Professional Documents
Culture Documents
Pikkula ABME 2004
Pikkula ABME 2004
pp1282-ambe-490790
17:49
Annals of Biomedical Engineering, Vol. 32, No. 8, August 2004 (2004) pp. 11311140
Department of Bioengineering, Rice University, Houston, TX; 2 Current address: G.R. Harrison Spectroscopy Laboratory,
Massachusetts Institute of Technology, Cambridge, MA; and 3 Department of Plastic Surgery, University of Texas
M.D. Anderson Cancer Center, Houston, TX
(Received 16 May 2003; accepted 31 March 2004)
INTRODUCTION
Laser radiation is currently utilized to remove unwanted cutaneous structures such as hypervascular lesions,
rhytides, and hair.13,20,30,39 The unwanted structures are
heated by prescribing a particular wavelength which is
preferentially absorbed by the targeted chromophore (e.g.,
hemoglobin, water, or melanin), and specifying an appropriately short laser pulse to limit thermal diffusion from
the chromophore. These targeted chromophores undergo
photothermolysis, in which the energy deposited to the desired chromophore results in its thermal destruction.3 Despite the spatial confinement of heat within the targeted
chromophore, light absorption by the overlying epidermal
melanin which takes place over a broad spectral range26 can
Address correspondence to Bahman Anvari, Department of Bioengineering, MS 142, Rice University, P.O. Box 1892, Houston, TX 77251.
Electronic mail: anvari@rice.edu
1131
C 2004 Biomedical Engineering Society
0090-6964/04/0800-1131/1
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
1132
17:49
PIKKULA et al.
2
dimpact vimpact
(2)
q(t)
,
Ts (t) Tdroplet
(3)
where Ts is the temperature of the substrate at z = 0 (computed by solving the forward heat transfer problem once q(t)
was estimated) and Tdroplet is the cryogen droplet temperature just prior to impacting the substrate, measured by inserting a 60-m bead diameter thermocouple (CHAL-001,
OMEGA Engineering, Inc., Stamford, CT) into the
cryogen spray. Since calculation of utilizes measured
and estimated temperatures [i.e., Tdroplet and T (z = 0, t),
respectively], all thermal resistances between the droplet
and substrate (i.e., at cryogen-substrate interface and within
the cryogen film on the surface) are taken into account.
The dynamic thermal conductance was time averaged over
the duration of the cryogen spurt, resulting in an average
value, .
Surface Tension
The single capillary rise method is a classical method
of quantification of surface tension.22 To simplify the
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
17:49
(4)
g( )(h 1 h 2 )
2(1/r1 1/r2 ) cos
(5)
Images of the capillaries were captured with a CCD camera and an attached a zoom-macro lens, described in the
next section, using ambient lighting, and NIH Scion Image (www.scioncorp.com) was utilized to measure h 1 h 2
and .
EXPERIMENTAL METHODS
Cryogen Type and Delivery Systems
The cryogen used in this study was refrigerant R134a, 1,1,1,2 tetrafluoroethane (National Refrigerants, Inc.,
Rosenhayn, NJ), which is a nontoxic, environmentally compatible, Freon substitute2,23 (boiling point 26 C at 1
atm) used clinically for epidermal protection and pain reduction during cutaneous laser therapy. Spurt duration was
set to 200 ms, and controlled by a programmable digital
delay generator (DG535, Stanford Research Systems, Sunnyvale, CA).
Droplet velocity and diameter were altered by three
methods. First, immersing the cryogen canister into a warm
water bath led to increased pressure within the canister,
causing a more forceful release of the cryogen at a higher
ejection velocity. The second method was to change the
spraying distance between the substrate and cryogen delivery device. Since droplets evaporate and cool in flight,
adjusting the spraying distance also alters droplet impact
temperature. However, droplet velocity and diameter could
be independently varied from the droplet temperature when
1133
changing the pressure of the cryogen canister. Third, different cryogen delivery devices were used to produce droplet
diameter and velocity dependent on the design of the device.
Three types of cryogen delivery devices were used. The
first device (Device 1) was a standard automobile fuel injector with a 1-mm diameter orifice and an attachment nozzle
with a length of 2 mm and 1.5-mm orifice diameter that
produced a uniform spray cone with the following spray
characteristics (standard deviation) at ambient conditions:
average droplet diameter of 26 2.1 m and Sauter mean
diameter of 66 17 m at a spraying distance of 100 mm.27
The Sauter mean diameter is commonly used to quantify
droplet size during spraying processes and is calculated as
the diameter of a single droplet whose volume to surface
area ratio is equal to the sum of the volume over the sum of
the surface area of all droplets in the droplet set:
N
SMD = i=1
N
i=1
Di3 n i
Di2 n i
(6)
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
1134
17:49
PIKKULA et al.
removal. Three surfactants (Table 1), Brij 30, Tween 20, and
polyethylene glycol (PEG) 300, were chosen for this study
based on their solubility in R-134a, current applications in
drug delivery, and low toxicity.7 These three surfactants are
among the highest for solubility in R-134a.7
Several concentrations (w/w) were used for each surfactant (Table 1). The upper limit of each surfactant concentration in R-134a was determined by the maximum reported
solubility.7 If excessive foaming was present after spurt termination, a decreased concentration was used.
Surfactants were added to the pressurized canisters by
cold filling which consisted of attaching a regulator to
the canister, and placing it in a cooler filled with dry
ice to reduce the vapor pressure of the cryogen to atmospheric levels. After several minutes, temperature of the
fluid within the canister was well below the boiling point of
R-134a allowing the removal of the regulator and exposing
the contents to atmospheric pressure without releasing its
contents. Surfactants were subsequently added via a calibrated pipette to create a surfactant/R-134a mixture in a w/w
proportion.
In some experiments, surfactants were spread on the substrate surface to reduce the cryogen surface tension. In these
cases, two drops (5060 l) of surfactant were placed onto
a Kimwipe (Kimberly-Clark, Roswell, GA), and spread
manually on the substrate to provide an even distribution of
surfactant on the surface (area 16 cm2 ). When the surfactant was spread on the surface, the resulting surface tension
was not measurable. Whether the surfactant was added to
the cryogen or spread on the substrate surface, a Kimwipe
was used to remove any residue from the substrate from the
prior spurt.
Normalized Cryogen Film Height
Although the absolute cryogen film height was not measured, we normalized the film height for a certain spraying
condition compared to those under different spraying conditions. The apparent cryogen film height (H ) on the substrate
was estimated as
m
H=
,
(7)
A
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
17:49
1135
Common name
Chemical name
Chemical
composition
% w/w Proportions
used
Brij 30
Polyoxyethylene(4)
lauryl ether
1.8%
0.3%, 1.8%
Tween 20
Polyoxyethylene(20)
sorbitan monolaurate
0.1%
0.02%, 0.1%
PEG 300
Poly(oxy-1,2-ethanediyl),
alpha-hydro-omega-hydroxy
4%
0.5%, 1%
spurts immediately preceding the laser pulse. Each cooling and irradiation combination utilized either of the two
cryogen delivery devices which produced droplets with the
greatest (Device 1) and least (Device 2) Weber numbers
with the cryogen canister at room temperature. Following
laser irradiation, 6-mm punch biopsies were excised and
fixed in 10% buffered formalin and hematoxylin & eosin
stained for histological analysis of thermal damage. Thermally mediated morphological damage was characterized
by a score of 0 (no observable damage), 1 (nuclei shrinkage), 2 (<10% dermal-epidermal separation), 3 (10% to
50% dermal-epidermal separation), and 4 (>50% dermalepidermal separation). The percentage of separation was
quantified by the linear distance ratio of the separated to intact tissue at the dermalepidermal junction for the portion
of the histological section exposed to laser irritation.
RESULTS
Laser Irradiation of ex Vivo Human Skin
in Conjunction with CSC
We sought to determine if altering the cryogen droplet
and velocity (hence, Weber number) could change the
amount of epidermal protection ex vivo in response to laser
irradiation. Skin samples were obtained from consenting patients undergoing autograft breast reconstruction using abdominal skin. CSC and laser irradiation were subsequently
performed on the samples. The study was carried out under
a protocol approved by the institutional review boards of
Rice University and The University of Texas M. D. Anderson Cancer Center.
One of each Fitzpatrick skin types V and VI (medium
brown and dark brown, respectively) at room temperature
were irradiated using the Candela (Wayland, MA) Vbeam
pulsed dye laser (595 nm wavelength). The temporal pulse
profile in this laser consists of a macropulse in which a
train of four 100-s micropulses are placed in equal time
intervals within the macropulse. In our experiments, the
macropulse was set to 6 ms; hence, delivering four 100-s
micropulses. Depending on the size of the excised skin,
one or two sites were cooled and irradiated using a combination of 10 or 15 J cm2 , and 100- or 200-ms cryogen
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
1136
17:49
PIKKULA et al.
Surface tension
SD (mNm1 )
0 (pure R-134a)
0.3 (Brij 30)
1.8 (Brij 30)
15.5 0.3
13.1 0.9
Not measurable
due to foaming
13.7 0.9
15.2 0.4
15.6 0.8
13 0.6
FIGURE 4. Loglog plot of time-averaged thermal conductance, , versus Weber number for a 200-ms spurt.
FIGURE 3. Loglog plot of heat removal, Q, versus Weber number for a 200-ms cryogen spurt.
(8)
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
17:49
FIGURE 6. Heat removal (black) and average thermal conductance, (grey), for a 200-ms spurt versus the normalized cryogen film height, .
1137
(standard deviation) of protect = 1.93 1.77 when using the highest Weber number device (Device 1) as compared to the lowest Weber number device (Device 2). For
instance, a decrease in the thermal damage score of approximately 2 would reduce epidermal damage from between
10 and 50% dermalepidermal separation to that of just nuclei shrinkage. Comparing protect to a null hypothesis (no
increase in epidermal protection when using a spray with
higher Weber number droplets), protect was statistically
larger ( p < 0.05). The relatively large standard deviation
from the mean value of protect is likely due to biological
variation and a digression of up to 3% from nominal laser
fluences. In all cases, the device that produced the highest
Weber number provided equal or greater epidermal protection when compared to the device that produced the lowest
Weber number droplets.
DISCUSSION
The Weber number is sensitive to the variations in droplet
velocity (especially at higher velocities), and slight changes
in velocity could completely mask the effects of reducing
the cryogen surface tension. On the basis of the results of
this study, increased heat removal was not observed with
decreased cryogen surface tension due to the effects of the
variations in velocity of the cryogen droplets. These variations appear to obscure the expected increase in heat removal by the altered cryogen surface tension; nevertheless,
we have observed that a thinner film generated by a high Weber number spray allows for increased heat removal (Fig. 6).
Additionally, at high Weber numbers (i.e., We > 10) such
as the values described in this text (We > 800), the changes
in kinetic energy dominate the changes in surface energy of
the droplet. When We 1, we would expect an equal effect
of the two energies, and if We < 1, the surface tension will
dominate.
The spray impacts the film and substrate, and its energy
is transferred in the radial direction creating a larger sprayed
area; therefore, film diameters increase with greater Weber
numbers, resulting in thinner films.11 Our results agree with
Ref. 11 and indicate that a thinner film is beneficial in increasing heat removal (Fig. 6), supporting the theory that the
cryogen film acts as an impediment to heat transfer (Fig. 4),
where thermal conductance increases with increased Weber
number (i.e., thermal resistance decreases with higher Weber number values). Possible methods to enhance heat removal by CSC would be to decrease the cryogen film height,
or equivalently, induce deeper droplet penetration into the
film, both of which may be occurring using droplets with
greater Weber numbers.
From Eq. (8), the relative contribution of the parameters
influencing heat removal can be determined. Considering
the range of dimpact and impact and their respective exponents, the influence of impact on heat removal is approximately six times that of dimpact , which was ascertained by
P1: KEE
Annals of Biomedical Engineering [AMBE]
1138
pp1282-ambe-490790
17:49
PIKKULA et al.
FIGURE 7. H&E histological sections of ex vivo Fitzpatrick type V skin cooled with a 100-ms cryogen spurt immediately prior to
pulsed laser irradiation with a fluence of 10 J cm2 for Weber numbers of (A) 1,100 (Q = 26.9 kJ/m2 ), and (B) 5,100 (Q = 33.2 kJ/m2 ).
Arrows indicate areas of slight nuclei shrinkage and dermal-epidermal separation.
P1: KEE
Annals of Biomedical Engineering [AMBE]
pp1282-ambe-490790
17:49
ACKNOWLEDGMENTS
This study was supported in part by the Institute
of Arthritis, Musculoskeletal, and Skin Disease (1R01AR47996) at NIH; Texas Higher Education Coordinating
Board, Candela Corporation to B.A.; and a Student Research Grant from the American Society for Lasers in
Medicine and Surgery to B.M.P.
REFERENCES
1
15
1139
P1: KEE
Annals of Biomedical Engineering [AMBE]
1140
34
pp1282-ambe-490790
17:49
PIKKULA et al.