Q.A. GUIDE
Indice : C
Page : 1/3
Modification date : 16/04/15
This guide has to be used together with the report sheet FRSATA051,
in order to perform the maintenance test (3 year, 6 year ...) or to perform the test after a repair.
Only fromSchiller autorized person/establishment can perform this test.
Version : Semiautomatic/Automatic - Ethernet/Online/SD card & Option : Manual Metronome - ECG display
VISUAL INSPECTION/ MECHANICS INSPECTION OF THE UNIT (for all versions) :
The unit must be turned off and the battery removed
1.Check the general aspect (specially damaged parts which can involve to tightness issues)
2.Check if the buttons are present and check their status
3.Check the labels (serial number, EC).
4.Shake the device and check if you hear any suspicious noise
5.Check the external screws aspect.
6.Check if the electrodes connector is clean and not deteriorated
15.Connect the Adult electrodes connector check if the character A (for Adult mode ) or a pictogram representing an adult
is present on the upper right-hand corner of the screen. Connect the Child electrodes connector and then check if the character C
(for Child mode ) or a pictogram representing an child is present on the upper right-hand corner of the screen.
18A.Push the analysis button (if it's necessary).After the analysis , a shock 18B.An automatic analysis is started, shock is
is recommanded and the shock
button is lighting up and the recommanded. Check if a warning about the imminent
shock appears after analysis.
message Press Orange Button appears.
19B.Check if the orange shock button blinks
19A.Disconnect the electrodes connector
Check if : the orange shock button switches off and the LED 20B.Disconnect the electrodes connector as soon as
Electrode lights up (+ internal discharge).
orange shock button blinks.
20A.Repeat 18th step, when Press Orange Button appears, wait 20 Check if the orange shock button switches off and
the LED Electrode lights up (+ internal discharge).
seconds 1 sec after the charging is done, (don't deliver the shock)
the shock button must switch off (internal discharge).
21B.Repeat steps 18B and 19B. After a 4 seconds
21A.Repeat step 18th step, when Press Orange Button appears,
deliver shock (deliver 3 shocks), use a Joulemeter (note the kind, the
management number and the deadline calibration) to check the 3
delivered energy values and if the delivered values correspond to the set
values. The values must imperatively be located between the tolerances
(setting value 15%). Use the following table to know nominal values
tolerances.
Number : GUSATA051
Q.A. GUIDE
Indice : C
Page : 2/3
Modification date : 16/04/15
11
18
34
57
80
103
126
149
172
Nominal Value J
15
30
50
70
90
110
130
150
12
26
43
60
77
94
111
128
Check if :
23C.The LED (on the blue box)
switches on when the unit is turned on
and
blinks
while
the
device
communicates with the server.
25B.You hear an audio signal and a warning 25C.The message Fred Easy Online
appears when transfer is done.
Ready appears.
OPTIONS INSPECTION :
Manual option :
Turn off the device for at least 5 minutes (to reset shock counter)
Turn on the device and, put it in manual mode : press simultaneously green and orange buttons (
prompted by a displayed message to press again both buttons. This must be done within 5 seconds.
Check if :
Connect an ECG simulator to the unit and check if you can see :
27. ECG waveform on the screen.
28.The selected energy and a prompt to push the green button.
Energy value Check :
Connect a Joulemeter (note the kind, the management number and the deadline calibration) to the unit.
). You will be
Number : GUSATA051
Q.A. GUIDE
Indice : C
Page : 3/3
Modification date : 16/04/15
to begin the charge of selected energy. When it's done, the orange button lights up and both a
voice prompt and a message ask you to deliver the shock (orange button).
Check if :
30.There is an internal discharge after 20 seconds 1 sec without delivering shock.
31.Repeat 29th step and deliver 3 shocks (by pressing on orange button
when unit asks it) using a Joulemeter to check the
delivered energies value. The values must imperatively be located between the tolerances (setting value 15%). Use the same
table than in steps 21A and 21B to know nominal values tolerances.
(To return to the semiautomatic mode, switch off the unit and leave it turned off for at least 5 minutes.)
Metronome option :
32.Check if the unit generates a steady beat during cardiopulmonary resuscitation (CPR) to give the frequency for chest
INSPECTION OF LEAKAGE CURRENTS (according to the international standard : IEC 60601-1 and 60601-2-4) :
34.Connect a leakage currents test device to the inputs/outputs of the device, note the kind, the serial number and the deadline
calibration of the test device. The values must imperatively be located between the tolerances
Has not to be performed during the 3 yearly maintenance, if the Easy has not been opened
Has to be performed during the 6 yearly maintenance after the obliged replacement of the memory back up battery
Has to be performed after a repair of the Easy (as soon as the Fred Easy has been opened)
100
100
Admissible maximum values (A)
100
100
Admissible maximum values (A)
5000
5000