Digital Imaging
Digital Imaging
html
Digital Imaging
It is vital for any radiologic technologist working with digital imaging systems to be educated on the equipment
they are using. This lesson covers the topic of digital imaging. The advantages and disadvantages of the different
systems available are mentioned. Basic digital imaging terminology, the process of capturing and viewing a digital
image, and basic quality control to be done in a digital imaging department are discussed. Also contained in the
lesson is a breakdown of a PACS. The lesson will also discuss possible ways to decrease patient dose with digital
imaging systems.
Author: Jennifer M. Schmidt BS, R.T. (R)(M), written on Friday April 10th 2009 - 6:20 PM
Credits: 0
-----------------
The A.R.R.T. Continuing Education Requirements for Renewal of Registration states that Internet
activities reported in a biennium may not be repeated for credit in the same or subsequent
biennium. If you have already completed this article, DO NOT take this CEU article again.
Upon completion of this course the participant should be able to do the following:
1.
2.
3.
4.
Understand why new exposure techniques can be used to decrease a patient's dose.
5.
6.
Outline:
I.
II.
Introduction
Digital Imaging
A.
B.
C.
III.
IV.
V.
Imaging Plate
2.
Reader
DR Equipment
2.
Direct Conversion
b.
Indirect Conversion
Comparison of CR vs. DR
1.
2.
Look Up Table
A.
Algorithm
B.
Automatic Rescaling
C.
Latitude
Image Manipulation
A.
Image Orientation
B.
Image Stitching
C.
D.
Shuttering
Collimation
2.
VI.
B.
Positioning Considerations
C.
Spatial Resolution/Detail
B.
C.
VIII.
IX.
X.
1.
Edge Enhancement
2.
Smoothing
Equipment
A.
VII.
Lead Masking
Imaging Plate
1.
Reflective Layer
2.
Active Layer
3.
Photostimulable Phosphor
Flat-Panel Detector
1.
Direct Conversion
2.
Indirect Conversion
Matrix
1.
Pixels
2.
Pixel Pitch
Exposure Indicator
B.
kVp
C.
ALARA
Monthly
B.
Weekly
C.
Daily
D.
PACS
A.
Basics
B.
Display Stations
C.
Archiving Data
Summary Points
Introduction
The world of x-ray is always is changing. The name of the field, radiologic technology, alone gives the indication
that the latest technology is used in the profession. In recent years the jump to using the latest technology has
never been truer. More and more radiology departments are converting from conventional radiography
departments to digital imaging departments with complicated computers. In the middle of all the conversion we
have radiologic technologist trying to keep up.
To insure that technologist have all the latest information to provide the best quality patient care, this publication
will break down how a digital image is formed, basic terminology used with digital imaging, possible ways to
decrease the patient's dose, and a brief discussion on quality control is provided. At the completion of this article it
is the intent of the author for a radiologic technologist to be able to offer the best quality care with their new
knowledge of how digital imaging works.
Digital Imaging
Digital imaging can be broken down into two categories: computed radiography and digital radiography. Digital
imaging is the proper term to use when collectively speaking about any type of image that is taken with digital
media, whether a cassette or no cassette is involved.
Computed radiography (CR) is a way to capture x-rays in a digital format using a cassette and laser reader
system. It is commonly referred to as CR imaging. The conversion to a CR system has many benefits, the biggest
being the use of the same x-ray table and generator, in most cases. The conversion to CR imaging requires the
radiology department to replace the conventional film-screen cassettes with digital imaging cassettes and
replacing of a processor with a laser reader. The digital cassettes can be referred to as imaging plates or image
receptors.
The cassette used for CR systems looks very similar to that of a conventional radiography cassette. The actual
cassette is made of a radiolucent material such as plastic in the front and a radiopaque material like aluminum in
the back. Inside the cassette of a CR system is where the big difference begins. The film inside the CR cassette
is called an imaging plate. Unlike conventional film, the imaging plate can be used over and over. It can also be
positioned the same as a conventional radiography cassette. This ability to position the imaging plate similar to a
conventional film-screen cassette can help in the transition to digital imaging.
Conventional film-screen imaging requires the use of wet processing with harsh chemicals to form the latent
(visible) image. A laser reader is used to produce the latent image in CR imaging systems. The harsh chemicals
have been replaced by a red laser beam.
Once an exposure has been taken using an imaging plate, the imaging plate must be "processed" to release the
manifest (invisible) image. Releasing of the manifest image is accomplished by placing the imaging plate into a
laser reader. The laser reader uses a red laser light to scan the information from the imaging plate. When the
red laser hits the imaging plate it (the imaging plate) releases a blue light which is visible. This blue visible
light is then captured to create the latent image.
CR offers smooth transitions for departments by requiring fewer equipment changes. Those being the ability to use
the same x-ray table and x-ray tube. There are still many other hurdles that can come up when implementing any
new system including digital imaging equipment.
Digital radiography (DR) another category under the umbrella of digital imaging does not use cassettes. Digital
radiography can be referred to as DR or a cassette-less system . A radiology department that decides DR is
the best method will need to replace all the equipment when converting from conventional screen-film.
Depending on the manufacturer of the DR system there are a few different set-ups available. One of the setups is similar to that of a fluoroscopy table where the table does not move up or down, but does move from
horizontal to vertical. There is U-shaped digital radiography equipment where the x-ray tube and film
capturing device are connected in a sort of "U" fashion. Which set-up is right for a radiology department
depends upon the department's needs.
The DR, cassette-less system captures the x-ray image with a flat panel detector, containing a photoconductor.
The flat panel detector takes the place of a traditional Bucky. Manufacturers of DR equipment have made two
versions of the flat panel detector. One version is direct conversion the other is indirect conversion.
The difference between direct and indirect digital conversion is very slight. The name of each basically gives it
away. With the direct conversion the x-rays are directly converted to a digital signal. With indirect conversion xrays are first converted to visible light, and then to a digital signal. Theoretically indirect conversion would lose
more information compared to direct conversion, if a person was following the rule of thumb that every time a
signal is converted a part of it is lost. However, most DR systems are manufactured with an indirect conversion flat
panel detector, so the loss of information, if any, is minimal.
There are advantages and disadvantages to each DR system. Each radiology department will need to reflect on
their specific needs to decide which is best for their situation. The largest advantage with a DR room is the
increase in speed. DR rooms eliminate all steps pertaining to the placement and processing of a cassette,
allowing for the manifest image to be seen very quickly. The biggest disadvantage of the DR system would
be the need to replace all the x-ray equipment which can be more expensive then a budget has room for.
The largest advantage of a CR system would be the ability to keep most x-ray tables and x-ray tubes in
a department that wants to convert. This would allow even a small budget to convert to digital imaging. The
largest disadvantage of a CR system would be it is not the fastest system available.
In the end, it is each departments individual decision which system is right. The important item to remember is
with either system over time they will be more cost-effective and environmentally safer than conventional filmscreen. This happens through eliminating the purchase of film and harsh chemicals to process the film.
Image Manipulation
What is it about a digital imaging system that allows for this wide latitude? Part of the reason is the image
manipulation tools. There are several tools available to manipulate the final image. There are simple tools
likeimage orientation that allow users to place the image in proper reading placement. Then there are more
complicated tools like image stitching that allows a user to piece several x-rays together to form one. Image
stitching is commonly used with scoliosis or leg length films.
In addition to these tools there are also tools available that can alter the contrast and density of the final image.
These tools help provide the wide latitude for digital imaging systems. Window and level functions are easily
manipulated with the movement of a mouse. Moving the mouse adjusts the gray tones on the image. Instead of
being stuck with whatever contrast you have from you exposure, you can now manipulate the contrast in an effort
to salvage a once repeatable exposure into an exposure of diagnostic quality. It is this piece that makes the
latitude of digital imaging wider than conventional film-screen.
Shuttering is another handy tool provided by a digital imaging system. Similar to the idea of masking in
mammography, to allow for better quality diagnosis digital imaging systems allow the background to be black
instead of white. So if you can visualize a view box with a hand film hanging you can visualize how difficult it can
be to spot hair line fractures because of the brightness around the image. This is where another x-ray or piece of
paper is held up around the hand to better visualize it. That is known as shuttering with a digital imaging system.
No more finding a sheet of paper, the computer does it for you.
Take for example a chest x-ray, the central ray is to be located on T7, if an image was taken with the central ray
located at T12 the computer does not know that and will process the image as if the central ray is at T7 leaving for
the distal portion of the lung field to be processed with the contrast and density of the mid-portion of the thorax.
This becomes a problem with heavier patients because the computer does not know to compensate for the missposition of the central ray, leaving the area around the costophrenic angles of poor diagnostic quality.
There are some exceptions to this idea, and that is when the digital imaging system is programmed to read two
images on one imaging plate. Not all systems are programmed to allow two or more images on one cassette, for
example a finger x-ray, whether the digital imaging system is programmed for two or more images on one
cassette would need to be asked of the installers of the system.
The amount of spatial resolution available on a digital image can vary just like that of conventional film-screen
images. Similar to film-screen images the properties of the emulsion; thickness of the active layer, number of
crystal-halides, and the size of the crystal halides the properties of the active layer of a digital imaging plate can
affect the spatial resolution.
The big difference between the spatial resolution available with conventional film screen combinations and that of
digital imaging is what you can do after the image is taken. All radiologic technologists who have worked with
conventional film know that once the image is processed there is nothing more to be done. Digital imaging on the
other hand does have some tools that can be used to help improve the spatial resolution of an image.
Edge enhancement and smoothing are two of the tools available to control spatial resolution. To help process
images faster and require less storage space pixel frequencies are averaged across an entire image. This can lead
to some areas being grainy. Edge enhancement makes the field of pixels being averaged smaller making the
image have higher resolution. Smoothing is very similar to edge enhancement it averages pixel frequencies, but
on a smaller scale. Making smoothing the appropriate tool to increase the spatial resolution of small structures,
such as a finger, on a digital image.
The negative side to these tools is once they are used and the image is saved you cannot retrieve the original
image. One would think no big deal, but it is a big deal. The computer monitors in the x-ray suite have
significantly less resolution available on them compared to that of the Radiologist reading station. Therefore what
looks like a better picture in regards to spatial resolution in the x-ray suite may be significantly different on a
reading station. The best practice to increase spatial resolution without manipulating the image is to use close
collimation, decrease the object-to-image distance, and use a small focal spot size whenever possible.
Equipment
The imaging plate, similar to that of conventional film has many layers. In both conventional film and the CR
imaging plate there are protective layers, a reflective layers, and a support layer. In the case of the reflective
layer, it is used to project light away from the imaging plate when it is being processed by the laser reader, where
in conventional film the reflective layer is used to keep light near the film.
The largest difference between conventional film and a CR imaging plate is the active layer. In conventional film
the active layer is the emulsion. Comprised within the emulsion are silver halide crystals in a gelatin suspension.
The active layer of the CR imaging plate used for computed radiography is termed thephotostimulable
phosphor. This photostimulable phosphor is made of barium fluorohalide.
How the x-rays pass through the patient has not changed with digital imaging. What happens after they go
through the patient is where the big change occurs. X-rays upon exiting the patient interact with the imaging
plates active layer made of barium fluorohalide. The x-rays excite the electrons in the barium fluorohalide to the
point where they become trapped within the conductive layer of the imaging plate. The conductive layer of an
imaging plate is meant to reduce static electricity and absorb the excited electrons.
The absorbed x-rays in the conductive layer are stuck here until they are released by the laser reader. The x-rays
will remain stuck here forever until they interact with a red laser, but the interesting part is that immediately after
the x-ray exposure the conductive layer actually starts to lose some of the recorded information. An imaging plate
should be processed as soon as possible. Probably more interesting than loosing information starting immediately
is the fact that a portion of the image will always remain on the imaging plate even after being struck by the red
laser and erased. The amount of the image that will remain is very minimal but over time an imaging plate will
need to be replaced because of the residual build up of x-rays in the conductive layer. Even with this need for
replacement of the imaging plate eventually, CR cassettes are still more economical when you compare them to
conventional radiography cassettes with film and how they need to be processed.
After the manifest image is captured on the imaging plate in the conductive layer it must be placed in a laser
reader to form the latent image. In the laser reader the imaging plate is slid out of the protective cassette at
which time a red laser beam is directed at the imaging plate. The red laser causes the information stuck in the
conductive layer of the imaging plate to be emitted as visible blue light. This blue light is than captured by a
photodetector, which in turn allows the manifest image to become the latent image.
Once the information has been captured by the photodetector the digital imaging system must convert it to a
digital signal so it can be viewed. To do this a number must be assigned to every blue light photon. The number
assigned will denote how many gray tones should be assigned to the photon to establish the latent image. These
numbers are associated with the histogram that has been preselected on the LUT. The numbers also determine
how many gray tones for each light photon will be available in the final image.
Histograms are programmed for each body part to aid in the processing of an image. As the laser translates
the information into a digital signal it will read from the center of the film to the collimated edge. The
information that is within the collimated field is put into a graph. This graph is called a histogram. The
histogram is being adjusted when a person adjusts the window and level settings of an image. Some digital
imaging systems allow a viewing of the graph or histogram while an image is being post-processed.
Digital radiography systems use a flat-panel detector to capture the x-rays once they have passed through the
patient. Direct converters absorb x-rays in an amorphous selenium material called a scintillator then convert the
x-rays into a digital signal. The scintillator used for in-direct conversion captures the x-rays and then converts
the x-rays to a visible signal. The visible signal then interacts with a photodetector which converts the visible
signal into a digital signal.
A big part of how the final image looks is to do with the matrix it is viewed on. A matrix is a collection of many
boxes that forms a digital image. An entire TV is considered a matrix as is the computer monitor considered a
matrix. What makes one matrix different from the other is the number of boxes, called pixels contained within the
matrix. A pixel is one box, the smallest visible unit of an image. As the number of pixels increase in a matrix
Take for example the new digital television sets available, they are sold in 720p or 1080p. The p denotes how
many pixels are contained within the television set or matrix. To enhance the amount of detail or resolution you
see on your television you would want to purchase the highest number of pixels available. The catch here is that
television stations only broadcast in 720p, the 1080p televisions are mainly for individuals that watch a lot of
movies or play video games.
The resolution of the monitor in the x-ray suite is usually around 1000p. The resolution of the Radiologist reading
station is anywhere from 2000p upwards to a 5000p. The resolution needs to be this high to allow for optimal
diagnosis of the x-ray images.
There is one other piece of the puzzle a person should look into before automatically picking the highest pixels
available and that is pixel pitch. Pixel pitch is how much space is in between each pixel in a matrix. The
information that takes up the space between each pixel is an average of all the other pixels around it affecting the
final resolution of an image. One can imagine what the image resolution would look like if the pixel pitch was an
inch or more big. The resolution would be poor, so a bigger matrix is not always better unless the pixel pitch is
minimal.
It is all about the exposure indicator the digital imaging system reads out. The goal of any radiologic technologist
should be ALARA with the production of exposure indicators the objectives to meeting this goal are more easily
achieved if the radiologic technologist is diligent in achieving it. X-rays images are given a range of acceptable
exposure indicators it is the job of the radiologic technologist to produce a diagnostic image with the least amount
of radiation as possible. Whatever the range given for a particular radiograph it should be the goal of the radiologic
technologist to be as close as possible to being underexposed but still being within the acceptable exposure
indicator range. For example with a Fuji system that is indirect, the exposure indicator would want to be as high as
possible without being outside the range.
So why are we seeing dose creep up in the radiology departments that have converted to digital imaging
systems? The main reason is because it is a known fact that if you slightly overexpose the image with a digital
imaging system you can window and levels the image back into a diagnostic range, similar to using a hot light on
over exposed conventional film. In contrast to an image being underexposed the digital image can not be
salvaged. With a better understanding of how digital radiography systems operate radiologic technologists should
be able to work with their physicist to adjust the techniques used and decrease patients radiation doses.
Equipment should also be looked over for any damage on a weekly basis. Cords and wires should be
evaluated for any cracks. The computer monitors should be cleaned weekly along with the keyboards and mouse.
This helps insure adequate viewing conditions and to help decrease the spread of infectious diseases. Laser
readers have an air intake vent that allows it to keep all the working mechanisms cool. This vent should be
cleaned on a weekly basis to ward off any malfunctions with the laser reader due to poor air quality or flow.
In addition to these simple weekly tasks the manufactures of each digital imaging system will have
recommendations to follow to insure a properly functioning unit. The quality control technologist will want to
diligently follow those recommendations to insure the best quality care and to extend the life of the digital imaging
system. Some of the manufacturers recommendations may consist of checking the integrity of the system with a
designated cassette and testing for proper image acquisition with a phantom.
On a daily basis every CR cassette should go through an erasing cycle to eliminate any amount of scatter
radiation. Each imaging plate does go through an erasing cycle in the laser reader after it has been read, however
in addition to this every cassette in a department should be erased on a daily basis. This helps eliminate any
unnecessary scatter radiation that the cassette may have absorbed since they are so much more sensitive
compared to conventional film.
CR cassettes should be wiped down after every patient and their hinges checked every day for proper
unloading into a laser reader. In addition to checking the hinges the barcode on the imaging plate should be
inspected to insure it can be properly read and recognized by the computers information.
The major point for quality control with digital imaging systems is to follow the manufacturers
recommendations. Most state health departments do not have laws in place yet because this technology is so
new they are not sure how to regulate it. To find out your states regulations on quality control contact your states
Department of Health. They are there to help you not hinder you.
In addition to the radiologic technologist monthly, weekly, and daily quality control responsibilities a service
engineer and physicist will do addition quality control tests on digital imaging systems similar to those
performed on conventional radiography systems. The service engineer usually does testing every 6
monthsto insure quality images are being taken and low exposure rates are being employed. The physicist,
depending upon the requirements of the facility, will do their checks for quality on a 6 month or yearly basis.
The other piece of PACS is the communication part. It is with the computer server that allows physicians the ability
to view an image from an authorized location this is considered a form of communication. Where ever a PACS
monitor is available a physician can log-in and view the image, this can even be done from home or cell phone in
some cases.
In order for images to be saved into a PACS the imaging equipment must contain a DICOM compatible with the
PACS. DICOM is Digital Imaging and Communications in Medicine. Since 1985 it has been standard for imaging
equipment to contain a DICOM, but the fine print must be read to insure compatibility.
On a very simple basis a PACS is broken down to image acquisition, a computer monitor and a server .
Of course it can get much more difficult than just image acquisition, one monitor and a server, but these basics
relatively remain the same.
Image acquisition is the taking and digital formatting of the x-ray. The final product after the digital formatting
is termed a soft copy. A soft copy is what is seen on the computer monitor. If one were to print a copy of the x-ray
this would be a hard copy. Soft copy image acquisition has been performed in special modalities for a long time so
the transition to viewing the images only as a soft copy was more like a natural step.
Now what needed to happen was for multiple computer monitors to be able to view the soft copy from various
locations. That is where display stations come into the picture. Display stations are computer monitors set up to
be able to view images on a PACS. The images available to view on a display station have been approved by the
radiologic technologist, so no repeats will be available to be seen. Display stations can be setup throughout a
hospital or clinic. They are however very expensive because of the need for high resolution (more pixels) to get a
proper view of the image. In some cases, if a physician is ok with poor resolution they may be able to view a soft
copy on their Personal Digital Assistant (PDA). It would not be recommended to do a diagnostic reading from a
PDA, but it could be used for quick glances. It is all through a PACS that makes any of this possible.
The PACS has really helped the speed of patient care in a radiology department without infringing on quality of
care. Before PACS hard copy films would need to be hung on a view box and go through several steps before a
report was ready for the ordering physician to read. Then if the ordering physician needed to see the images they
would either need to go to the radiology department or wait for the images to be delivered to them. Now with
PACS diagnostic imaging has become almost instantaneous.
Additional to being able to almost instantaneously view diagnostic images a person can somewhat manipulate the
image on a display workstation. There are applications available that one can use to evaluate the image. The
image can be magnified, the contrast can be adjusted, and there are even measuring tools available. In most
cases these applications can not be saved to the image, the image will convert back to its original state.
There are also several different types or grades of display workstations. Due to the cost of high resolution
monitors usually only those of the Radiologist reading station will be the highest resolution. An ordering or
referring physicians display workstation will more likely be of a lesser resolution compared to that of the
Radiologist. The radiologic technologist display workstations and QC stations will be of similar resolution of the
ordering physicians, usually about 1000 pixels, but will have different applications available to be used; tools that
can delete repeated images and tools to add text to the image.
The radiologic technologists display workstation comes with a minimal amount of memory available to archive
images. The length an image remains available on these display workstations depends upon how busy of a
department they are in. Some of the display workstations will have images from a month ago, while others will
only have the days images available.
In order to view the images on the display workstation they (the images) must be saved in an archive. The
archive for PACS is computer servers. These computer servers replace the customary file room. The volume of
images needing to be saved will determine how big or how many servers are needed for a facility.
There are a few different ways a PACS can communicate with the display workstations and each comes with its
own set of pros and cons. There is a client/server-based system which immediately sends the images to an
archival server upon acceptance. Distributed systems are also available where images are sent to specific reading
stations. For example if MRIs were only read from one location they could always only be sent to that specific
location. There is also a web-based system available. Teleradiology uses a web-based system.
A small hurdle in the beginning of a transition from conventional film to soft copy is what to do with old images for
a patient that may be needed for comparison or hard copy images a patient may bring from another facility. The
simplest solution is to digitize the hard copy films and add a soft copy to the patients PACS file. It is not advisable
to digitize all old films only those necessary for comparison. The digitizing of any unneeded hard copy would take
Summary Points
Reference:
Carter C. Digital R