Operation manual
DSR Bitumen Rheometer
measuring place RHEOTEST RN 4.3
Contact
RHEOTEST Medingen GmbH
OT Medingen
Rdertalstrasse 1
D-01458 Ottendorf-Okrilla
Germany
Telefon: +49 35205 58-180
Fax:
+49 35205 58-297
E-Mail: info@rheotest.de
www.rheotest.de
Contents
1
2
3
4
5
Introduction ..................................................................................................................................................7
Safety Instructions .......................................................................................................................................7
Scope of Supply...........................................................................................................................................8
Unpacking and Checking.............................................................................................................................8
Technical manual.........................................................................................................................................9
5.1
Technical specifications ......................................................................................................................9
5.2
Rheometer basic instrument (figure 1)..............................................................................................10
5.3
Control unit (figure 3) ........................................................................................................................11
6
Startup .......................................................................................................................................................15
7
Test preparation.........................................................................................................................................16
8
Software RHEOTEST RN 4 Manager......................................................................................................18
8.1
System requirements ........................................................................................................................18
8.2
Installation of the software RHEOTEST RN 4 Manager ................................................................18
8.2.1 Installation of USB-COM-Converter..............................................................................................18
8.2.2 Login .............................................................................................................................................18
8.2.3 Set up the serial interfaces ...........................................................................................................19
8.3
Software concept and structure ........................................................................................................21
8.3.1 Job ....................................................................................................................................................22
8.3.2 Sample..............................................................................................................................................22
8.3.1 Method ..............................................................................................................................................23
8.4 User surface.............................................................................................................................................23
8.4.1 Inlays.................................................................................................................................................24
8.4.2 Tree for jobs, samples and methods ................................................................................................24
8.4.3 Help system ......................................................................................................................................25
8.5 Job............................................................................................................................................................25
8.5.1 Definition ...........................................................................................................................................26
8.5.2 Settings .............................................................................................................................................27
8.5.3 Sample..............................................................................................................................................28
8.5.4 Actions ..............................................................................................................................................28
8.5.5 Information ........................................................................................................................................29
8.5.6 Main Menu Job .................................................................................................................................30
8.5.6.1 New Job.....................................................................................................................................30
8.5.6.2 Copy Job ...................................................................................................................................31
8.5.6.3 Insert Job...................................................................................................................................31
8.5.6.4 Delete Job .................................................................................................................................31
8.5.6.5 Samples -> New ........................................................................................................................31
8.5.6.6 Samples -> Copy.......................................................................................................................31
8.5.6.7 Samples-> Insert .......................................................................................................................31
8.5.6.8 Samples -> Delete.....................................................................................................................32
8.5.6.9 Table Settings -> show ..........................................................................................................32
8.5.6.10 Table Settings -> copy.............................................................................................................32
8.5.6.11 Table Settings -> Insert ...........................................................................................................32
8.5.6.12 Diagram Settings -> Show...................................................................................................32
8.5.6.13 Diagram Settings -> Copy .......................................................................................................32
8.5.6.14 Diagram Settings -> Insert ......................................................................................................32
8.5.6.15 Report Settings -> Show......................................................................................................32
8.5.6.16 Report Settings -> Copy ..........................................................................................................33
8.5.6.17 Report Settings -> Insert .........................................................................................................33
8.5.6.18 Unit settings -> Show...........................................................................................................33
8.5.6.19 Unit settings -> Copy ...............................................................................................................33
8.5.6.20 Unit settings -> Insert ..............................................................................................................33
8.5.6.21 Methods for Change................................................................................................................33
8.5.6.22 Export / Import -> Export Job ..................................................................................................33
8.5.6.23 Export / Import -> Import Job ..................................................................................................34
8.5.6.24 Export / Import -> Export Jobgroup .........................................................................................34
8.5.6.25 Export / Import -> Import Jobgroup .........................................................................................34
8.5.6.26 Explorer for transfer directory..................................................................................................34
8.6 Table ........................................................................................................................................................34
8.6.1 Table Display ....................................................................................................................................34
3
Introduction
Quality problems during construction of roads in the whole world as well as qualitatively new bitumen mixtures
require also new, rheological justified measuring methods. Using a DSR-Rheometer is prescribed in different
binding standards.
DSR Bitumen Rheometer RHEOTEST RN 4.3 including Software RHEOTEST RN 4 Manager completely
meets requirements to the DSR-Rheometer stated in following binding standards:
DIN EN 14770,
ASTM D7175-08, ASTM D 4402 / D4402 M, ASTM D7405-10a,
AASHTO T350-14, AASHTO M332-14, AASHTO T315-12, AASHTO M320-10
Contact:
RHEOTEST Medingen GmbH
Medingen
Rdertalstr. 1
01458 Ottendorf-Okrilla
Phone: +49 (035205)-58 0
Fax: +49 (035205)-58 297
www.rheotest.de
info@rheotest.de
Safety Instructions
-
Take into consideration that, depending on the temperature set value, some parts of the casing of the
measuring plate are heated partially to a high temperature. It might result in burns in case of a contact.
Therefore it is recommended to put on heat-insulating gloves during operating, especially during
applying of investigated material on the plate and replacement of the measuring system.
The measuring system must only be operated with a sufficient amount of cooling liquid. The level of liquid
can be read at the coolant inlet (see (7)). During operation, the coolant inlet should be filled to lower edge
of of screwing-on cap threading. The device must be switched off if there are any leaks!
The measuring system consists of the stainless steel brands 1.4305, 1.4057. The material of sealings
(cylinder measuring systems) consists of Fluorine rubber (FKM) with trading designation VITON. For
special cases it is possible to request the use of other sealing materials.
Attention!
The instrument should not be put into service in case of considerable external damages or heavy
noise during operation.
Scope of Supply
The DSR Rheometer measuring place RHEOTEST RN 4.3 consists of:
-
basic instrument Rheometer RHEOTEST RN 4.1 (measuring head, stand, control unit);
plate-plate-measuring system, peltier temperated consists of:
peltier control unit RHEOTEST PP 4.1
peltier temperated basic plate, adapter, basic plate support, exchange grip,
exchangeable plate ( 25mm);
measuring plate P3 ( 25mm);
measuring plate P4 ( 8mm);
set of trimming tools (1 spatula, 1 stripper gripper);
CD with software RHEOTEST RN 4 Manager for CR-, CS-, OSC-Tests (by concrete order) and/or
modul for temperature control;
2 main supply cable for Rheometer control unit and Peltier control unit
interface cable (Rheometer PC);
USB-Adapter incl. extension cable (Rheometer PC);
interface cable (tempering system PC);
USB-Adapter incl. extension cable (Tempering system PC);
packaging;
user manual (Rheometer and Peltier heating);
set of rubber moulds (has to be ordered separately);
calibration kit (has to be ordered separately);
PC and accessories (have to be ordered separately);
After removing the device and accessories of packaging, please check their completeness according to the
attached delivery note.
In case of any transport damage, the carrier and RHEOTEST must be notified and receive a damage report.
Technical manual
5.1
Technical specifications
Rheometer
torque
torque resolution
speed range
speed resolution
angle range
angle resolution
frequency
temperature range
supply voltage
electrical performance maximum
13,5 kg
9,6 kg
0,5 kg
11,5 kg
3,7 kg
Peltier
temperature maximum
temperature minimum
temperature accuracy
interface
dimensions Peltier control unit (W x D x H)
dimensions basic plate (W x H)
length cable
150C
-10C
0,1 K in the range of -10C to 150C
RS 232 or USB 2.0
180 mm x 300 mm x 460 mm
110 mm x 110 mm
approx. 2 m
measuring drive
knurled screw
quick connecting coupling
coupling
ball handle
adapter
basic plate
knurled screw
knurled screw
spring pin
basic plate support
basic plate guide
microindicator
micrometer screw
stand
10
5.3
The on/off switch (1) is located at the front panel of the instrument. At the backside are located: interface
Pt100 (2), interface RS 232 for PC (3), both pin-plugs for Rheometer (4) and mains supply (5).
11
12
One could scroll in the selected choice with the help of this key. It means in menu (see Appendix 1)
one step downward. Jump to the beginning is fulfilled at the end of the choice.
ESC
It means in menu (see Appendix 1) one level (step) to the left or to the end left to the beginning.
Entry of data on display does not fulfil/ confirm.
ENTER
It means in menu (see Appendix 1) one level (step) to the right, one choice is selected or fulfilled.
It is activated, as well as confirmed digits input in the choice where this value could be set up.
MENU
13
Note: Using this key combination, the device can be restarted from memory data even after a possible loss.
UP + DOWN
Note: The peltier control unit may only be operated with a sufficient amount of cooling liquid. The level of the
liquid can be read in the coolant expansion tank (7). During operation, the coolant inlet should be filled
to at least 50%.
14
Startup
PC-Connections
For a PC with serial interfaces the compounds via a serial data cable according to the standard RS232 are to be
produced. The necessary cables are included.
For a PC with USB interfaces on the included USB adapter is connected to the USB port of the PC. The connection
from the USB adapter to the controller or thermostat is made with the supplied serial data cables.
If it does not fit into the USB port of the USB adapter, because it is too large, the included USB extension cable can
also be used.
To connect peltier control unit with PC please use the cable labeled Peltier.
15
Test preparation
See figure 1
1.
Before the first use, fix the temperature controlled basic plate support (11) by knurled screw (8) (at the
backside) at the basic plate guide (12) . The basic plate support (11) has to lie on the left and right
stopper.
2.
Clean the measuring cone and the basic plate with appropriate solvent and dry them.
3.
Insert basic plate (7) into the basic plate support (11) in the following manner: insert the basic plate (7)
(thermostat branch pipes look forward) into the basic plate support, turn 90 (thermostat branch pipes
look left) and put it onto the joint spring pins (10).
4.
Close the opening of adapter (6) by keep turning the window Square (opposite the stainless steel
base) until the opening if the ring window is closed. Turn the closed opening in direction of stand (15)
and place the adapter on the upper ring of basic plate (7). In showing to the left ball handle (5),
push the basic plate guide (12) uncl. Basic plate and adapter up. Lock it by turning the ball handle (5)
into to the right direction.
Thereafter, the opening of the adapter can be rotated by hand to the front, as shown on fig. 5.
Figure 5 Adapter
5.
Again move down the measurement plate guide without the adapter (6).
6.
Couple cone with measuring drive - to do this, lift the bushing of the quick connecting coupling (3) ,
insert the ball head of the cone into the cone opening of the quick connecting coupling and pull down
the bushing.
7.
Set the basic plate (7) into its lower position, turning the micrometer screw (14) left.
8.
9a.
Lift the plate guide on rails (15) until it stops at the adapter.
Fix measuring head (1) with the help of knurled screw (2). Then press the plate guide up to the adapter
(6) to ensure that joint spring pins (10) are imbedded into the plate support (11) (displacement about 3
mm) and fix by the knurled-head screw (9) or:
9b.
Fix the basic plate (7) by the knurled screw (9). Put down the whole measuring head (1) until the joint
spring pins (10) are imbedded (displacement about 3 mm) and fix by the knurled screw (2).
10.
Adjusting gap between cone (or plate) and basic plate: The microindicator (13) serves to control plate
feed value in m range. (13). In this case it is necessary to ease off clamp screw on micro-indicator
shaft, to lift indicator along longitudinal axis in order to use measuring range completely.
Move the basic plate (7) by micrometer screw (14) carefully upwards (turning right), until basic plate
will nearly touch the cone (or plate).
The gap has to be set only for completely thermostated measuring device, because the gap can
change (thermal expansion). In the case of large temperature changes (thermal expansions), the cone,
plate or measuring shaft bearings might be destroyed.
16
Note: The touch of cone (or plate) and basic plate can not be felt at the micrometer screw (14), but only to
existing torque on the cone/plate. The cone (or plate) does not longer rotate freely.
Set the display of microindicator (13) by turning the outer ring to zero. The specified distance is set by turning
the micrometer screw (14) counterclockwise (to the left). In order to take out backlash, first lower too much
about 10 m and then adjust the precise gap by rotating clockwise (to the right).
In the software RHEOTEST RN 4 Manager, the value for the trimming distance (e.g.: 0.05 mm) is
predetermined. This has to be in addition to the gap ("window adjust distance: distance: 1.05 mm") set.
Note: The gap is to be checked after every 10th measurement or once a day.
9.
Unlock knurled screw (9) and move basic plate support upwards (11).
10.
Place the sample centric on the basic plate. All bitumen samples have to extracted according to
DIN EN and prepared according to DIN EN 12594. We recommend tu use the rubber moulds for sample
preparation including in scope of supply).
11.
Run the measuring head (1) and the basic plate (7) together again, for it carry out steps 8 and 9a or 9b.
The sample has to be be trimmed with a spatula or a trimming tool - the excess material to be
measured removed. For trimming start job preparation_trimming in the software RHEOTEST RN 4
Manager. After finishing the job, please trimm the sample.
The sample may optionally be trimmed with a spatula or trimming tool. The trimming tool is used, as
shown on the Fig.6. Lay the stripper gripper flat on the basic plate. The inward facing ends are applied
to the measuring plate. By simultaneously pressing the brackets and pulling forward the tool, the excess
material is removed.
To ensure all-sided trimming of the sample, the opening of the adapter can be rotated forward (see fig.
6). After trimming, the window is to turn back again to make ready for measurement. After trimming, the
opening has to be turned back again to make ready for measurement.
16.
After measurements are finished, move down the basic plate (7) with coupling bushing in upper position
(the measuring cone (or plate) remains "stick" on the basic plate (7), pull the measuring cone forward,
clean the basic plate (7) and the cone.
Subsequently, new sample is applied and after the cone (or plate) was hitched again, a new
measurement can be started.
17
For installation and working with the software, the following requirements have to be met:
- computer operating system Windows XP SP3, Windows 7 or above
- Microsoft.NET Framework 3.5 or above (normally included in computer operating system)
- processor: at least Pentium 4 or matchable, frequency of 1,7GHz or above
- working storage at least 1 Gbyte
- free memory space on harddisk of 10 MB
- 1 CD-ROM device
- 2 COM or USB interfaces for rheometer and thermostat / cryostat
- option printer port with printer
Login
To be able to login the software RHEOTEST RN 4 Manager, different logins with appropriate access rights
available:
user name
Operator
Engineer
password
none
engineer
Administrator
rheotest
rights
execution of tests and access to created jobs and methods
additional to operator, additional set up of communication and creating
of jobs and methods
additional to engineer, additional configuration of measuring systems,
creating of new measuring systems as well as change the system
language
18
8.2.3
57600
8
0
1
Setting of Com-Port
1. click Check Port" all available PC interfaces are added to the selection list.
2. Select the interface on which RHEOTESTRN 4.1 is connected.
3. click Init Port" the interface is initialized, that means the hardware receives the
communication parameters.
4. click INIT send" now, a communication to Rheometer RHEOTESTRN 4.1 is established
by sending a special code (send).
5. Typical, the Rheometer RHEOTESTRN 4.1 responds with a special code (receive). If this
code appears, the communication is correctly installed and configured. The user can start
measurements now.
6. If no code appears but an error message, then the communication is not correct. Please
check the cable connections as well the Rheometer RHEOTESTRN 4.1. The following
issues should be reviewed:
selected COM-Port is wrong
RHEOTESTRN 4.1 is not switched on
wrong type of cable
defective cable
(COM USB converter not installed correctly)
defective COM-Port of PC, possibly test on another computer
If communication has once been working correctly, after re-starting the software again no settings are
necessary (if the configuration has not changed).
Integration of a temperature control system for temperature control:
Activate in menu Communication settings the tab Temperature-controller
For the peltier system PP no settings of communication are necessary. All parameters are specified:
Baudrate:
9600
Data bits:
8
Parity:
0
Stop bits:
1
19
Setting of Com-Port
1. Click Check Port all available PC interfaces are added to the selection list.
2. Select the interface on which the peltier control unit is connected.
3. Click Init Port - the interface is initialized, that means the hardware receives the
communication parameters.
4. Select type of temperature control and click Load Types.
5. Now, in the tab Temperature a communication to peltier control unit can be started:
cur. Temp. [C]: by click Read the actual temperature is displayed
settemp. [C]: setting of a target temperature
click Set: send the signal to peltier control unit
click Read: displaying the actual temperature
The peltier control unit as well the temperature control can be started now. Additional,
temperature ramps are possible.
If an error message appears, the following issues should be reviewed:
selected COM-Port is wrong
RHEOTESTRN 4.1 is not switched on
wrong type of cable
defective cable
(COM USB converter not installed correctly)
defective COM-Port of PC, possibly test on another computer
If communication has once been working correctly, after re-starting the software again no settings are
necessary (if the configuration has not changed).
Other types of temperature control systems:
Select controller type, transfer parameters and serial interface for temperature control system. (Before start,
you have to know the transfer parameters of controller type (please see manual or configuration of device).
Note USB interface: Even by using an USB interface, a serial interface must be selected. After installation of
the USB driver serial interfaces are provided in the WINDOWS system. Those serial interfaces can be
determined and checked in WINDOWS system control after installation of the driver software.
During the installation of the USB driver default parameters are setted. If the communication is unstable by
using a USB interface, we recommend to set the parameter waiting time [ms] to 4. (Advanced Connection
Settings).
Exit the program, all settings are saved.
20
21
8.3.1 Job
A job contains all components necessary for processing of
one theme, such as:
1. Method for reasonable data generation
(analysis method) as a proper tests sequence
with corresponding test parameters.
2. Settings for data output as tables and
diagrams.
3. Settings for data analysis such as report on
evaluation for determination of rheological
characteristics.
4. Documentation of results and researches such
as printing and archiving.
In job Definition information is contained that describes
the job exactly, e.g. name, commentary.
Further setups are in job Settings as methods of
presentation and calculations for oscillation measurements
and possibilities and print options that will be taken into account during job printing.
Jobs are organized in jobs groups. One jobs group should contain jobs that relate to the upper theme.
It is possible to copy jobs from one group into another. In such a way reorganization of jobs is possible in new
created jobs group when, e.g. boundary conditions have changed or for thematic archive.
One could export jobs groups in order to give excess to them for other users. They could see data, analyse
them or add own measurements.
Besides there is in a job control of measuring results (samples). Any number of samples could be in one job.
A job could be created in any place in jobs group.
Note
It is better to copy available job with the similar theme for a new theme of analysis without measuring results
(samples) and contained inside settings should be adjusted for a new theme.
In this case there is a functional start point that should be optimized only.
8.3.2 Sample
The sample contains measuring results and some information
that is necessary for data analysis.
The sample does not contain information about conditions of
data obtaining or how they
should be represented and analyzed.
Information about representation and analysis should be
controlled in the job because samples in one job relate to one
theme.
In such a way all samples will be represented and analyzed in
the same manner.
One could copy samples from one job to another. After this
settings of the new job are in effect. It means that data will be
represented and analyzed in another way according to setup
of table, plot, report, units of measurement.
22
8.3.1 Method
An important constituent part of a job is method that
should be used for analysis.
Methods are saved in a library because task settings
are repeated mainly.
Due to this it is possible to use already developed
methods time and again. Later on the available
methods could be used as a starting point for
development of own methods.
A method contains a tests sequence or one test only.
This test contains information about operation with
material (shear / deform) and which data should be
measured as a response to load applied.
So, e.g. it is possible to set shear rate during defined time at certain temperature and to measure shear stress
necessary for this.
Method contains parameters necessary for data transfer from rheometer to computer as well as for reliable
measurement. Such as, e.g. time raster, boundary values and interruption value.
Further method contains units of measurement for all parameters. These units could be brought in
correspondence at any time and represented at once in tests table and at a plot.
Methods, as well as jobs, are organized in groups.
Methods could be grouped in accordance with the measurement conditions in the following way:
- Measurements with controlled shear stress - CS
- Measurements with controlled deformation - CD
- Oscillation measurements
-
Or according to analysis themes as:
- Material A
- Material B
-
- Goods supply
- Monitoring
-
It is possible to copy methods from one group into another. In such a way it is possible to reproduce new
thematic situations.
One could export methods in order to render to other users.
Information is in method definition that describes the method exactly, e.g. name, commentary, and sequence
of tests. Printing parameters are in methods settings that will be taken into account during method printing.
23
1 Main menu
2 Inlays
3 Job tree
4 Sample tree
8.4.1 Inlays
Presentation is organized in inlays of settings and analysis results. They could be selected by pressing name
of an inlay.
Job: It is shown certain job and settings of the actual selected sample, control keys to begin an analysis
and detailed information about selected sample at the moment.
Table: Table of data is shown in accordance with the settings.
Plot: Plot is shown in accordance with the settings.
Report: Report data are shown as statistic evaluation or calculated rheological parameters.
Method: It is shown job method or methods in methods library.
8.4.2 Tree for jobs, samples and methods
Designators for two trees are at the left side of window in RHEOTEST RN 4 Manager. They
are independent of the selected inlay for job, table, plot or report.
Through the upper tree it is possible to select a job in jobs group that should be presented.
The corresponding image will be activated and active sample will be shown from the selected
job after click.
It is possible to select active sample in the job in the low structural tree.
After selection the corresponding image will be active and data for all tests will be shown.
It is necessary to click + sign in front of sample name, if it is necessary to display only one
test. After this individual tests will be listed. If only one test is selected, so only those datas
are shown in table or graph.
Width and height of structural trees could be adjusted, if necessary.
Methods groups with their methods will be shown at upper tree, if display is activated of
methods library. Method with its settings will be shown after click.
24
8.5 Job
1 Job tree
2 Sample tree
3 Active job definition; inlay behind that
contains further job settings
4 Name of active sample
5 Field of job execution
6 Information about active sample
25
8.5.1 Definition
Job definition contains the job description and information about necessary components such as measuring
system and method.
1 Job name: the job name should describe jobs theme. It makes easier the further search.
In principal it is possible to give the same name to many jobs. However it could lead to confusion very quickly.
Therefore it is not recommended.
2 Method: it contains all necessary information for analysis carrying out. For this see Section Method.
Display of method list shows all available methods in the library at the moment. At that method name will be
shown and methods group after it. At present methods groups available in the library are set in General
Settings.
Method data are saved in the job directly, that is changes in a method in the method library do not influence
the next measurement directly. Neither it is required that set method is available in the library. At this it is
possible to operate with a setting that was obtained and imported from external source, e.g. from engineer of
Rheotest Company.
It is necessary to press R key (see paragraph 4 R (Reboot)), if you have changed method in the library and
want that this change works during the next measurement.
If a new method is selected from the list, request appears if it is necessary to take over the methods units of
measurement. Each job and each method have their own settings of the units of measurement that should be
in accordance with the given measuring task.
The request makes it possible a selection, if it is necessary to keep the settings of the job units of
measurement or it is necessary to take over the settings of the job units of measurement from the method.
If you would like to change method that was imported from outside, it is necessary to copy it in the library
through Method for Change.
3 Measuring System: measuring system that is used for sample measuring. Measuring system data are
copied at start of each measurement from the library of measuring systems. At that changes will be hold true
by every new measurement, e.g. calibration.
Message will appear and it is necessary to select another measuring system, if specified system is not in the
library.
4 R (Reload): At that actual methods settings will be copied from the library in the job. It has an influence on
all other changes.
Button is active only in that case when the specified method could be accessed in the library.
5 Comment Job: Always it is necessary to record here substantial comments concerning the job. It might be
information about the jobs task such as incoming inspection from supplier A, product B development or
monitoring of the production line XY.
It gives possibilities for other users of quick classification of jobs for their own tasks.
Besides that this information could be used for documentation in table heading, plot heading or in report
heading.
6 Product: Product name that should be analysed with the help of this job. However this field could be used
also for equipment, supplier etc. This information could be used for documentation in table heading, plot
heading or in report heading.
7 Created: Date and time when the job was created or when the job was imported.
26
8 Changed: Date and time when the job was changed last time. It relates to all jobs settings.
9 Last used: Date and time when the job was used last time. It relates to the start of a measurement.
It is unimportant, if the measurement of a sample was interrupted or deleted.
8.5.2 Settings
Further jobs settings.
1 Suppress Wait Time at the beginning: The first measured value is transferred from rheometer
RHEOTEST to computer after defined time depending upon methods parameters used Trigger time and
Time raster. Measured values could be also presented / displayed starting from the moment when time = 0
with the help of Suppress Wait Time at the beginning.
2 Show Sample Collection: Data from the data list will be also shown at a plot in parallel with the actual data.
Data of the data list are set in diagram settings.
Data of the data list will be always shown with the line thickness XS and appeared in legend with their set
names. Colours and symbols correspond to colours and symbols of the actual data.
3 Show Complete during Analysis: This option is active in that case only when data list for this job will shown
also. This option acts during one measurement.
Diagram will be shown and scaled in accordance with the actual data, if this option is inactive. This concerns
X- and Y- axes.
Diagram presentation and scaling is defined by the data from the data list before beginning of the
measurements, if this option is active. It means that, if the data list includes measuring results during 300 s the
measuring parameters will be shown completely during this 300 s. Actually measured data will be added to
this image after analysis.
4 Osci-Correction: There is possibility to correct measuring data during data processing of oscillation tests;
in particular J-correction assumes inertia correction.
5 Standard Temperature: This temperature is set at the following conditions:
27
8.5.3 Sample
Sample for analysis is described by the following data.
1 Sample Name: This name will be shown in sample tree and could be used for documentation in table
heading, diagram heading or in report heading. Name should be indicated before analysis start. Checking is
carried out at every analysis begin. However the name could be changed later.
In order to simplify work the name could be formed automatically, especially if there are a lot of similar
samples. For this see the next Section.
2 Name Mode: It defines how sample name will be formed. Automatic name formation is necessary for quick
start.
The following possibilities are at the disposal:
Manually: Sample name input at every analysis start. Quick start is impossible.
Charge+Index (_000): Here the name will be formed from the charge names and increasing
index. For that the charge should be indicated necessarily. Index begins from 000 and the next free index is
used. If e.g. there are all samples up to the sample _006 and the sample _003 is deleted, the next sample will
be sample _003 and it will be located after the sample _006.
DateTime: The sample name corresponds to actual date and actual time in format YY-MMDD-HH-MM-SS (for example 2012-03-18-10:42).
Charge+DateTime: The sample name is formed from the indicated charge, actual date and
actual time (for example CD-Glue 2012-03-18-10:42). One obtains very good documented sample name with
this option.
3 Comment: Comment concerning the sample. Comment to the sample might be changed after successful
analysis.
4 Charge: Name of analyzed charge. At that the charge could describe any other object such as for example
device, supplier, products line, monitoring name The charge is necessary obligatory in case of using names
mode Charge+Index and Charge+DateTime. The charge could be used for documentation in table
heading, diagram heading or in report heading.
5 Operator Name: setting of operator name. It could be set manually. New recording is impossible before
analysis in case of quick start. The recording could be changed after analysis. It might be indicated in settings
of user control, if this information should be filled automatically at every analysis start.
6 Sample Created: Date and time of sample creation.
8.5.4 Actions
Job actions are related always to the actually selected job.
1 New sample: New sample will be prepared for analysis. It is
possible only in that case when other sample is not activated.
Request appears after click, if it would be necessary to use available
data. It relates first of all to the sample name, charge and comment.
These records will be deleted, if one answers No to the request.
Mode of sample name and user are saved.
Now new records could be done, for example a new charge should
be analysed.
These records will be saved and will be valid from the beginning for
the new sample also, if one answers Yes to the request.
28
It is possible to know about situation that a new sample was created by indication Active near button New
Sample. Now changes made in sample data will be related to the new sample.
All changes in sample data will be in force for the actual sample, if this mode is inactive at the present
moment.
Request follows, if it is necessary to use this sample for a new analysis in special case when the last sample
does not contain any data. For new analysis will be used the last empty sample, if the answer to the request
will be Yes.
2 Check + Create Sample: Input data and system will be checked after this button pressed. New sample will
be created and data structure for this, if the check was successful. In case of good creation an inscription OK
will be shown near the button Check + Create Sample. It appears in information message Press button
Start Analysis and it appears new sample in the sample tree.
In other case it appears error message with information what should be changed.
The following will be checked:
Does exist the inputting sample name already and relates to manual input
Was the charge indicated and relates to the name input mode Charge+Index and
Charge+DateTime
Request about distance between plates. It relates to measuring systems Plate-plate only.
3 Start Analysis: Analysis begins after this button was pressed. Check is carried out. If a thermostat is used:
Connection with thermostat will be checked, if this is indicated in the method.
Then the analysis task will be sent to RHEOTESTRN 4 control unit. After that it is necessary to wait as
RHEOTESTRN 4 control unit sends data and analysis would be completed.
One could follow the sequence of this procedure at information display.
4 Easy Start: This version of analysis start is right for a lot of similar samples that are tested one after
another in a sequence. The sample will be prepared and analysis will be started. However start takes place in
that case only when all checks were successful. In other case error message appears.
5 Cancel Analysis: Active analysis will be interrupted at once after pressing. It is possible to cancel analysis
at the stage of analysis preparation. It should be beard in mind that at this error message could appear that
communication is impossible.
After analysis interruption it would be proposed to delete it.
If communication between RHEOTEST RN 4 Manager and RN control unit has been interrupted but the
reliable data of analysis were already obtained, it is possible to interrupt the analysis and one could use data
obtained before this further.
It is possible to identify loss of communication by that feature that data transmission does not take place
during a long period of time. One could check it at information display.
8.5.5 Information
Analysis status is shown at display or sample content is shown for measured samples. Analysis status begins
with pressing button New Sample or Easy Start. The actions fulfilled and their results will be displayed. If
one begins analysis in manual mode, but not easy start, so after each operation instruction appears what to
do next. For example after pressing Check+Create Sample instruction appears that it is possible to start
analysis with the help of Press button Start Analysis.
Tests that are carried out and already obtained data packages are displayed during analysis.
The sample content will be displayed from the point of view
of available data in case of already measured sample.
29
Table
Diagram
Report
Number of columns 3
Column 1 Time
Column 2 Torque
Column 3 Number of revolutions
All tests will be shown
There is no tables heading
Tables settings would not be printed
Mean value/standard deviation of torque for all tests, it is not shown at a diagram
There is no reports heading
Diagram will be added during report printing.
30
Insert complete job It will be inserted the job with all samples. This possibility is useful, if you
want to reorganize the jobs or place them in other jobs groups.
Insert empty job It will be inserted the actual job only. Jobs samples will not be inserted
together with it. This possibility is useful, if you want to start available job with new samples. It
could be for example in that case when monthly you lay together samples for storage in one
job. At the beginning of the month the job is copied without samples and in name new month
is added (For example, Product_A_May2014).
Now all samples for May could be analyzed.
Note: Jobs are inserted after the marked job. In order to obtain the job at the first place in the jobs group it is
necessary to mark the jobs group.
31
Delete all in job all samples will be deleted in the actual job. This possibility is used, if for
example copied job should be used for a new samples series.
Note: Inquiry follows for deletion in that case only, if it is indicated in users settings. Cancel
function is inaccessible.
32
33
8.6 Table
Table view is used to display measured values and calculated parameters in the form of values list.
8.6.1 Table Display
If inlay Table is active, the measured values and calculated parameters of the active sample will be shown
in the form of a table. Data could not be changed.
Selection of the displayed parameters is fulfilled in Table Settings.
Table dimensions are adjusted to the dimensions of RHEOTEST RN 4 Manager window.
Columns width could be varied with the help of a mouse and will be saved. For that it is necessary to put
cursor above the column right border. As soon as arrow left-right appears one could set the column width left
mouse button pressed and hold.
Note: If moves mouse above the right column border, right-left arrow appears. If at this one does two clicks,
the column width is adjusted automatically for optimal size.
34
35
8.6.3.2 Tests
1 Tests: it is possible to select here tests that should
be represented. This setting is active in that case only
when sample is marked in sample tree. It means that all
available tests could be represented.
If only one test was selected in sample tree, this setting
is ignored.
The setting is active during one analysis and in the
following representation. It means that for example in
one analysis and inactivated the 1st test any data will not
be shown until the 1st test is carried out.
For example conditional steps could be hidden with this option.
2 All: All tests will be selected.
3 None: Selection of all tests will be cancelled; data are not represented until the sample is selected.
Representation is still active in case of selection of single tests.
8.6.3.3 Head
Table heading will be shown above the table itself. It could contain additional information to analysis.
This table heading could be configured for every job. In this case it is valid for all samples that are in the job.
Table heading makes possible to display 9 different information which is located as a matrix of 3 columns and
3 rows.
1 Column/Row: Information could be assigned from
drop-out list for each field. See in Application Heading
Data details of this selection.
2 Name: If this option will be marked, then the selected
marking will be shown before information itself. So it will
be displayed for example Sample: Sample Date during
selection of the sample date.
3 Free texts: All texts could be saved here that
afterwards will be shown as information in table heading.
It makes possible good documentation of analysis
results.
Settings shown at the figure will be displayed in the table heading in the following way:
Printing
If option Add Settings is marked, then settings will be presented during table printing. See Table Printing for
that.
8.6.3.4 Printing
If option Add Settings is marked, then settings will be presented during table printing. See Table Printing for
that.
36
8.7 Diagram
8.7.1 Diagram Display
If bookmark Diagram is active, it will be displayed measured values and calculated parameters of the active
sample. Any data could not be changed.
The following main settings are possible: displayed parameters, colours, lines thickness, marking, coordinate
grid, scaling, linear/logarithmic and much more.
These settings take place in Diagram Settings that could be displayed by a simple mouse click at a diagram or
could be called through the main menu.
Diagram dimension is adapted to the dimension of RHEOTEST RN 4 Manager window.
Image is divided in 2 areas, diagram heading is above and diagram itself is below.
Diagram heading will be shown in that case only when this information has to be displayed.
If any information was not defined for displaying in diagram heading, this area will be used for diagram square.
The diagram itself is depicted under the diagram heading. It is possible to represent 4 parameters. The
corresponding scale is depicted to the right and to the left of the diagram area. Displayed sequence of data
names appear above the diagram square under the diagram heading.
8.7.2 Main Menu Diagram
8.7.2.1 Diagram Settings -> Show
Diagram settings of the selected job are shown. New setup window will be opened.
See settings details in Section Diagram Settings.
This function acts in main menu as with the help of pressing Diagram -> Settings, as well as by pressing
submenu Show.
One could not control the main window of RHEOTEST RN 4 Manager until setup window is active.
37
8.7.2.5 Settings
They serve for setting of content in form of a diagram. This setup is active for the actual job and with it for all
samples in this job. The settings are divided in 6 tab pages: Axes, Grid, Tests, Heading, Collection, Print.
All changes made will be rejected with the button Cancel and Menu Settings is closed. The changes will be
accepted and they become actual ones with the help of button OK. Menu Settings is closed.
8.7.2.5.1 Axes
It serves for coordination of displayed parameters with
diagram axes as well as for definition of that how they
have to be displayed.
1 Type: Setting of displayed parameters from the list
of possible variants for X axis.
Available parameters and their meaning one could find
here: Application Parameters X Axis.
2 Scaling: It will be defined scaling regime of X axis.
In case of Auto scaling will be fulfilled on the basis of
displayed data. Scaling is carried out from 0 to
maximal data value.
In case of Min-Max displayed limits could be set. Data that are outside these limits will not be displayed. It is
possible in this mode display fixing and its zooming.
Min: Minimal value of X axis. Data smaller than this value, will not be displayed. This value is active
only in the scaling mode Min-Max.
Max: Maximal value of X axis. Data will not be displayed that are higher than this value. This value is
active only in the scaling mode Min-Max.
log: Logarithmic scale is displayed. If this option is not selected, linear scaling is displayed.
3 Sort: Sorting takes place of all data (all displayed parameters) according to X axis parameter in increasing
order.
At that it is unimportant at what time and in which test data were obtained. Sorting occurs by all actual
displayed data (either one single test or all tests of a sample).
4 Definition of all Y axes: At the left, at the right, 2. at the left and 2. at the right.
5 Type: Definition of displayed parameters from the selection list for the corresponding Y axis. One could
find available parameters and their meaning here: Application Parameters Y axis.
Note: One parameter could be displayed at several axes. It might be used for example in order to have one
parameter with different scaling.
6 Active: It sets, if this parameter will be displayed (marked) or not.
1. Parameter (at the left) is displayed always independent from this setting.
7 Scaling: It will be determined mode of Y axis scaling. In case of Auto scaling will be realized on the basis
of displayed data. Scaling occurs between 0 and maximal data value.
In case of Min-Max displayed limits could be determined for each axis. Data beyond these limits will not be
displayed. It is possible display fixing and its zooming in this mode.
Min: Minimal value of X axis. Data smaller than this value, will not be displayed. This value is active
only in the scaling mode Min-Max.
Max: Maximal value of X axis. Data will not be displayed that are higher than this value. This value is
active only in the scaling mode Min-Max.
log: Logarithmic scale is displayed. If this option is not selected, linear scaling is displayed.
38
8 Colours: Colour setting for this parameter (Axis). This setting is displayed also in data collection, see for
that Collection.
There are following colours for selection:
Black
Brown
Violet
Blue
Green
Red
Pink
Orange
Yellow
Grey
White
9 Thickness: Lines thickness could be of 5 different values. The following values are possible with the
thickness in pixels:
XS
S
M
L
XL
1
2
4
6
7
The line thickness relates to displaying of actually selected sample. Data collections lines will be presented
with thickness = XS always.
Signs form for lines marking could be set by User also. It is possible selection from the following list for lines
marking:
None
Square
Diamond
Triangle
Triangle at the right
Triangle at the left
Triangle inverted
Circle
The mark form relates to displaying of the selected sample as well as to data collection displaying.
Note
It is recommended to select black colour for line for all parameters and set difference in parameters with the
help of various marking in case of using diagram in black and white presentation.
One could vary the marking size also. The following dimensions and values in pixels are possible:
XS
S
M
L
XL
3
5
6
8
9
39
8.7.2.5.2 Grid
1 X grid: one sets that vertical grid will be shown
corresponding to X axis. At that values range will be
optimally divided in order that appear about 7 to 12
portions.
X grid is displayed with width of 2 pixels.
X subgrid: one sets that vertical grid will be shown
corresponding to X axis. At that values range will be
optimally divided in order that appear about 14 to 24
portions.
X subgrid is displayed with width of 1 pixel.
2 Y grid: one sets that horizontal grid will be shown
corresponding to left Y axis. At that values range will
be optimally divided in order that appear about 7 to 12 portions.
Y grid is displayed with width of 2 pixels.
Y subgrid: one sets that horizontal grid will be shown corresponding to left Y axis. At that values range
will be optimally divided in order that appear about 14 to 24 portions.
Y subgrid is displayed with width of 1 pixel.
3 Grid colour: Setting of grid colour. Colours are possible that are available by colour selection for
parameters. See for that section Colours.
8.7.2.5.3 Tests
1 Tests: here tests could be marked that should be
displayed. This setting is activated in that case only
when sample is marked in sample tree, that is to say
all available tests could be represented.
This setting is ignored, if in the sample tree one
selects single test only.
The setting is in effect during analysis and further
displaying. It means that, for example, at one analysis
and one nonactivated 1. test, any data will not be
displayed until 1. test is fulfilled.
Conditional steps could be hidden, for example, with
this option.
2 All: all tests will be selected.
3 None: Selection of all tests will be cancelled; data are not represented until the sample is selected.
Representation is still active in case of selection of single tests.
8.7.2.5.4 Head
Diagram heading will be shown directly above the diagram. It could contain additional information about
analysis.
This diagram heading could be configured for each job. It is valid afterwards for all samples that are in the job.
Diagram heading makes possible to display 9 different informations which are located as a matrix of 3
columns and 3 rows.
Since the heading specification corresponds to the table heading, see for that Section Table Settings
Heading.
40
8.7.2.5.5 Collection
RHEOTEST RN 4 Manager proposes possibility to show samples that had been already analyzed as a
background for analyzing or displaying sample. It is unimportant from what job background samples were
taken or they were imported from external source.
This possibility could be used, for instance, that case when during material changes the desired characteristic
of the material always presents against the background as a target. So it is possible to compare different
samples very quickly and effectively on the basis of their analysis.
Besides it would be possible to show permissible limits in routine analysis against the background. At that
two samples were shown against the background, one of them represents the lower limit and another
represents the upper limit.
Then actual analyzed sample must be presented during measurement between these two samples. At that
there is possibility to check very quickly, if the actual sample represents a deviation.
1 Name: New name could be assigned for a
sample. At first original name will be inserted after
sample addition in the samples collection. In order
better to distinguish and identify samples at diagram
later, a new name could be assigned according to
the themes.
2 Active: Here one could select which samples will
be actually displayed.
Sample collection could contain more samples than
it must be shown. With the help of this it is possible
quick switching over or turning on/ turning off of
different samples.
Marked = Sample will be shown.
3 Line type: Displaying parameters of background samples will be presented at a diagram with the same
colour, such as actual sample. It is possible to set here line type in order to distinguish the background
sample from the actual sample.
There are four lines types:
Dot
Dash
Chain line
Dotted line
If it should be displayed more than four samples, it will not be possible to distinguish them by lines type.
4 Sample details: Sample definition data will be shown. It gives possibility for better samples classification.
The following data will be presented:
Original sample name
Date of sample creation
Field recording Charge
User name
Comment to sample
5 Sample details: Sample definition data will be shown. It gives possibility for better samples classification.
The following data will be presented:
Measuring system in use
Factor KTau of measuring system
Factor KD of measuring system
Correction factor for rotary tests
Correction factor for oscillation tests
Number of tests carried out
41
6 Copy: Selected sample from samples collection will be copied in the clipboard. Windows clipboard is not
used.
Afterwards this sample could be inserted in another place of samples collection.
It gives possibility for sorting, if the previous sample is deleted later.
Later on one could insert this sample in the clipboard I any job, in any place after closing of diagram setting.
7 Insert: Sample from the clipboard is inserted after the selected sample in the samples collection. At this it
is unimportant, if this sample was copied from the job before this or it is earlier copied sample from the
samples collection.
There are not limitations concerning number of samples that will be saved in the sample collection.
8 Delete: Marked sample is deleted from the sample collection.
9 Delete all: All samples are deleted in the sample collection.
8.7.2.5.6 Printing
1 Add Report: Actually set report will be added
after diagram printing.
2 Add Table: It will be added actual table after
printing of diagram or as well as a report
(optionally).
3 Add Settings: It will be added diagram
settings after printing of diagram or as well as a
report (optionally) or a table.
All 3 options could be selected in any
combination. See for that Diagram Printing also.
8.8 Report
8.8.1 Report Display
Certain evaluations of the actual sample will be shown, if inlay Report is active. These could be statistical
parameters such as: Mean value, Standard deviation, Minimal and maximal value. Besides, these could be
calculated, rheological parameters for analysis such as flow limit, models parameters etc.
Calculated data could not be changed.
Report is subdivided into reports heading and
report itself.
One could see in the report certain data as
grey lines. One grey line exists for each
evaluation. Results are below the grey line.
1 Heading: Reports heading according to
Settings.
See for that also Section Table Table
Settings Heading.
2 Index: Current evaluations index. It serves
exceptionally for orientation in case of too long reports.
3 Symbols: Certain type of analysis results evaluation is displayed in the grey line. Results parameters are
displayed including. There are a different number of results parameters depending upon the type of
evaluation. Indication will appear, if reasonable results evaluation is impossible.
42
4 Parameters / Results: Certain parameter to which the report relates is displayed in grey line. Results will
be shown as a list according to symbols that are in previous column.
5 Test / Units of measure: Certain text of evaluation is displayed in grey line. Units of measure including
provided that measuring results have units of measure.
6 Comment: Certain comment for evaluation is displayed in grey line. Margins under that could be described
with additional information to results. Then they appear in printing.
Attention: Saving of this information does not take place. This information is lost again in case of selection of
another sample.
Reports length depends on number of certain evaluations. Selection of displayed evaluation takes place in
Report Settings.
8.8.2 Main Menu Report
8.8.2.1 Report Settings -> Show
Report settings of the selected job are shown. New setup window will be opened.
See settings details in Section Report Settings.
This function acts in main menu as with the help of pressing Report -> Settings, as well as by pressing
submenu Show.
One could not control the main window of RHEOTEST RN 4 Manager until setup window is active.
8.8.2.5 Settings
They serve for setting of report content. This setup acts for the actual job and at the same time for all samples
in this job. The settings are divided in three tab pages: Report, Heading, Print. All changes made will be
rejected with the button Cancel and Menu Settings is closed. The changes will be accepted and they become
actual ones with the help of OK button. Menu Settings is closed.
8.8.2.5.1 Report
It serves for setup of necessary results evaluations, for which tests they should be implemented and should
this results be presented at diagram as additional information.
43
44
8.8.2.5.3 Printing
1 Add diagram: diagram will be added after the report.
Note Diagram dimensions are determined by available
place at the page. The larger is the report the smaller
will be diagram. If report dimensions exceed the bounds
of the page, diagram printing will not be done.
2 Add table: Actual table will be added after report or
diagram (optionally).
3 Add settings: Report settings will be added after
report, diagram (optionally) or table.
All three options could be selected in arbitrary
combination. See for that Print report.
8.9 Method
Method will be shown, if inlay Method is
active. It includes definition, settings and
certain single tests of a method.
The inlay serves for displaying method of
the actual selected job and active method of
the method library.
One method contains at least one test. One
test describes conditions of sample
measurements, as for instance number of
revolutions, time and temperature.
1 Method tree.
2 Specification of displayed method. Inlay
after it contains further settings for the
method.
3 Single tests are presented in form of a
table.
4 Single tests are presented at a diagram.
5 Tests displacement in a table or at a diagram.
8.9.1 Definition
Method definition contains information concerning description of analysis carried out.
45
1 Name: Method name should describe the method theme. This facilitates subsequent search.
In principle it is possible to give the same names for a lot of methods. However it could lead to confusion
quickly and it is not recommended to do so.
2 Group: Group displays methods group in which displayed methods which are saved or has been saved. If
display Method from the library is active, group will be shown where this method was saved.
If display Setup Method is active, group will be shown from which this method was copied into the job.
Therefore it could occur that this group is no longer in the method library. It might be for example renamed,
deleted or is inactive. The job could be imported also and thus the method in it.
3 Comment: Always it ought to write here useful and factful instructions to analysis. It could be information
about the task of analysis such as: input control of supplier A, product B development or monitoring of
products line XY. It is possible to set name according to rheological sense, such as measurements with
controlled shear rate, temperature dependence, determination of flow limit.
During method development it is useful also to comment development steps in this area, i.e. to fix changes in
comparison with previous version. It gives possibility for other Users for quick method classification for their
tasks. Besides this information could be used for documentation in table heading, diagram heading or in report
heading.
4 Method created: Date and time of method creation or when the method was imported.
5 Method changed: Date and time when method was changed last time. It relates to all method settings.
6 New test: New test will be inserted in the last position of the tests table. At that settings of active test at the
moment will be applied. Thereupon these settings could be corrected.
7 Copy test: Actual test is transferred to the clipboard. Windows clipboard is not used.
8 Insert test: Test from the clipboard will be inserted after the actual test.
9 Delete test: Selected test will be delayed.
Note: Request arises about confirmation of deletion, if this is defined in User settings. Cancel function is not
foreseeing.
10 Test details: it shows actual test settings in details. See for that Test Settings. One cannot use inlay
methods while test settings window is active.
11 Display mode: Inlay Method serves for imaging of the actual job method and actual method from the
methods library.
This display serves for distinction of imaging modes. Besides, this mode is displayed in status line of
RHEOTEST RN 4 Manager.
Display mode at
inlay Method
Jobs method
Tree display
Jobs display
Job x from y
Methods display
Method x from y
Display mode switching: The jobs display is active after program start. If now one switches to inlay Method,
jobs method will be displayed.It is necessary to select choice Method in RHEOTEST RN 4 Manager main
menu in order to switch imaging mode. At that switching takes place to the inlay Method and method display
is activated. This will be active until one of the choices or one of the inlays Table, Diagram or Report will not
be selected in the main menu. After this display mode is switched again to the jobs display.
Note: It is necessary to press and hold the left button of mouse in the main menu, if you are in jobs display
and want to see parameters of the jobs method in details. With pressed mouse button one could now move to
the menu choice Parameters -> Show
Now parameters will be displayed of jobs method.
46
8.9.2 Settings
Further methods settings.
1 Special function: one could see this choice only during method look through and when registered User
has rights of administrator.
Special function is used for start of certain evaluations after analysis fulfilment.
There are following possibilities for choice:
This setting is also used for check-up of possibility of the special evaluation in case of calibration selection in
the main menu. Here could be used the samples for Pt100 calibration or inertia calibration only that were
measured with the correspondingly adjusted method.
2 Printing parameters: These settings determine printing, if inlay Method is displayed and if one selected
File -> Page view or File -> Printing. One could find details of these settings here Print Method.
8.9.3 Test-Table
It is certain tests presentation in form of a table according to the order of their fulfilment.
Maximum 65 tests could be defined.
Main data of tests will be displayed. Further settings could be done in Test Settings.
1 Tests table: The most important parameters of a single test are displayed. Values that are obligatory
necessary for test are shown as white areas for input. Values that are not necessary for the corresponding
test are marked in grey. Showed here values do not have influence on the test.
The necessary values are determined by test type selection.
Data could be marked by one click and changed afterwards.
Note: However it is recommended to carry out changes in input mask Test settings.
47
It will not be shown unit of measure and it is not necessary to input it in the changes mode.
Column width is optimal, fixed and could not be changed. So the width of test table is fixed also.
One could find in Section Test settings values of columns together with set values.
2 Tests displacement in table or diagram: Marked test is moved in the limits of the table one line upwards or
downwards. In such a way it could be corrected the tests sequence.
8.9.4 Test-Diagram
Graphic display of certain tests.
1 It displays time variation of the
set values.
Here it is used start or end values
depending upon test type. Initial
conditions
could
not
be
determined previously for some
types of tests.
It becomes apparent in that case
when, for example a transfer
takes place from one CS-test to
another CD-test in form of a ramp.
Here the last condition in CS-test
is decisive where the start value is
of CD-ramp. However it could
differ from reality.
The following 4 physical values are displayed:
Blue: Shear stress and/or torque, it is valid also for oscillation measurements.
Red: Shear rate and/or number of revolutions.
Green: Deformation.
Orange: Frequency by oscillation measurements.
2 Estimated time: this is time necessary for test carrying out. Set tests time will be used.
Development of a process in time in some cases is only an approximation since in some tests, for example in
equilibrium experiments when time is not fixed.
3 Set values range: Physical values used in tests will be presented. Scaling is defined by the highest
encountered value and could not be changed.
The same scale is used during simultaneous display, for example of shear stress and number of revolutions.
Recalculation does not take place of shear stress in number of revolutions.
4 Displaying of single tests areas. Each test is limited by vertical lines at time axis. If there is enough room
for image, so it will be indicated test type.
The actual test, i.e. marked test in a table will be designated with some other colour and arrow before test type
name.
In such a way the marked test is moved then during Tests Displacement. Here displacement takes place to
the left (upwards key) or to the right (downwards key).
48
7 Measuring points: It is number of measuring points at test types without definite behaviour in time, for
instance oscillation tests or equilibrium measurements.
12,00
In case of oscillation tests this value gives number of oscillations
that will be done. Then result of measurements contains also time
linear
10,00
geometrisch
information about oscillations duration and duration of complete
geomtretic
8,00
test.
In case of equilibrium measurements this value determines
6,00
number of measuring points.
4,00
2,00
8 Geometric division: At sweep-tests ramp is not divided into
linear portions but into geometric segments. As an example at
0,00
0
2
4
6
8
10
12
figure ramp has been divided into 9 segments from 1 to 10.
X
Linear division is shown in blue colour, all portions are equal.
Geometric division shows unequal portions on Y axis. Each
individual value is a x-fold of the previous value. This setting is recommended for sweep tests in very wide
ranges.
9 Temperature control: It sets with which temperature control test should be carried out.
The possibilities are as follows:
None: Temperature control is unnecessary for this test. Measured temperature is displayed along with
measuring result.
Const: Temperature control is activated, set temperature will be specified as a new required value.
After that one waits achievement of the required value and then test is executed. This temperature is
maintained constant during the whole test.
Jump: Temperature control is activated, set temperature will be specified as a new required value.
Test starts at the same time. Temperature approaches to the set value as soon as possible and for
this reason is not constant during the test. Achievement of set value is independent of the carried test.
Ramp: During test temperature will be reduced to the set target value. Temperature control is
activated. It will be preset the earlier actual target value. It could be target temperature value from the
previous test. Waiting takes place until this value stabilized. Only after that a new target temperature
value is set and test starts up. Now temperature changes so rapidly in order to achieve a new target
temperature exactly at the end of the test. Temperature ramp and test run synchronous during test
period.
49
All temperature ramps depend upon temperature control techniques. So it is technically impossible for
example to realize a set temperature ramp from 20C to 100C in 10 s with the help of liquid thermostat.
Especially it has to do with temperatures below zero (cooling).
At the beginning of test it is checked if temperature control is necessary and is it possible to reach set
temperature values with the help of the specified temperature controller. Error message appears, if it is
impossible.
10 Temperature: Temperature value will be preset which is necessary to set at constant temperature or
temperature value that should be reached at temperature ramps or temperature jumps.
11 Interruption criterion: it defines when the test should be interrupted. The following settings are possible:
None: Interruption criterion has not been defined.
Angle: Test will be interrupted at achievement of the preset angle. The angle value is set in methods
parameters.
This criterion is valid for CD-tests only.
Viscosity: Test will be interrupted at achievement of the preset viscosity. The viscosity value is set in
methods parameters.
Speed: Test will be interrupted at achievement of the preset number of revolutions. The speed value
is set in methods parameters. This criterion is valid for CR- and CS-tests only.
12 Use end value of the previous test as initial value. If this option is active, initial value of this test will be
taken automatically from the previous test. Initial value could be also a constant value in test with constants.
You obtain automatic combination of the transmitted values in such a way that during changing of the previous
test this value will be changed also.
This option is possible from the second test only, if the previous test has the end value and tests types match
to each other, for instance both tests are CR tests.
13 Backwards: Transfer to the previous test in the limits of the tests row, if it is possible.
14 Further: Transfer to the next test in the limits of the tests row, if it is possible.
15 Display test: It will be shown which test from the test row is active at the moment.
16 Cancel: All changes will be rejected and input display is closed.
17 OK: All changes will be accepted and input display is closed.
Any recordings could not be done in the inlay Methods until window for test settings is active.
8.9.6 Parameter-Settings
Method parameters
1 Triggertime: This time in s causes delayed data recording after the start of the tests. This can be used to
hide transients or transient phenomena.
2 Time Raster 1: This time in s determines the
sampling of data during the test. The smallest
theoretically possible value is 0.004 s (4 ms).
Formally, only multiples of this value are
possible. Different inputs are adapted to the
required value. To ensure a secure and
continuous data transfer, we recommend a
minimum value of 0.04 s (40 ms).
50
3 Time Raster 2: corresponds to that of Time Raster 1. A second time intervals are specified to have two
different sampling rates available within a method here.
4 Viscosity Raster: This percentage value determines the next data transfer, provided that "viscosity" was
specified as the grid type.
5 Temperature Raster: This percentage value determines the next data transfer, provided that "temperature"
was specified as the grid type. This value is valid only for temperature jumps or temperature ramps.
6 Averaging: If this option is active, floating mean value will be formed in the control unit of RN. This option
allows obtain result with small noise.
7 Angle Abort Limit: This value in gives
possibility to interrupt test at achievement of
the definite angle. It is necessary to select
interruption criterion Angle in test settings.
8 Viscosity Abort Limit: This value in Pas
gives possibility to interrupt test at
achievement of the definite viscosity. It is
necessary to select interruption criterion
Viscosity in test settings.
9 Speed Abort Limit: This value in 1/s gives
possibility to interrupt test at achievement of
the definite speed. It is necessary to select
interruption criterion Speed in test settings.
10 Cancel: All changes will be rejected and input display is closed.
11 OK: All changes will be accepted and input display is closed.
Interruption conditions could be used for protection of material to be measured, the instrument and User from
damages or dirt.
So it would be possible at ramp of a torque obtain such high speed that the sample under study will be thrown
out of the instrument. This could be prevented with the help of function Interruption speed.
8.9.7 Unit settings
Parameters units are set with the help of units of measure settings. It relates to the method definition as well
as to displaying of measuring results in table, diagram or report.
Each settings change has a direct influence on the presented numeric values, if these changes are confirmed
with OK key.
No.
Parameter
Possible Units
1
Torque
mNm, Nm
2
Speed
1/s, 1/min
3
Time
ms, s, min
4
Temperature
C, K, F
5
Shear rate
1/s, 1/min
6
Shear stress
Pa, mPa, dyn/cm
7
Viscosity
Pa s, mPa s
8
Deformation
arb.unit*1, %
9
Angle
, rad
10
Frequency
Hz, 1/s
11
Angular frequency
Hz, 1/s
12
Modulus
Pa, kPa, MPa
13
Elasticity
1/Pa, 1/kPa, 1/MPa
*1
arb.unit represents dimensionless value relative to the initial value
51
52
Units
Torque
Speed
Time
Temperature
Shear rate
Shear stress
Viscosity
Deformation
Angle
Frequency
Angular frequency
Modulus
Elasticity
mNm
1/min
s
C
1/s
Pa
Pa s
%
rad
1/s
1/s
Pa
1/Pa
8.10 Settings
Measuring systems could be edited in this choice of the main menu, as well as specific settings. Besides,
calibrations could be fulfilled in case of corresponding rights.
8.10.1 Measuring Systems
RHEOTESTRN 4 could be used with different measuring systems. In selection one should follow samples
under study and analysis purpose. Various measuring systems are necessary in order to cover a wide range
of measurements for very different tasks.
For RHEOTEST RN 4 Manager are necessary certain characteristics that describes the measuring system in
order to calculate rheological parameters.
Control of these data is fulfilled in measuring systems library. One could select from available measuring
systems the measuring system in use at the moment. One does this before analysis in actually used job (see
also Job Measuring System).
For specific special jobs one could determine and add new measuring systems also.
55
Type:
Constant Tau:
Constant D:
Inertia moment:
Rotation correction factor:
Oscillation correction factor:
Rotor diameter:
Rotor length:
Measuring vessel diameter:
Cylinder S
1.0
1.0
0
1.0
1.0
0
0
1.0
57
Calculation rules correspond to the regulations in force and depend upon measuring system inserted. Also
see for that Applications Measuring Systems.
8.10.1.2.2 Calculation -> Constant D
It will be calculated D constant on the basis of the geometric data of measuring system. It is possible for User
with administrator rights only.
Calculation rules correspond to the regulations in force and depend upon measuring system inserted. Also
see for that Applications Measuring Systems.
8.10.2 General Settings
In order to achieve optimal use of RHEOTEST RN 4 Manager some view options could be adjusted to
individual requirements.
These settings are valid for all users of the software and hence have global character.
1 Inlays: Jobs, Methods, Print and System. Content
is described in details below.
Inlay Job
2 It is a list of actually available jobs groups. This list
includes all jobs groups that are in the system
catalogue of RN-Manager at present.
In case of General Settings invocation checking is
fulfilled which jobs groups are available.
If new jobs groups will be found, message appears
with indication and they will be added to this list.
If one jobs group from the list will not be found in RHEOTEST RN 4 Manager system catalogue, it will be
informed and this jobs group will be deleted from the list.
This jobs group is displayed in jobs tree of the main display with the help of a mark before jobs group name.
Due to this possibility unnecessary at present jobs groups could be hidden in jobs tree. It was done for better
image clarity.
3 New group: New empty jobs group will be created. It contains preset name from input area under this
button. If the name was not set, message appears with indication.
4 Input area for jobs group: Here one could set name of jobs groups which afterwards is used to create a
new jobs group or to rename available jobs group.
5 Rename group: Existing jobs group will be renamed. For that it is necessary to input a new name in the
input area (3). Then it is necessary to select jobs group from the name list (6) which it is necessary to rename.
Now name of jobs group could be changed with the help of Rename Group button.
6 List of groups names: it is a list of available jobs groups. Selection in this list is necessary for functions
Rename Group and Delete Group.
7 Delete group: Selected jobs group will be deleted in the list of groups names. It takes place in that case
only, if this jobs group is empty, i.e. there are no more jobs in this jobs group. It will be informed, if selected
jobs group is not empty.
However if this jobs group should be deleted, at first it is necessary to escape from this setting. After that it is
necessary to select in the main RHEOTEST RN 4 Manager view window a job in the group that has to be
deleted. It is necessary to fulfil Job -> Delete Job -> Delete all in Group through the main menu.
Now jobs group is empty and could be deleted completely in general settings. For that in inlay Job one
selects group name in names list (6) and then the jobs group is deleted with the help of Delete Group.
8 Cancel: All changes will be rejected and input display is closed.
9 OK: All changes will be accepted and input display is closed.
58
Note: By creation, renaming or deletion of the jobs group in every case only one function could be fulfilled for
the jobs group. It means for example that new created jobs group could not be renamed or deleted at once.
For that it is necessary to escape from General settings and recall them again. In this case only it might be
fulfilled the second function.
It refers to the inlay Methods also.
Inlay Methods
59
Inlay Print
One defines heading form for printing.
It will be determined how should be represented heading
of each printout. This includes text information as well as
company logo.
1 User heading for printing. If this function is active, it will
be output for each printout heading in accordance with the
following settings.
2 Free text 1: Main text in heading. It is output with large
type in the first line of the printout. It might be for example name of an enterprise.
3 Free text 2: as addition to the heading. It will be considerably smaller than the main text and is situated in
the second line of the printout. This might be for example a department.
Note to 2 and 3: Texts will be aligned to the left edge and be in a frame. If text should be wider than the frame
or page, so the text will be shown as it is possible.
It is recommended to check texts image through Page View and if necessary correct text in accordance with
the page parameters.
4 Logo name: here it could be determined name of a graphic file (*.bmp, *.jpg) that represents logo. It might
be logo of enterprise or branch.
If possible graphics should have size 100 pixels x 100 pixels. Other dimensions are used but it could result in
images distortion.
If file was set that was not found, so the logo would not be presented. Message with indication does not
appear.
Logo appears in heading at the right side.
5 Show heading at all pages: If this function is active, heading will be presented at every printed page. If it is
not active, so it will be at the first page only.
Inlay System
It serves to set or display of system significant settings.
1 Identification: These are codified numbers to check
compatibility of RN Manager and the instrument.
Identification numbers are set in RHEOTEST and could
be changed by administrator only.
It is necessary to inform Rheotest about this number in
order to identify instrument and software correctly.
2 Language: Settings of software language. After new
language has been installed it is necessary to reset the
software to activate these settings.
3 Pt100 calibration: It will be displayed saved in the system data about Pt100 calibration. They will be
updated if fulfil Service-Job xPt100-Calibration in the jobs group xJobs-for-Service.
Reference temperature will be shown at which calibration was carried out. Correction factor is a calibration
itself. Measured temperatures will be corrected with this value.
4 Time correction: This value is used for correction of measuring data by oscillation measurements. The
value is inserted in Rheotest and could be changed with the rights of administrator only. One wrong change
could lead to distortion of measuring data.
60
61
Inlay Admin
It is possible to control registered users in inlay Admin.
1 Users list: All registered users are listed in the list.
One might start RHEOTEST RN 4 Manager with one
of these names only.
2 New: New user is created. For that it will be used
name for login from the input field (4). If no name was
indicated, message appears about this.
3 Delete: Actually selected user will be deleted.
Afterwards he could not launch RHEOTEST RN 4
Manager. User with higher status could delete other
user only. Also it is impossible to delete oneself by
oneself.
4 Input field Name for login: Here one sets name for login of a new user. New user will be registered, if one
presses New (2).
5 Name: Full name of user. It appears then as operator name during sample creation. Operator name could
be output later in headings of table, diagram or report.
6 Department: Input of user department. It serves for user identification exclusively. This information does
not use later.
7 Phone: input of phone number of user. It serves for user identification exclusively. This information does
not use later.
8 User level or status also: These settings serve for determination of software use. It is reasonable to permit
to some users not to do all changes. It could assure prevention of errors, accidental data deletion, error
measurements and in such a way incorrect analysis.
It is possible 4 users levels, see for that the following Section Operation Status.
Operation Status
Designation User level
1.
Operator 0: Measurements (jobs) could be started and results could be printed.
2.
Engineer 1: It could be created and changed methods and jobs including formation of displaying and
output of measured values and report settings.
3.
Admin 2: It could be done settings connected with the system, for example measuring constants,
parameters, constants of measuring systems and control of user. These settings require detail
knowledge as in such a way measurements precision and rheometer functions are determined.
Incorrect input data could lead to erroneous measurements results or to instrument damage also.
4.
Super-Admin 3: here it could be set special parameters which determine instrument functioning.
These settings are fulfilled by Rheotest before delivery and then they are saved by the service of
Rheotest.
User status change is possible only in such volume that he could change other user status that is smaller than
his own status.
This new user will be added to the users list with the help of pressing New. Now additional information could
be input such as full name, department, phone number and user status.
It is possible to do settings of specific user in inlay User.
However password could not be entered. Password input is always related to the password itself.
Afterwards as all settings were done it is possible to exit from the input form with the help of OK. All settings
will be saved.
Now new user has to login with his registration name.
As far as a password was not created it is impossible to enter any password.
After program was launched it is necessary to define password in input form Settings -> User Settings.
The user opens the input form enters twice his password in inlay User and confirms it with the help of
Set/Change.
After exit from the input window using OK this password is active. It is necessary to enter login and password
at the next registration during program launching.
Note: Only registered user could change own password also.
One could refuse from password entering, if only one person works with the program or it is unnecessary to
use different users rights.
Simply password is not determined for new users. Password could not be deleted directly for existing users.
Here it is necessary to give possibility to delete user and then create a new one.
8.10.4 Communication Settings
Prerequisite for functioned working place is correct communication between computer, RHEOTESTRN 4
and connected temperature controller.
Communication is carried out through interface COM. Connection of RN-control unit and temperature
controller is fulfilled with RS232 cable or with RS232 cable and transducer USB-Serial.
One could do necessary settings for this in 3 inlays.
1 Inlays:
RN
Plate control
Temperature controller
Temperature
RN
Used data protocol between PC and RN control unit was developed specially for this task. It is impossible to
set communication with the help of other software. Settings of used interfaces are saved constantly in the
program. It must be determined only COM-port number.
2 COM port: Number of used COM-port. If one uses USBadapter here, it is necessary to input number of virtual COMport of USB transducer.
3 Data bit: Number of data bits for one sign 8
4 Baud rate: 57600
5 Stop bit: Number of stop bits in data protocol 1
6 Parity: Bits state as parity characteristic 0
7 Transmission: Transmitted signs sequence from RHEOTEST RN 4 Manager to RN-control unit.
63
Startup RN-interface
The following sequence exists in order to setup and commissioning of interface:
1. Press Check port: all available COM-ports are entered into selection list.
2. Read in PC to which COM-port is connected the RN-control unit. It is not so easy in case of use of
USB cables. Here one should see and check equipment settings in PC system.
3. Select from the list COM-interface.
4. Press Init. Port: Interface is initialized with fixed parameters.
5. Press Transmission INIT to check communication: now definite command is sending to RNcontrol unit. Sent text one could read on display Transmission.
6. If communication was set correctly, it appears text with answer on display Receiving.
7. If error arises, it means that communication was set incorrectly.
At launching the program tries to use the last communication settings. So it is not necessary to carry out
commissioning again in case of changeless complete set of working place for measurements.
Temperature Control
Used data protocol is adjusted according to producer of temperature controller. There are some commands or
data structures that are the same for all temperature controllers but every controller type has its own special
commands. It is related with technical progress and different functionality.
Only definite commands are used in the software in order to correspond to this circumstance. They meet
requirements of rheological studies. Moreover there is freedom of actions due to this and it is possible to use
practically every temperature controller with this software.
Settings for link with temperature controller are made in inlay Temperature Controller.
1 COM-port: Number of used COM-port. If one uses USBadapter here, it is necessary to input number of virtual
COM-port of USB transducer.
2 Controller type: Selection of available thermocontroller. It
will be shown thermocontrollers that are recorded in file
ThermoControl.set.
3 Data bit: Number of data bits for one sign 8
4 Baud rate: 9600. This value is in Operating Manual of the
thermocontroller.
64
5 Stop bit: Number of stop bits in data protocol 1. This value is in Operating Manual of the
thermocontroller.
6 Parity: Bits state as parity characteristic 0. This value is in Operating Manual of the thermocontroller.
7 Transmission: Here signs are represented that should be transmitted. These symbols could not be entered
directly but they are determined with the software. This display is necessary for service or for experienced
user in case of troubleshooting.
8 Receive: Here received signs are presented. This signs are sent from RN-control unit to RHEOTEST RN
4 Manager and are answer to command sent. This display is necessary for service or for experienced user
in case of troubleshooting.
9 Check port: All available COM-port will be detected that could be used for communication.
10 Initialize port: Actual COM-port is initialized with the standard settings. It means that setup data are
transmitted to interface.
11 Types loading: Available thermocontrollers will be read from file ThermoControl.set and put into the list
(2).
Parameters and commands for thermocontrollers are indicated in file ThermoControl.set.
File is in the same catalogue where the software is. If this file does not exist, the program could not be started.
All available in the file thermocontrollers are loaded at program launching. If the file contains wrong data, the
loading will be stopped and there will not be available thermocontroller.
Startup Thermocontroller
The following sequence exists in order to setup and commissioning of thermocontroller:
1. Press Types loading: file ThermoControl.set is loading. If file contain errors, it will be shown. If
load took place without faults, then message appears Available.
2. Select connected thermocontroller from the list. Now parameters and commands are transferred
to display.
3. Press Check port: all available COM-interfaces are put in the list for selection.
4. Read in PC to which COM-port is connected the thermocontroller. It is not so easy in case of use
of USB cables. Here one should see and check equipment settings in PC system.
5. Select from the list COM-interface for thermocontroller. COM-interface that is used by RN-control
unit could not be selected.
6. Press Init. Port: Interface is initialized with fixed parameters.
7. Communication check is carried out in inlay Temperature.
Commands and replies are shown on displays Transmission and Receive, if communication took place.
They are necessary for service in troubleshooting.
At launching the program tries to use the last communication settings. So it is not necessary to carry out
commissioning again in case of changeless complete set of working place for measurements.
65
Temperature
It is possible remote control of thermocontroller in inlay Temperature, if communication was set correctly.
Commands are divided in two groups: fixed temperature and temperature ramps.
The following commands are for fixed temperatures:
1 Enquiry: Actual temperature will be displayed. If it
concerns the internal controller, temperature or the
external temperature is determined by Init-commands or
through a command. Actual temperature appears on
display Act.Temp.[C].
2 New: setup of new target temperature. Input of new
value in input field Target Temp.[C]. Afterwards this
value will be transmitted to thermocontroller as new set
value.
3 Enquiry: It is enquiry of actual set target temperature. Enquired value will be shown in input field Target
Temp.[C].
4 - START: Temperature control starts, i.e. the set temperature is reached by cooling or heating. At that it is
necessary to pay attention to the fact that at command START very high, dangerous temperatures could arise
in the measuring system.
5 STOP: Temperature control is off. Heating or cooling takes place no more.
Note It could be still very high temperatures in the measuring system under certain circumstances!
The following commands are for temperature ramp:
6 Set Ramp: It is transferred information necessary for temperature ramp. At that value on display Target
Temp.[C] is transferred as a target temperature. The value on display Time[min] is the ramp time. Then
ramp slope is calculated in thermocontroller from these two values.
7 Time [min]: It is field for input of temperature ramp duration. Setup of this value is used for temperature
ramp settings (6).
8 - Status: Thermocontroller status is read out. Reply is different for each controller type and must be taken
from the Operating Manual of the thermocontroller.
9 - START: Temperature ramp starts, i.e. new temperature will be reached in preset time from the actually set
value. At that it is necessary to pay attention to the fact that at command START very high, dangerous
temperatures could arise in the measuring system.
10 - STOP: Temperature control is off. Heating or cooling takes place no more.
Note It could be still very high temperatures in the measuring system under certain circumstances!
11 Init.1: Optionally commands could be sent to the thermocontroller with the help of this. These commands
could fulfil certain settings in the thermocontroller. This command is fulfilled one time at program launching.
The commands could be the following: readings precision, internal / external temperature, pump ON/OFF. The
commands must be stored in setup file of the thermocontroller.
12 Init.2: It is such as Init.1 for further commands.
Note: Enquiry is a simple command to check communication. Act.Temp.[C] (actual thermocontroller
temperature) appears on display as a reply on this command.
At launching the program tries to use the last communication settings. So it is not necessary to carry out
checking again in case of changeless complete set of working place for measurements.
All settings will be accepted by pressing OK. If one pushes Cancel, all changes will be rejected.
66
67
Job name
. . .
Measuring
system
. . .
Method of
analysis
. . .
Date of
change
. . .
Number of
samples
. . .
Date of creation
Date of changing
Date of the last use
Setting
Settings from the inlay Settings
Suppress wait time
Show sample collection
Active job
Measuring system
Samples list
Method
Table settings
Report settings
68
Measuring system
Here are data of measuring system that is actually indicated in the job. It must not be obligatory the measuring
system that was used for previous measurements.
Measuring system name
Commentary
Constant Tau
Inertia moment
Correction factor
The following geometric data are based on the type of displayed measuring system. Here is example
for cylinder H
Rotor diameter
Measuring vessel diameter
Rotor length
Samples list
Here is list of samples that are in the job.
Samples group
Charge
. . .
. . .
Date
creation
. . .
of
Editor name
Tests number
. . .
. . .
Charge
Editor name
Tests number
Inertia moment
Constant D
Test type
. . .
. . .
Number
points
. . .
of
data
Speed
Temperature
Shear stress
Angle
Angular frequency
Elasticity
Method
Here is method that is stored in the active job at the moment. It must not be the method used for previous
measurements. The following is listed:
Test type
Beginning
End
Time
Points
Geometric
Fixed
parameter
Temperature
type
Temperature
Raster type
Interruption
type
Use data of
previous test
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
Date of creation
Date of change
. . .
Method name
Methods group
Commentary
List of tests in the method
Method parameters
Here is list of parameters for RN control.
Method parameters
Trigger time
Time raster 1
Time raster 2
Viscosity raster
Temperature raster
Mean value
Limit angle
Limit viscosity
Limit speed
Reference temperature
Extended parameters
-DynamicsParameter 0
Parameter 1
Parameter 2
Parameter 30
-RheologySampling filter
Tolerance
Mean number
Interruption time
Table settings
List of all table settings
Number of columns
Parameter 1
Parameter 2
Parameter x
. . .
Test11
Test12
Test13
Test14
Test15
Test16
Test17
Test18
Test19
Test20
Data 13
Data 23
Data 33
Show
designation
13
Show
designation
23
Show
designation
33
Print
Add setting
Diagram settings
List of all diagram settings
Axis X
Type
Scaling
Min.
value
Max.
value
log
Axis X sorting
Axes Y
Left
Type
Scaling
Right
Type
Scaling
2.left
Type
Scaling
2.right
Type
Scaling
Grid Settings
X-grid
Y-grid
Grid colour
Min.
value
Min.
value
Min.
value
Min.
value
Max.
value
Max.
value
Max.
value
Max.
value
log
Colour
Width
Marking
Size
log
Colour
Width
Marking
Size
log
Colour
Width
Marking
Size
log
Colour
Width
Marking
Size
X-small grid
Y
Displayed tests
Test1
Test6
Test2
Test7
Test11
Test12
Test16
Test17
70
Test3
Test4
Test5
Test8
Test9
Test10
Test13
Test14
Test15
Test18
Test19
Test20
Line 1
Column 3
Line 2
Column 3
Line 3
Column 3
Show
designation
13
Show
designation
23
Show
designation
33
Print
Add report
Add table
Add setting
Parameter 1
Test 1
Parameter2 1
. . .
. . .
. . .
. . .
Show at
diagram 1
. . .
Commentary 1
. . .
Line 1
Column 3
Line 2
Column 3
Line 3
Column 3
Show
designation
13
Show
designation
23
Show
designation
33
Print
Add diagram
Add table
Add setting
Print Table
For that it is necessary to select pointer Table.
Table of the active job is displayed always. At that tables actual settings are taking into account such as:
Displayed parameters (number of columns)
Selected tests
Table heading settings
Columns width
See for that menu Table settings.
It is recommended to use paper format A4 across for better presentation depending upon number of
parameters and number of columns set. It is necessary to use preview function in order to obtain optimal
printing result. At that it is possible to see necessary corrections of columns width or page format.
71
Parameter 2
Parameter x
. . .
Test11
Test12
Test13
Test14
Test15
Test16
Test17
Test18
Test19
Test20
Data 13
Data 23
Data 33
Show
designation
13
Show
designation
23
Show
designation
33
Print
Add settings
Print Diagram
For that it is necessary to select pointer Diagram.
Diagram of the active job is displayed always. At that diagrams actual settings are taking into account such
as:
Displayed parameters (number of Y axes)
Selected tests
Diagram heading settings
Sample collection settings
Report parameters settings
See for that menu Diagram settings.
Printed diagram size is adjusted to the page format used. It will be printed maximal possible format, without
changing proportions of the diagram page. It means that page width is decisive for A4-portrait. In contrast with
this page height is dominant for A4-album. One could setup pages proportions with the help of window
change for diagram preview.
It is necessary to use page preview function in order to obtain optimal printing result. At that it is possible to
see necessary corrections of page format.
The following printing options are possible:
Report
Report is displayed directly under the diagram. So it is possible visual presentation of results on paper.
Here the report is output without its heading. Settings correspond to ones stated in the Section Report.
Table
Table is displayed directly under the diagram or under the report, if it was selected. So it is possible visual
presentation of all results on one paper sheet.
Here the table is output without its heading. Settings correspond to ones stated in the Section Table.
72
Diagram settings
List of all diagram settings
Axis X
Type
Scaling
Min.
value
Max.
value
log
Axis X sorting
Axes Y
Left
Type
Scaling
Right
Type
Scaling
2.left
Type
Scaling
2.right
Type
Scaling
Grid Settings
X-grid
Y-grid
Grid colour
Min.
value
Min.
value
Min.
value
Min.
value
Max.
value
Max.
value
Max.
value
Max.
value
log
Colour
Width
Marking
Size
log
Colour
Width
Marking
Size
log
Colour
Width
Marking
Size
log
Colour
Width
Marking
Size
X-small grid
Y
Displayed tests
Test1
Test6
Test2
Test7
Test3
Test8
Test4
Test9
Test5
Test10
Test11
Test12
Test13
Test14
Test15
Test16
Test17
Test18
Test19
Test20
Line 1
Column 3
Line 2
Column 3
Line 3
Column 3
Show
designation
13
Show
designation
23
Show
designation
33
Print
Add report
Add table
Add setting
Table
Table is displayed directly under the report or under the diagram, if it was selected and there is enough room.
So it is possible visual presentation of all results on one paper sheet.
The table is output without its heading. Settings correspond to ones stated in the Section Table.
Print Method
For that it is necessary to select pointer Method.
One could select print active method or list of available methods at the moment.
Methods list
List of all methods is displayed. The methods are taken into account that one could actually select in the
selection tree. Methods from the methods groups that are not shown will not be taken into account.
It is recommended to use paper format A4 across for better presentation.
List content:
Methods group
Method name
. . .
. . .
Date
creation
. . .
of
Date
changing
. . .
of
Tests number
. . .
Date of creation
Date of changing
Settings
Output of settings from the pointer Method:
Active method
Definition
Test diagram
Methods parameters
Method
Methods list
Settings
Tests list
Units
Methods parameters
Test diagram
Tests diagram is output. Diagram size is fixed well-defined.
Tests list
List is output with defined tests.
Test type
Beginning
End
Time
Points
Geometric
Fixed
parameter
Temperature
type
Temperature
Raster type
Interruption
type
Use data of
previous test
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
. . .
Mean value
Interruption criterion Angle
Interruption criterion Viscosity
Interruption criterion Speed
Reference temperature
-RheologySampling filter
Tolerance
Mean number
Interruption time
Units of measure
Here is list of actually used units in the method for:
Torque
Time
Shear rate
Viscosity
Deformation
Frequency
Modulus
Speed
Temperature
Shear stress
Angle
Angular frequency
Elasticity
Logo size is fixed and is 50 x 50 pixels. If the picture in file is of other size, it will be recalculated to this size.
Possible files types are .bmp or .jpg. This file must be saved in the same place as file RN Manager.
Show heading at all pages
This option permits to display at every sheet specific for User print heading.
The heading is output at the first sheet only, if this option is inactive.
8.11.3 End
End of RN-Manager operation. All settings are saved.
Note: Current analysis must be interrupted before end of RHEOTEST RN 4 Manager operation.
76
9. Appendix
9.1 Bitumenwizard
9.1.1 General Information
The Bitumen Wizard is a handling instruction to carry out the measurements to determine the deformation
behaviour of bitumen according to:
T
DIN EN 14770 Bitumen and bituminous binders Determination of complex shear modulus and phase
angle using a Dynamic Shear Rheometer (DSR) (German version EN 14770:2012);
Instruction sheet DSR-Test (T-Sweep) edition 2014 and additional descriptions of TL Bitumen-StB 07/13;
ASTM D7175-08 Standard Test Method for Determining the Rheological Properties of Asphalt Binder
Using a Dynamic Shear Rheometer;
ASTM D7405 Standard Test Method for Multiple Stress Creep an Recovery (MSCR) of Asphalt Binder
Using a Dynamic Shear Rheometer;
AASHTO T 315-12 Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear
Rheometer (DSR) and AASHTO M 320-10 Performance-Graded Asphalt Binder;
AASHTO T 350-14 Multiple Stress Creep Recovery (MSCR) Test of Asphalt Binder Using a Dynamic
Shear Rheometer (DSR) and AASHTO M 332-14 Performance-Graded Asphalt Binder Using Multiple
Stress Creep Recovery (MSCR) Test.
The Bitumen Wizard is used for explaining a bitumen measurement including sample preparation, handling
instructions and test procedure with the Rheometer RHEOTEST RN 4.3
77
Enter the desired temperature (start temperature of following bitumen test) and click the button OK: the
screen will display:
Follow the instructions on the display; after clicking OK, the plate P3 starts rotating at 10 rpm;
78
After running of the first readings, click (to see in window Information, right: receive data:) on the tab
Chart;
Move the basic plate by rotating the micrometer screw counterclockwise carefully up;
You recognize the contact point by increasing the torque (about 0,1 ... 0,3 mNm);
Then, adjust the gap (including trimming distance);
Cancel/finish the Job by pressing the button Cancel Analysis under the tab Job.
Jobs like OSC_T_Sweep_StB50_70 (the Jobs are named after the type of bitumen, e.g. Job for road
pavement bitumen 50/70) for execution of temperature sweeps of different bitumen types according to the
instruction sheet DSR-Test (T-Sweep). They are used for determination of complex shear modulus and
phase angle with DSR at different temperatures under defined oscillating stress according to DIN EN
14770 and instruction sheet DSR-Test (T-Sweep). The temperatures are rising during measurement
stepwise from 30C to 90C in 10C step. The sample is oscillating stressed at a frequency of 1,59 Hz.
Jobs like MSCR_AL_FGSV are used for determination of deformation behavior of bitumen and
bituminous binders according to FGSV-instruction sheet for execution of MSCR-Test (DSR). A MSCRTest (Multiple Stress Creep Recovery Test) is the repeated short-term static shear stress and subsequent
recovery of the test sample to determine the percent recovery of the binder sample as well its notrecoverable content.
The MSCR-Test according to FGSV_AL is executed in three load levels as follows: with the shear stress
levels of 100 Pa; 1600 Pa and 3200 Pa.
Jobs available in the Job group AASHTO according to binding standards of ASTM / AASHTO:
Jobs OSC_RTFO, OSC_Binder, OSC_PAV_S, OSC_PAV_H (the Jobs are named after the type of
bitumen, e.g. RTFO=Rolling Thin-Film Oven Residue) for execution of temperature sweeps of different
bitumen types according to ASTM D7175; AASHTO T 315-12; AASHTO M 320-10 are used for
determination of complex shear modulus and phase angle with DSR at different temperatures under
defined oscillating stress. The start temperature depends on the Performance Grade (e.g. PG64: start
temperature 64C). The binder sample is oscillated at a frequency of 1,59 Hz, the temperature is
increased in 6C increments. After crossing of the termination criteria, the Job is automatically terminated.
Job MSCR_RTFO_DETERMINE is used to determine the deformation behavior independent from the
type of bitumen. The PASS requirements only for the most significant bitumen type (S, H, V or E) will be
shown in the Report.
Job MSCR_ RTFO_GENERAL is used to determine the deformation behavior independent from the type
of bitumen. All PASS/FAIL requirements for all types of bitumen (S, H, V or E) will be shown in the Report.
The MSCR-Test according to ASTM / AASHTO is executed in two stress levels as follows: with the shear
stress levels of 100 Pa and 3200 Pa.
79
9.1.3.4 Measurement
9.1.3.4.1 DSR-Test (T-Sweep) according to DIN EN 14770, AL DSR-Test (T-Sweep) and MSCR-Test
according to AL FGSL
-
Clean the measuring and basic plates with a suitable solvent and dry before the measurement;
Select an appropriate Job (e.g. OSC_T_Sweep_StB50_70 or MSCR_AL_FGSV) in the Job tree Bitumen
(column left);
For easy start select the Name Create Mode DateTime (YYMMDDHHMM) in the Sample Definition and
click the button Easy Start in Actions for Job;
or:
Choose in the Name Create Mode manually, enter consecutive New sample, then give the Sample
Name, press Check+Create Sample (waiting until the sample name appears in the Job tree) and click
the button Start Analysis;
The program reminds you of the setting of zero point and will display:
After clicking OK, the program reminds you of the settings of the gap and trimming distance and will
display:
After clicking OK, the program asks you to place and trim a sample and will display:
Put the sample on the basic plate and trim it using e.g. spatula or trimming tool. After clicking OK, the
program asks you to set the distance between the plates (e.g. 1,00 mm) and will display:
80
After clicking OK the measurement starts. The progress of the measurement can be seen in the
information window (bottom right).
9.1.3.4.2 Determination of complex shear modulus and phase angle according to ASTM D7175, AASHTO T
315-12, AASHTO M 320-10 and MSCR-Test according to ASTM 7405, AASHTO T 350-14, AASHTO M 332-14
Clean the measuring and basic plates with a suitable solvent and dry before the measurement;
Select an appropriate Job (e.g. OSC_BINDER or MSCR_RTFO_S) in the Job tree AASHTO (column left);
For easy start select the Name Create Mode DateTime (YYMMDDHHMM) in the Sample Definition and
click the button Easy Start in Actions for Job;
or:
Choose in the Name Create Mode manually, enter consecutive New sample, then give the Sample
Name, press Check+Create Sample (waiting until the sample name appears in the Job tree) and click
the button Start Analysis; the following notification will be displayed:
Enter the temperature corresponding to the Performancee Grade of the binder type (PG) and click OK;
the Peltier temperatured basic plate starts tempering to the predetermined temperature;
The program reminds you of the setting of zero point and will display:
After clicking OK, the program reminds you of the settings of the gap and trimming distance and will
display:
81
After clicking OK, the program asks you to place and trim a sample and will display:
Put the sample on the basic plate and trim it using e.g. spatula or trimming tool. After clicking OK, the
program asks you to set the distance between the plates (e.g. 1,00 mm) and will display:
After clicking OK the measurement starts. The progress of the measurement can be seen in the
information window (bottom right).
All Jobs start with appropriate tempering of the measured material. During this time (10 min) there is no
measurement. After 10 min the first measured data arrive. When the measured data of the second partial test
came, they can be also displayed in the Chart and in the Table.
The completion of the analysis will be displayed in the status line. Now, the measured values can be viewed
and / or evaluated in the Report.
If the termination criteria are exceeded (for OSC_Jobs), the Job is automatically terminated. The
evaluation/output of the measured results is stored in the Report according to the appropriate AASHTO
Standards.
After the measurement is completed, loosen the knurled screw (2), simultaneously lift the quick connecting
coupling and move the measuring drive upwards. Clean the measuring and basic plates.
82
83
9.2 Tests
Two test methods are available: rotary tests and oscillating tests.
In the rotation tests, the rotational direction is not changed. The default value is either torque or speed or the
angle. The other value is measured as the response of default value then.
In the oscillating test, the rotational direction is changed. Generally for an oscillation dataset, a sine wave is
executed and evaluated. In doing so, the torque is given as a sine wave and the angle is measured in
response.
test
CR const
CR sweep
CR equilibrium
CS const
CS sweep
CS equilibrium
CS OSC const
CS OSC fsweep
CS OSC asweep
CD OSC const
CD OSC fsweep
CD OSC asweep
CD const
CD sweep
Creep
Speed const
Speed sweep
Torque const
Torque sweep
description
constant shear rate
shear rate sweep
shear rate equilibrium
constant shear stress
shear stress sweep
shear stress equilibrium
oscillation at a constant frequency and
amplitude
oscillation at a frequency sweep and
constant amplitude
oscillation at an amplitude sweep and
constant frequency
oscillation at an constant frequency and
amplitude
oscillation at an frequency-ramp and
constant amplitude
oscillation at an amplitude ramp and
constant frequency
constant deformation
deformation sweep
constant shear stress
constant speed
shear rate sweep
constant shear stress
torque sweep
set value
shear rate
shear rate
shear rate
shear stress
shear stress
shear stress
shear stress
response
shear stress
shear stress
shear stress
shear rate
shear rate
shear rate
deformation
shear stress
deformation
shear stress
deformation
deformation
shear stress
deformation
shear stress
deformation
shear stress
angle
angle
shear stress
shear rate
shear rate
shear stress
shear stress
shear stress
shear stress
deformation
shear stress
shear stress
shear rate
shear rate
9.3 Result
The results of tests are divided into three categories: raw measuring datas, rheological measuring datas and
model parameters.
The Rheometer RHEOTESTRN 4 generates data sets, which belong to the category raw measuring datas.
These are the basis of all other values.
The rheologial measuring datas are derived from the raw measuring datas. In principle, these can be
distinguished into two groups:
1. CR/CS-, measuring data: result from rotary tests
2. Oscillating measuring datas: result from Oscillating tests, that is the torque is changed sinusoidal, the
rotation direction changed, too / (see Tests)
The availability depends on the performed test. Thus, a value of the group two only is generated by an
oscillation test. The raw measuring datas and some internal parameters are generated at each test
(Gr. = 1,2).
84
data
Gr.
meaning
calculation
unit
time- t
1,2
1,2
raw measuring
data
raw measuring
data
s (SI), min
speed N
or
angle-
torque -M
1,2
temperature - T
1,2
shear stress -
1,2
shear rate - D
deformation -
1,2
raw measuring
data
raw measuring
data
= KTau * M
(KTaumeasuring
system)
D = kD * N
(KD measuring
system)
= kD *
(KD measuring
system)
=/D
dynamic viscosity -
complex viscosity -
imaginary part of
complex viscosity -
shear modulus G
complex modulus
Modul G*
storage modulus G
1
2
x
x
rotation
oscillation
oscillation
loss modulus G
oscillation
compliance J
rotation
complex compliance
J*
storage compliance J
oscillation
oscillation
loss compliance J
oscillation
phase angle -
angular frequency -
torque_max M
angle_max -
test
index
index cumulative
2
2
2
1,2
1,2
1,2
x
x
x
time cumulative t
system result 0 SR
1,2
1,2
status Mode - SM
result Mode - RM
1,2
1,2
= _max /
_max *
= _max /
_max * *
sin()
= _max /
_max * *
cos()
G=/
G = _max /
_max
G = _max /
_max * cos()
G = _max /
_max * sin()
J=/
J = _max /
_max
J = _max /
_max * cos()
J = _max /
_max * sin()
phase shift
between torque
and angle
Pa s (SI), mPa s
f=1/T
= 2 f
85
9.4 Analysis
Overview of available evaluations to determine statistical and rheological characteristics of a sample.
name
MW/StdAbw
Min / Max /
spreading
upper limit
lower limit
target value
correlation
yield point Tau
thixotropy
Newton
function / formula
average and
standard deviation
minimum value,
maximum value,
spreading (Max
Min)
application
statistics
chart
statistics
Ostwald de
Waele
viscosity
empirical function for formal adjustment to the
flow curvature
readout:
De Haven
- consistency factor k
- flow exponent n
polymer melts
readout:
-
Carreau
polymer solution
readout:
-
Cross
viscosity _0 0
taumax m
exponent n
viscosity infinite
viscosity _0 0
shear rate Db
exponent n
generally valid
readout:
-
viscosity infinite
viscosity _0 0
shear rate Db
exponent n
generally valid
Golub
readout:
-
viscosity infinite
viscosity _0 0
exponent factor k
86
Reiner Phillippoff
generally valid
readout:
-
viscosity infinite
0
- viscosity _0
- taumax m
generally valid
Meter
readout:
-
viscosity infinite
viscosity_0 0
taumax m
exponent n
paste, cream
Bingham
readout:
- yield point 0
- viscosity
empirical function for formal adjustment to the
flow curvature
Herschel Bulkley
readout:
-
yield point 0
- factor k
- exponent n
suspensions (paint, varnish, chocolate)
Casson
readout:
- yield point 0
- viscosity
suspensions
Casson allgm.
readout:
yield point 0
- viscosity
- exponent denominator n
Tscheuschner
allgm.
OSC-T-Sweep
MSCR-Test
MSCR-AASHTO
chocolate
readout:
-
G* and as a
function of
temperature
determination of
percent recovery and
Non recovery creep
compliance at the
creep stress levels
100 Pa; 1600 Pa and
3200 Pa
determination of
percent recovery at
the creep stress
levels 100 Pa and
3200 Pa
bitumen
yield point 0
viscosity infinite
viscosity str
exponent n
shear rate Dr
bitumen
bitumen
As a prerequisite for application of the evaluations of "Newton" to "Tscheuschner allgm." is the presence of a
sample which was measured as a flow curve. Flow curves arise from CR or CS test, executed a sweep or as
an equilibrium. Necessary are datas which describe the dependence of the shear stress as a function of shear
rate ( = f (D)).
87
Meaning
nothing is displayed
job name from Job definition
group and job name from Job definition
record from the field Product from Job definition
name of method used
group and name of method used
commentary to product / Setting from Job definition
measuring system used
sample name from Sample definition
record from the field Charge from Sample definition
user from Sample definition
date of sample creation
commentary to sample from Sample definition
it will be shown free text 1
it will be shown free text 2
it will be shown free text 3
Description
CR sweep: during 120 s with shear rate from 0 s-1 to 250 s-1
CR const: 30 s at constant rate 250 s-1
CR sweep: during 120 s from 250 s-1 to 0 s-1
Method parameters:
Trigger time: 1 s
Time raster1: 1 s
None interruption criteria
Method control
Job control
Table settings
88
Diagram settings
Report settings
1.
2.
3.
1.
89
5. Diagram setup
Diagram setting through settings
menu Diagram
6. Report setup
Report setting through settings
menu Report
1.
2.
3.
1.
Corresponding operations are displayed in the information window. During data obtaining they could be shown
with the help of inlays activation Table or Diagram. All single tests are displayed together in case of
activation of the whole sample (in sample tree). Only one single test will be shown in case of activation of this
single test. In state row (inlay Job is active) status is displayed and measurement end (end of analysis)
respectively.
90
Calibration fluid
NF 5000000
(appr. 5.000.000 mPas, at 25C)
NF 5000000
(appr. 5.000.000 mPas, at 25C)
Job
KalRotP3
KalRotP4
Calibration fluid
NF 5000000
(appr. 5.000.000 mPas, at 25C)
NF 5000000
(appr. 5.000.000 mPas, at 25C)
Job
KalOSCP3
KalOSCP4
9.8.1.1 Heating
105
85
Temperatur [C]
Jump
Start T = -10 C to Tsoll = 105 C
Set value with the maximum possible temperature
65
45
25
Sprung
-10C auf 105C
Jump from 10C to 105C
to 105C
Ramp from10C
Rampe
von -10C
auf 105C
Ramp
Start T = -10 C toTsoll = 105 C in 600 s
Controlled temperature increase within a predetermined time.
This corresponds to a ramp rate of appr. 0,19 K/s.
-15
0
100
200
300
Zeit [s]
400
500
600
91
HHeating
eizen
2
130
90
0,5
70
50
-0,5
30
-1
10
-1,5
-10
-30
200
1,5
110
Temperatur [ C]
-2
600
400
Zeit [s]
deviation [K]
0,1
-10
-0,009
0,08
30
-0,002
0,06
105
-0,012
0,04
9.8.1.3 Cooling
Temperatur [C]
T [C]
Constant
0,02
0
-0,02
-0,04
Deviation at - 10C bei -10C
Abweichungen
Abweichung
Deviation at 30Cbei 30C
Deviation at 105C bei 105C
Abweichungen
-0,06
Jump
Start T = 105 C to Tsoll = -10 C
set value with the maximum possible temperature
-0,08
-0,1
0
200
400
600
800
1000
Zeit [s]
105
85
Temperatur [C]
Ramp
Start T = 105 C toTsoll = -10 C in 600 s
Temperature controlled descent within a specified time.
This corresponds to a rampenrate of appr. -0,19 K/s.
65
45
25
5
-15
0
100
200
300
400
500
600
Zeit [s]
Khlen
Cooling
3
130
2,5
110
90
1,5
70
50
0,5
30
10
-0,5
-10
-30
0
200
400
Temperatur [C]
-1
600
Zeit [s]
92
Set values, Profiles and the commands START and STOP are specified by the software, it is not
necessary to operate manual on the control unit.
Before starting the software RHEOTEST RN 4 Managers, all cables have to be connected, the control
unit as well as the mains have to be switched on.
After finishing th e software RHEOTEST RN 4 Manager, place the control unit into STOP mode. There is
no temperature control / regulation anymore, now, the control unit can be switched off.
93
94
SENSOR BROKEN!
Error message:
Temperature is monitored permanently inside the device and error message appears after excess of a limit
value.
Cause of trouble:
Inadmissible heat source was emerged in the control unit device failure
Device is in operation at too high ambient temperature
Device fan is closed
Troubleshooting:
Assure acceptable environment conditions and necessary cooling
Inform service of RHEOTEST
95
Error message:
Incorrect Value!
Range of permissible values is prescribed for each menu choice where input of numerical values is possible.
This error message appears and input value is ignored if input value is above or below this range.
Cause of trouble:
Input numerical value was rejected during checking of plausibility.
Troubleshooting:
Input permissible value.
96