Certificate of Calibration: Instruction Manual
Certificate of Calibration: Instruction Manual
Graduation
Minimum (1 piece)
0.005 kg
0.01 kg
0.02 kg
Part Weight
0.025 kg
0.05 kg
0.1 kg
SUPPLIED ACCESSORIES:
AC Adapter
#AC-1 -For use in USA/Canada (AC 120V 60Hz)
#AC-2 -For use in Europe (AC 240V 50Hz)
#AC-3 -For use in U.K. (AC 240V 50Hz)
#AC-4 -For use in Japan (AC 100V 50/60Hz)
#AC-5 -For use in AUS/NZ (AC 240V 50Hz)
Specifications
Models
SV-30
SV-50
SV-100
Capacity (kg, lbs)
30 kg/66 lbs
50 kg/110 lbs 100 kg/220 lbs
Readability (kg, lbs) 0.005 kg/0.01 lbs 0.01 kg/0.02 lbs 0.02 kg/0.05 lbs
Repeatability
3d
Weighing Modes
kg, lb & parts counting presets of 5, 10, 25 & 50 pieces
Tare Range
To capacity by subtraction
Calibration (NBS Class F)
30kg
50kg
100kg
3
Stabilization Time (sec)
Maximum Overload
200.0%
Overload Indication
E with audio tone
Underload Indication
Negative value
Auto Off
4 minutes no activity (battery only)
Optimum Operating
Temperature
64 - 77 F (18-25 C)
Optimum Operating
Humidity
Less than 80% RH
Display Type
LCD/.75 high w/sement test
Audio Tone
On, start up, function keys, overload, low battery
Low Battery Indication
LO w/10 beep signal
Power Requirements
9V alkaline or nicad battery, or included AC adapter
Platform Size/Type
12 x 12.5 stainless steel (305 x 315mm)
Dimensions HxWxD
4 x 14.5 x 18.5 (100 x 370 x 470mm)
Net Weight (lb/kg)
12/5.4
Optional Accessories
Calibration mass
Limited Warranty
Two Year Limited Warranty (effective February 1, 2002):
Models
Certificate
of Calibration
We hereby certify that this instrument
was manufactured in the United States
and that it has been calibrated using
appropriate reference standards traceable
to the
National Institute of Standards
and Technology.
This instrument has been fully
inspected and has passed all quality tests
as specified by
Acculab quality procedures.
To assure proper measurements, it is
necessary to periodically recalibrate the
instrument using appropriate standards
as specified in the operating manual.
Duration: The Acculab balances and scales are warranted as a consumer product to the original consumer
purchaser for a period of two (2) years from the original purchase date.
Instruction
Manual
PLEASE READ COMPLETE INSTRUCTIONS BEFORE USE
Platform
Detachable Cord
AC Adapter Input
LCD Display
Membrane
Coverage: The Acculab balances and scales are warranted against defects in material and workmanship.
The warranty includes parts, labor and return shipping to customer (ground freight).
This warranty does not cover batteries and shall not apply to any product damaged due to accident, misuse, abuse, lightning strike, power surges, improper service, improper packaging or other causes not arising out of defects in material or workmanship.
Disclaimer: The original purchasers sole remedy, Acculab and its Authorized Distributors sole liability are
set forth herein. In no event shall Acculab or its Authorized Distributors be liable for any damages, incidental or consequential, expenses, lost profits, lost savings, or any other claims due to the use or inability
to use the product/s. Acculab shall limit its obligations under warranty to a time not to exceed the original product warranty period set forth herein. Acculab reserves the right to either repair or replace the
balance or scale. This warranty gives you specific legal rights and you may have other rights which may
be different from state to state.
In the event that service is required, the Customer/Distributor should call Acculab (800-656-4400 or
267-988-8702) for technical support and proper return authorization if needed. The product must be
delivered to Acculab Service Center within the warranty period, freight charges prepaid. The customer is
responsible for product removal, installation and proper packing in order to prevent damage while shipping to Acculab Service Center.
This limited manufacturer warranty supersedes all other warranties pertaining to the Acculab products,
whether expressed or implied. Claims may be made by calling toll free 1-800-656-4400, local 267-9888702, by fax toll free 1-800-356-0338, local fax 303-940-2250 or in writing to:
Acculab
67 Buck Road, Suite 131
Huntingdon Valley, PA 19006
Acculab Service Center
6542 Fig Street
Arvada, CO 80004
Phone:
Fax:
Performance specifications measured within optimum temperature range
after recommended warm up and proper calibration. Specifications subject
to change without notice.
as applicable
Sales Office:
800-656-4400
267-988-8702
800-356-0338
303-940-2250
67 Buck Road, Suite 131
Huntingdon Valley, PA 19006
www.acculab.com
400258.1 Rev. B
Models
Thank you for purchasing an Acculab Electronic Balance. With proper care and treatment it will provide years of reliable service.
Please read all operating instructions carefully and be sure to activate your warranty per the instructions listed under the WARRANTY
section of this manual.
BEFORE YOU BEGIN
Avoid lengthy exposure to extreme heat or cold. Your balance
works best when operated within the temperature limits listed in
the specifications. Always allow the unit to acclimate to a stable
temperature before calibration.
Allow at least 30 minutes of warm up time before starting calibration to give internal components sufficient time to stabilize.
Keep your operating environment as clean as possible.
Electrostatic fields, gravitational fields, dust, dirt, moisture, vibration, air currents, and proximity to other electronic equipment
can all have an adverse effect on the reliability and accuracy of
your unit.
Handle with care. Carefully set weighing tray onto top of unit
until level. Gently apply all items to be weighed onto the center
of tray top. Although designed to be quite durable, avoid rough
treatment as this may permanently damage the internal sensor.
OPERATION
AC Adapter Operation:
1) The unit will operate on 115V/AC or 220V/AC, depending on
the type of AC Adapter provided. CAUTION: Use only a 12V
Adapter with a negative tip, as provided, to supply AC power.
Use of another type of adapter may permanently damage the
unit and will void the warranty.
2) Insert the female plug into the receptacle at the rear of the unit,
and then into the desired electrical outlet. Check your use of any
extension cord which may affect power output. Unplug AC
adapter when not in use for extended periods.
FEATURES
Power Up Segment Test: When first turning on the unit, all
segments of the LCD display will appear as shown below.
This display will remain for approximately
+ 88.8.8.8
Instruction Manual
Battery Operation:
1) One nine (9) Volt ALKALINE or 9V NICAD RECHARGEABLE BATTERY is required. Your balance contains a 9V recharging circuit,
which will continuously charge an approved 9V rechargeable
battery whenever the AC adapter is used.
Key Tone: Pressing any of the four keypads on the front panel
will emit a tone indicating registry of the function.
Overload: When an applied load exceeds the units capacity,
E will appear on the display accompanied by intermittent
beeps. Remove excessive load immediately. The unit should
return to normal operation.
Negative Value: Any tared value or a value left in memory
will be displayed as a negative number once all weight is
removed from the unit. Press tare to return to normal operation.
Tare: The tare feature is designed to allow the user to reset the
balance to zero at any time. Tare can also be used to eliminate
the value of a pan liner, scoop, beaker, etc. from a sample or
procedure, when the weighed material needs to be held in a container and only the net weight of the material is desired. Place
the container only (without material) on the weighing tray, wait
for a stable reading and then press tare. The unit returns to zero
and the weight value of the container is permanently removed
from the remainder of the procedure.
NOTE: When all weight is removed from the weighing tray, the
tared value of the container will be displayed as a negative number.
Press the tare key again to return the balance to zero.
Memory: The memory function is designed to allow the user
to store values and to then progressively add these values into a
sample or procedure. Unlike the tare feature, the unit returns to
zero and a memory indication appears at the upper left of the
display. After a subsequent weight is added, press memory
again. The original value is added to the current value giving the
total gross weight of the current sample. The memory feature is
useful when multiple weighing of different samples is needed
and a total gross weight of all samples is required.
Auto Off: An auto shut off feature is provided which is operational only while the unit is operating under battery power. The
unit will automatically turn off after approximately four minutes if
no active weighing takes place. Auto shut off will not occur during operation with AC Adapter.
KEYPAD FUNCTIONS
ON/MEMORY: Press to turn on balance. Once the unit is on,
the same key operates the additive memory feature. The following steps outline the procedure for cumulative weighing of samples.
1) If needed, place container on tray top, wait for stable reading,
then press tare to return balance to zero and permanently
remove container weight from the additive procedure.
2) Add first sample until desired weight is achieved.
3) Press memory to reset unit to zero. The memory symbol
( ) appears in the upper left corner, indicating the previous
weight value has been stored.
4) Add second sample until desired weight is achieved.
5) Press memory again. The ( ) symbol will disappear and the
display will show the combined weight of the first and second
samples. Repeat steps 2 through 4 for additional items.
NOTE: When the entire load is removed from the balance, the display will show a negative value equal to the tared value of the container if used. To reset the balance to zero, press the tare key.
OFF: Press to turn off balance. The off key is also used to perform the calibration procedure as outlined under CALIBRATION.
CAL/MODE: Press and hold this key to begin calibration procedure as outlined under the CALIBRATION section. After unit is
on or calibrated, briefly pressing the same key will change the
weighing mode. When balance is turned on the unit initially
defaults to the kilogram (kg) mode. The following mode may
also be selected: pounds (lbs).
TARE: Press tare to reset the balance to zero. Tare can be used
to eliminate the weight value of a container from a sample or procedure. The container weight is permanently removed from the
remainder of the procedure.
NOTE: When all weight is removed from the weighing tray, the
tared value of a container will be displayed as a negative number.
Press tare again to return the balance to zero.