Manufacturer

Abaxis Inc North


America
Piccolo Xpress

Techno Medica Co
Ltd
GASTAT-navi

Nova Biomedical
Corp
Stat Profile pHOx
Basic

WHERE MARKETED
FDA CLEARANCE
CE MARK (MDD)
TEST MENU
Blood gases/electrolytes

Worldwide
Yes
Yes

Worldwide
No
Yes

Worldwide
Yes
Yes

K+, Na+, Cl-, TCO2

pH (6-8), pCO2 (5250 torr), pO2 (5-760


torr), cNa+ (80-200
mmol/L), cK+ (1-12
mmol/L), cCa++
(0.25-5 mmol/L)

pH (6.5-8), PCO2 (3200 mm Hg), PO2 (0800 mm Hg), SO2


(30-100%) Optional:
Na+ (80-200
mmol/L), K+ (1-20
mmol/L), Cl- (50-200
mmol/L), Ca++ (0.12.7 mmol/L)

Chemistries

Albumin, ALP, ALT,


No
amylase, AST, BUN,
cholesterol, CA,
cholesterol/HDL ratio,
CK, CL, CRE, CRP,
DBIL, eGFR, GGT,
GLU, HDL, K, lactate
dehydrogenase, LDL,
Mg, NA, PHOS,
tCO2, total bilirubin,
total protein, TP,
triglycerides, UA
(total/direct), VLDL

Optional glucose (15500 mg/dL), lactate


(0.3-20 mmol/L)

Others

Not specified

Hct (15-75 %)

METHOD USED

Spectrophotometry

ISE, others

SAMPLE TYPE

Whole blood, serum, Whole blood,


plasma
aqueous solution
100
>50, aqueous
solution 200
~10-12
0.45
Dry reagent beads
Cartridge

Hct (12-70%), Hgb


(4-24 g/dL)
ISE, optical,
conductivity
Whole blood

SAMPLE SIZE, L
ANALYSIS TIME, min
REAGENT TYPE
REAGENT PREPARATION
CALIBRATION
QC materials
QC lockout
Mfr supplied
ALERT INDICATORS
DISPLAY, TYPE

None
Self calibrated
Built-in continuous
iQC
Not specified
No
Audible, visual
Touchscreen

None
Automatic, 1-point
Solution
Not specified
Yes
Yes
Color LCD
touchscreen

70, 45 micromode
0.75
Single pack (4 sizes
available)
None
Automatic
35-day snap-in QC
pack
Yes
Yes
Not specified
Color

Data displayed
DATA MANAGEMENT
Programmable
Data stored

Printer

Bar-code reader
Reagent
Sample
COMPUTER INTERFACE
HIS

LIS

Various

Yes

Not specified

With Data Manager No


5,000 patient records 1,000 tests

Yes
Patient and QC data
storage, LeveyJennings, CLIA
reports, maintenance
logs/reports,
cumulative reports

Internal printer on
sticky-backed rolltape

Yes

Built-in

Optional via USB port Not specified

Not specified

Optional via USB port Not specified

Optional

USB, TCP/IP
USB

RS232C, USB
Not specified

RS232
Yes, through data
management system

USB

Not specified

Yes, through data


management system

CLIA CLASSIFICATION

Depends on test, 11 No
of 15 reagent discs
are CLIA waived

Nonwaived

CONFIGURATION
ACCESSORIES

Mobile
Battery pack, field
deployment kit

Handheld
Not specified

Benchtop, mobile
Not specified

100-210
Yes
Not specified
Not specified
Not specified
Not specified
Not specified
32.4 x 15.2 x 20.3
(12.8 x 6 x 8)
5.1 (11.2)

100-450
Battery
Built-in
Not specified
Yes
Not specified
No
9.6 x 25 x 12 (3.7 x
9.8 x 4.7)
1.4 (3)

90-264
Optional UPS
Not specified
30 min
Not specified
Not specified
Not specified
38 x 31 x 38 (15 x 12
x 15)
8.2 (18.1)

$16,500

$15163 (1,500,0000 $12,000

$13-20

15,000 for Sensor Varies


Card 091; 19,000
for Sensor Card 092;
20,000 for Sensor
Card 093

NA
1 year
Included

NA
1 year
Not specified

POWER REQUIREMENTS
Line power, VAC
Backup
Battery, type
Operating time, hr
Rechargeable
Recharge time, hr
Low-battery indicator
H x W x D, cm (in)
WEIGHT, kg (lb)
PURCHASE INFORMATION
List price
Reagents

QC, per test


Warranty
Training & support

Varies
1 year
Training included

Service contract
Year first sold
Number sold
Delivery time, ARO
GREEN FEATURES

Not specified
2006
8,000
1 day
None specified

Not specified
2010
Not specified
Not specified
Rechargeable battery

OTHER SPECIFICATIONS

Saves files in ASCII,


ASTM, XML, or PDF
on Windows PC;
optional carrying
case.

Disposable cartridge Tankless gas


with built-in sensor
calibration; automatic
and calibrator;
onboard QC; sealed
automatic pre-heating waste system;
mode.
automonitoring of QC
and reagent packs;
optional cart. Meets
requirements of CSA
(NRTL), EN 46001,
ISO 9001, and TUV.

UMDNS CODE(S)

18509, 18510,
18511, 18853
August 2013

18509, 18510,
18511, 18853
December 2014

LAST UPDATED
Supplier Footnotes
Model Footnotes
Data Footnotes

Optional
2002
Not specified
4-8 weeks
Standby mode

18509, 18510,
18511, 18853
December 2014

You might also like