PD205 PD206 Manual
PD205 PD206 Manual
Thank you very much for purchasing Hongyi Precision Industrial Inc digital pressure gauge
Preface
Please refer to this manual carefully,
installation and maintaince of PD205 digital
pressure gauge should be carried out after
reading this manual.
It is not allowed to remodel PD205 digital
pressure gauge. If there is an accident owing to
remodeling, it will not be Hongyi company
responsibility.
Regarding detailed specifications, figure and
size of the item, please refer to the following
information.
Safety Notice
Please pay attention to and obey all the safety notices and relevant safety notices listed here. Regarding
safety notice, there are two classes: warning and caution. However, even the sign is Caution,
there may be dangerous consequences in certain situations. Therefore, both of the two classes of safety
are vitally important, and the user should strictly obey them.
Caution
Warnings
It is a digital pressure gauge with no explosion prevention. Do not use it in the environment of
explosive gas. Otherwise it will cause serious accidents such as explosion and fire.
Pressure
value display
Max/min value
display
Max/min
Battery sign
Units display
Description
ON/OFF
Display Pressure units e.g. Kpa, bar, Mpa, Psi, Kgf, mbar,
MIN/MAX
P/UNIT/ZERO
RESET/ENT
The gauge can work normally when the indicating light of power sign is on.
There is a need to change battery when the battery sign is flashing!
Display table2
Table 2
Characteristic
Description
Unit
Display Pressure units e.g. Kpa, bar, Mpa, Psi, Kgf, mbar,
Pressure
Display
Process
Battery Sign
The gauge can work normally when the indicating light of power sign is on.
There is a need to change battery when the battery sign is flashing!
Press the key [MIN/MAX] to shift the maximum value or minimum value of
history pressure.
CIP
When the sign appears, it is indicated that the measured pressure exceeds
the valid measuring range.
Panel Operating
Display unit change
In the condition of pressure measuring, press
the keyP/UNIT/ZEROfor one second and then
loose it, the user could choose the pressure unit.
Repeat this operation, the pressure unit could be
changed among Psi Bar Mpa Kpa. Meanwhile,
the according sign of the pressure unit will appear
on the display screen.
Display ON/OFF
Parameter Operating
PD2XX-XXX series have some program parameters; working status and functions of the gauge can
be modified by changing these parameters. See Table3
Table 3
No
.
Para.
name
Description
Rang
e
Default
Unit
Memo
CO
Cycle of scanning
00-- Always updating
00-99
Sec.
LC
00-99
Min.
ZER
Zero adjustment
Y/n
N: no Zero adjustment;
Y: carry out Zero adjustment
Parameter setting
Parameters setting
1. In the condition of pressure measuring,
press the keyP/UNIT/ZEROfor the first time five
seconds and enter the mode of parameter setting,
and display CO_XX. In this condition, adjust the
parameter by the keyRESET/ENT. Press the key
RESET/ENT shortly for one time, the sampling
time will be one larger in number. Press the key
RESET/ENTfor 3 seconds, the sampling time
will add continuously.
2. Press the keyP/UNIT/ZEROfor a second
time shortly, it will begin to set the next
parameters and display LC_XX. In this condition,
adjust the parameter by the keyRESET/ENT.
Press the keyRESET/ENTshortly for one time,
the sampling time will be one larger in number;
press the keyRESET/ENTfor three seconds, the
finishing
parameter,
press
the
key
P/UNIT/ZEROshortly for one time and it will
display Zer_n. Press the keyP/UNIT/ZERO
again for long, exit setting to the mode of display.
Notice:
When the value of sampling time exceeds 99,
the sampling time will add from 00. 00 stands for
continuous sampling. The bigger the value of the
sampling time is and the lower the power
consumption is, the longer the battery longevity is,
and the default is 05 sec.
Adjustment of Zero
In the condition of pressure measurement,
press the key P/UNIT/ZEROfor five seconds, it
will enter the mode of parameter setting and
display CO_XX; at the same time, it displays the
instruction of setting. In this condition, press the
keyP/UNIT/ZEROagain for two times, it displays
Zero parameter Zer_n. Press the key
RESET/ENT to decide whether do Zero
adjustment operation. When shows Zer_n, it
does not carry Zero adjustment; when shows
Zer_Y, does Zero adjustment. After press the
key P/UNIT/ZERO for long, exit setting to the
mode of display.
Notice:
Zero adjustment must be done before the
installation of the gauge, or Zero adjustment is
done after the installation of the gauge but in the
condition of non-measurement. This operation is
fobidden in the course of using or measuring to
avoid wrong actions. If this operation is done
when the gauge is working, the device must be
stripped down or Zero adjustment must be done
in the condition of non-measurement.
Battery Installation
Operation preparation
There is no battery installed in the gauge; the users should install battery themselves, battery type:
AAA, 4pcs.
Method:
Installation instrument
A. Position
According to the measuring position to choose a
suitable place and direction to install the gauge.
Note:
1. The gauge should be mounted in a place that is
well ventilated, dry and shady and without
corrosion. It should be mounted with the
protection cover if it is mounted in the open air in
case of the sun shining and rains. This will help to
avoicd faults or performance degradation to the
gauge.
2. Avoid the dregs sedimentation in conduit. The
gauge should not touch with corrosive and
superheat material.
3. When measure the liquid pressure, the
pressure port should in the side face of pipeline in
case of depositing dregs.
4. When measure the gas pressure, the pressure
port should b on the top of the pipeline. Also the
gauge should mount on the upside of the pipeline
so that the accumulated liquid will pour into the
pipeline.
5. The connection pipe should be mounted in a
place where the temperature changes little.
6. When choose the measuring pressure point,
there should be a straight pipeline at the pressure
point. The length of straight pipe of gauges both
sides should be greater than 5xD(D means inside
diameter of pipeline);
When adopt pressure on the steam pipeline,
there should be a guiding pressure conduit on the
side face;
Fig.3
Fig.4
Calibration
2. Inaccurate pressure display due to improper
PDXX-XXX
usage.
This series of digital pressure gauge could be
3. Special require of calibration in certain
calibrated. However, all the gauges have been
condition.
calibrated before they are dispatched to
The calibration operation is in the attachment.
customers and customers do not need to
<<The Instruction of Calibration Operation>>
calibrate again. Only in the following conditions,
(This attachment is not with product)
you could calibrate the gauge.
1. Customers can calibrate the gauge after 2
years usage, in normal condition.
Nomination of PD2XX-XXX
There are three parts of the nomination of digital pressure gauge. PD2XX means the type of digital
pressure gauge. XXX means the connecting ports and pressure range of the gauge. In XXX, 1st X means
the connecting ports (See table 2), 2nd X means the output type (See table 2), and 3rd X means the
pressure range (See table 3).
Connecting Port
Table 1
Connection
Code
Connection Port
AXX
1/4-NPT
BXX
CXX
DXX
1/2-NPT
1/2-BSP
M20x1.5
Pressure range
EXX
FXX
G1/2
1/4 BSP
GXX
HXX
KXX
Option
diaphragm clamp 2
diaphragm G1/2
LXX
MXX
M16x1.5
M18x1.5
Output
Table 2
OUTPUT
X0X
X1X
X2X
X3X
X4X
X5X
X6X
Connection Port
None
SPDT
4-20mA
0-20mA
0-10V
0-5V
SPDT / 0-10V
X7X
SPDT/ 4-20mA
X9X
Option
RangeCode
Unit
XXa
Pressure
Range
0-20
XXb
0-35
Kpa
XXc
XXd
0-70
0-100
Kpa
Kpa
XXe
XXf
XXg
XX1
XX2
0-250
0-400
0-600
0-1
0-1.6
Kpa
Kpa
Kpa
Mpa
Mpa
XX3
XX4
XX6
XX7
XX8
XX9
XXA
XXB
XXC
0-2.5
0-4
0-6
0-10
0-25
0-40
0-60
0-80
0-100
Mpa
Mpa
Mpa
Mpa
Mpa
Mpa
Mpa
Mpa
Mpa
Kpa
Specification
Characteristic
Measure
Media
Case material
SUS304
Case Size
3" diameter
Accuracy
PD205-XXX 0. 5%FS
PD206-XXX 0.25%FS
PD505-XXX 0.1%FS
Operating Temperature
-10 to 60C
Storage Temperature
-20 to 70C
Relative Humidity
0 to 90% non-condensing
Medium Temperature
-10-80 C
Pressure Range
See table
Temperature
compensation
-10-70C
Temperature stability:
0.05%FS/C
Power:
Display
Burst Pressure
3xFS
Over pressure
SUS316
Connecting Port
See table 3
Port Size
See table 4
Lifetime
1200h(Min)
Weight
13
Sealing
IP65
Case
Panel
Ring
Pressure value
Battery
Setup
Unit mark
Key Unit/Zero
Key On/Off
Wetted
3.
Pressure
sensor
Wetted
Pressure process
Pressure value
display
display
Max/min
value
Max/min
display
Units display
Battery sign
Fig.3
Fig.4