KT88-3200 Opration 090911
KT88-3200 Opration 090911
This document contains information that is protected by copyright. All rights are reserved. No part
of this document may be photo copied, reproduced or translated to another language without prior
written consent of Contec Medical Systems Co., Ltd. The information contained in this document
is subject to change without notice.
KT88-3200 digital brain electric activity mapping should be used by the physicians who have
received professional training.
The user should read this operational manual carefully before operating KT88-3200, and
operate it in accordance with the manual.
The user should observe KT88-3200 and the patient when operating it.
The user is responsible to supply the use situations of KT88-3200 to Contec Medical Systems
Co., Ltd.
Contec Medical Systems Co., Ltd supplies the standard product to the user.
Contc Medical Systems Co., Ltd installs and debugs KT88-3200 and trains the physicians.
KT88-3200 should be repaired by Contec Medical Systems Co., Ltd.
Contec Medical Systems Co., Ltd should use original accessories to repair KT88-3200.
Contec Medical Systems Co., Ltd is responsible to answer the questions about KT88-3200 on
time.
Warranty Information
Your CONTEC Warranty
Contec Medical Systems Co., Ltd (hereinafter referred to as Contec) hereby warrants that
Contec products (hereinafter referred to as Products) shall be free from defects in material and
workmanship under normal use, service and maintenance for the warranty period of such Product
from Contec or an authorized distributor or representative of Contec Normal use, service and
maintenance means operation and maintenance in accordance with appropriate instructions and/or
information guides. This Warranty does not apply to the Products caused by any or all of the
following circumstances or conditions:
Freight damage;
I
Part and/or accessories of the Products not obtained from or approved by Contec;
Misapplication, misuse, abuse, and failure to follow the Product instruction sheets and/or
information guides;
Accident, a disaster affecting the Products;
Alterations or modifications to the Products not authorized by Contec;
Other events outside of Contec s reasonable control or not arising under normal operating
conditions.
The remedy under this warrenty is limited to the repair or replacement without charge for labor or
materials, or any productrs found upon examination by contec to have been defective. This
remedy shall be conditioned upon receipt of notice by contec of any alleged defects promptly after
discovery thereof within the warranty period. Contec s obligation under the foregoing warranty
will further be conditioned upon the assumption by the purchase of the products ( ) of all carrier
charges with respect to any products returned to contecs principal place or any other place as
specifically designated by contec or an authorized distributor or representative of contec, and ( )
all risks of loss in transit. It is expressly agreed that the liability of contec is limited and that
contec does not function as an insurer. A purchase of a product, by its acceptance and purchase
thereof, acknowledges and agrees that contec is not liable for loss, harm or damage due directly or
indirectly to an occurrence or consequence therefrom relating to the products. If contec should be
found liable to anyone under any theory (except the expressed warranty set forth herein) for loss,
harm or damage, the liability of contec shall be limited to the lesser of the actual loss, harm or
damage, or the original purchase price of the product when sold.
Excluded from the limited warranty set forth above are consumable items such as paper, batteries,
electrodes, patient cable, lead wires and magnetic storage mediums.
Except as set forth herein with respect to reimbursement of labor chages, a purchases sole
exclusive remedy against contec for claims relating to the products for any and all losses and
damages resulting from from any cause shall be the repair or replacement of defective products to
the extent that the defect is notice and contec is notified within the warranty period. In no event,
including the claim for negligence, shall contec be liable for incidental, special or consequential
damages, or for any other loss, damage or expense of any kind, including loss of profits, whether
under tort, negligence or atrict liability theories of law, or otherwise. This warranty is expressly in
lieu of any other warranties, express or implied, including, but not limited to the implied warranty
of merchant ability and the warranty of fitness for a particular purpose.
Preface
This operation manual introduces you the total function of the KT88-3200 digital brain electric
activity mapping and the detail operation procedure.
Please read the manual carefully before use the machine, which makes you easier to use it.
II
Table of Contents
Chapter 1 Introduction....................................................................................................................1
Chapter 2 The main technical index..............................................................................................1
Chapter 3 Technical device..............................................................................................................2
Chapter 4 Software and Hardware setup Wizard........................................................................3
Chapter 5 The system operation...................................................................................................13
Chapter 6 Notice.............................................................................................................................37
III
Chapter 1 Introduction
The features of digital brain electric activity mapping:
The digital brain electric activity mapping transit head bioelectrical activity into marked
intuitionistic image, it becomes an advanced, no wound technical instrument for skull sickness,
qualitative analysis for head enginery. Characters for this equipment are as follows:
Help to bulge and recognize EEG partial features.
Recognize the subtle features which was easy to be ignored ago.
Help to academic communication, especially for the amateurs.
Evaluate the unusual conditions in the traditional EEG and the unexplained situations by eye
measure.
17. Product registration number: Hebei drug possession of firearms (quasi-) character No.
2,210,019, No. 2003
18.Medical Device Manufacturing Enterprise License: Hebei Food Drug Administration
20,050,526 Number of firearms production
19. Transport and storage
aThe range of Environment temperature-4055
bRelative Humidity range90%
cAtmospheric pressure range86kPa106kPa
20. Product Safety Type: II Class BF-based Application Part
This operation manual is used for KT88series digital brain electric activity mapping. Because of
the different setup in every series of machines, this manual is accordance to the largest
configuration.
This operation manual is used for KT88-3200 series digital brain electric activity mapping.
Because of the different setup in every series of machines, this manual is accordance to the largest
configuration.
chart 1
To continue, click next, if you would like to select a different folder, click Browse. Click
continue 3 times
c
h
a
rt
2
c
h
a
rt
3
chart 4
There will be shown "ready to install", click to install, pop-up "finish install wizard
KT883200", the installation is complete.
ch
ar
t
5
chart 6
Chart 7
Select language
and video and then c
lick the button of
"OK" to enter the
KT883200 system.
Uninstalling the
software:
Method 1.
Click "Start" menu in the
"All Programs" to find the installation package to install KT883200, click the following iconto
uninstall the software.
Method 2.
Find KT883200's installation directory (usually installed by default under the C drive), open the
installation directory to find and double-click the icon below to unload
5
Hardware Wizard:
1Amplifier and Flash lamp
After installing the software, Link the EEG amplifier and the USB of computer main frame, then
the computer will be prompted to find a new hardware:
chart 8
Choose the third one:no not this time, click to next to continue, then will show a Notice, click to
continue.
ch
ar
t
9
chart 10
There will be shown Completing the Wizard, click Finish.
NOTICE :
after click "finish",the computer will be prompted to find a new hardware again. And you
can setup the hardware according to the steps above again. The hardware will finish the
setup when setup is complete.
chart 11
(2) Video Optional Function
(1)Link the video class and one of USB port in computer , then the computer will setup the
driver automatically.
2}Please the softdog inset into one of USB port in computer, the softdog`s driver will be setup
as same as the ampliter`s driver.
Electrode Wizard:
1 The placement of electrode
This system use silver chloridize channel electrode, drip in brine liquor at 5% at work.. Wash
down with clean water after using and airing everyday. Cover it to avoid exposal to protect the
chlorination on electrode.
All
the
name
of
the
poles
are
FP1,
FP2,
F3,
F4,
C3,
C4,
P3,
P4,
O1,
O2, F7,
F8, T7, T8,
P7, P8 16
Fz, Pz, Cz, PG1, PG2, AFz,
FCz,
CPz,
CP3,
CP4,
FC3,
FC4,
TP7,
TP8,
FT7,
FT8 16
The pole location on the head shows int the follow picture
3.Connection lead
At least two electrodes for EEG was recorded, an electrode linking with the first gate of ECG
equipment, another one in second gate, Between the two electrode potential difference recorded
EEG was formed.
The electrode Placed in scalp is named active electrode or detection electrodes, the electrode
Placed in ear or zero potential is named unrelated electrode or reference electrode. EEG leads have
2ways: one is unrelated electrode unipolar leads method, another way is only use the bipolar lead
of the activity electrode to instead of the unrelated electrode.
(1) Unipolar lead (Reference lead): The activity of the scalp electrodes and unrelated to the
ipsilateral electrode is linked; Its depiction of the EEG recorded activity for the electrode and
unrelated to the potential difference between electrodes.
(2) Bipolar lead: Its the way of described and record two Active electrode in scalp and the two
gate in ECG is linked to instead of use unrelated to the electrod.
10
11
Picture 1
Click the New Patient; the physician input the parameters according to the patients condition.
The name item could input English or Chinese. Press Ctrl+ Space in keyboard would switch
into Chinese input mode, and then input Chinese name. The physician can choose the acquired
method (Roll View or Movie View) by mouses arrowhead. Click Gather and enter the picture
3.
12
Picture 2
The physician can choose the gather mode(16 montages or 19montages or 32 montages or Userdefined), Click OK and enter the picture 4.
13
Picture 3
The video setup
Click OK and enter video setup(as picture 4). The physician can set up the quality, differentiate
rate and other parameters of acquired picture.
14
Picture
4
The
15
Picture 5
The acquisition control panel
In the acquisition interface, the left is the acquired wave to display the patients EEG at real time
and the right is the acquisition control panel to display various of premiums events and control the
acquired methods.
Click
button, the physician
could chose acquired velocity
in practice.
Click button, the physician could
adjust the plus of EEG in practice.
Click
button, the frequency disturb
would get rid of AC work frequency disturb which is recommended to use for batter images.
16
Picture 6
Click end to stop EEG acquisition.
17
Picture 7
This interface is for user to replay EEG in sequence.
button is used to display 32-channel
EEG in same screen.
button is used to display 16-channel
EEG in same screen..
button is used to display 8channel EEG in same screen.
Physicians can be where it is needed to
add tag information, specifically as follows:
Click the toolbar button, click left mouse button where you want to add a tag, pop-up shown in
picture Picture 8 window, the corresponding location displays the flag icon .
18
Picture 8
The
physician can
choose
a
segment
of
EEG
to
analyze.
The
method is as
follow:
remove
the
mouse to the
beginning,
click the left
button inside the time rectangle, you can see a small red triangle, then select the end of the EEG
with the right button of the mouse in the same way. so as to choose the EEG segment.
The chosen EEG
segment turns into
white. Remove the
mouse to EEG, Click the
right key of the mouse, and
appear items. Move
mouse to Display Speed,
it changes to , The physician
could chose display speed in
practice. Move mouse to display Amp, it changes to the physician could chose display plus in
practice. Move mouse to Auto Play, and then the physician can choose to play EEG actually.
If you want to cancel the segment of EEG,
you can just click buttont. Finally, Click
button to dispose the chosen EEG segment to FFT exchange and insert value operation.
After the data disposal is over, the EEG is to be analyzed. Click button enter menus as follows:
This menu is to observe power chart distributing map. Click left button of mouse on any power
chart map could magnify this special lead power chart map.
19
20
absolute
relief
map
and
Picture 10 BEAM
Click button as picture11:
Picture 11
The physician
can
design
compressed
chart
map
according
to
menu tips:
After
the
design is over,
21
22
23
24
25
26
Picture 19
In this menu, the physician can choose the channel method.(monochannel or double-channel).
After the choice, Click yes and display the accordant EEG.
Lead setting instruction: click right of the mouse to customize the any lead from 3 to 9 lead (t
he first two are default leads which can not be modified), click the left of the mouse to drag any le
ad out of the channel list box, and this operation finishes the delete of one lead. If you need to add
lead only by using the mouse to click any two electrodes in the dialog box to complete a operation
of add Settings, click the button of "ok" after finish 32 leads setting. The next time you check lead
s, the setting will show to you.
Click button and enter the picture 20.
28
Picture 20
In
this
menu, the
physician
can design
filter
according to
the patients
EEG
to
clear EEGs
disturbance
and
false
balance.
Click OK
button and
appear EEG
which has being filtered.
Play
Click button in the toolbar would play EEG
button in the toolbar to stop play.(picture 21)
29
Picture 21
Double click the left of mouse in the EEG waveform area, dialog box wil
l appear as shown in the figure 4-19, and select the any one of the three re
d line in the dialog, after that drag the mouse and the gain of waveform, pe
riod and frequency will be changed.
30
Picture 22
In "under control" menu select "the system Settings" menu items, as shown in figure, 23 dialog
box will appear.
Picture 23
As the menu on the left side:
1Select" infromation setting", you can edit the information of hospital.( as shown figure 23);
2Select"parameter setting", you can set the parameters in the process of waveform replay.
31
( as
shown figure 24).
3Select "color setting", you can set color of background the induced event and so on.( as
shown figure 25);
The operation of this function is following:
Click the left of the mouse to the button for color after the color change in advance.
4Select" event add setting", the user can add the event himself , name it and select color ( as
shown in figure4-26).
The operation is following:
Click the left of mouse to the edit fram after "name" and input the name of the
adding event, click the color button below, click the "add" button to finish the
adding of the user-defined induced event; if you need to delete the induced event,
you can select the induced event item with the left of the mouse and click "delete"
button, in this way the delete is completed.
Click" default setting", all the information will be defaulted and show in this window. Click
"definite" to save it and if you open the case again, all the information is the same as the defaulted
information.( each time you set the parameters, and it will come into effect the next time you
open it).
Picture 24
32
Picture 25
33
Picture 26
Click you will come into the figure
27video replay control panel will appear on the
right side of the replay graph. The doctor
can control replay in this panel. "play" will
make the waveform show from beginning to end consequently and "stop" will make it stop. Under
the situation of no play, beginning and ending are those of EEG, and "increase degree" and
"decrease degree" is the gain which can adjust the EEG.
Picture 27
Click to come into the calibrate interface
picture 28
34
Picture 28
You can choose the different model of printing in the " printing model" of the "printing out"
menu. You can also choose the "model editor" to edit the report model and save it.
Graph printing :
According to picture 5,pictures 7-15, user can print out the graph according to the need.
The operation is following in the any
interface of picture 5,pictures 7-15clickand
the menu 29will appear and then click definiteyou can print out the topographic map
and EEG graph.
As show in the figure 29, you can set the printing area range and you can input the number
of the first page and the last page, if you want the current page, set the range to 1-1.
35
Picture 29
Select patients :
Click Select patient in the menu as picture 1 and enter picture 30.
36
Picture 30
As shown in the figure of 30, click any one case with one click of the left mouse, the
information of the patient will be shown in the right side of the column; if there are many cases
you need to choose, you can also do conditional fast search to find the case you need.
The operation is following:
Select the "select search" with the left of mouse,click the left of mouse to select in the small rect
angular box before screening case conditions, input condition content and click the "search" . T
he content that accords with the case condition will be displayed in the list on the left,and for th
ose cases do not meet the conditions of the case, the system will point out, Click the "clear up",
the content of the user input will be delect.
by clicking the left of mouse to select cases, and clicking the "open" button ( or double click
cases), you can enter the cases of patient according to the figure of4-5. The rest operation is the
37
Picture 31
As shown in the menu of figure 4-31, you can choose cases(You can also make use of the
function of research to find the case you need.), and modify it (them) in the right side of the
column. After that , click "save" and the prompt will appear, and you click "yes", you can save the
modified case. You can also delete the cases, click and select the number of the case you need to
delete, and them click "delete", you delete the case.
As shown in the figure 4-1, click " data format conversion", you can come into the figure 32.
With the tools of format conversion, you can transform the case into EDFASCII form in order
to data or academic exchange.
Click the button of "Data Type" to select the product model and click "Browse"Button to find
the cases file path (* eeg) on the second line left side, click the button "Data Transform" to begin
the data format conversion, if the conversion is successfully, the system will offer prompt.
Picture 32
Segment saving:
Open the patient case to examine,
physician can choose a segment of given
waveform to save according to the
requirement. And the steps are as follows:
below the last one lead waveform of the beginning place of the waveform segment selected ,click
the left key of mouse , and there will be the red triangle sign "". Correspondingly , at the end of
the waveform.,click the right key of mouse ,and you will pitch on a segment of waveform. As
shown in the figure4-1, you select " save segment" in the menu, the figure 33 will appear.
input the name of the segment,
click " OK" to save it. If you
need to view the old cases, you
can choose one case with left of
the mouse, click the "segment:"
shown in the figure 30 and
prompt of the segment selection
according to the figure 34.select
the segment
and open the
waveform.
Pict
ure 33
39
Picture 34
Event list :
Click the press in the tool column, and come into the event list as shown in the
figure35, you can find the name of the induced events and time of those events happen
with the mouse or the up/down
arrow
to select the event, and click
P
the right forward arrow to switch
icture 35
into the waveform of the induced
event.
Multilingu
al
interface :
In the Chinese
interface,
choose
"system
control" - > "Language Select" - > "English" switch to English interface state if System is in
English interface state, choose the "System Configure" - > "Language the Select ->"Chinese"
switch the system to Chinese interface state. The same operation ia also applicable to other
languages.
Exits :
Choose the item of "Exit" in the manu
corner of the menu to exit from KT88
topography.
Chapter 6 Notice
The notice for installation
Dont hang anything on the power wire and be treaded by people.
The machine needs good ground cable. The electric resistance isnt more than 4 . dont use
the heat pipe or water pipe as the temporary ground cable.
All the holes at the back of the machine should not be covered. Otherwise, the machine will
be damaged because its too heat.
The notice for application
The power voltage and frequency should be 220V and 50Hz.
Dont cover anything on the holes at the back of the machine.
Dont put the machine in a humid environment. When cleaning the machine, please take the
plug away from the socket. Dont use the liquid cleaner or the alcohol to wash it. Please use
the wet cloth to wipe it. When the machine meets thunderstorm or it stays for a long time,
make sure that the plug is taken away from the socket.
40
Cause
Correction
41
not clear.
The printer doesnt work.
1.The
print
connected.
line
isnt
1.The ground
connected well.
cable
isnt
Appendix
The principle of EEG relief map
EEG relief map is a new method to analyze head bioelectricity in accordance with computer
technology and computer picture disposal technology. As a new method around the world right
now , it transforms huge complex but abnormity head bioelectricity to a form which could reflex
quantitative change and orientation by computer. It displays complexion about diversification of
head energy, it becomes an excellence , no wound and scientific method for skull sickness
diagnose and research on head energy quantitative analysis. It is useful to analysis correctly on
head energy using this technology .
Principle and method:
Placement of electrodes: lay 16 lead on outboard head according to international 10/20 system
method standard, collect head wave signals and make A/D (modules transform) transform.
F-transform:
Transform the collected head wave signals to power chart map form to display the relationship
about head wave frequency flied and power.
Frequency assort:
frequency flied: 0.8~3.8Hz
frequency flied: 4.0~7.8Hz
1 frequency flied: 8.0~9.8Hz
2 frequency flied: 10.0~12.8Hz
1 frequency flied: 13.0~19.8Hz
2 frequency flied: 20.0~29.8Hz
42
43