Professional Documents
Culture Documents
Campimetro Octopus 101 SERNEW100
Campimetro Octopus 101 SERNEW100
No. 100
Page:
Subject:
1/6
Replaces:
To:
File:
C:\Programme\Adobe\Acrobat
4.0\Acrobat\plug_ins\OpenAll\Transform\temp\SERVICE-NEWS100.doc
Date:
15.05.2001
Below is a list of the most common errors in regard to the OCTOPUS 101. If you encounter one of
the errors below then please proceed as described.
1)
SERVICE NEWS
No. 100
Page:
Subject:
2/6
Replaces:
SERVICE NEWS
No. 100
Page:
Subject:
3/6
Replaces:
Proceed as follows:
1. Start an examination to see whether their is an eye image on the LCD screen of the
OCTOPUS 101. Is their an eye image on the LCD screen?
YES: Proceed to step 2.
NO: Check IR illumination and camera. The easiest way to check whether the IR
illumination or camera is defective is by doing the following: In the Test Programs switch on
the camera (Mainboard CCD Camera) and shine light into the cupola by means of a flash
light. If the LCD screen does not light up then it is most likely that the camera is defective and
needs replacement (see Service News No. 93). If however the LCD screen lights up then the
camera is recognizing the light. This means that the IR panels are defective and need to be
replaced.
After replacement check that the eye image is visible both on the LCD screen as well as on
the monitor of the PC.
2. In Perimetry Programs (on the PC) check that the video interface (Definitions Setup) is
activated, if a Frame Grabber Board is being used. It must be deactivated if the USB function
is being used. Check the video cable (USB cable on perimeters as from Serial No. 940).
Note: It is quite common that video cables have been disconnected by moving the PC or that
the cable has been damaged, therefore check the cable carefully. Is it connected correctly?
YES: Proceed to step 3.
NO: Connect the cable, check for any damages and run an examination to check that the
eye image is displayed on the monitor of the PC.
3. On Perimeters running with a Frame Grabber Board: Check the setup of the Frame
Grabber Board (video board inside the PC). If the setup is ok, it is most likely that the Frame
Grabber Board is defective and needs to be replaced. Replace the Frame Grabber Board and
install the correct software for the Frame Grabber Board as described in the Service Manual.
Run an examination to check that the eye image is displayed on the monitor of the PC.
On Perimeters running with the USB function: Check the installation, see Service News
No. 95 from pt. 9 - 11. Also check the settings described in Service News No. 96 and No. 97.
After the settings have been made, run an examination and check if the eye image is
displayed on the monitor of the PC. If the error still occurs, contact Service Department at
INTERZEAG AG.
SERVICE NEWS
No. 100
Page:
Subject:
4/6
Replaces:
SERVICE NEWS
No. 100
Page:
Subject:
5/6
Replaces:
5)
Proceed as follows:
1. Open Perimetry Programs V3.16 or higher (on PC) and check in Definitions Setup that the
correct COM port has been selected, it must be the one to which the connection cable to the
OCTOPUS 101 is connected (usually COM1).
Is the setting correct?
YES: Proceed to step 2.
NO: Choose the correct COM port in the program and check that the cable is connected to
the correct port. Run an examination to be sure that the connection is correct.
2. Check that no patient file (*.PVD) has more than 1000 examinations. If files have close to
1000 examinations stored in it, the file should be split up. Too large files can prevent an
examination from being downloaded from the OCTOPUS 101 perimeter to the PC. The result
will be an error message which can also refer to a communication problem. Are any files too
large?
YES: Split up all the large files into smaller files. Run an examination and save it in the new
file. If their is still a communication error, proceed to step 3.
NO: Proceed to step 3.
3. Reset the perimeter by doing the following:
Switch the OCTOPUS 101 off
Press and hold down the patient response button
Switch the OCTOPUS 101 on (response button is still pressed)
After 1 or 2 short beeps (about 2 seconds after turning the OCTOPUS 101 on), release
the patient response button
Run calibration procedure and start an examination. Does the error still occur?
YES: Proceed to step 4.
NO: Problem has been solved.
4. Check the serial RS 232 interface communication on the OCTOPUS 101 without the PC. To
do this follow the steps described in the Service Manual on page 57 under pt. 7.6.1. Note: If
you do not have a test plug you can connect pins 2 and 3 together, as well as pins 7 and 8.
Are 'Sent' and 'Received' messages (abcdefghijkl......) identical?
YES: Proceed to step 5
NO: There is an internal problem. First check the 5V DC and 12V DC on the power
distribution board which goes to the connector panel via J13 (J13: Pin 1 = 5V, Pin 2 = GND,
Pin 3 = 12V). If the voltage is not present (obviously the perimeter is switched on during the
measurement), the power distribution board needs to be replaced.
SERVICE NEWS
No. 100
Page:
Subject:
6/6
Replaces:
If the voltage is present, the connector panel needs to be replaced. After replacement check
the communication again.
5. If the function of the internal RS232 interface communication is correct, do the same
communication test as described in pt. 4 but this time connect the test plug to the end of the
cable (computer side). If the 'Sent' message does not appear under 'Received, replace the
RS232 cable. If the above mentioned points have been checked but you still have a
communication error, contact Service Department at INTERZEAG AG.
NOTE: