Professional Documents
Culture Documents
JUrnal Von Willwbrand
JUrnal Von Willwbrand
Abstract
A new, automated assay for von Willebrand factor
(vWF) activity has recently become commercially
available (HemosIL vWF activity assay,
Instrumentation Laboratories, Lexington, MA). We
prospectively studied 61 specimens from 58 patients
undergoing laboratory testing for suspicion of von
Willebrand disease with this new method, in
comparison with the established ristocetin cofactor
method. Assays for factor VIII and vWF antigen were
also performed using an established method on an
MDA-180 coagulation analyzer (bioMrieux, Durham,
NC) and a new method on an ACL TOP coagulation
analyzer (Instrumentation Laboratories). Blood types
were determined. The results showed no significant
difference between the assays for factor VIII (mean,
97% for MDA-180 and ACL TOP; P = .494) or vWF
antigen (mean, MDA-180, 109%; ACL TOP, 111%; P =
.766). The mean result for the ristocetin cofactor assay
was 106% vs 93% with the automated vWF activity (P
= .007). The automated activity assay was 100%
sensitive and 86% specific for detecting vWF
abnormalities and seems to be a suitable screening test.
Abnormal results should be followed up with a
ristocetin cofactor activity assay for confirmation.
Further study is recommended to confirm these
conclusions.
730
730
DOI: 10.1309/CEPND3LFHQ87XU4D
731
731
Results
The mean age of patients was 37.0 years (range, 2-89
years), and 61 specimens were analyzed. Two specimens were
excluded owing to interference by hemolysis (n = 1) or hyperlipidemia (n = 1) in the ACL TOP automated vWF activity
assay. Five specimens were excluded because the vWF level
was elevated, and there was insufficient specimen to repeat the
ACL TOP vWF activity assay at a higher dilution. In the
Bland-Altman plots and t tests, data from these 7 specimens
were included if appropriate (ie, factor VIII and vWF antigen
results were valid for comparison in these specimens). For the
remaining 54 specimens, the data set was complete.
Assay Comparisons
The mean values for factor VIII, vWF antigen, and vWF
activity are given in Table 1. Results for factor VIII and vWF
antigen obtained from the MDA-180 were similar to the results
obtained from the ACL TOP, with no significant differences.
50
ACL
TOP
97
109
106
0.96
97
111
93
0.88
P
.494
.766
.007
.003
30
20
10
0
10 0
20
30
40
50
100
200
300
100
50
0
0
50
100
150
Reference
Method
B
Difference (MDA ACL)
Table 1
Assay Comparisons: Mean Values for Factor VIII Activity,
vWF Activity, and vWF Antigen on the ACL TOP vs the
Reference Method in 54 Cases
100
50
0
50
50
100
Ristocetin Cofactor (%)
Figure 1 Bland-Altman plots comparing ACL TOP with reference method results. Bland-Altman plots show no bias between
the 2 factor VIII methods (A) or the 2 von Willebrand factor (vWF) antigen methods (B). The ristocetin cofactor result tended to
be higher than the ACL TOP automated activity result, indicating that the automated assay tends to underestimate the vWF
activity (C). MDA ACL, result from the MDA-180 analyzer minus that from the ACL TOP analyzer; Risto ACL, ristocetin
cofactor result minus the ACL TOP result. For proprietary and other information about the methods, see Table 1.
732
732
Table 2
Clinical Classification of Results With ACL TOP vs Reference Method in 54 Cases*
Results
Reference Method
Result Category
Normal (n = 42)
Nondiagnostic
abnormality (n = 7)
von Willebrand
disease (n = 5)
No. of
Cases
Reference
Method
ACL TOP
36
Normal
Normal
Normal
NDR
2
4
Normal
ND60
ND60
ND60
1
1
1
1
ND60
ND60
ND60, NDR
Type 1
ND60, NDR
Type 2
ND60
Type 2
O, 1 (100%)
O, 1 (100%)
O, 1 (100%)
A, 1 (100%)
Type 2
Type 2
O, 1 (100%)
1
1
1
Type 2
Type 2
Type 1
ND60
Type 2
Type 1
O, 1 (100%)
B, 1 (100%)
A, 1 (100%)
Comment/Details
Blood Type
ND60, nondiagnostic abnormality consisting of vWF of 60% but not low for blood type; NDR, nondiagnostic abnormality consisting of a ratio of <0.7; vWF, von Willebrand
factor activity.
* Normal was defined as vWF activity of >60% and a ratio of >0.7; and von Willebrand disease as follows: type 1, vWF low for blood type with a normal ratio; type 2, vWF low
for blood type with an abnormal ratio. For proprietary and other information about the methods, see Table 1.
Respectively refers to the reference method and the ACL TOP.
There were 6 (11%) total false-positive results.
DOI: 10.1309/CEPND3LFHQ87XU4D
733
733
Discussion
Screening assays should have high sensitivity, even at the
expense of specificity, so that no cases of disease are missed.
The presence of disease can then be confirmed or excluded
by a confirmatory assay, such as the ristocetin cofactor assay.
In this study, the automated vWF activity assay performed
very well as a screening assay, with 100% sensitivity and
86% specificity. Use of the automated vWF activity assay
would have successfully eliminated the need for 67% of the
ristocetin cofactor assays. Many clinical laboratories send out
ristocetin cofactor assays rather than perform the assay in
house owing to the skilled labor and specialized equipment
required and the labor-intensive nature of the assay. As an
alternative, the results of the present study suggest that laboratories could perform the automated vWF assay in house
(along with a vWF antigen and factor VIII assay), reducing
the need for ristocetin cofactor send outs by 67%. To detect
all possible cases of vWD, we recommend that a ristocetin
cofactor assay be performed if the automated vWF activity
result is 60% or less and/or if the vWF activity/vWF antigen
ratio is less than 0.7.
Laboratories that currently perform ristocetin cofactor
assays in house often perform the assay relatively infrequently in batches. The automated vWF activity assay is easier to
perform, and, therefore, it would be easier to offer on a more
frequent or even a stat basis, and it could also reduce the volume of ristocetin cofactor assays.
Although the present study involved 61 specimens with a
broad range of results, including type 1 and type 2 vWD, nondiagnostic abnormalities, normal results, and elevated values,
further study should be performed on an even larger number
of specimens to confirm the findings. The automated assay
assesses the ability of vWF to bind to platelet glycoprotein Ib.
734
734
Perform ristocetin
cofactor activity
No further testing
at this time
References
1. Rodeghiero F, Castman G, Dini E. Epidemiological
investigation of the prevalence of von Willebrands disease.
Blood. 1987;69:454-459.
2. National Committee for Clinical Laboratory Standards.
Collection, transport, and processing of blood specimens for
testing plasma-based coagulation assays. 4th ed. Wayne, PA:
National Committee for Clinical Laboratory Standards; 2003.
Approved Guideline H21-A4.
DOI: 10.1309/CEPND3LFHQ87XU4D
735
735