Operating Instructions
VOLUMEASURE
Model:

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

29-4522 (CN-980A)

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

TABLE OF CONTENTS
I.

GENERAL INFORMATION

II.

USER RELATED DOCUMENTATION

III.

ASSEMBLY PROCEDURE

IV.

CALIBRATION

V.

OPERATING PROCEDURE

VI.

MAINTENANCE

VII.

ACCESSORIES

FIGURE 1 29-4522 (CN-980A)

29-4522 (CN-980A) Parts List

FIGURE 2, 29-4524 (CN-978) AND PARTS LIST

SPECIFICATIONS/MODEL

Rev.3:10/04

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

VOLUMEASURE
MODEL 29-4522 (CN-980A)

I.

GENERAL INFORMATION
The 29-4522 (CN-980A) Volumeasure is a direct reading instrument for making
field and laboratory in-place density tests of soils. The graduated cylinder (Model
29-4522/11, (CN-983B) used in this apparatus has the graduations permanently
printed on the cylinder walls.
In case of damage or breakage of the cylinder, replacement calibrated cylinders
are available. No further calibration of replacement parts is necessary in the
field; just insert the new cylinder in the apparatus, fill, and it is ready for
operation.
Each Volumeasure is supplied complete with a package of 12 density balloons
and a field density plate. It is shipped in a unitized carton for maximum shipping
protection. Should damage occur, necessary claims should be filed immediately
with the delivering carrier. The unitized carton may be used to store and
transport the Volumeasure to and from the job site.

II.

USER RELATED DOCUMENTATION


These operating instructions do not contain all the necessary information on the
specific test procedures. Please refer to ASTM D-2167 or AASHTO T-205.

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

III.

ASSEMBLY PROCEDURE
The Volumeasure is packed in a partially assembled condition. The only tools
needed for assembly are a 3/16" allen wrench and a 9/16" wrench.
A.

Remove the unit from the packing box and set it upright. Set aside the
field density plate (Figure 1,#3). Detach the thin brass pressure tube and
save for assembly later.

B.

Hold the cylinder support #2 and remove the two knurled screws holding
the base #1 in place. Remove cylinder and cylinder support.

C.

Prepare base.
1.

Assemble the gasket #11a in the base, if necessary, by fitting it into


the groove provided.

2.

Attach a density balloon to the lip provided in the top center of the
base. Check that it is securely attached.

3.

Install the pressure tube into the threaded hole in the base near the
sealing lip of the balloon. Tighten by hand.

4.

Close the control valve #18 so that handle is at right angles to the
valve.

5.

Set the base aside for now.

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

D.

Prepare cylinder
1.

Remove the graduated cylinder from the cylinder support. The


rubber hold down pad #11b and the steel pressure plate #13 should
be included with the graduated cylinder. Set these pads aside.

2.

Loosen the allen screw #20 on the cylinder support until it is flush
with the inside of the casting.

3.

Fill the graduated cylinder with water 1 1/2" (38.1 mm) from the top.
Note: High heat and fluids containing methyl alcohol are
detrimental to the plastic cylinder. Temperatures above 120oF
(48oC) and certain solvents cause cylinder deformation under
seating pressure. Water and ethylene glycol are the only
recommended liquids for the Volumeasure.

E.

4.

Hold the cylinder support by the handle, open end up. Insert first
the steel pressure plate, then the rubber hold down pad inside on
the surface against the allen screw.

5.

Insert the cylinder with water into the cylinder support, turning the
desired graduations toward the opening facing you.

Assemble base and cylinder.


1.

Hold the base by the coupling and valve assembly and insert it into
the graduated and support cylinder. The case handle and the
coupling assembly should be on opposite sides of the assembly.

2.

Insert the 2 knurled screws #22 and tighten, holding the base and
cylinder together. Tighten the screws finger tight.

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

IV.

3.

Holding the assembly on the edge of a table, tighten the allen


screw against the cylinder. Center the cylinder in the support. The
open end of the cylinder should be snug against the gasket in the
base, making the assembly water tight. Tighten the setscrew.

4.

Turn upright; inspect for leaks. If the unit is tight, lock the jam nut
on the allen screw.

5.

The Volumeasure is now ready for the calibration of the first


reading for density measurements.

CALIBRATION
A.

Position the field plate #3 on a flat surface. Set the Volumeasure in the
recessed hole in the field plate. Place the rubber bulb assembly (294522/13, (CN-984C) in the quick coupler #17 so that it will produce
pressure in the cylinder when actuated.

B.

Open the control valve #18 by turning the handle 90o and pump the
balloon down until the water level in the graduated cylinder has reached
its lowest point. The Volumeasure must be held firmly down in position
during this operation. Record the volume shown on the graduated
cylinder. This is the initial base reading.

C.

Release the rubber bulb from the quick coupler. Turn the bulb 180o and
insert the vacuum end of the bulb in the coupler. Pump the balloon back
into the cylinder and close the control valve. The Volumeasure is now
ready for making density measurements.

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

V.

VI.

OPERATING PROCEDURE
A.

With the field plate in a level position on the material to be tested, dig
through the hole in the plate to the desired depth. Retain all the material,
which is removed from the hole for weighing.

B.

Set the Volumeasure in the recess in the density plate, open the control
valve and pump the balloon into the hole. (Hold the Volumeasure down
firmly while pumping). Record the lowest point reached by the water in
the cylinder.

C.

Invert the pressure-vacuum bulb assembly and pump the balloon back
into the cylinder.

D.

Subtract the initial base reading from the final reading. This is the volume
of the hole in cubic feet.

E.

Make density calculations based on the volume of the test hole, and dry or
wet weight (as required) of materials removed (Density = Weight).
Volume

MAINTENANCE
The Volumeasure requires very little maintenance. Naturally, it should be kept
clean and protected when not in use. Do not store in direct sunlight.

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

VII.

ACCESSORIES
29-4522/10 (CN-982)

One dozen replacement density balloons

27-4522/11 (CN-983B)

Replacement graduated cylinder assembly for


29-4522

29-4522/12 (CN-984A)

Replacement gasket set, 2 gaskets

29-4522/13 (CN-984C)

Replacement rubber bulb assembly

88-7950 (CN-503)

Density Pick

88-7700 (CN-995)

Sampling Spoon

88-7042 (LT-96)

Sample can, round, 1 gallon

88-7960 (CN-998)

Chisel

88-7940 (CN-999)

Rubber mallet

78-7140 (L-415)

Triple Beam Scale, 2610 grams x 0.1 gram

88-1600/10 (L-770B)

Field Scale, 36 pounds by 0.01 pound

88-0210/02 (L-5B)

Double Wall Oven, chamber sizes 17" x 11 1/2" x 16,


115 volts, 50/60 cycles, A.C.

29-4524 (CN-978)

Volumeasure Pressure Gauge

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International
FIGURE 1 29-4522 (CN-980A)

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International
29-4522 (CN-980A) Parts List
Part No.

Item No.

Description

No. Req'd.

4950-0001

Base

4950-0002

Cylinder Support

4950-0003

Field Plate

4950-0004

Male Bulb Connector

4950-0005

Female Adaptor

4950-0006

Male Adaptor

4950-0007

Female Bulb Connector

29-4522/11
(CN-983B)

Replacement Graduated
Cylinder Assembly

4950-0008

Rubber Valve Disk

4950-0009

10

Pressure Tube

4950-0013

11(a,b)

Set of Gaskets

---------

--

------------------

4950-0011

13

Pressure Disk

9409-0011

14

"O" Ring 7/16 x 5/8 x 3/32

9429-0005

15

Rubber Pressure Bulb

29-4522/10
(CN-982)

16

Balloon (12 per pkg.)

8303-0027

17

Quick Coupler

8920-0060

18

Control Valve

8301-0027

19

Reducing Elbow

9518-0027

20

Allen Head Set Screw

9601-0060

21

Jam Nut 3/8-16 UNC THD

9519-0013

22

Knurled Screw

29-4522/12 (CN-984A)

Replacement Gasket Set

29-4522/13 (CN-984C)

Replacement Rubber Bulb Assembly

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

Pkg.

International

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

International

SPECIFICATIONS / MODEL
Specifications
Graduated Cylinder

Heavy-duty plastic; dual graduated

Balloons

Resilient latex; integral retaining collar

Guard

Aluminum with integral carrying handle

Guard Base

Aluminum; with fitting, outlets, valves and vent tube for


balloon and actuator bulb

Actuator Bulb

Double acting type for pressure or vacuum;


fitted with adapter, quick coupler and
valve assembly

Density Plate

Aluminum with flanged sides;


9" x 9" x 3/4"h. (229 x 229 x 19 mm)

Operating Liquid

Water

Weight

29-4522 (CN-980A): Net 12 lbs. (5.4 kg).

Model
29-4522 (CN-980A)

ELE International, Inc.


PO Box 608
Loveland, CO 80539
USA
Phone: (970) 663-9780
Fax: (970) 663-9781
E-Mail: soiltest@eleusa.com
Website: www.eleusa.com

1/20 cu. ft. (1596 ml) capacity

Distributor:

ELE International, Ltd.


Chartmoor Road Chartwell Business
Park
Leighton Buzzard, Bedfordshire
LU7 4WG England
Phone: + (0) 1525 249200
Fax: + (0) 1525 249249
E-Mail: ele@eleint.co.uk
Website: www.eleint.co.uk

10

You might also like