Professional Documents
Culture Documents
Contrast Reduction Using Energy Dependent Intensifying Screens
Contrast Reduction Using Energy Dependent Intensifying Screens
Department of Radiologic Sciences, Faculty of Allied Health Sciences and Nursing, Kuwait University, Kuwait
Abstract
The speed of intensifying screens, in particular rare earth screens, increases with beam energy. Using such a
screen, the measured contrast of an aluminium disc, with optical densities in the linear region of the
characteristic curve, was found to be significantly and consistently less than that calculated from the gamma
of the film-screen combination for all tube potentials from 40 to 140 kVp. This contrast reduction effect can
be explained in terms of the filtering action of the disc which increases the effective energy of the beam.
Hence, the film-screen combination is exposed to X-rays which have different effective energies. This results
in two characteristic curves of different speed. The contrast of the disc image is not dependent on the gamma
of either curve, but an effective gamma which links the two curves. This effective gamma is less than the
gamma of either curve. Contrast calculated using the effective gamma gave good agreement with the measured
contrast of the disc image over the tube potential range examined.
The energy dependent attenuation of a heterogenous the exposures Ei and E 2 are of different effective energies
X-ray beam leads to beam hardening as the beam passes since the X-ray beam passing through the anatomical
through an attenuating medium. Therefore, in conven- structure has been hardened, i.e. effective energy of Ex is
tional radiography the beam transmitted through ana- greater than the effective energy of E 2 . As the effective
tomical structures and onto the screen-film system has energy of the beam increases, the shape of the character-
a higher effective energy than the incident beam. Rare istic curve remains essentially unchanged but it shifts to
earth as well as calcium tungstate screens exhibit signi- the left [ 7 ] . This leads to the situation illustrated in
ficant energy dependency; speed increases with beam Figure 2. Exposure E 2 operates on curve A and exposure
energy [ 1 - 5 ] . However, this effect is more marked for Ei on curve B. Therefore, instead of a contrast C given
rare earth screens [5]. by OD 2 —OD 1; the resultant contrast C is given by
A recent observation has highlighted the fact that these OD 2 — ODi'. As a consequence, the contrast achieved is
energy dependent processes might interact in such a way less than would be expected from the value of y. In effect
that would lead to a reduction in the expected contrast.
Theoretical considerations
If we consider the film-screen response in terms of the
characteristic curve shown in Figure 1, then gamma of
the system (slope of the characteristic curve in the linear
region) is given by the following equation [6]:
£ OD2
Gamma = (OD 2 -OD 1 )/(log 1 0 E 2 - l o g 1 0 E t ) (1)
In the absence of scatter, the radiographic contrast, C, Q
of a structure of thickness t and attenuation coefficient O
pi is given by [6]: O
Attenuation coefficient
At each tube potential the half value layer (HVL) was
determined using type 1100 aluminium sheets and
narrow beam geometry, as shown in Figure 3. The linear
attenuation coefficient was calculated for each tube
potential using the following equation [6]:
C = 0.693 (3)
d)
a
"o
o Results
Characteristic curves of the Lanex/T Mat G system
O with no additional aluminium in the beam for different
OD, tube potentials are shown in Figure 4. As may be seen,
the range of exposures chosen was sufficiently large to
cover the gamma region of the film. The energy depen-
dency of the screen may be clearly seen, the speed increas-
ing by a factor of 3.4 (at an optical density of 1.0 above
base plus fog) between 40 kVp and 140 kVp. The effect
log E, log E2 on the film response of the additional 2 mm of aluminium
Log Exposure is illustrated in Figure 5, which shows the characteristic
curves obtained at 140 kVp with and without the alu-
Figure 2. Schematic shift of the characteristic curve for the minium. It can be seen that a substantial shift in response
attenuated beam to the left, y and y' defined. occurs as a result of beam hardening. A similar shift was
Methods
A gadolinium oxysulphide rare earth screen (Lanex
Detail) was used in combination with T-Mat G Film
(Kodak, Rochester, USA). All exposures were made with
FILTERS
a high frequency X-ray unit (Picker Clinix E, Cleveland,
USA). The inherent filtration of the X-ray tube was
0.5 mm Al equivalent and the added filtration was
2.5 mm Al. The film was processed using a Kodak RP
X-Omat, model M6B processor operating at 35 °C and
a cycle time of 90 s.
Disc contrast
A 2 mm aluminium disc was placed on the film-screen 200cm
cassette and imaged at each tube potential. Optical den-
sity values OD 2 and ODj were measured with a densi-
tometer, X-Rite model no. 301X (X-Rite, Grandville,
USA).
Gamma of the screen-film
The characteristic curves of the screen-film combi-
nation were determined from 40 to 140 kVp in 20 kVp DETECTOR
steps. Tube current was kept constant; only exposure
time was varied to obtain the required optical density.
Exposures were measured with a Victoreen 660-1 elec- LEAD SHIELD
trometer and a Victoreen model 660-3 ionization BACKINGv
chamber (Victoreen, Cleveland, USA). The detector res-
ponse is linear over the range of exposures measured
and its sensitivity is essentially unchanged for photon
energies from 30 to 140 keV. Two characteristic curves FLOOR
were obtained at each tube potential with and without Figure 3. Schematic diagram of the experimental arrangement
an additional 2 mm aluminium sheet placed in the beam. used to determine the half value layer.
2. VYBORNY, C J, METZ, C E, DOI, K and ROSSMAN, K, 5. KODERA, Y, DOI, K and CHAN, H P, Absolute speeds.of
Screen/film system speed: its dependence on X-ray energy, screen-film systems and their absorbed-energy constants,
Radiology, 125, 811-816 (1977). Radiology, 151, 229-236 (1984).
3. VENEMA, H W, X-ray absorption, speed, and luminescent 6. WOLBARST, A B, Physics of Radiology (Appleton & Lange,
efficiency of rare earth and other intensifying screens, Norwalk)(1993).
Radiology, 130, 765-771 (1979). 7. VYBORNY, C J, H and D curves of screen-film systems:
4. THUNTHY, K H and WEINBERG, R, Effect of kilovoltage factors affecting their dependence on X-ray energy, Med.
on the relative speed of rare earth screens, Dentomaxillofac. Phys., 6, 39-44 (1979).
Radiol.,15, 27-30 (1986).