Professional Documents
Culture Documents
U309 Ta10zg1b Dop e
U309 Ta10zg1b Dop e
ZG1b
Sensor TA10 ... ZG1b (left) for use with transducer U10a (above right) and hand-held flowtherm NT, HTA and HTA-Ex
(below right)
Measurable variable
standard velocity Nv, standard
volume flow NV/t, mass flow
proportional
norm:
temperature tn = +21 C,
pressure pn = 1014 hPa
temperature t
(hand-held flowtherm NT, HTA,
HTA-Ex)
Medium
clean gases, gas mixtures:
air, nitrogen, methane, natural
gas, argon, carbon dioxide,
helium, sulphur hexafluoride,
landfill gas ...
measuring
(heat transfer method)
laminar flows in clean rooms
Design
or machines
probe
- in outgoing air, burner supply
air and draughts
- in climatic applications
Advantages
- in air in the rough vacuum
high measuring dynamics Nv
range with pressures greater
(up to 1 : 1000)
than 200 hPa abs.
small starting value: 0.2 m/s
minimum reaction time
low measuring uncertainty, even Particles, condensation,
at lowest flow velocities
humidty in the gas
direct air/gas mass flow Charges in the gas caused by
proportional measuring, largely
particles such as dust and fibres
irrespective of working pressure
do not affect the measurment,
and temperature
as long as abrasion and
sensor has no moving parts
agglomeration do not occur on
greater working temperature
the sensor
and pressure ranges
relative gas humidty of less than
high time yield
100 % does not affect the
corrosion resistant
measuring uncertainty if the
working temperatures are less
lower pressure drop due to small
than +40 C
dimensions
U309_TA10ZG1b_DoP_e_120412
http://www.hoentzsch.com
1/6
Thermischer Strmungssensor
Thermal
Flow Sensor TA10 ... ZG1b
TA10
... ZG1b
140
(5)
p16
(6)
Ex
(7)
ZG1
(8)
Basic types
Type
TA10-185 GE 140 / p16 ZG1b
Article No.
b013/300
b013/301
b013/302
b013/303
(3) Medium
Air, clean gases, gas mixtures with ratio of mixture consistent
When calibrating/justifying sensor and evaluation unit for gases other than air the slightest possible
measuring uncertainty is only achievable by ensuring fixed allocation of sensor to evaluation unit.
Measuring ranges
2/6
0.2 ...
30 m/s
Article No.
v_ta10_1b_30
0.2 ...
0.2 ...
60 m/s
120 m/s
v_ta10_1b_60
v_ta10_1b_120
0.2 ...
0.2 ...
150 m/s
180 m/s
v_ta10_1b_150
v_ta10_1b_180
0.2 ...
200 m/s
v_ta10_1b_200
http://www.hoentzsch.com
U309_TA10ZG1b_DoP_e_120412
'200 m/s'
77
220
356
513
912
1425
2050
3210
5700
12820
22800
35600
142500
256
730
1180
1710
3040
4750
6840
10600
19000
42750
76000
118800
475000
154
440
713
1030
1820
2850
4100
6410
11400
25650
45600
71200
28500
192
550
890
1280
2280
3560
5130
8020
10700
32060
57000
89100
356300
231
660
1070
1540
2740
4280
6160
9620
17100
38480
68400
106900
427600
Standard volume flow measuring range specifications with centric positioning of the sensor,
irrotational afflux and amply-dimensioned input and output section (see Instruction Manual).
The profile factor PF describes the ratio of average flow velocity in the measurement cross section
and the flow velocity measured from the sensor. The afore-mentioned operating conditions apply.
Measuring uncertainty / Time constant
Measuring uncertainty for flow velocity Nv
less than/equal to 40 m/s
: 2 % of test value + 0.02 m/s
greater than 40 m/s
: 2.5 % of test value
Time contstant
: in seconds
Article No.
Ex ia IIC T4
Category 2G (Zone 1)
necessary when using with HTA-Ex
Ex ia IIC T4
Category 1/2G (Zone 0)
necessary when using with U15-Ex
Ex nA IIC T4
Category 3G (Zone 2)
Ex tc IIIC T135C
Category 3D (Zone 22)
in combination with U10a and flowtherm NT
ta10_1b_ex1
ta10_1b_ex0
taex2
(8) Design
Probe; as in Drawing ZG1b (see Page 1)
U309_TA10ZG1b_DoP_e_120412
http://www.hoentzsch.com
3/6
transducer U10a
hand-held HTA
flowtherm NT
for Ex applications
4/6
transducer U15-Ex
Ex ia IIC T4 Category 1 (Zone 0)
hand-held HTA-Ex
Ex ia IIC T4 Category 2 (Zone 1)
http://www.hoentzsch.com
U309_TA10ZG1b_DoP_e_120412
Accessories
Calibration certificate
probe guide piece
SFB 10 E-35 / G " ZG5
as in Drawing 5
Description
Article No.
klbneu
b004/503
b004/600
b004/601
Drawing 5
Drawing 6
Drawing 7
U309_TA10ZG1b_DoP_e_120412
http://www.hoentzsch.com
5/6
Accessories (cont.)
HG10/18A-130
VS18A-350
Ball valve
Description
Article No.
b099/948
Hntzsch GmbH
Gottlieb-Daimler-Strae 37
D-71334 Waiblingen (Hegnach)
Telefon
+49 7151 / 17 16-0
Telefax
+49 7151 / 5 84 02
E-Mail
info@hoentzsch.com
Internet www.hoentzsch.com
6/6
b099/947
b099/010
b099/951
b004/900
U309_TA10ZG1b_DoP_e_120412