CewePrometer. The incoming characters that the meter registers are presented on the
display. The function can be chosen from the system menu, accessible by pressing the sealed button
for about two seconds. Use the step buttons to choose one of the three communication ports: trace
opto, trace port 1 or trace port 2. Confirm with the Menu/Apply button and an empty display with a
cursor at the upper-left corner will be shown.

Layout for Trace to file for communications with the meter.

To check if communication is working:

Open an empty display.

Connect a PC or other meter reading unit to the meter.

Check the display after the call to the meter: /?sssssss!CRLFAK0Z1CRLF. The sssssss
string represents the meters serial number and it is only included when the meter must be
addressed. Z can be a digit between 0 and 7.

Functioning communications with the meter.

If you do not attain these results, try one or more of the following:

56

Check the cabling and connections between the meter and reading unit.

Check that the meter and the reading unit have the same baud rate.

With RS422 communications, it is important that there is a termination resistor installed on


the communication cable. Problems arise if the resistor is not installed or incorrectly located.
For more information, see the handbook for the RS422 converter.

Appendix A Display Layouts


This appendix presents the layouts that can be viewed in the CewePrometers display sequences.

Instant values

57

Energy registers

58

Graphs

Maximum demand values

From top to bottom, the highest to the third-highest maximum


demand value for the period is displayed. R1 indicates the rate
that was active when the maximum demand value occurred.

Historical registers

Example of historical register


active energy import. Title
line with date and time shows
that it is a historical register.

Press Apply for about two


seconds to finish the period.
#5 shows the no. of periods
that have been finished.

Historical TOU register.

Press Apply for about 2


seconds to switch historical
period.

Press Apply once more to


change rate.

Time of Use

TOU register 2 of 8.

Current season S1. Active rate for energy


Enrg and for maximum demand value
MD. Active day types for energy and
maximum demand value.
59

Miscellaneous

One of four display layouts


for general information texts
with label Owner: and
text Cewe Instrument.

Total events.

Press Apply for about two


seconds to browse through the
events.

Constants for calibration


LEDs presented as secondary
values.

Constants for calibration


LEDs presented as primary
values.

Correct voltage order and


current direction.

Incorrect voltage order.

Ratios for current and voltage


transformers CT and VT.

60

Incorrect energy direction


on L1.

Appendix B - Events
This appendix lists the events that can appear in the CewePrometer event log.
ID indicates the code that is displayed for an event on the meter's display. Category indicates one of
the following:
H
A
I

Meter event
User-defined alarm
Internal error

%1 and %2 represents data that is stored for the event.


Event
Data
Phase (%1) has reversed energy direction. %1 indicates phase with reversed
energy direction.
The meters internal clock is changed to %1 indicates the time that the
%1.
meter was changed to. The time
stamp for the event indicates the
time before adjustment.
Energy registers have been reseted.
None
Logger #%1 is reset.
%1 indicates logger that is reset.
Auxiliary power loss on meter
None
Historical register reset.
None
Registers are transferred to historical
None
period. Maximum demand value reset.
Maximum demand registers has been re- None
configured and therefore reset.
Logger #%1 has been re-configured and %1 indicates logger that has been
therefore reset.
re-configured and reset.
Measurement voltage interrupt (duration %1 indicates duration of the
%1).
interrupt.
Sag (duration %1).
%1 indicates sag duration.
Swell (duration %1).
%1 indicates swell duration.
Alarm for low measuring voltage.
None
Alarm for high measuring voltage.
None
Alarm for low power factor.
None
Alarm for voltage unbalance.
None
Alarm for current unbalance.
None
Alarm for low active power.
None
Alarm for high active power.
None
Alarm for high THD, voltage.
None
Alarm for high THD, current.
None
Alarm for high voltage harmonic #%1 on %1 indicates harmonics order
phase %2
number.
%2 indicates the phase on which
the harmonic exceeded the limit

Category ID
H
1
H

H
H
H
H
H

3
4
5
6
7

14

H
H
A
A
A
A
A
A
A
A
A
A

15
16
17
18
19
20
21
22
23
24
25
26

61

Alarm for high current harmonic #%1 on


phase %2

Alarm for long pulse at input #%1.


Alarm for short pulse at input #%1.
Alarm for missing voltage phases %1.
Configuration changed.
Calibration changed.
Initiation changed.
Energy direction is opposite of normal
direction on phase %1.
Meters firmware has been upgraded.
Internal error. Requires factory service.
Configuration parameters in the
communication module have been reset
after discovering that they were corrupt.
Configuration parameters in the I/O
module have been reset after discovering
that they were corrupt.
Configuration parameters in the
measurement module have been reset
after discovering that they were corrupt.
Initiation parameters in the measurement
module have been reset after discovering
that they were corrupt.
Calibration parameters in the
measurement module have been reset
after discovering that they were corrupt.
Configuration parameters in the display
and register module have been reset after
discovering that they were corrupt.
A historical period has been reset after
discovering that is was corrupt.
MD registers have been reset after
discovering that they were corrupt.
Measurement module missing or cannot
be started.

62

value.
%1 indicates harmonics order
number.
%2 indicates the phase on which
the harmonic exceeded the limit
value.
%1 indicates inputs order
number.
%1 indicates inputs order
number.
%1 indicates one more missing
voltage phases.
None
None
None
%1 indicates one or more phases
where the energy direction is not
the normal.
None
None
None

27

28

29

30

H
H
H
H

42
43
44
45

H
I
I

46
1000
1001

None

1002

None

1003

None

1004

None

1005

None

1006

None

1007

None

1008

None

1009

Appendix C Serial Communication Port


RS232/422
Serial communication port RS232
The PIN configuration is compatible with a standard PC COM port, 9-pin, D-sub connector with a
few exceptions. Some of the handshaking pins are not used.
Hardware
Contact in meter
Communication protocol
Handshaking
Data format
Baud rate

RS232 serial communication port


D-Sub, 9 pin, male contact
IEC62056-21/IEC1107 Mode C, programming mode
RTS always active, CTS never active
1 start bit, 7 data bits, 1 stop bit, even parity
Port #1 30019200 bps, Port #2 120019200 bps
1
2
3
4
5

RD
TD
Signal GND

6
7
8
9

RTS
CTS
-

PIN configuration for RS232 connector on meter

Serial communication port RS422


A twisted pair of wires is used as the receiving bus that all meters monitor, and another pair is used
as a transmitting bus that all meters send on. The communication equipment uses an RS422 port or
an RS232 port through an RS232-to-RS422 converter. The converter must be designed for 4-wire
communications (full duplex). Recommended converters are Westermo MA45, MD45 and
MDW45.
Hardware
Connector in meter
Communication protocol
Data format
Baud rate

RS422 serial communication port


5-pole connector (Phnix)
IEC62056-21/IEC1107 Mode C, programming mode
1 start bit, 7 data bits, 1 stop bit, even parity
Port #1 30019200 bps, Port #2 120019200 bps
1
2
3
4
5

RX+
RXTXTX+
Signal GND

PIN configuration for RS422 connector on meter

63

IEC62056-21/IEC1107 optical communication port optical port


Place the optical head so that the IR receiver and transmitter are in the optimum positions for
exchanging signals. This position is attained when the optical head is centred over the optical port
with the cable hanging straight down. Visible ambient light (daylight, light bulbs and fluorescent
lamps) can interfere with the transfer of IR signals if the light is strong, and cause communications
to fail. Visible light from most light sources contains a certain amount of IR light that can cause
interference. To ensure reliable communications, avoid installing the meter where it can be
subjected to strong ambient light when communications via the optical port are required.
Hardware
Communication protocol
Data format
Baud rate

64

IEC62056-21/IEC1107 optical communication port


IEC62056-21/IEC1107 Mode C, programming mode
1 start bit, 7 data bits, 1 stop bit, even parity
3009600 bps

Appendix D - Module block diagram

Communications between modules


The intelligent function modules that make up the CewePrometer are autonomous units with their
own micro-controllers and own firmware. These are connected in an internal network through an
internal bus. For internal communications, information transfer is via a CAN bus. This bus is
designed for reliable operation in harsh electrical environments, primarily in motor vehicles, and
has high transfer reliability and error detection.

65

Appendix E Frequently Asked Questions


Display sequences
Which information is not available via the CewePrometer's display?
Graphs for discrete harmonics, vector graphs on 2-element meters, logged values and counters for
short sags and swells. This information cannot presently (meter version 1.2.0) be read via the
meter's display. The information is accessible via CeweConfig and via the communication protocol.

Meter clock
What happens if a new successive adjustment is made when one is already underway?
The current adjustment will be stopped and the new begun.
What happens if the meter time is set within an hour before the change from daylight savings
time to standard time?
If the meter time is set to daylight savings time before the change, it will be set back to standard
time automatically at the approaching time change. If the meter time is set to standard time before
the change, it will not be affected at the approaching time change.
What happens if more than one input is used for clock synchronisation?
Several inputs can be used simultaneously to receive synchronising pulses even if this is unlikely.
How are logged values affected by the daylight savings time change?
According to the example:
Change from standard to daylight savings time
2005-03-28 01:30 3467.4 kWh
2005-03-28 03:00 3523.2 kWh
2005-03-28 03:30 3602.5 kWh
Change from daylight savings to standard time
2005-10-31 02:30 3345.4 kWh
2005-10-31 02:00 3756.7 kWh
2005-10-31 02:30 4132.8 kWh
2005-10-31 03:00 4583.1 kWh
Logged values during daylight savings time are labelled with the flag S.
What are the possible consequences of adjusting the meter clock immediately?
Because CeweConfig does not permit the clock to be adjusted immediately without simultaneously
resetting the logging memory, this is purely hypothetical. If it were to be allowed, then if the clock is
adjusted forward over one or more interval limits, there will be empty spaces in the logging
memory and values will be lacking for one or more intervals.
It the clock is adjusted backward over one or more interval limits, one or more values will exist
with the same time stamps. In the worst event, this can result in data read from the meter not
matching the expected period.

66

Appendix F Calculation Principles


Current and voltage
RMS values for current and voltage are calculated as the root of the sum of squares for the
harmonic components up to the 31th harmonic.
I1
I2

The currents first harmonic component (fundamental) specified as peak value.


The currents second harmonic component specified as peak value, has doubled frequency
compared to the first harmonic.
( I 12 + I 22 + ...I 312 )

I RMS =

Calculated phase to phase voltage

Phase to phase voltage is calculated from the phase voltages fundamental vectors. This calculation
is only made in the 3-element meter; in the 2-element meter, the phase to phase voltage is
measured.

U 121 = (U 112 + U 212 2 U 11 U 21 cos(U 11 p U 21 p)


Calculated I2

In the 2-element meter, I2 is not measured but is calculated for monitoring purposes. It is calculated
from the currents fundamental vectors.
I 21

Current I2s fundamental harmonic.

I 21 = ( I11 + I 31 cos( I11 p I 31 p )) 2 + ( I 3 sin( I11 p I 31 p)) 2

Power
Harmonic component power

The calculations below are for active power, the calculations for reactive are identical except for
that cos-functions are replaced with sin-functions.
P1n
Pn

Active power in L1 is calculated for harmonic component n.


Total active power is calculated for harmonic component n.
Phase angle between harmonic component U1n and I1n

3-element meter:
P1n = U 1n I1n cos( n )
2-element meter:
For 2-element meters, only the total power is calculated in each harmonic component.
67

1n
2n

Phase angle between harmonic component U12 n and I1n


Phase angle between harmonic component U 32 n and I 3 n

Pn = U 12 n I1n cos(1n ) + U 32 n I 3n cos( 2 n )


Active and reactive power

Active and reactive power is calculated as the sum of harmonic component power up to 31st
harmonic. The calculation is made with plus and minus signs, where negative phase angle
represents export direction and positive represents import direction.

P
P1
Q
Q1

Total active power


Active power in L1
Total reactive power
Reactive power in L1

P = P1 + P 2 + P3
Q = Q1 + Q 2 + Q3
For 2-element meters, two elements are added instead of three.
Apparent power

S
S1

Total apparent power


Apparent power in L1

S = P2 + Q2
S1 = P12 + Q12

Energy
Energy is calculated by integrating power (P, Q and S) over time.
Definition of quadrants
The term phase angle is described under its own heading below.

Quadrant I: phase angle 190


Quadrant II: phase angle 90180
Quadrant III: phase angle -180(-90)
Quadrant IV: phase angle (-90)0
Active energy

Active energy is calculated for import and export. The direction is controlled by the sign for active
power (+ import, export).
Active energy import: quadrant I and IV
Active energy export: quadrant II and III

68

Reactive energy

Reactive energy is calculated for four quadrants. The quadrant is controlled by the sign for active
and reactive power (e.g., active power >= 0 and reactive power >= 0 corresponding to quadrant I).
Reactive energy import: quadrant I and II
Active energy export: quadrant III and IV
Reactive energy inductive: quadrant I and III
Reactive energy capacitive: quadrant II and IV
Apparent energy

Apparent energy is calculated for import and export. The direction is controlled by the sign for
active power; apparent energy is registered for the direction that the active energy has during the
same period.
Apparent energy import: quadrant I and IV
Apparent energy export: quadrant II and III

Power factor
pf ( L1) = P1 / S1
pf (Tot ) = P / S

The power factor is calculated without signs and is thus always positive.

Phase angle
( L1) = U 1 p I1 p
Phase angle for an element is calculated from the fundamental phase angles.

(Tot ) = arctan(P1 fund / Q1 fund )


Total phase angle is calculated from fundamental power.
Phase angle values specified between 180 and 180.

THD
Total harmonic distortion
THDeur =

I 22 + I 32 + ...I n2
I 12 + I 22 + ...I n21

100%

Where I 1 I n are the currents harmonic components. The calculation is made in the same ways
for current and voltage.

69

Transformer compensations
Instrument transformer compensations

Instrument transformer compensations are specified as error in percent for amplitude, and as error
in minutes (one-sixtieth of a degree) for phase angle. These errors can be specified separately for all
voltages and currents.
AmpErr
PhaseErr
u meas
u

Amplitude error as percent


Phase angle error in minutes
Measured voltage
Compensated voltage

Amplitude compensation

u = u meas /(1 + AmpErr )


Phase angle compensation

up = up meas PhaseErr
In 3-element meters, phase voltages are compensated and in 2-element meters, phase to phase
voltages. The same calculation is used for all currents and voltages.
Power transformer losses

Power transformer losses consist of copper and iron losses. They are expressed as percentages of
nominal power. One value is specified for active loss and one for reactive.
Iron loss
FeLoss
P1meas
P1
Nom

Active iron loss as percent of nominal power.


Measured power.
Compensated power.
Nominal power per phase

P1 = P1meas + FeLoss Nom


Copper loss
CuLoss
i
inom

Active copper loss as percent of nominal power.


Phase current.
Nominal current.

P1 = P1meas + (i / inom ) 2 CuLoss Nom


Corresponding calculations made for active and reactive power.

70

Definition of phase order

Correct phase order (123) corresponds to phase position:


U1 p

Phase position for U1

U 1p = 0
U 2 p = 120
U 3 p = 120
The same system is represented graphically below. The vectors rotate in an anti-clockwise
direction.

71

Appendix G Material Declaration


CewePrometer-R
Enclosure
Front plate
Rear panel
Side plates
Top & bottom plates
Handle

Steel sheet, galvanised, chromated


Steel sheet, galvanised, chromated
Steel sheet, galvanised, chromated
Steel sheet, galvanised, chromated
Steel, chrome plated

Plastic parts
Inner window
Outer window
Side pieces
Battery cover

Polycarbonate V0
Polycarbonate V0
Polycarbonate/ABS V0
Polycarbonate/ABS V0

CewePrometer-W
Enclosure
Bottom plate
Meter cover
Inner window
Outer window
Battery cover
Terminal cover

Polycarbonate/ABS V0
Polycarbonate/ABS V0
Polycarbonate V0
Polycarbonate V0
Polycarbonate/ABS V0
Polycarbonate/ABS V0

Connection terminal, current/voltage


Terminal body
PPO + 10 % GF V0
Terminals
Brass
Screws
Steel

72

Appendix H Connection Data


Safety
Isolation voltage according to IEC 62052-11, IEC 62053-22, IEC 62053-23
WARNING: Live parts inside meter cover. Always disconnect all wires carrying dangerous
voltages before opening the meter.

Current circuit
Measurement current (I in)

Burden

1(2) A
5(10) A
1(6) A
<0.4 VA/phase

Voltage circuit
Measurement range (Un)

3-wire system:
4-wire system:

Frequency range
Burden
Max. overvoltage

3x57120 V
3x120-240 V
3x57/99120/208 V
3x120/208240/415 V

4565 Hz (Class 0.2S 4555 or 5565 Hz)


<0.2 VA/phase
1.3 x Un (limited by protective varistors)

Auxiliary power
Separate auxiliary voltage Uaux
Power Consumption

40130 VDC/35100 VAC


100300 VDC/85265 VAC
<6 VA

Digital outputs
Type
Relay performance

Solid-state relay (MOSFET, bi-directional)


0.2 A, 110 V AC/DC (varistor-protected)

Digital inputs
Type
Voltage (AC or DC)
Incoming resistance

Optical switch
48220 V
20 kohm

73

Notes

Notes

Cewe Instrument AB
Box 1006 SE-611 29 Nykping SWEDEN
Tel: +46 155 775 00 Fax: +46 155 775 97
e-mail: info@ceweinstrument.se www.ceweinstrument.com
A0172e-1a

You might also like