H P (0,01%) P T S 33 X S 35 X: Ighly Recise Ressure Ransmitters Eries Eries
H P (0,01%) P T S 33 X S 35 X: Ighly Recise Ressure Ransmitters Eries Eries
H P (0,01%) P T S 33 X S 35 X: Ighly Recise Ressure Ransmitters Eries Eries
mathematically compensated
Series 33 X
Series 35 X
/ programmable
graphic display on the PC. Up to 128 transmitters can be hooked together to a Bus-system.
Transmitter with Analog Output
Integrated in the XEMICS processor is a D/A converter of 16 bit for analog signal outputs of
420 mA or 010 V. The output rate is 400 Hz. The accuracy is diminished by this converting
process by 0,05 %FS. The digital output is available on all transmitters with analog output.
Programming
With the KELLER software READ30 and PROG 30, a RS485 converter (i.e. K102 or K107 from
KELLER) and a PC, the pressure can be displayed, the units changed, a new gain or zero set.
The analog output can be set to any range within the compensated range.
Series PD-33 X
Series 35 X
G1/2, flush diaphragm
PIN ASSIGNMENT
35 mm
75
83
22
67 mm
22
110
HEX 27
97
11,5
30
10
22
12 18.9
HEX 27
G 1/2"
10,5
G 1/4
Female
Port
Function
MIL
C-26482
Binder
723
DIN
43650
420 mA
OUT/GND
2 Wire
+Vcc
010 V
GND
3 Wire
OUT
+Vcc
RS485A
RS485B
Digital
2
18
59,1
G1/4"
Serie 33 X (G1/4)
G 1/4
Female 40
Port
Output
Serie PD-33 X
with DIN 43650 Connector
Serie 35 X (G1/2)
Subject to alterations
08/2009
KELLER AG fr Druckmesstechnik
CH-8404 Winterthur
Schwarzwaldstrasse 17
D-79798 Jestetten
www.keller-druck.com
KELLER
Specifications
Standard Pressure Ranges (FS) and Overpressure in bar
PR 33 X / PD 33 X / PR 35 X
1
3
10
30
PA(A) 33 X / PA(A) 35 X
0,81,2
1
3
10
30
100 300
700 1000
Overpressure
20
60
20
200
400
1000
1000
(analog, 2-wire)
420 mA
828 V
0,1 %FS
0,15 %FS
(analog, 3-wire)
010 V
1328 V
0,1 %FS
0,15 %FS
02,5 V / 05 V
628 V / 828 V
0,1 %FS
0,15 %FS
0,12,5 V
3,512 V
0,1 %FS
0,15 %FS
400 Hz
0,002 %FS
Gauges: 1 mbar or 0,05 %FS
Absolute: 0,5 mbar or 0,025 %FS (1040 C)
Load Resistance ()
Electrical Connection
Insulation
Storage-/Operating Temperature Range
Pressure Endurance
Vibration Endurance
Shock Endurance
Protection
CE-Conformity
Material in Contact with Media
Weight
Dead Volume Change
Remarks:
- Disturbance of the 4...20 mA signal can occur during communication through RS485
- All versions are also available for use in hazardous areas (Ei-versions); see sep. data sheet
- Options:
- Calculations such as density, differential pressure, flow, absolute value, etc.
- Different housing-material, oil filling, pressure thread or connector
Polynomial Compensation
This uses a mathematical model to derive the
precise pressure value (P) from the signals
measured by the pressure sensor (S) and the
temperature sensor (T). The microprocessor in
the transmitter calculates P using the following
polynomial:
P(S,T) = A(T) S0 + B(T) S1 + C(T) S2 + D(T) S3
x
=
=
=
=
A0 T0
B0 T0
C0 T0
D0 T0
x
x
x
x
+
+
+
+
A1 T1
B1 T1
C1 T1
D1 T1
x
x
x
x
+
+
+
+
A2 T2
B2 T2
C2 T2
D2 T2
x
x
x
x
+
+
+
+
A(T)D(T)
A3 T3
B3 T3
C3 T3
D3 T3
x
x
x
x
Accessories Series 30
Each Series 30 transmitter also integrates a digital interface (RS485 halfduplex) which you can make use of: Connect the transmitter to a PC or
Laptop via a converter RS232-RS485 (i.e. K102 or K107) or USB-RS485 (K104 or K104B). Two programs are offered for free:
READ30: Data collection with graphs
PROG30: Instrument Settings
Call up of information (pressure- and tempe- Fast read-out and viewing of the pressure
signals in a graph
rature range, version of software etc.)
Documentation of dynamic measurements
Indication of actual pressure value
Up to 16 transmitters on one serial
Selection of the units
connection (Bus-operation)
Setting of a new zero and gain for the
transmitter
Reprogramming of the analog output
(i.e. different unit, other pressure range)
Setting of the instrument address
(for Bus-operation)
Low-Pass Filter adjusting possibility
Software PROG30
You can also tie up the transmitters into your own software. You have then a documentation,
a DLL and numerous examples at your disposal.
Changing the plug connector
Laboratory applications require the same transmitter to be used at different measurement points with different electrical connection arrangements. To accommodate such applications, Keller can supply different connectors matching with the internal standard plug. This makes it
easy to exchange the electrical connector of the transmitter.
Subject to alterations
08/2009
KELLER AG fr Druckmesstechnik
CH-8404 Winterthur
Schwarzwaldstrasse 17
D-79798 Jestetten
www.keller-druck.com