SPECIFICATION OF HEAMODIALYSIS MACHINE

STT

SPECIFICATION

Maximum
Score
15
2
2
1

I
1

General specification (15 Score)


Origin country
- G7 countries
- Other country

Year of manufacture
- 2010
1
- < 2010
0
Certificate
- ISO; FDA or CE
2
- No certificate ISO, FDA or CE
0
Manufacture must has 300 haemodialysis machines which are
installed in Viet Nam
- 300 machines
2
- < 300 machines
0
Maintenance at least 3 months /time during warranty time
- > 24 months
2
- > 18 months
1
- < 18 months
0
Time of good delivery
- < 90 days
1
- > 90 days
0
Installing, testing and training for end-Users

3
4

6
7

9
II

Minimum
Score
12
1

Bidder has ability to supply spare pare after warranty time over
1
- > 05 years
1
- < 05 years
0
Requirement of specification
Bt buc
- 15 inch LCD ( touch screen ), high resolution and monitor can
rotate to face the operation allowing doctor, nurse, technician or
patient to easy read information any position without moving the
monitor.
- Automatic Blood Pressure measuring ( ABPM )
- Automatic Blood Pressure Stable by Biologic RR mechanism
- Automatic mixing Acetate and Bicarbonate concentrate
- The machine can using many profiles : UF profile , temperature
profile, conductivity profile, Natri profile, HCO3 profile, Dialysate
profile, Heparin Profile.
- Disinfection mode include : heat and chemical
- Evaluation of efficiency of haemodialysis by Kt/V calculation
( K: clearance of Urea, t : time of dialysis, V: Blood volume)

- Oversea trip
2
- In the hospital which install machine
1
Bidder must has engineers with Certification of service training
about repairing Hemodialyis machine
- > 04 engineers
2
- < 04 engineers
0

Bt buc

Calculate Kt/v (efficiency haemodialysis) during dialysis.

Warning when really Kt/V value < setting Kt/V value.

- The machine can save all data treatment information patient by


chip card.
The patient use chip card to identify themselves
and their medical data are automatically
displayed on the screen, with this software the
hospital staff no longer need to input data
manually because individual treatment profiles
can be pre-set.
- The machine still is working at least 20 minutes by backup
battery after failure power supply.
- Power supply : 220 V , 1 phase , 50/60 Hz.
Accessories and consumables for each machine:
+ Dialyzer holder : 01 pc
+ Infusion Venous pole : 01 pc
+ Tray canister disinfection : 01 pc
+Dialyzer : 5 Pcs
+ Bloodline : 5 sets
+ Fistula Arterial and venous : 5 sets
+ Concentrate dialysis : 5 can/ 10 L ( including : 02 Acetate
canisters, 01 canister Acid concentrate : 02 canisters Bicarbonate)
+ Transducer protector : 5 sets
+ Supply full set of document including:
Users manual by English and Vietnamese : 01 set

III

Detail of specification (85 Scores)

85

78

- Battery power supply 20 minutes after power failure


+ Working 20 minutes
+ No battery or working < 20 minutes
2

5
0

Many different therapies in 01 machine :

Volume Controlled through balancing chambers


Sequential ultrafiltration
Using double needle on single pump machine ( standard
therapy)
Single needle
Ultrafiltration rate : From 0 n +3000ml/h or t: 0 to
+5000ml/h option HDF

- All above items


- Not enough items

5
0

There is at least 10 UF programmable default profile (ultrafiltration


profile )
- 10 Uf default profile
- 7 Uf default profile
- < 7 Uf default profile

10

11

12

10

10

2
1

Warning when treatment parameters differents with setting


parameters
- Yes
- No

5
0

Automatic mixing Acetate and Bicarbonate concentrate


- Yes
- Acetate or Bicarbonate only

10
0

Calculation Kt/V during the dialysis


- Yes
- No

1
0

15 inch LCD ( touch screen ), high resolution and monitor can


rotate to face the operation allowing doctor, nurse, technician or
patient to easy read information any position without moving the
monitor.
- Yes
- No

1
0

heparin volume profile


- Yes
- No

1
0

Dialysate flow profile


- Yes
- No

1
0

HCO3 conductivity profile


- Yes
- No

1
1
0

Na conductivity profile
- Yes
- No

3
1
0

temperature profile
- Yes
- No

2
0

13

Displays treatment parameters and setting parameters


- yes
- No

14

15

16

17

18

19

20

21

22

1
0

Safety air detector by ultrasound measurement in the tube


- Yes
- No

2
0

Venous pressure (PV) from 20 to +390mmHg.


- Yes
- No

2
0

Arterial pressure (PA) From -400 to +400mmHg


- Yes
- No

2
0

Dialyzer inlet pressure ( PBE ) From 0 to +700mmHg


- Yes
- No

5
0

Dialysis fluid preparation


Total conductivity : 12.5 16 mS/cm,
Conductivity bicarbonate : 2 - 4mS/cm or t 4 to 7mS/cm
- Yes
- No

2
0

Chemical and heat disinfection


- yes, both chemical and heat
- Chemical or heat only

5
0

Material of machine can prevent oxidize and corrosion


- Yes
- No

5
5
3

Heparin pump for 10 to 30 ml syringes


- Yes
- No using 10 ml syringe

2
1

Blood pump : 02 roller, peristaltic pump with delivery rate 50 to


600 ml/ min
- 02 roller, peristaltic pump
- 03 roller, peristaltic pump

1
0

23

Degassing system by mechanical via controlled negative pressure


- Yes
- No

24

25

26

27

28

Total score:
Total:

- Maximize score = 100 score


- minimize score = 90 score

100

90

5
0

Upgrade and networking capability


- Yes
- No

5
0

Automatic Blood Pressure Stable by Biologic RR


- Yes
- No

5
0

Automatic Blood Pressure measuring ( ABPM )


- Yes
- No

2
1
0

Saving all data treatment information patient by chip card.


- Yes
- No

1
0

Displays the warning signals by sound and light also displays alarm
reason help healthcare staff solve the problem fast with confidence.
- Yes
- Warning or alarm by light signal but no displays alarm reason
- No warning or alarm

3
0