Professional Documents
Culture Documents
Transport X Manual English PDF
Transport X Manual English PDF
Operation Manual
Version 1.5
KELMAN LTD.
Lissue Industrial Estate East
Lisburn
Co. Antrim
N. Ireland
Tel : +44 (0)28 92622915
Fax : +44 (0)28 92622202
Email: transportx@kelman.co.uk
Website: www.kelman.co.uk
(EN) Notes
Mode of Operation of TRANSPORT X
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
Exhausted Gases
Gases exhausted by TRANSPORT X are at concentrations which are non flammable and non toxic. All
exhausted gases are quickly diluted into surrounding atmosphere.
Warnings
There are no serviceable parts inside the equipment.
For continued fire protection, use only approved and recommended fuse.
Do not use equipment if power cord is visibly damaged or worn out or missing or damaged PE pin.
Use only safety approved power cord.
Do not operate TRANSPORT X (top cover open) in rain conditions.
Unit should be stored with lid closed when not in use to avoid accidental damage during transit
Equipment should not be left operating unsupervised.
Gaz dchappement
Les gaz produits par le TRANSPORT X sont non inflammables et non toxiques. Tous les gaz
dchappement sont rapidement dilus dans l'atmosphre environnante.
Consignes de scurit
-
Aucune pice l'intrieur de l'quipement nest rparable par lutilisateur. Contacter un technicien approuve
pour tous travaux de maintenance et de rparation
utiliser uniquement le model de fusible approuv par le constructeur
n'utiliser pas l'quipement si le cordon dalimentation de secteur ou la prise secteur est endommage
Utiliser uniquement le cordon dalimentation approuv par le constructeur
Ne pas utiliser TRANSPORT X (couvercle suprieur ouvert) par temps de pluie
Le matriel doit tre stock avec couvercle ferm afin dviter tout endommagement durant le transport
Le matriel ne doit pas tre laisser sans surveillance durant son fonctionnement
Section
Page
Introduction
System Description
2.1
TRANSPORT X
2.2
2.3
Accessories Kit
2.4
Thermal Printer
11
2.5
Specification
12
TRANSPORT X operation
13
3.1
Turning ON / OFF
15
3.2
Preventing Contamination
15
3.3
Unit operation
17
17
23
28
30
3.3.5 Settings
31
32
4.1
33
Sampling Procedure
36
Cleaning Apparatus
40
43
(Buchholz) relays
7.1
Introduction
43
7.2
Apparatus
43
7.3
Procedure
43
47
Introduction
essential in this area for many years, allowing faults to be detected in their early
stages, costly outages to be avoided and assets to be optimised.
System Description
2.1
TRANSPORT X
These include a sample bottle with connections and pipes and a syringe for
extracting a 50ml oil sample from the equipment and injecting it into the bottle.
See Section 2.2
The
incorporated software contains instructions to guide the user through the operation
of the system and algorithms to assist the diagnosis of the equipment. The PC
has an internal database that can store over 16000 records. Communication with
external PCs is possible via USB connections, allowing databases to be imported
or downloaded.
A thermal printer is provided enabling the user to maintain hard copy records of all
samples tested on the instrument.
System Check Kit is available to allow the user to check the operation of the
TRANSPORT X. See Section 8.
Optional extras are sampling of direct gas samples (see Section 7), and also for
the forced cooling of oil samples.
2.2
In-line Teflon
filters
Airtight gas
inlet and
outlet fittings
Airtight sample
injection valve
Temperature
probe
Sample
bottle
However, Kelman
It is essential that the bottle be kept upright and not be inverted while containing
oil. The oil should be disposed of suitably as soon as the sample has been tested
(see section 6).
The bottle should be cleaned thoroughly after each test with a clean cloth. It is
also recommended that the bottle be washed with new transformer oil or the next
oil to be sampled if possible (see section 6).
It is important also to ensure there is no residue of oil left in the oil injection port in
the lid assemble or the connections valve for the syringe, as this could cause
contamination of the next sample. For cleaning purposes it is recommended that
the user inject a sample of new, clean oil (or the oil that s next to be sampled) into
the bottle as normal.
TRANSPORT X prepared as normal for the next sample. This will ensure that no
unwanted residue of oil remains from a previous sample.
Care should also be taken to prevent any contamination from the syringe (see
section 4 & 6).
A Teflon coated magnetic stirrer is provided and should be placed in the bottle
before the lid is screwed on during assembly of the pipe work. This magnet is
rotated within the bottle during the analysis procedure, and it is essential for the
correct operation of the TRANSPORT X. The magnetic stirrer should also be
provided.
An in-line Teflon filter is fitted in both the return and inlet gas pipes. Luerlock
fittings are used to connect these filters and these fittings allow for quick
replacement of the filters at regular intervals or if oil should get into the pipe work
by accident. These filters are impervious to oil and function to protect the
instrument from aerosol and accidental spills. It is imperative that both these filters
are present at all times. If a spill occurs and oil makes its way into the gas analysis
chamber of the photo-acoustic module, the instrument will require a full service
and recalibration by the manufacturer.
It is recommended that the Teflon filters are changed after every 20 samples. If
any oil residue is present in the pipes, or these filters appear discoloured, then
they should be replaced before taking the next sample.
2.3
Accessories Kit
The TRANSPORT X comes complete with a full range of accessories and some
replacements. These include:1
Bottle lid-assembly
Magnet retriever
12
sample bottles
All of the accessories are supplied in a light carry case, with protective foam
inserts. The figure overleaf identifies each of the Accessories in the carry case.
50 ml
Syringe
Teflon
Coated
Magnets
Sample
Bottles
Magnet
Retriever
Teflon
Filters
Quick
Connection
Valve
115/230v
Power Cable
Bottle LidAssembly
10
2.4
Thermal Printer
The TRANSPORT X is equipped with an inbuilt 2 thermal printer that can provide hard
copies of results on site. If the paper should need changed it is a very simple procedure
to do this. The paper holder is released by holding in the catch on the left hand side,
shown in Figure 3.
The new roll of paper can then be easily fitted into the holder and then fitted back into
place. The button on the top left corner allows the user to feed the paper through the
printer when fitting, and also to check the function of the printer.
Note: If the printouts from the thermal printer begin to show a red tint, it can be an
indication that it may need a new roll of paper soon.
11
2.5
TRANSPORT X Specification
Fault Gas
Hydrogen, H2
Carbon Dioxide, CO2
Carbon Monoxide, CO
Methane, CH4
Ethane, C2H6
Ethylene, C2H4
Acetylene, C2H2
Water, H2O
3ppm
Accuracy
Power Supply
Fuse
F6.3AH250V 5x20mm
Battery
Output Digital
USB
Output Hardcopy
2 thermal printer
Temperature Range
5 - 40C
Operating Altitude
Max 2000m
Operating Pressure
Weight:
Dimensions:
IP Rating
IP20
12
3.0
TRANSPORT X Operation
The TRANSPORT X has been designed to provide the user with a reliable device
that is easy to use in the field. The embedded PC and touch screen display stepby-step instructions, and there is also an Instruction Card inserted in the lid of the
case.
A syringe and valves are supplied for the oil extraction and injection.
Provided the easy steps are followed the TRANSPORT X will provide reliable
DGA results.
Notes:
use a clean syringe, lid-assembly and bottle (See section 3.2 and section 6)
one magnetic stirrer piece must be placed in the bottle for the sample to be
analysed correctly.
sample bottle must not be moved until the analysis is complete. If the bottle
is moved the results could be adversely affected.
sample bottle must be kept upright to avoid ingress of oil into the pipe-work
of the lid-assembly.
13
ensure the pipes are connected correctly and securely, as per instructions.
when removed, the bottle lid-assembly should be held upright then cleaned
carefully with a cloth after use and before storage to avoid any ingress of oil
into the pipe-work (see Section 6 for details).
always empty the sample bottle after a test is completed. Also, clean the
apparatus after every test (see Section 6 for instructions).
The Purge time can be increased (max 10 mins) to help prevent carry-over
from a heavily gassed sample to the next sample.
once the TRANSPORT X is unpacked the user should run a System Flush
operation to help clear the gas paths from potential contamination from any
packing material. See Section 3.2
14
3.1
Turning ON / OFF
mains and the unit, then use the red power switch at the top right of the front
panel. The touch screen will then display the TRANSPORT X software. See
Figure 1.
To turn the unit OFF press the Turn Off button on the Main Menu, shown in
Figure 1. The system will then advise when it is ready to be switched using the
red power switch.
3.2
Preventing Contamination
the lid-assembly should be cleaned with air and clean oil (see section 6)
15
It is recommended that after unpacking the unit the user implements a System
Flush. This flushes ambient air through the TRANSPORT X for 20 minutes, and
will help clear any potential contamination from packing material. This function
can also be used to clear the TRANSPORT X if desired after concentrated
sampling.
To access this function from the Main Menu: Settings -> System -> Start
System Flush
3.3
TRANSPORT X Operation
The instructions displayed on the touch screen are used to control the operation of
the TRANSPORT X. Upon power-up, the screen shows the display Figure 4.
Turn Off:
PC Sync:
Settings:
After selecting Start New Measurement the user must select the Equipment
Type from the options available highlight the desired option and press Next.
This is vital for the database and the diagnosis options for the sample.
The TRANSPORT X then will request that the information for the database is
entered.
This information is vital, as it is the store and search criteria for all
records.
previous test, then the user can select this from the existing database list, see
Figure 5. The unit will use this information and store it in the new record with the
appropriate date stamp.
If, however, the information does not already exist in the database then the user
will have to enter it. For this the user selects Add New from Figure 5. This will
bring up the keypad on the screen, see Figure 6. The software will guide the user
17
through entering all the necessary details: Location, Equipment ID, Equipment,
Manufacturer and Sampling Point.
The user must then select Oil Sample or Gas Sample. The default is Oil Sample,
but if it is a direct gas sample, for example from the transformer headspace or the
Buchholz Relay, then the user should select Gas Sample. Press Next when the
correct option is highlighted.
The user has the option of adding the Oil Sampling Temperature, i.e. the
temperature of the oil at the time it was sampled. This is be used to calculate the
% Relative Saturation of the sample. If the temperature is not added then the
results will not include a %RS figure.
The user also has option to add to a Comments field for the sample data. This
can be used to record additional information for a particular sample or equipment.
Once all the necessary information has been entered, the user checks all the
information before continuing.
18
After all the sample information has been selected or entered, the software guides
the user through the testing process.
It is strongly recommended that a clean dry bottle be used for each test.
Figure 9 shows the initial instruction screen.
19
The unit prepares for the oil sample by purging with ambient air, and then
automatically taking a zero reference measurement. The user selects the purge
time between 2 and 10 minutes. The default is 2 minutes, but longer purges can
help prevent contamination from previous samples which are very heavily gassed,
for example from Tap Changers.
20
Once the sample has been injected the temperature of the oil is measured. This is
to avoid condensation in the internal pipe-work of the system. Therefore it is
recommended that hot oil not be tested, and that it has time to cool.
The
NOTE: If the oil temperature is higher than the internal cell temperature, then the
sample does not continue. The screen will display this temperature difference and
its countdown. Do not remove the bottle from the holder during this time. When
the temperature difference figure reaches 0.0C the unit will automatically start the
magnetic stirrer. If the bottle is not in place at this time, the stirrer will not operate,
meaning the results will not be correct.
After the temperature measurement is complete the unit begins the process of
measuring the dissolved gases in the oil sample. This takes 18 minutes in total.
21
When the measurement procedure is complete, the results are shown on the
touch screen, as illustrated in Figure 15.
Pressing Next will provide the options window shown in Figure 16. The user can
then select Print to get a hardcopy from the in-built thermal printer, or select
Advanced in order to use the embedded DGA diagnostic tools to help in analyzing
the result. See Section 3.3.6.
NOTE:
Fig.16
22
Note: These tools are not the opinion of Kelman, rather based upon thresholds
set by the user and also on established standards and algorithms.
The first diagnostic tool is based upon gas level thresholds settable by the user.
Default levels are initially programmed into the Transport X based upon the
program TOA 3 from Delta X Research. Each gas level has preset levels for both
Caution and Warning levels. If the value of any gas is over their Caution or
Warning level then it will be highlighted in yellow (if above Caution level) or red (if
above Warning level). See Figure 17. A Transformer Condition, Normal,
Caution or Warning, is then given based upon these alarm levels.
23
24
Analysis Algorithms
The operator can select the Diagnosis option shown in Figure 18 to give access
to Rogers Ratios, Key Gas method and Duvals Triangle analysis tools.
Note: These algorithms are based on established algorithms and are designed to
aid in the diagnosis of transformer faults, not in the detection of transformer faults.
If the gas levels are very low and the transformer does not appear to have a fault,
then these algorithms can give misleading results and should not be applied. The
user settable thresholds are very useful for this purpose. This is explained fully in
the standards and references stated above. For example, Duvals Triangle will
always provide a fault code, even if there is no fault in the transformer, so should
only be used if a fault is already suspected.
25
IEEE C57.104-1991
This technique allows visual comparison of the result to 4 typical fault results and
examine which one matched most closely, with the TRANSPORT X result, which
is shown in a blue bar graph. The user can then scroll through the 4 typical
options using the arrow keys, and they are shown in pink alongside the
TRANSPORT X result. The user can select the option most like his result, or else
select None if there is nothing applicable. See Figure 16.
Figure 19. Key Gas Method, showing Overheated Cellulose selected by the
Operator
Rogers Ratios
IEEE C57.104-1991
This method is based upon established ratios and uses these to determine the
type of fault. See Figure 17.
26
The method uses a triangle to plot the intersecting point of the gas values. Where
this point falls indicates the type of fault the transformer may have.
27
28
A Hardware code that is a Caution (highlighted in Yellow) prompts the user to take
certain action, such as change the in-line filters. If this code persists the user
should contact Kelman Customer Support.
A Sample code that is a Caution (highlighted in White) tells the user that there is
an issue with the sample, but the results are still valid, but may have a larger
margin for error. These may be caused by high target gas in the ambient (the user
can potentially take the TRANSPORT X unit to another location for further testing).
If any of these error codes persist, or there are any further questions please
contact Kelman Customer Support.
29
This option allows the user to examine results stored in the database.
By
selecting View Previous Results from the Main Menu (Figure 4) the options for
searching the database are given. See Figure 24. The database is organised
based upon information entered by the user when doing the DGA tests. If the user
selects the desired search criteria and presses Next the database is displayed. If
Location is used for example, all the various locations in the database are listed.
Using the scroll buttons the operator can select the desired location and press
Next and all the records for this location will be listed. The appropriate record can
easily be found using the arrow keys and by pressing View Record the result will
be displayed as previously shown in Figure 15. The results can be printed and
analysed as described in the previous sections.
30
3.3.5 Settings
The Settings option from the Main Menu brings up the screen shown below.
Touchscreen
Language
Gas Limits
Set the gas thresholds for the diagnostic tools see section
3.3.2
Normalization
System
31
This sampling procedure has been adapted from International standard IEC 567
for the sampling of oil from oil-filled electrical equipment for the analysis of
dissolved gas.
The method described below ensures that a secure sample is obtained and is
ready for injection into the TRANSPORT X without further risk of contamination,
post sampling.
The technique assumes that there is an available fitting on the equipment being
tested that will allow connection of a 2-way lure lock valve directly to the
equipment (see valve B, Figure a), either in-line in a sampling pipe, or as a fitting
that can be connected directly to the sampling point. Many different configurations
are possible but the requirement is the same for all.
32
4.1
Sampling Procedure
Connect syringe to lure lock valve (B) on equipment sampling point (Figure
a). Turn valve B to allow oil to flow out to waste. Let approximately 1 litre of
oil flow to a waste container.
Turn valve B and gently draw oil fully into the syringe (Figure b). At this
stage bubbles of air from the dead volume in the valves will be drawn into
the syringe and some of the gas present will be dissolved into the oil. Also
some of the gas in the oil will escape into these bubbles. This oil and gas
must be rejected as unrepresentative of the oil in the transformer.
Turn valve B and allow oil to flow out of this valve to a waste container
(Figure c). The amount of oil run off at this point will depend upon the size
of the equipment being tested but could vary between 1 and 4 litres. At the
same time disconnect the syringe from valve B and, holding it upright, expel
all of the bubbles and almost all of the oil from the syringe (Figure d).
Turn valve A when most of the oil has been expelled, leaving approximately
2ml of oil remaining in the syringe (Figure e).
Connect the syringe back onto valve B. Turn valve B and allow a small
volume of oil to flow out through the open arm of valve A (Figure f).
Turn valve A and gently draw 50ml of oil into the syringe (Figure g). Flush
this oil out of the syringe into the waste container (this is important to
ensure no contamination from the previous oil in the syringe).
You now have a representative sample of oil from the equipment ready for
injection into the TRANSPORT X.
33
Valve A
Valve B
Oil Pipe /
Sampling
point
Figure a
Oil Pipe /
Sampling
point
Figure b
Figure c
Figure d
Figure e
34
Oil Pipe /
Sampling
point
Valve B
Valve A
Oil Pipe /
Sampling
point
Figure f
Oil Pipe /
Sampling
point
Figure g
Oil Pipe /
Sampling
point
Figure h
35
Inspect the oil sample to ensure there are no large bubbles present. If there
are bubbles present inspect the three-way stopcock on the end of the syringe
to ensure it has not worked loose or that the valve is orientated correctly. If a
problem is found the equipment should be sampled again. If the valve is
secure and closed correctly then the plunger should be put under a slight
positive pressure and shaken vigorously until the bubble dissolves back into
the oil.
Note: If the oil being sampled is hot and the plunger is slightly tight in the
barrel of the syringe, it is possible for the oil to cool and contract, forming a
partial vacuum which will then draw gas from the sample into it and form a
bubble. Shaking with a slight positive pressure on the syringe will dissolve
this gas back into the oil and the sample will be okay for analysis.
36
Once the sample is confirmed to be okay the male quick connect fitting
should be attached to the three way stopcock.
Fig. 25
Connect the syringe to the bottle. The outer sleeve of the female fitting on
the lid-assembly should be held between thumb and forefinger and pulled
downwards as far as it will go. See Fig. 26. Insert the syringe with the
male quick connect fitting firmly into the female connector, until a slight
click is heard. Release the outer sleeve of the female connector.
37
38
Note1: Do not turn the three-way valve to disconnect. By not turning the
three-way valve, gas is prevented from escaping through the side arm and
also the syringe will not drip oil when syringe is being removed.
Note2: Be careful not to separate the male quick release section from the
valve/syringe until the syringe is disconnected from the bottle at the quick
release valve as that would provide an open channel for the gases to
escape to atmosphere.
39
Cleaning Apparatus
It is important that the correct clean-out procedure is used to avoid any contamination
from one sample to the next. Also, thorough clean out is important for maintenance of
the correct volumetric ratio, which calls for 50 ml of oil. A small amount of residual oil
left in the sample container or injection pipe-work may have an impact on the results
from the TRANSPORT X for the next sample. Cleaning also helps to keep the unit
and its accessory case in good condition.
Note: DO NOT use any solvents to clean any part of the TRANSPORT X or its
accessories as this can seriously affect results.
If cleaning of the sample bottle and syringe is deemed necessary then warm
soapy water can be used.
The syringe, male quick connect valve and the oil injection tube in the lid-assembly
should be cleaned out after oil injection. This can be done with air, some clean oil or
a sample of the oil that is to be tested next. How this is done and The sequence in
which it is done can often depend on the circumstances of the testing at that time.
The best method is with a combination of air and new / next oil.
This entails
i)
clearing the syringe, valve and lid-assembly injection tube with air
ii)
iii)
injecting a syringe full of new oil into the bottle via the lid-assembly
iv)
cleaning bottle and lid-assembly with clean cloth (see below instructions on
how to do this).
By following these best-practices the TRANSPORT X is able to easily clear the affects
of heavily gassed samples for the next test.
40
Syringe
It is imperative that the syringe is kept clean in order to prevent cross-contamination of
results.
turn 3-way valve to take air into syringe through side arm of valve
hold syringe with the quick release valve pointing vertically down
push air through syringe and valves, flushing out any oil residue
repeat 4 - 5 times
The syringe should also be rinsed out with oil 3 -5 times when drawing a sample to
ensure there is no contamination in the new sample (see section 4)
Disconnect the pipe work from the top panel of the TRANSPORT X.
41
Unscrew the lid assembly from the sample bottle, holding the assembly upright
to avoid any oil ingress into the tubes.
Clean the lid assembly underside and pipes with a clean cloth.
Connect the syringe to the lid assembly with the male quick connect valve.
Using the same procedure as above, connect the syringe to the opened lidassembly and expel any oil residue from the oil injection tube by flushing
through with air 4 -5 times.
Give the bottle a thorough clean with an oil absorbent rag or clean cloth.
For best results (for example if the user is concerned about potential carry-over from a
heavily gassed sample) the user should flush the bottle and lid-assembly with new oil.
To do this the user should:
i)
screw the lid-assembly into the bottle (which has been cleaned with a clean
cloth);
ii)
iii)
inject a syringe full of new oil into the bottle via the lid-assembly, as normal;
iv)
v)
clean the bottle and lid-assembly cleaned as normal with a clean cloth.
This ensures there is not oil residue from previous samples in the injection tune or
connection valves of the system. This is particularly important after testing a highly
gassed sample, for example from a tap changer.
42
7.1
INTRODUCTION
Gas samples from relays should be taken from the equipment with minimum
delays after gas accumulation has been signalled. Changes in composition
caused by the selective re-absorption of components may occur if free gases are
left in contact with the oil.
7.2
APPARATUS
5ml Gas Tight Syringe with luer lock termination and integrated valve.
7.3
PROCEDURE
This should be
43
5. Open the gas relay valve (5) and carefully turn stopcock (4) to position X
so gas from the gas collecting relay can flush oil and air from valve, tubing
and stopcock.
6. When gas reaches the stopcock (4) turn it to position Y to allow gas from
the gas relay to fill the syringe (1) by the hydrostatic pressure in the relay,
taking care that gas pressure does not eject the plunger completely.
7. Turn the stopcock (4) to position Z and expel gas from the syringe through
the exhaust port of stopcock by pushing plunger home.
8. Turn the stopcock (4) to position Y and fill syringe to the 5ml mark
9. Close the valve of the gas collecting relay (5).
10. Close the integrated valve (2) on the syringe (move to position B)
11. Remove the plastic tubing from the relay valve (5) and stopcock (4)
12. Remove the plastic stopcock from the syringe.
13. Analyze sample as soon as possible after collection and shade from light
in the interim. This will avoid excess gas loss, and also preserve the
integrity of the sample taken.
44
OFF
Position of
Integrated
syringe
valve
2
Position of
Stopcock
Y
OFF
OFF
B
Z
2
1
6
6
4
6
1: 5ml Syringe
2: Syringe Integrated valve
3: Gas impermeable tubing
4: Three way stopcock
5: Relay sampling valve
6: Gas collecting relay
7: Waste vessel
7
6
45
1. Select Start New Measurement from the Main Menu and follow the on
screen instructions as normal to enter all the data for your sample. For
example when prompted to enter the Sampling Point of your sample, you
could enter Buchholz relay. Ensure to select Gas Sample when
prompted by the on-screen menu to chose Sample Type
5. Leaving syringe and valve in the same position, draw the plunger back to
the 5ml mark (be careful not to remove it completely) and depress again.
Repeat this procedure approximately 5 times in quick succession to rinse
the entire sample into the sample bottle of the TRANSPORT X. End with
the plunger in the fully depressed position.
6. Pulling down on the outer sleeve of the quick connect fitting, disconnect
syringe and quick connect from sample bottle, similar to disconnecting
syringe when testing an oil sample
46
Finish.
System Check
This kit consists of a cylinder of gas containing all the DGA target gases at
known concentrations, together with a length of Nafion tubing, used to connect
the gas cylinder to the inlet connector on the top panel of the TRANSPORT X.
Procedure
1. Power up the unit.
2. Follow the options from the Main Menu -> Settings -> System -> Start
System Check. Follow on screen instructions.
Note: do not put the oil sample bottle into its holder and do not connect
the bottle lid assembly to the top panel.
47
3. Connect the Nafion tubing to the cylinder fine control valve by pushing
the flexible black viton tubing over the barb connector in the cylinder mini
flow control valve outlet. See Figure 18.
4. Connect the other end of the Nafion tubing, which is fitted with a male
quick release connector, to the top panel female quick connector as
shown in Figure 19 below.
5. Carefully open gas flow from cylinder by turning the knob on the Mini Flow
Controller slowly in an anticlockwise direction.
Tee
Junction
Barb Connector
Mini
Flow
Control
Knob
Viton tubing incorporating
open Tee junction fitting.
Figure 28. Details of Viton tubing connection to Mini-Flow Controller and flow
indicator setting
48
Note: The regulator can be sensitive so careful attention is required when setting
the flow rate.
24 long
Nafion Tube
Figure .19. This shows the cylinder of gas mixture connected to the
TRANSPORT X via a 24" length of Nafion tubing.
8. When the test has been completed the TRANSPORT X will return either a
Pass or Fail report.
10. In the event of a fail the user should re-do the test.
49
11. Before re-doing the test ensure that there is no obstruction in the Nafion
tube assembly : This can be checked as follows :
Connect the Viton end of the Nafion tube to the barb connector of
the Mini Flow Controller on the cylinder.
Block the side arm of the Tee junction in the Nafion tube.
Using the Mini Flow Controller set the flow to 500ml/min .
Place an obstruction momentarily over the open end of the Nafion
tube and note the effect this has on the flow rate.
If the flow rate drops this indicated that there is still an open
channel in the Nafion tube .
The user should also ensure the flow remains steady at 500ml/min or
more throughout the test by observing the position of the flow indicator
ball throughout this 2nd test.
12. If a second fail occurs the user should re-do the test one more time and
report all the results to Kelman.
After studying the results Kelman may report a faulty unit and request return
or may make other suggestions for confirming functionality of the
TRANSPORT X by the user.
Notes
1. The cylinder of Gas Mixture is non flammable and non toxic. It should however only be
used in a well ventilated area by trained personnel.
2. When empty the cylinder can be disposed of by recycling or as normal domestic or
industrial trash.
o
3. The cylinder is pressurized\ed to 300psi at 20 C when full and contains 20 L of gas.
4. Each test consumes at least1L of gas and so a full cylinder should be good for up to 20
system checks.
5. The gas mixture in the cylinder has a shelf life of 2 years from the date of manufacture marked on the bottom on the cylinder.
50