Professional Documents
Culture Documents
LCR Meter and Its Application 30july13
LCR Meter and Its Application 30july13
LCR Meter and Its Application 30july13
30 July 2013
Outlines
Electrical Club
30 July 2013
1st-level
Working
Standard
Capacitance bridge
uncertainty 10-6
2nd-level
Working
Standard
Capacitance standard
uncertainty 10-5
General
Measuring
Equipment
Electrical Club
LCR Meter
uncertainty 10-4
30 July 2013
Set of Inductance
standard
100 H to 10 H
uncertainty 10-6
Working
Standard
Set of Inductance
standard
100 H to 10 H
uncertainty 10-5
General
Measuring
Equipment
LCR Meter
uncertainty 10-4
Electrical Club
30 July 2013
CMC
Capacitance
1 pF, 10 pF, 100 pF
1000 pF
Inductance
100 H, 1 mH, 10 mH,
100 mH, 1 H, 10 H
Electrical Club
Range
1000 Hz and 1592 Hz
1000 Hz and 1592 Hz
Uncertainty (k=2)
0.8010-6
3.510-6
Range
Uncertainty (k=2)
1000 Hz
0.1610-3
30 July 2013
Bridge method
Resonant method
I-V method
Network analysis method
Auto-balancing bridge method
DC to 300 MHz
10 kHz to 70 MHz
10 kHz to 100 MHz
300 kHz and above
20 Hz to 110 MHz
Electrical Club
30 July 2013
Electrical Club
30 July 2013
Measurement
function
Cp
Cp-D
Cp-Q
Cp-G
Cp-RP
CS
Lp
LS
Electrical Club
CS-D
CS-Q
CS-RS
Lp-D
Lp-Q
Lp-G
Lp-RP
LS-D
LS-Q
LS-RS
Definition of D, Q
and G
D=
1
1
=
2fC P RP Q
G=
1
RP
D = 2fC S RS =
Q=
1
(1 + D 2 )
LS = LP
1
(1 + D 2 )
1
Q
1
RP
=
2fLP D
G=
Q=
CP = CS
1
RP
2fLS 1
=
RS
D
30 July 2013
Capacitance Measurement
Large value
Small value
Rp
Large C
Less significant
Low Z
Rs
More significant
Electrical Club
Small C
High Z
RP
More significant
RS
Less significant
30 July 2013
Inductance Measurement
Large value
Small value
Rp
Large L
Less significant
High Z
Rs
More significant
Electrical Club
Small L
Low Z
RP
More significant
RS
Less significant
30 July 2013
10
Electrical Club
Cs
Ls
Cp
Lp
30 July 2013
11
Terminal configuration
Two terminal
Three terminal
Four terminal
Four terminal pair
Electrical Club
30 July 2013
12
Electrical Club
30 July 2013
13
Electrical Club
30 July 2013
14
Electrical Club
30 July 2013
15
Electrical Club
30 July 2013
16
Terminal configuration
Type : GR 1409
Type : GR 1482
Electrical Club
2-terminal
3-terminal (binding post)
2-terminal
30 July 2013
17
Terminal configuration
Type : IET 1404
3-terminal (coaxial)
Type : HP 16380
4-terminal
Electrical Club
4-terminal pair
30 July 2013
18
Electrical Club
30 July 2013
19
Electrical Club
30 July 2013
20
Electrical Club
30 July 2013
21
Electrical Club
30 July 2013
22
Electrical Club
30 July 2013
23
Electrical Club
30 July 2013
24
Thanks
Electrical Club
30 July 2013
25
Traceability chart
Electrical Club
30 July 2013
26
Reference
Electrical Club
30 July 2013
27
Electrical Club
30 July 2013
28